High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors
Applicant
Foshan Coxo Medical Instrument Co., Ltd.
Product Code
EFB · Dental
Decision Date
Sep 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
High-Speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Straight Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Geared Angle Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Air Motors: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Device Story
Hand-held dental instruments; air-motor driven; used to hold rotary attachments (burs/drills) for cutting, grinding, and drilling. Operated by dental professionals in clinical settings. Inputs: compressed air; some models include electric connection for LED illumination. Outputs: mechanical rotation of dental burs. Device facilitates tooth preparation and restoration procedures. Benefits: enables precise removal of dental material and finishing of restorations.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10, -23, -11), electrical safety (IEC 60601-1, -1-2, 80601-2-60), and performance testing per ISO 14457.
Technological Characteristics
Materials: copper (chromium plated), aluminum, rubber, stainless steel. Sensing/Actuation: air-driven turbine/motor. Connectivity: some models include electric connection for LED. Sterilization: moist heat (ISO 17665-1). Standards: ISO 14457, ISO 9168, ISO 17664, ANSI/AAMI ES60601-1.
Indications for Use
Indicated for use by trained dental professionals for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations, and polishing teeth.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
High-Speed Air Turbine Handpiece / Low-Speed Air Turbine Handpiece (K181691)
Reference Devices
Dental Low-speed Handpieces and Accessories (K202786)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 22, 2023
Foshan COXO Medical Instrument Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China
Re: K220179
Trade/Device Name: High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB, EGS Dated: September 1, 2023 Received: September 1, 2023
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K220179
## Device Name
High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors
## Indications for Use (Describe)
High-Speed Air Turbine Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Straight Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
## Geared Angle Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Air Motors:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
**Type of Use (Select one or both, as applicable)**
| <span style="font-size: 12px;"> <span style="text-decoration: overline;">☑</span> Remediation Use (Part 21 CFR 601.3 Subpart D)</span> |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12px;">☐ Own-Use Exception (21 CFR 601.3 Subpart C)</span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K220179
- Date of Preparation 1.
09/21/2023
- 2. Sponsor
# Foshan COXO Medical Instrument Co., Ltd.
BLDG 4, District A Guangdong New Light Source Industrial Base, South of Luocun Avenue Nanhai District Foshan 528226 Guangdong China
Contact Person: Zheng yongjian Position: Legal Representative Tel: 13702544788 Fax: 0757-81800058 Email: 13702544788@163.com
- 3. Submission Correspondent
Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
{4}------------------------------------------------
- 4. Identification of Proposed Device
Trade Name: High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/ Air Motors Common Name: Dental Handpiece and Accessories Model(s): High-Speed Air Turbine Handpieces CX207, CX207-G, CX207-2, CX207-A, CX207-A-2, CX207-B, CX207-B-2, CX207-C, CX207-C-2, CX207-W, CX207-W-2
Straight Handpieces CX235-2, CX235-2A, CX235-2B, CX235-2F, CX235-2G, CX235-2C, CX235- 2S2
Geared Angle Handpieces CX235-1B, CX235-1C, CX235-1F, CX235-1G, CX235C1, CX235C2, CX235C3, CX235C4, CX235C5, CX235C6, CX235C7, CX235C8, CX235-2S, CX235-2S1
Air Motors: CX235-3B, CX235-3F, CX235-3C
Regulatory Information: Regulation Name: Dental Handpiece And Accessories Classification: I; Product Code: EFB; Additional Product Code: EGS; Regulation Number: 21 CFR 872.4200; Review Panel: Dental;
Indications for Use Statement:
### High-Speed Air Turbine Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
## Straight Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
#### Geared Angle Handpieces:
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Air Motors:
{5}------------------------------------------------
It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
#### Device Description 5.
The proposed device is used to hold rotary instruments such as burs and drills for cutting, grinding, and drilling operations. The dental handpieces are hand-held instruments driven by air motor, thus driving the rotation of dental attachments for the purpose of drilling and tooth cutting. Some models include an electric connection to power an LED light
#### 6. Identification of Predicate Device
Primary Predicate Device: 510(k) Number: K181691 Product Name: High-Speed Air Turbine Handpiece / Low-Speed Air Turbine Handpiece Manufacturer: Foshan Wenjian Medical Instrument Co., Ltd.
Reference Device: 510(k) Number: K202786 Product Name: Dental Low-speed Handpieces and Accessories Manufacturer: MicroP Technology (Taiwan), Inc.
