K220168 · Shenzhen Kaiyan Medical CO , Ltd. · OHS · May 17, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220168
Device Name
Skin Care Beauty Mask (Model: MJ-06)
Applicant
Shenzhen Kaiyan Medical CO , Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
May 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Skin Care Beauty Mask (Model: MJ-06) emits energy in the red and blue region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne.
Device Story
Facemask-shaped device applies light to face skin surface; utilizes 150 blue LEDs (465nm) for mild to moderate inflammatory acne and 150 red LEDs (640nm) for wrinkle improvement. Irradiance is 30mW/cm2. User-operated in home environment (OTC). Device features selectable 5-minute or 10-minute treatment timers with automatic shutdown. Benefits include non-invasive cosmetic treatment of acne and wrinkles.
Clinical Evidence
Bench testing only. Evaluated for safety and performance per ANSI AAMI ES60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Facemask form factor; 300 total LEDs (150 blue @ 465nm, 150 red @ 640nm); 30mW/cm2 irradiance. Powered by external adapter (100-240V AC input, 5V DC output). Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57. Biocompatibility per ISO 10993-5 and 10993-10.
Indications for Use
Indicated for individuals with full face wrinkles and/or mild to moderate acne.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 17, 2022
Shenzhen Kaiyan Medical Co Ltd Alain Dijkstra CEO 40A Fuxin Road Fuyong Subdistrict BaoAn District, Shenzhen, Guangdong 518101 China
Re: K220168
Trade/Device Name: Skin Care Beauty Mask (Model: MJ-06) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP Dated: January 17, 2022 Received: January 20, 2022
Dear Alain Dijkstra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220168
Device Name Skin Care Beauty Mask (Model: MJ-06)
Indications for Use (Describe)
The Skin Care Beauty Mask (Model: MI-06) emits energy in the region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K220168
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 878.4810.
## 1. Submitter's Information
Company Name: SHENZHEN KAIYAN MEDICAL CO LTD Address: 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong, China Contact Person (including title): Alain Dijkstra (CEO) Tel: +86 755 82129361 E-mail: regulation@kaiyanmedical.com
### Distributor
Company Name: GODDESS'S MESSENGER LTD Address: 1624 Market St Ste 226, #79525 Denver CO 80202 USA Phone: +86 17310376031
### Application Correspondent:
Contact Person: Alain Dijkstra Company: SHENZHEN KAIYAN MEDICAL CO LTD Address: 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong, China Tel: +86 755 82129361 Email: regulation@kaiyanmedical.com
### 2. Subject Device Information
Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over-The Counter Wrinkle Reduction (OHS) Trade Name: Skin Care Beauty Mask Model Name: MJ-06 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Regulatory Class: II
### 3. Predicate Device Information
Sponsor: SHENZHEN KAIYAN MEDICAL CO LTD
{4}------------------------------------------------
## K220168
Trade Name: Aduro Light Beauty Mask, MK-660, MK-66USBO, MK-66USBA, MK-66USBB, MK-020, MK-02A, MK-02B Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over-The Counter Wrinkle Reduction (OHS) 510(K) Number: K202390 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Regulation Class: II
### 4. Device Description
The Skin Care Beauty Mask (Model: MJ-06) is a facemask-shaped device, which makes use of specific light spectral characteristics and directly applies light onto the face skin surface. The Skin Care Beauty Mask can emit two different kinds of wavelengths. The blue light (465nm) is intended to help reduce the appearance of mild to moderate inflammatory acne. Red light (640nm) is intended to improve the appearance of wrinkles. There are 150 LED lights with a center wavelength of 465nm and 150 LED lights with a center wavelength of 640nm, and the energy provided by each mode is 30mw/cm2 . The user wears the mask on their face for the treatment, and the device will shut down automatically after finishing a 5-minute treatment or 10-minute treatment selected.
### 5. Intended Use / Indications for Use
The Skin Care Beauty Mask (Model: MJ-06) emits energy in the red and blue region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne.
