SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SonoFinder Wireless Probe Type Ultrasound Scanner (Model: SF14L25)
Applicant
Bionet Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Oct 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
SonoMe and SONOFINDER are indicated for examining the adult, pregnant woman, and children. These products are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. An appropriately trained healthcare professionals can have operator qualifications. The device use settings are intended in hospital clinic, and medical office settings. The general clinical applications include fetal/ obstetrics(OB), gynecology(GYN), abdominal, small organ and peripheral vessel imaging. The modes of operation are B-mode, PDI(Power Doppler Imaging) mode, PW(Pulse Wave Spectral Doppler) mode and Harmonic mode.
Device Story
Wireless ultrasound system using pulsed-echo technology; consists of battery-operated hand-held transducer and software app running on commercial off-the-shelf (COTS) iOS, Android, or Windows mobile devices/tablets. Transducer transmits ultrasound waves into body; reflected echoes converted to electrical signals; processed and displayed as anatomical images on mobile device screen. Used in hospital, clinic, and medical office settings by physicians or trained healthcare professionals. Provides diagnostic imaging to assist in clinical decision-making for fetal, OB/GYN, abdominal, small organ, and vascular applications. Benefits include portability and wireless connectivity for diagnostic ultrasound imaging.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility, electrical, and mechanical safety standards.
Technological Characteristics
Hand-held wireless ultrasound transducer using piezoelectric material. Modes: B-mode, PDI, PW Doppler, Harmonic. Connectivity: Wireless to COTS mobile devices (iOS, Android, Windows). Biocompatible patient-contact materials. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1/5/10.
Indications for Use
Indicated for adult, pregnant, and pediatric patients. Used for fetal/OB, GYN, abdominal, small organ, and peripheral vessel imaging. Intended for use by or under supervision of licensed physicians or trained healthcare professionals in hospital, clinic, and medical office settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
October 11, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Bionet Co., Ltd. % ChuelWon Lee RA Manager 5F, 61 Digital-ro 31-gil Guro-gu Seoul, 08375 REPUBLIC OF KOREA
Re: K220169
Trade/Device Name: SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SonoFinder Wireless Probe Type Ultrasound Scanner (Model: SF14L25) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: September 19, 2022 Received: September 20, 2022
Dear ChuelWon Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220169
#### Device Name
SonoMe Wireless Probe Type Ultrasound Scamer (Model: 5C, 5CB, H5C, 10L, 10LB, H10L, H5C10L) and SONOFINDER Wireless Probe Type Ultrasound Scanner (Model: SF14L25)
#### Indications for Use (Describe)
SonoMe and SONOFINDER are indicated for examining the adult, pregnant woman, and children. These products are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. An appropriately trained healthcare professionals can have operator qualifications. The device use settings are intended in hospital clinic, and medical office settings. The general clinical applications include fetal/ obstetrics(OB), gynecology(GYN), abdominal, small organ and peripheral vessel imaging.
The modes of operation are B-mode, PDI(Power Doppler Imaging) mode, PW(Pulse Wave Spectral Doppler) mode and Harmonic mode.
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) | |
|----------------------------------------------------------------|--|
|----------------------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Bionet. The logo consists of the word "bionet" in lowercase letters, with a stylized blue butterfly above the word. The word "bionet" is in a sans-serif font and is black. The butterfly is a light blue color and has a simple, elegant design.
