The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
Device Story
Battery-powered biofeedback and neuromuscular electrical stimulation (NMES) device for pelvic floor muscle dysfunction. Inputs: myoelectric signals from pelvic floor muscles via sensors. Processing: device evaluates EMG signals; provides multimedia biofeedback training; triggers electrical stimulation (ETS) when muscle contraction reaches a threshold; offers passive pulse stimulation programs. Output: visual/multimedia biofeedback and electrical pulses to pelvic floor muscles. Used in home environment by patients. Healthcare providers use device output to monitor rehabilitation progress. Benefits: strengthens pelvic floor muscles, treats incontinence, and maintains continence.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), biocompatibility (ISO 10993-1), nerve/muscle stimulator performance (IEC 60601-2-10), usability engineering (IEC 60601-1-6, IEC 62366-1), and software verification/validation.
Technological Characteristics
Battery-powered (7.4V DC/1200mAh Li-ion) NMES/biofeedback device. Features LCD touch screen. Output: 2-100Hz frequency, 50-450µs pulse duration, max current 94.4mA @ 500Ω. Connectivity: standalone. Software: embedded firmware for EMG signal processing and stimulation control. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for women with weak pelvic floor muscles requiring rehabilitation for stress, urge, or mixed urinary incontinence and maintenance of urinary continence.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
July 20, 2022
Shenzhen Konmed Technology Co.,Ltd. % Cassie Lee Manager Guangzhou Glomed Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, Guangdong 510530 China
Re: K220161 Trade/Device Name: Biofeedback Nerve and Muscle Stimulator. Models: KM530B, KM531B, KM536, KM537 Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI, HCC Dated: June 14, 2022 Received: June 21, 2022
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220161
Device Name
Biofeedback Nerve and Muscle Stimulator, Models: KM530B, KM531B, KM536, KM537
Indications for Use (Describe)
The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510 (k) Summary
## 1.Submitter Information
| 510 (k) submitter: | Shenzhen Konmed Technology Co., Ltd. | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | 601, Building B4, Shenchengtou Creative Factory Life<br>Science Park, Julongshan A Road, Xiuxin Block,<br>Kengzi Street, Pingshan District, Shenzhen,<br>Guangdong, CHINA, Postal Code: 518118 | |
| Contact Person Name: | Shuishan Yin | |
| Contact Title: | General manager | |
| Contact Person's email: | 2519021651@qq.com | |
| Contact Person's Phone<br>Number: | +86 755 8670 4556 | |
| Contact Person's Fax<br>Number: | +86 755 8670 4556 | |
| Preparation date: | July 19, 2022 | |
### Application Correspondent
| Contact Person Name: | Ms. Cassie Lee |
|-----------------------------------|--------------------------------------------------------------|
| Contact Title: | Manager, Guangzhou GLOMED Biological Technology Co.,<br>Ltd. |
| Contact Person's email: | regulatory@glomed-info.com |
| Contact Person's Phone<br>Number: | +86 20 8266 2446 |
## 2.Device Information
| Trade Name of the Device: | Biofeedback Nerve and Muscle Stimulator,<br>Models: KM530B, KM531B, KM536, KM537 |
|---------------------------|----------------------------------------------------------------------------------|
| Common Name: | Nonimplanted electrical continence device |
| Classification Name: | Nonimplanted electrical continence device,<br>Biofeedback Device |
| Classification Number: | 21 CFR 876.5320, 21 CFR 882.5050 |
| Regulatory Class: | II |
| Review Panel | Gastroenterology/Urology, Neurology |
| Product Code | KPI, HCC |
# 3.Predicate Devices
| 510(k) number | K202648 |
|--------------------------|-----------------------------------------|
| Trade name of the Device | Biofeedback Nerve and Muscle Stimulator |
{4}------------------------------------------------
The predicate has not been subjected to a design-related recall.
### 4.Device Description
The Biofeedback Nerve and Muscle Stimulator (Model: KM530B, KM531B, KM537) is a type of biofeedback and neuromuscular electrical stimulation therapy device for patients with pelvic floor muscle dysfunction. Through the evaluation of myoelectric signal acquisition, multimedia biofeedback training, electromyography triggered electrical stimulation, passive electrical stimulation training and treatment it helps to strengthen weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women.
The device is battery-powered with a touch screen Liquid Crystal Display (LCD) and offers the user a choice of electromyography (EMG) triggered stimulation (ETS) mode (The device provides a passive pulse stimulation when the level of active contraction of the pelvic floor muscles reaches a threshold by means of electrical feedback from the pelvic floor muscles. 17 (KM536, KM537) or 22 (KM530B, KM531B) pre-set sow-frequency pulses stimulate pelvic floor muscles Electrical Stimulation (STIM) programs with 6 pre-set biofeedback response-based EMG games and the EMG therapy are available for the users. The differences between the model KM536 are only the model's name and the number of STIM Programs. The differences between the model KM531B and KM537 are also only the model's name and the number of STIM Programs.