### 6. Test Summary
Skin Care Beauty Mask (Model: MJ-06) has been evaluated the safety and performance by lab bench testing as following:
| Standards No. | Standard Title | Version | Date |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------|
| ANSI AAMI<br>ES60601-1 | Medical electrical equipment - Part 1:<br>General requirements for basic safety and<br>essential performance | 2005/(R)2012 and<br>A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R)2012 | 07/09/2014 |
| Standards No. | Standard Title | Version | Date |
| IEC 60601-1-11 | Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: Requirements for medical<br>electrical equipment and medical electrical<br>systems used in the home healthcare<br>environment | Edition 2.0 2015-01 | 06/27/2016 |
| IEC 60601-2-57 | Medical Electrical Equipment - Part 2-57:<br>Particular requirements for the basic safety<br>and essential performance of non-laser<br>light source equipment intended for<br>therapeutic, diagnostic, monitoring and<br>cosmetic/aesthetic use | Edition 1.0 2011-01 | 03/16/2012 |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: Electromagnetic disturbances -<br>Requirements and tests | Edition 4.0 2014-02 | 09/17/2018 |
| ISO 10993-5 | Biological evaluation of medical devices -<br>Part 5: Tests for in vitro cytotoxicity | Third edition 2009-<br>06-01 | 12/23/2016 |
| ISO 10993-10 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin<br>sensitization | Third Edition 2010-<br>08-01 | 07/26/2016 |
{5}------------------------------------------------
# K220168
## 7. Comparison to predicate device and conclusion
The subject device is very similar in design principle, indications for use, functions, materials, and applicable performance, when compared to the above models of the K202390 predicate. Differences
{6}------------------------------------------------
## K220168
between the subject device and predicate do not raise new types of questions with regard to the safety and effectiveness of the subject device for the proposed indications for use.
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Company | SHENZHEN KAIYAN MEDICAL CO LTD | SHENZHEN KAIYAN MEDICAL CO LTD | -- |
| Trade Name | Skin Care Beauty Mask | Aduro Light Beauty Mask | -- |
| Classification<br>Name | Over-The-Counter Powered<br>Light Based Laser For Acne (OLP), Light Based Over-The<br>Counter Wrinkle Reduction (OHS) | Over-The-Counter Powered Light<br>Based Laser For Acne (OLP), Light<br>Based OverThe Counter Wrinkle<br>Reduction (OHS) | -- |
| 510(k) Number | K220168 | K202390 | -- |
| Product Code | OHS, OLP | OHS, OLP | SE |
| Intended Use /<br>Indications for Use | The Skin Care Beauty Mask<br>(Model: MJ-06) emits energy in<br>the red and blue region of the<br>spectrum, specifically indicated<br>to treat full face wrinkles and/or<br>mild to moderate acne. | For MK-660, MK-66USBO, MK-<br>020:<br>The Aduro Light Beauty Mask<br>(Model: MK-660, MK-66USBO,<br>MK-02O) emits energy in the red<br>and blue region of the spectrum,<br>specifically indicated to treat full<br>face wrinkles and/or mild to<br>moderate acne. | SE |
| Wavelengths | Red: 640nm±10nm<br>Blue: 465nm±10nm | Red: 640nm±10nm<br>Blue: 465nm±10nm | SE |
| Power supply | Adapter:<br>Input: 100-240Va.c., 50/60Hz,<br>0.5A<br>Output: 5.0Vd.c., 2.0A | For model MK-660, MK-66USBO:<br>3.7Vdc 2600mAh Lithium battery<br>Adapter for charging only:<br>Input: 100- 240Vac;<br>Output: 5Vdc, 2A.<br>For model MK-020: 3.7Vdc<br>2600mAh Lithium battery | SE |
| Irradiance source | LEDs | LEDs | SE |
| Treatment time | 5 or 10 minutes/day, 3 times per | 10minutes/day, 3 times per week | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
| | week | | |
| Power Density<br>(mw/cm²) | 30mW/cm² | 30mW/cm² | SE |
| Location for Use | Face | Face | SE |
| Environment of<br>Use | OTC | OTC | SE |
| Safety and EMC | IEC 60601-1<br>IEC 60601-1-2<br>IE C60601-1-11<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IE C60601-1-11<br>IEC 60601-2-57 | SE |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | SE |
{7}------------------------------------------------
#### 8. Conclusion:
The proposed device utilizes the same technological characteristics as the above listed model configurations of K202390, and is also for use in the treatment of mild to moderate acne and treatment of wrinkles. After an analysis of the proposed device's design, functionality, safety, materials, and performance testing results, the sponsor believes that the device's technology does not raise new types of questions regarding safety and efficacy of the subject device for the proposed indications, and performance testing supports that the device can be used safely and effectively. The subject device Skin Care Beauty Mask (Model: MJ-06) is considered to substantially equivalent to the above-specified devices of K202390.
### 9. Date prepared: May 2022
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.