K220156 · Tung Keng Enterprise Co., Ltd. · ITI · Oct 7, 2022 · Physical Medicine
Device Facts
Record ID
K220156
Device Name
Electrically Powered Wheelchair, Joy Rider
Applicant
Tung Keng Enterprise Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Oct 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.
Device Story
Joy Rider is a portable, foldable, electrically powered wheelchair for indoor use by disabled or elderly adults. Input is provided by the user via a joystick control panel; the device transforms this input into motion using two brushless DC motors and an electrical controller. It features electromagnetic brakes that engage automatically when the joystick is released. The device is powered by a 24 VDC Lithium-ion battery. It is designed for maneuverability and portability, with a lightweight 6063 aluminum alloy frame. The healthcare provider or user does not require complex clinical training; the device is intended for personal mobility assistance. By providing a motorized, easily maneuvered, and foldable transport solution, it enhances the user's ability to travel and maintain independence. Safety is managed through standard braking, stability, and obstacle-climbing performance, with the device capable of climbing obstacles up to 1.96 inches and operating on inclines up to 10 degrees.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including ANSI/RESNA WC-1/WC-2, ISO 7176 series (stability, braking, energy consumption, dimensions, speed, obstacle climbing, electromagnetic compatibility), and biocompatibility testing (ISO 10993-5, ISO 10993-10).
Indicated for adults (over 22 years) with mobility limitations requiring a seated transportation vehicle for indoor use.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2022
Tung Keng Enterprise Co., Ltd. % Jen Ke-Min Contact Person Roc Chinese-European Industrial Research Society No. 58. Fu Chiun Street Hsin Chu, Taiwan 30067 Taiwan
Re: K220156
Trade/Device Name: Electrically Powered Wheelchair, Joy Rider Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 30, 2022 Received: September 7, 2022
Dear Jen Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in large, bold, dark blue font. To the right of the letters are three blue arrows pointing to the right, with each arrow being a slightly lighter shade of blue. To the right of the arrows is the word "City" in a smaller, black, italicized font.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
# 510(k) Summary (per 21 CFR 807.92)
# 510(k) number: K220156
| Submitter's Name: | Tung Keng Enterprise Co., Ltd. |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Date summary prepared: | October 6, 2022 |
| Trade Name: | Joy Rider |
| Common or Usual Name: | Powered Wheelchair |
| Classification Name | Powered Wheelchair, Class II, 21 CFR 890.3860 |
| Product Code: | ITI |
| Company contact person: | Dr. KE-MIN JEN |
| Email: | ceirs.jen@msa.hinet.net |
| TEL: | +886-3-5208829 |
| Predicate Device | Manufacturer: Nanjing Jin Bai He Medical Apparatus Co., Ltd.<br>Device name: Powered Wheelchair, DYW30A(D09)<br>510(k) number: K170787 |
#### ● Indications for Use:
The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.
#### ● Intended User Population:
The target population of the device is Adults only. (FDA defines 'an adult' as anyone over the age of 22 years.)
#### . Device Description:
The TUNG KENG Electrically Powered Wheelchair, Joy Rider, is suitable for indoor uses. It is characterized by high portability in comfortable travel, small storage and convenient delivery in daily life. As folding of Joy Rider only takes just a few seconds, so this model increases opportunities for travel and provides small storage. To use Joy Rider makes your travel in any place and at any moment! It is an individual transportation vehicle for disabled and elderly people that experience difficulties in walking. The Joy Rider comes as preassembled, designed to be lightweight, easily maneuvered, highly portable, and most importantly, safe and comfortable.
It is driven by two brushless DC motors, and contains two foldable armrests, a seat belt, a
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font, followed by a series of blue chevrons that get progressively lighter in color. The word "CITY" is in a thinner, black font and is separated from the chevrons by a forward slash.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
backrest, a seat cushion and small with light texture, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, one Lithium-ion battery, an off-board battery charger, a control panel, and an electrical controller.
The device is powered by one 24 VDC/ 10.5 Ah Lithium-ion battery with 9.3 miles cruising range that can be recharged by an off-board battery charger, rated at input: 100-240 VAC / 50-60 Hz, and output: 24 VDC/ 2 A, that can be plugged into an AC outlet, when the device is not in use. Its overall dimensions are 33.6' x 22.6' x 35.8" (855 x 580 x 910 mm). The user can activate the joystick to move in the direction of the joystick is actuated. When the user releases the joystick, the device slows to stop and the brakes are automatically re-engaged. The maximum weight capacity of Joy Rider is 264 lbs. (120 kg) and its maximum speed is 3.75 mph (6 km/h).
The following surfaces are recommended NOT to operate on:
- Sand surface �
- Wet or icy surface �
- Road maintenance hole metal cover �
- � Avoid going up multiple steps.
- � Avoid using escalators. Use the elevator.
- Too steep incline over 10 degrees.