#### 5.Intended Use
The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
| | K220161 (Subject) | K202648 (Predicate) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Biofeedback Nerve and<br>Muscle Stimulator | Biofeedback Nerve and Muscle Stimulator |
| Indication For<br>Use | The device is intended to<br>provide electrical stimulation<br>and<br>neuromuscular<br>re-<br>education for the purpose of<br>rehabilitation of weak pelvic<br>floor muscles for the treatment<br>of stress, urge and mixed<br>urinary incontinence in women<br>and to maintain urinary<br>continence in women. | As a powered muscle stimulator the Biofeedback Nerve and<br>Muscle Stimulator is indicated for the following conditions:<br>• Relaxation of muscle spasm<br>• Prevention or retardation of disuse atrophy<br>• Increasing local blood circulation<br>• Muscle re-education<br>• Immediate post-surgical stimulation of calf muscles to<br>prevent venous thrombosis<br>• Maintaining or increasing range of motion<br>As a biofeedback device the Biofeedback Nerve and Muscle<br>Stimulator is indicated for the following conditions:<br>• Biofeedback, relaxation and muscle re-education purposes<br>As a nonimplanted electrical continence device the<br>Biofeedback Nerve and Muscle Stimulator is indicated for<br>the following conditions:<br>• Acute and ongoing treatment of stress, urge or mixed<br>urinary incontinence and where the following results may<br>improve urinary control: Inhibition of the detruser muscles<br>through reflexive mechanisms and strengthening of pelvic<br>floor muscles.<br>• Incontinence treatment for assessing EMG activity of the<br>pelvic floor and accessory muscles such as the abdominal |
### 6.Comparison of The Technological Characteristics with Predicate Device:
{5}------------------------------------------------
| | | and the gluteus muscles. |
|-------------------------------------|---------------------------------------------------|-------------------------------------------------|
| Prescription or<br>OTC | OTC | Rx Only |
| Power source | 7.4V DC/1200mAh<br>rechargeable lithium battery | 7.4V DC/1200mAh<br>rechargeable lithium battery |
| Maximum<br>Output Current<br>(±10%) | 94.4mA @ 500Ω<br>54mA @ 2kΩ<br>15 mA @ 10kΩ | 94.4mA @ 500Ω<br>54mA @ 2kΩ<br>15mA@ 10kΩ |
| Maximum<br>Phase Charge | 51.4μC@500Ω | 51.4µC @ 500Ω |
| Maximum<br>Current Density | 6.01mA/ cm²@ 500Ω<br>Surface = 15.7 cm² | 6.01mA/ cm² @500Ω |
| Maximum<br>Power Density | 0.2814mW/ cm²@ 500Ω | 0.2814mW(12mW) /<br>cm²@ 500Ω |
| Frequency (Hz) | 2-100Hz | 2 -100Hz |
| Pulse Duration<br>(usec) | 50-450µs | 50- 450 µs |
| Net charge | For pulsed symmetric,<br>biphasic:<br>0µC @ 500Ω; | For pulsed symmetric,<br>biphasic: 0µC @ 500Ω; |
As evidenced by the above table, both the subject and the predicate devices have the same intended use. Additionally, the subject device has identical technological characteristics to the predicate except the predicate device was cleared for prescription only use and the subject device is an over the counter (OTC) use device. However, this difference does not raise different questions of safety or effectiveness and the testing mentioned below showed that the subject is substantially equivalent with the predicate.
## 7. Performance Data
In support of this premarket notification, Shenzhen Konmed Technology conducted the following safety and performance testing on the subject device-
- Electrical safety test according to IEC 60601-1, Medical electrical equipment Part 1: ● General requirements for basic safety and essential performance
- Electromagnetic compatibility test according to IEC 60601-1-2, Medical electrical ● equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . Performance testing according to IEC 60601-1-11, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- Biocompatibility test according to ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Performance test according to IEC 60601-2-10, Medical electrical equipment -- Part 2-● 10: Particular requirements for the basic safety and essential performance of nerve and
{6}------------------------------------------------
#### muscle stimulators
- Usability Study according to IEC 60601-1-6, Medical electrical equipment Part 1-6: ● General requirements for basic safety and essential performance - Collateral standard: Usability and IEC 62366-1, Medical devices - Part 1: Application of usability engineering to medical devices
- Software verification and validation test according to the FDA guidance document . "Guidance for Pre-Market Submissions and for Software Contained in Medical Devices"
All testing results confirmed that the products described in this submission met the necessary specification.
#### 8. Conclusion
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.