K220169
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided.
| 1. Submitter | |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Bionet Co., LTD<br>5F, 61 Digital-ro 31-gil<br>Guro-gu, Seoul, Republic of Korea [08375] |
| Connect: | Primary Contact Person Name: ChuelWon Lee<br>5F, 61 Digital-ro 31-gil<br>Guro-gu, Seoul, Republic of Korea [08375]<br>Tel: +82.70.7585.6459 or +82.10.2067.2442<br>Email: cwlee@ebionet.com<br><br>Secondary Contact Person Name: Robbin<br>5F, 61 Digital-ro 31-gil<br>Guro-gu, Seoul, Republic of Korea [08375]<br>Tel: +82.10.2067.2442<br>Email: shpark @ebionet.com |
| Date Prepared: | January 17, 2022 |
# 2. Proposed Device
| Device Name | Wireless Probe Type Ultrasound Scanner<br>(Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Trade Name | SonoMe |
| Regulation Number | 21 CFR 892.1550, 21 CFR 892.1560, 21 CFR 892.1570 |
| Regulation Name | Ultrasonic pulsed doppler imaging system<br>Ultrasonic pulsed echo imaging system<br>Diagnostic ultrasonic transducer |
| Product Code | IYN, IYO, ITX |
| Regulatory Class | Class II |
| Device Name | Wireless Probe Type Ultrasound Scanner<br>(Model: SF14L25) |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Trade Name | SONOFINDER |
| Regulation Number | 21 CFR 892.1550, 21 CFR 892.1560, 21 CFR 892.1570 |
| Regulation Name | Ultrasonic pulsed doppler imaging system<br>Ultrasonic pulsed echo imaging system<br>Diagnostic ultrasonic transducer |
| Product Code | IYN, IYO, ITX |
| Regulatory Class | Class II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Bionet. The logo consists of the word "bionet" in a sans-serif font, with a stylized blue butterfly above the word. The butterfly is positioned slightly to the left of the word "bionet". The overall design is simple and modern.
### 3. Predicate Device
| Type | Proposed Device | Predicate Device: |
|--------|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Convex | Model: 5C, 5CB, H5C | Wireless Probe Type Ultrasound Scanner /<br>Model: UProbe-C (K172750)<br>SONON Ultrasound Imaging System / Model:<br>SONON 300C (K151339)<br>C5 Diagnostic Ultrasound System (K171926) |
| Linear | Model: 10L, 14L, 10LB, H10L, SF14L25 | Wireless Probe Type Ultrasound Scanner /<br>Model: UProbe-L (K172750)<br>SONON Ultrasound Imaging System / Model:<br>SONON 300L (K170085)<br>C5 Diagnostic Ultrasound System (K171926) |
| Dual | Model: H5C10L | Wireless Probe Type Ultrasound Scanner /<br>Model: UProbe-C (K172750)<br>SONON Ultrasound Imaging System / Model:<br>SONON 300C (K151339)<br>C5 Diagnostic Ultrasound System (K171926)<br>Wireless Probe Type Ultrasound Scanner /<br>Model: UProbe-L (K172750)<br>SONON Ultrasound Imaging System / Model:<br>SONON 300L (K170085)<br>C5 Diagnostic Ultrasound System (K171926) |
### 4. Device Description
The SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SONOFINDER Wireless Probe Type Ultrasound Scanner (Model: SF14L25) are a wireless ultrasound system that uses pulsed-echo technology to transmit ultrasound images via wireless communication to a Tablet PC or Mobile Phone that utilizes the iOS, Android OS or Windows System.
The SonoMe Wireless Probe Type Ultrasound Scanner and SONOFINDER Wireless Probe Type Ultrasound Scanner are a portable, general-purpose, software-controlled, hand-held diagnostic ultrasound system that consists of (i) a commercial off-the-shelf iOS, Android OS or Windows System, (ii) the Wireless Probe Type Ultrasound Scanner software that runs as an app on the mobile device, (iii) the battery-operated, hand-held Wireless Probe Type Ultrasound Scanner transducer that communicates wirelessly with Tablet PC or Mobile Phone which supports by iOS, Android OS or Windows System and (iv) User Manual, USB charging cable and Wireless Charger (H5C10L only).
The Wireless Probe Type Ultrasound Scanner software can be downloaded to Tablet PC or Mobile Phone which supports by iOS, Android OS or Windows System and utilizes an icon touch-based user interface.