- Non-Clinical Test Conclusion 7.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ANSI/AAMI ES60601-1:2005/(R)2012 And A1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests:
- > IEC 80601-2-60: 2019. Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment;
- ISO 10993-5:2009, Biological Evaluation of Medical Device-Part 5: Tests for Vitro > cytotoxicity;
- > ISO 10993-10:2021, Biological Evaluation of Medical Device-Part 10: Test for irritation and delay-type hypersensitivity:
- > ISO 10993-23:2021, Biological Evaluation of Medical Device – Part 23: Tests for irritation:
- > ISO 10993-11:2017, Biological Evaluation of Medical Device – Part 11: Test for systemic toxicity;
- > Pyrogen Study in Rabbits using USP General Chapter <151> Pyrogen Test;
- A ISO 14457 Second edition 2017-10 Dentistry - Handpieces and motors;
{6}------------------------------------------------
- > ISO 17664 Second edition 2017-10 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices
- > ISO 17665-1 First edition 2006-08-15 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- > ISO 9168 Third edition 2009-07-15 Dentistry - Hose connectors for air driven dental handpieces
- 8. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Primary Predicate Device | Reference Device | Remark | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------|
| Manufacturer | Foshan COXO Medical Instrument Co., Ltd. | Foshan Wenjian Medical<br>Instrument Co., Ltd. | MicroP Technology (Taiwan), Inc. | -- | |
| Device Name | High-Speed Air Turbine Handpieces/ Straight Handpieces/ Geared Angle<br>Handpieces/ Air Motors | High-speed air turbine handpiece<br>Low-speed air turbine handpiece | Dental Low-speed Handpieces and<br>Accessories | -- | |
| Model | High-Speed Air Turbine Handpieces:CX207, CX207-G, CX207-2,<br>CX207-A, CX207-A-2, CX207-B, CX207-B-2, CX207-C, CX207-C-2,<br>CX207-W, CX207-W-2<br>Straight Handpieces: CX235-2, CX235-2A, CX235-2B, CX235-2F,<br>CX235-2G, CX235-2C, CX23 5-2S2<br>Geared Angle Handpieces: CX235-1B, CX235-1C, CX235-1F,<br>CX235-1G, CX235C1, CX235C2, CX235C3, CX235C4, CX235C5,<br>CX235C6, CX235C7, CX235C8, CX235-2S, CX235-2S1<br>Air Motors:CX235-3B, CX235-3F, CX235-3C | Dental High-speed Turbine<br>Handpiece (WJ-114, WJ- 112,<br>WJ-124, WJ -122, WJ- 134, WJ<br>-132, WJ - 144, WJ -142, WJ<br>-154, WJ - 152, WJ -164, WJ<br>-162)<br>Dental Low-speed Turbine<br>Handpiece (WJ-414, WJ- 412,<br>WJ-424, WJ-422) | Contra-angle Handpieces: G25, G25L,<br>G23<br>Straight Handpieces: G65, G65L, G63 | -- | |
| Product Code | EFB, EGS | EFB, EGS | EFB, EGS | SAME | |
| Regulation No. | 21 CFR 872.4200 | 21 CFR 872.4200 | 21 CFR 872.4200 | SAME | |
| Class | I | I | I | SAME | |
| Indication | High-Speed Air Turbine Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>Straight Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth. | Dental High-speed Turbine<br>Handpiece (WJ-114, WJ- 112,<br>WJ-124, WJ -122, WJ- 134, WJ<br>-132, WJ - 144, WJ -142, WJ<br>-154, WJ - 152, WJ -164, WJ<br>-162) is intended for removing<br>carious material, excess filling<br>material, cavity and crown<br>preparation, finishing tooth<br>preparations and restorations, root | MicroP Dental Low-speed Handpieces<br>and Accessories is intended for removing<br>carious material, excess filling material,<br>cavity and crown preparation, finishing<br>tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>All the devices are designed for use by a<br>trained professional in the field of<br>general dentistry | SAME | |
| | Geared Angle Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>Air Motors:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth. | canal preparations and polishing<br>teeth.<br>Dental Low-speed Turbine<br>Handpiece (WJ-414, WJ- 412,<br>WJ-424, WJ-422) is intended for<br>removing carious material, excess<br>filling material, cavity and crown<br>preparation, finishing tooth<br>preparations and restorations, root<br>canal preparations and polishing<br>teeth. | | | |
| Air/water ports | 2/4/6 ports | 2/4 ports | up to four | DIFFERENT 1 | |
| Fiberoptic Glass Rod | CX207-G, CX235-1C, CX235C7 | Included in model WJ-162,<br>WJ-164, WJ-424<br>and WJ- 422Disposable | Included in model G25Land G65L | SAME | |