- Obstacle climbing ability: 1.96" (50 mm)
#### ● Non-clinical performance tests
- ANSI/RESNA WC-1:2019 Requirements and test methods for wheelchairs ● (including scooters)
- ANSI/RESNA WC-2: 2019 Additional requirements for wheelchairs (including scooters) with electrical systems
- ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability, (FDA Recognition Number: 16-195)
- ISO 7176-2:2001 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs, (FDA Recognition Number: 16-202)
- ISO 7176-3:2013 Wheelchairs - Part 3: Determination of effectiveness of brakes, (FDA Recognition Number: 16-192)
- ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range, (FDA Recognition Number: 16-162)
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space, (FDA Recognition Number: 16-163)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font, followed by a series of blue chevrons that get progressively lighter in color. The word "CITY" is in a black, italicized font.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Fax: +886-4-25325000 Tel: +886-4-25321000 Website: www.dkcity.com
- ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed. acceleration and deceleration of electric wheelchairs, (FDA Recognition Number: 16-204)
- ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions, (FDA Recognition Number: 16-196)
- ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength, (FDA Recognition Number: 16-197)
- ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchair. (FDA Recognition Number: 16-167)
- ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs, (FDA Recognition Number: 16-164)
- ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies, (FDA Recognition Number: 16-190)
- ISO 7176-13: 1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces, (FDA Recognition Number: 16-25)
- ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods, (FDA Recognition Number: 16-165)
- ISO 7176-15:1996 Wheelchairs -- Part 15: Requirements for information disclosure, documentation and labeling, (FDA Recognition Number: 16-27)
- ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devices, (FDA Recognition Number: 16-191)
- ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, (FDA Recognition Number: 16-166)
- ISO 7176-22:2014 Wheelchairs - Part 22: Set-up procedures, (FDA Recognition Number: 16-198)
- ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, (FDA Recognition Number: 2-245)
- ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, (FDA Recognition Number: 2-174)
- IEC 60601-1:2005 Medical electrical equipment - Part 1 General requirements for basic safety and essential performance, (FDA Recognition Number: 19-4)
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, (FDA Recognition Number: 19-8)
- IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. (FDA Recognition Number: 19-13)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font, followed by a series of blue chevrons that get progressively lighter in color. The word "CITY" is in a black, italicized font, with the "C" appearing to be connected to the chevrons.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
#### ● Biocompatibility of patient-contacting parts
| Patient-contacting Parts | Material's name | Contact classification & contact duration | Tests conducted | Rational for the tests conducted | Corresponding test reports |
|-------------------------------------------|-----------------|----------------------------------------------------|------------------------------------------------------------------------------|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.<br>Seat cover<br>Backrest<br>Seat belt | Vinyl Fabric | Surface contacting-<br>Less than 24 hours duration | Cytotoxicity<br>Sensitization +<br>Skin Irritation testing | Per ISO 10993-1:2009 | *Vinyl Fabric ISO 10993-5 (Cytotoxicity Test)_Joy rider<br>*Vinyl Fabric ISO 10993-10 (Maximization Sensitization Test)_Joy rider<br>*Vinyl Fabric ISO 10993-10 (Skin Irritation Test)_Joy rider |
| 2. Armrest | PU Foam | Surface contacting-<br>Less than 24 hours duration | Cytotoxicity +<br>Maximization<br>Sensitization +<br>Skin Irritation testing | Per ISO 10993-1:2009 | *PU Foam ISO 10993-5 (Cytotoxicity Test)_Joy rider<br>*PU Foam ISO 10993-10 (Maximization Sensitization Test)_Joy rider<br>*PU Foam ISO 10993-10 (Skin Irritation Test) Joy rider |
| 3. Joystick | TPU | Surface contacting-<br>Less than 24 hours duration | Cytotoxicity +<br>Maximization<br>Sensitization +<br>Skin Irritation testing | Per ISO 10993-1:2009 | *TPU Joystick ISO-10993-5 in vitro Cytotoxicity TEST_Joy rider<br>*TPU Joystick ISO-10993-10 Maximization Sensitization TEST_Joy tider<br>*TPU Joystick ISO-10993-10 Skin Irritation TEST Joy rider |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font. To the right of the letters are a series of arrows in shades of blue and purple. The word "CITY" is in a black, italicized font.