#### 5. Indications for use
SonoMe and SONOFINDER are indicated for examining the adult, pregnant woman, and children. These products are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. An appropriately trained healthcare professionals can have operator qualifications. The device use settings are
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Bionet. The logo features a light blue butterfly above the word "bionet" in a lowercase, sans-serif font. The butterfly is stylized with rounded wings, and the overall design is clean and modern.
intended in hospital clinic, and medical office settings. The general clinical applications include fetal/obstetrics(OB). gynecology(GYN), abdominal, small organ and peripheral vessel imaging, The modes of operation are B-mode, PDI(Power Doppler Imaging) mode, PW(Pulse Wave Spectral Doppler) mode and Harmonic mode.
### 6. Technology
The SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SONOFINDER Wireless Probe Type Ultrasound Scanner (Model: SF14L25) employ the similar fundamental scientific technology as its predicate device(s).
#### 7. Determination of Substantial Equivalence
The SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SONOFINDER Wireless Probe Type Ultrasound Scanner (Model: SF14L25) are substantially equivalent to the predicate devices with regards to intended use, principles of operation, technological characteristics and safety and effectiveness.
As described below, the system has been evaluated for acoustic output, biocompatibility, thermal, electrical, and mechanical safety, and has been found to conform to applicable standards and product specifications that demonstrate that the SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SONOFINDER Wireless Probe Type Ultrasound Scanner (Model: SF14L25) are substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Bionet. The logo consists of the word "bionet" in a sans-serif font, with the "b" in lowercase and the rest of the letters in lowercase. Above the word "bionet" is a stylized butterfly with a light blue color. The butterfly is positioned slightly to the left of the word "bionet".
# Comparison Table:
# Model: 5C, 5CB, H5C
| Comparison<br>Items | Subject Device | | | Predicate Device | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name /<br>Model Name | Wireless Probe Type<br>Ultrasound Scanner /<br>Model: 5C<br>(Under Review) | Wireless Probe Type<br>Ultrasound Scanner /<br>Model: 5CB<br>(Under Review) | Wireless Probe Type<br>Ultrasound Scanner /<br>Model: H5C<br>(Under Review) | Wireless Probe Type<br>Ultrasound Scanner /<br>Model: UProbe-C<br>(K172750) | SONON Ultrasound<br>Imaging System /<br>Model: SONON 300C<br>(K151339) | C5 Diagnostic<br>Ultrasound System<br>(K171926) |
| Manufacturer | Bionet Co., Ltd. | Bionet Co., Ltd. | Bionet Co., Ltd. | Guangzhou Sonostar<br>Technologies Co.,<br>Ltd. | Healcerion Co., Ltd. | Guangzhou Sonostar<br>Technologies Co.,<br>Ltd. |
| Product Code | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX | IYO, ITX | IYO, ITX | IYN, IYO, ITX |
| Classification | Class 2 | Class 2 | Class 2 | Class 2 | Class 2 | Class 2 |
| Indication for<br>Use | SonoMe is indicated<br>for examining the<br>adult, pregnant<br>woman, and children.<br>This products is<br>intended for use by, or<br>by the order of, and<br>under the supervision<br>of, a licensed<br>physician who is<br>qualified for direct use<br>of medical devices. An<br>appropriately trained<br>healthcare<br>professionals can have<br>operator<br>qualifications. The<br>device use settings are<br>intended in hospital<br>clinic, and medical<br>office settings. The<br>general clinical<br>applications include<br>fetal/obstetrics(OB),<br>gynecology(GYN) | SonoMe is indicated<br>for examining the<br>adult, pregnant<br>woman, and children.