| Accessories | NA | NA | NA | SAME | |
| Composition of material | Copper (chromium plating), aluminum , rubber and stainless steel | Stainless steel, Brass, Aluminum,<br>Titanium | Stainless steel, Aluminum Alloy | DIFFERENT 2 | |
| Chuck | Push button, Screw | Push button, Screw | Push button | SAME | |
| Bur extraction force | 22N | 28N | / | DIFFERENT 3 | |
| Maximum air pressure | High-speed Air Turbine Handpieces:0.2-0.25Mpa<br><br>Straight handpiece & Geared Angle handpiece & Air Motors:0.245MPa<br>~0.3925MPa | Dental High-speed Turbine<br>Handpiece: 200kPa ~ 250kPa<br>(29.01 psi~ 36.25 psi) Dental<br><br>Low-speed Turbine Handpiece:<br>245 ~ 392 kPa (35.53 psi~ 56.85<br>psi) | 0.248-0.296MPa<br>(36psi~43psi) | SAME | |
| Maximum<br>pressure | water | High-Speed Air Turbine Handpieces:<br>0.2-0.25Mpa<br><br>Straight handpiece & Geared Angle handpiece & Air Motors:0.245MPa<br>~0.3925MPa | Dental High-speed Turbine<br>Handpiece: 200kPa ~ 250kPa<br>(29.01 psi~ 36.25 psi)<br><br>Dental Low-speed Turbine<br>Handpiece: 245 ~ 392 kPa (35.53<br>psi~ 56.85 psi) | 0.248-0.296MPa<br>(36psi~43psi) | SAME |
| Speed in rpms | | High-speed Air Turbine Handpieces:280000 ~<br>450000<br><br>Straight handpiece &Geared Angle handpiece<br>&Air Motors: <40000rpm | Dental High-speed Turbine<br>Handpiece: 300,000rpm ~<br>400,000rpm<br><br>Dental Low-speed Turbine<br>Handpiece: 18,000 ~ 20,000 rpm | <40000rpm | DIFFERENT 4 |
| Conformance<br>with<br>standards for shanks | | Type 3 per ISO 1797-1 | Type 3 per ISO 1797-1 | / | SAME |
| Coupling dimensions | | Complied with ISO 3964 | Complied with ISO 3964 | / | SAME |
| Hose connections | | Type 1 for 2-hole model / Type 2 for 4- hole<br>model/Type 3 for 6-hole model | Type 1 for 2-hole model / Type 2<br>for 4- hole model | / | SIMILAR |
| Lubricant | | The lubricants approved by FDA, such as DO-ALL Dental Handpiece<br>Lubricant<br>(K073353). | The specified lubricant, type<br>"PANA SPRAY Plus"<br>manufactured by NAKANISHI<br>INC (cleared in K163483), must<br>be used during routine<br>maintenance. | MicroP Halley Spray | SIMILAR |
| Compliance Standards | | Complied with ISO 10993- 5, ISO 10993-10, ISO 10993-23, ISO<br>10993-11, USP <151> Pyrogen Testing, IEC 60601-1, IEC 60601-1-2, IEC<br>80601- 2-60, ISO14457, ISO 17665-1, ISO 17664, ISO 9168 | Complied with ISO 10993- 5, ISO<br>10993-10, IEC 60601-1, IEC<br>80601- 2-60, ISO14457, ISO<br>17665, ISO 11138, ISO 11607,<br>ISO 17664 | Complied with ISO 10993- 5, ISO<br>10993-10, ISO14457 | SAME |
# Table 7-3 General Comparison
Page 5 of 8
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
### 10. Substantially Equivalent (SE) Conclusion
## DIFFERENT 1
The difference between proposed device and predicate device is Air/water ports, which is use for connection, the difference does not affect the safety and effectiveness of proposed device. Based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.
#### DIFFERENT 2
The Handpiece is tissue-contacting, so it is tested accordance with the requirements of ISO 10993-1 and guidelines. The test results meet the requirements. Based on the biological test results, the proposed device is as safe, and performs as well as the legally marketed predicate device.
## DIFFERENT 3
The difference between proposed device and predicate device is bur extraction force. The ISO 14457 required the bur extraction force shall be at least 22 N for Type 5. Based on the nonclinical tests performed, the proposed device meets the standard and is as safe, as effective, and performs as well as the legally marketed predicate device.
## DIFFERENT 4
The difference between proposed device and predicate device is Speed. The proposed device's speed range is covered by the Predicate Device's speed range. Therefore, the proposed device's speed range can meet the use requirement. We tested as ISO 14457, based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.
Conclusion: Based on the comparison of technological characteristics, demonstrated through bench testing and intended use, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.