<br>This products is<br>intended for use by, or<br>by the order of, and<br>under the supervision<br>of, a licensed<br>physician who is<br>qualified for direct use<br>of medical devices. An<br>appropriately trained<br>healthcare<br>professionals can have<br>operator<br>qualifications. The<br>device use settings are<br>intended in hospital<br>clinic, and medical<br>office settings. The<br>general clinical<br>applications include<br>fetal/obstetrics(OB),<br>gynecology(GYN) | SonoMe is indicated<br>for examining the<br>adult, pregnant<br>woman, and children.<br>This products is<br>intended for use by, or<br>by the order of, and<br>under the supervision<br>of, a licensed<br>physician who is<br>qualified for direct use<br>of medical devices. An<br>appropriately trained<br>healthcare<br>professionals can have<br>operator<br>qualifications. The<br>device use settings are<br>intended in hospital<br>clinic, and medical<br>office settings. The<br>general clinical<br>applications include<br>fetal/obstetrics(OB),<br>gynecology(GYN) | Intended for<br>diagnostic ultrasound<br>echo imaging,<br>measurement, and<br>analysis of the human<br>body for general<br>clinical applications<br>including obstetrics<br>(OB), gynecology<br>(GYN) and general<br>(abdominal) imaging. | Intended for<br>diagnostic ultrasound<br>echo imaging,<br>measurement, and<br>analysis of the human<br>body for general<br>clinical applications<br>including obstetrics<br>(OB), gynecology<br>(GYN) and general<br>(abdominal) imaging. | C5 Diagnostic<br>Ultrasound System is<br>a general-purpose<br>ultrasonic imaging<br>instrument intended<br>for use by a qualified<br>physician for<br>evaluation for Fetal,<br>Abdominal, Pediatric,<br>Small Organ (breast,<br>testes, thyroid), Trans-<br>vaginal, Peripheral<br>Vascular, Musculo-<br>skeletal (Conventional<br>and Superficial),<br>Cardiac (neonatal and<br>adult), OB/Gyn and<br>Urology. |
| Comparison<br>Items | Subject Device | | | Predicate Device | | |
| | abdominal, small<br>organ and peripheral<br>vessel imaging.<br>The modes of<br>operation are B-mode,<br>PDI(Power Doppler<br>Imaging) mode,<br>PW(Pulse Wave<br>Spectral Doppler)<br>mode and Harmonic<br>mode. | abdominal, small<br>organ and peripheral<br>vessel imaging.<br>The modes of<br>operation are B-mode,<br>PDI(Power Doppler<br>Imaging) mode,<br>PW(Pulse Wave<br>Spectral Doppler)<br>mode and Harmonic<br>mode. | abdominal, small<br>organ and peripheral<br>vessel imaging.<br>The modes of<br>operation are B-mode,<br>PDI(Power Doppler<br>Imaging) mode,<br>PW(Pulse Wave<br>Spectral Doppler)<br>mode and Harmonic<br>mode. | Hospital, clinic, and<br>medical office settings | Hospital, clinic, and<br>medical office settings | |
| Environment<br>of Use | Hospital, clinic, and<br>medical office settings | Hospital, clinic, and<br>medical office settings | Hospital, clinic, and<br>medical office settings | Track 3 | Track 3 | |
| Acoustic<br>Output Levels | Track 3 | Track 3 | Track 3 | B | B, M, **PW, Color,<br>*PDI, Compound<br>Imaging | |
| Display mode | B, B/M, Color, *PDI,<br>**PW | B, B/M | B, B/M, Color, *PDI,<br>**PW | For use in all patients | For use in all patients | |
| Patient<br>Population | For use in all patients | For use in all patients | For use in all patients | General clinical<br>applications, including<br>fetal/obstetrics,<br>gynecology,<br>abdominal | Fetal, Abdominal,<br>Pediatric, Small Organ<br>(breast, testes,<br>thyroid), Trans-<br>vaginal, Peripheral<br>Vascular, Musculo-<br>skeletal (Conventional<br>and Superficial),<br>Cardiac (neonatal and<br>adult), OB/Gyn and<br>Urology. | |
| Clinical<br>application | General clinical<br>applications, including<br>fetal/obstetrics,<br>gynecology,<br>abdominal | General clinical<br>applications, including<br>fetal/obstetrics,<br>gynecology,<br>abdominal | General clinical<br>applications, including<br>fetal/obstetrics,<br>gynecology,<br>abdominal | Healthcare<br>professionals | Healthcare<br>professionals | |
| Users | Healthcare<br>professionals | Healthcare<br>professionals | Healthcare<br>professionals | Piezoelectric material<br>in the transducer is<br>used as an ultrasound<br>source to transmit<br>sound waves into the | The basic principle is<br>that system transmits<br>ultrasonic energy into<br>patient body and<br>implements post | |
| Principle/<br>Method of<br>Operation | Piezoelectric material<br>in the transducer is<br>used as an ultrasound<br>source to transmit<br>sound waves into the | Piezoelectric material<br>in the transducer is<br>used as an ultrasound<br>source to transmit<br>sound waves into the | Piezoelectric material<br>in the transducer is<br>used as an ultrasound<br>source to transmit<br>sound waves into the | | | |
| Comparison<br>Items | Subject Device | | | | Predicate Device | |
| Image Display<br>Unit | body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as image of anatomic structures.<br>Mobile device (4 to 13 inches approximately) | body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as image of anatomic structures.<br>Mobile device (4 to 13 inches approximately) | body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as image of anatomic structures.<br>Mobile device (4 to 13 inches approximately) | body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as image of anatomic structures.<br>Mobile device (4 to 13 inches approximately) | body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as image of anatomic structures.<br>Mobile device (4 to 10 inches approximately) | processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.<br>15-inch LCD monitor |
| Probe<br>Characteristics | Convex, 3.2/5MHz frequency | Convex, 3.2/5MHz frequency | Convex, 3.2/5MHz frequency | Convex, 3.5MHz frequency | Convex, 3.5MHz frequency | Supporting Linear probes (7.5MHz & 6.5MHz) and convex array probes (3.5MHz & 5.0MHz) |
| Probe<br>Connection to<br>Display | Wireless | Wireless | Wireless | Wireless | Wireless | Non wireless |
| Off-the-shelf<br>operation<br>system | iOS / Android / Windows | iOS / Android / Windows | iOS / Android / Windows | iOS | iOS / Android | Linux |
| Software | Run as an app on off-the-shelf mobile device | Run as an app on off-the-shelf mobile device | Run as an app on off-the-shelf mobile device | Run as an app on off-the-shelf mobile device | Run as an app on off-the-shelf mobile device | Based on an embedded Linux Operating System |
| System<br>Components | Commercial off-the-shelf iOS mobile device<br>Wireless Probe Type Ultrasound Scanner software that runs as an app on the mobile device<br>Wireless Probe Type Ultrasound Scanner battery-operated, hand-held ultrasound diagnostic transducer that communicates | Commercial off-the-shelf iOS mobile device<br>Wireless Probe Type Ultrasound Scanner software that runs as an app on the mobile device<br>Wireless Probe Type Ultrasound Scanner battery-operated, hand-held ultrasound diagnostic transducer that communicates | Commercial off-the-shelf iOS mobile device<br>Wireless Probe Type Ultrasound Scanner software that runs as an app on the mobile device<br>Wireless Probe Type Ultrasound Scanner battery-operated, hand-held ultrasound diagnostic transducer that communicates | Commercial off-the-shelf iOS mobile device<br>Wireless Probe Type Ultrasound Scanner software that runs as an app on the mobile device<br>Wireless Probe Type Ultrasound Scanner battery-operated, hand-held ultrasound diagnostic transducer that communicates | Commercial off-the-shelf iOS or Android mobile device<br>SONON Ultrasound Imaging System software that runs as an app on the mobile device<br>SONON Ultrasound Imaging System battery-operated, hand-held ultrasound diagnostic transducer that communicates | The C5 Diagnostic Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging system |
| Comparison<br>Items | Subject Device | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.