#### ● Comparison table
## TUNG KENG ENTERPRISE CO., LTD.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
| Devices | | | Comparison<br>analysis |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Specifications | Subject device | Predicate device | |
| Manufacturer | Tung Keng Enterprise Co., Ltd. | Nanjing Jin Bai He Medical<br>Apparatus Co., Ltd. | -- |
| Proprietary name | Electrically Powered Wheelchair | Powered Wheelchair, | -- |
| Model name | Joy Rider | DYW30A(D09) | -- |
| 510(k) No. | K220156 | K170787 | -- |
| Indications for<br>use | The device is a motor-driven, and<br>indoor transportation vehicle with<br>the intended use to provide<br>mobility to a disabled or an<br>elderly person limited to a seated<br>position. | The device is a motor-driven, and<br>indoor transportation vehicle with<br>the intended use to provide<br>mobility to a disabled or an<br>elderly person limited to a seated<br>position. | Same |
| Intended user<br>population | The target population of the<br>device is Adults only. (FDA<br>defines 'an adult' as anyone over<br>the age of 22 years.) | The target population of the<br>device is Adults only. (FDA<br>defines 'an adult' as anyone over<br>the age of 22 years.) | Same |
| Motor | LINIX, 120ZWN24-200,<br>Brushless DC motor<br>200 W x 24 VDC x 2 pcs<br>Electromagnetic brake | Chongqing Mao Tian Machinery,<br>Brushless DC motor<br>250W x 24 VDC x 2 pcs | Different |
| Battery | TOP Best Energy Storage Corp.<br>Lithium-ion battery<br>10.5 Ah x 24 VDC x 1 pcs | Jiangsu Feng Chi Green Power<br>Supply Co., Ltd.,<br>Lithium-ion, ITP2406<br>6 Ah x 24 VDC x 2 pcs | Similar |
| Charger model &<br>rating | High Power Technology<br>HP0060W(L2)<br>Input: 100-240 VAC, 50/60 Hz<br>Output: 24 VDC, 2 Amp | High Power Technology<br>HP0060W(L2)<br>Input: 100-240 VAC, 50/60 Hz<br>Output: 24VDC, 2 Amp | Same |
| Number of<br>Wheels | 4 | 4 | Same |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | Same |
| Brake System | Electromagnetic brake | Smart Electromagnetic brake | Same |
| (Forwards) | | | |
| Speed control method | Joystick control method | Joystick control method | Same |
| Main frame material | 6063 aluminum alloy | 6063 aluminum alloy | Same |
| Use environments | Indoor use | Indoor use | Same |
| Front wheel size & type | 7" x 1.77" (180×45 mm)<br>(PU Solid tire) | 7" x 1.77" (180 x 45 mm)<br>(PU solid tire) | Same |
| Rear wheel size & type | 10" x 2.25" (254 x 57 mm)<br>(PU solid tire) | 12.5" x 2.25" (320 x 57 mm)<br>(PU solid tire) | Different |
| Maximum safe operational incline | 10 degrees | 8 degrees | Different |
| Dimensions unfolded (LxWxH) | 33.6" x 22.6" x 35.8"<br>(855 x 580 x 910 mm) | 37.4" x 22.6" x 36.2"<br>(960 x 580 x 930 mm) | Different |
| Dimensions folded (LxWxH) | 43.8" x 22.8" x 12.5"<br>(1115 x 580 x 320 mm) | 22.6"* 12.6"* 30.4"<br>(580 * 325 *780 mm ) | Different |
| Weight, without battery | 49.8 lbs. / 22.6 kg | 51.8" lbs. / 23.9kg | Similar |
| Cruising Range | 9.3 miles (15 km) | 11.2 miles (18 km) | Different |
| Max Loading | 264 lbs. (120 kg) | 264 lbs.(120 kg) | Same |
| Max Speed backward | 2.4 mph (3.84 km/h) | 1.86 mph (3.0 km/h) | Different |
| Turning Radius | 33.5" (850 mm) | 32.5" ( 833 mm) | Similar |
| Maximum obstacle climbing | 1.96" (50 mm) | 1.36" (34.5 mm) | Different |
| Controller | PG nVR2 BLDC, Brushless<br>double drive joystick controller | Changzhou Billon Electronic<br>Appliance Co., Ltd., WS-1 | Different |
| Minimum braking distance | 1 m | 1 m | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font. To the right of the letters are three blue arrows pointing to the right, with each arrow a slightly lighter shade of blue. The word "CITY" is in a black, italicized font to the right of the arrows.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font. To the right of the letters are a series of progressively lighter blue chevrons. The word "City" is in a black, italicized font to the right of the chevrons.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
#### Substantial Equivalence discussion ●
We will not discuss the same or similar items in the substantial equivalent discussion. We will discuss the different items: motor used, Rear wheel size & type, Maximum safe operational incline, Dimensions unfolded (L x W x H), Dimensions folded (L x W x H), Travel range, Max speed backward, Maximum obstacle climbing, and Controller.
The predicate device uses 2 pieces of Chongging Mao Tian Machinery, Brushless DC motor (250W x 24 VDC) to drive the predicate device. The subject device uses 2 pieces of LINIX, 120ZWN24-200, Brushless DC motor (200 W x 24 VDC) to drive the Joy Rider. The differences between the motors' specifications lead to the different performances of both devices, i.e., cruising range, max loading weight. There are no any new safety and effectiveness concerns raised for the subject device due to the different motors used.
Rear wheel size & type are different for the two devices, which are 10" x 2.25" (254 x 57 mm) and pneumatic tire for subject device with 30 psi and 12.5" x 2.25" (320 x 57 mm) for predicate device (PU solid tire). As seen from the data. the differences for the rear wheel size and type are small. These differences will not raise any new safety and effectiveness concerns for the subject device.
Maximum safe operational inclines for the devices are 10 degrees for the subject device and 8 degrees for the predicate device. According to the data, the subject device can move on the more inclined slope than the predicate device. There are no any new safety and effectiveness concerns raised by the different operational inclined slope for the subject device.
Unfolded dimensions for the subject device are 33.6" x 22.6" x 35.8" (855 x 580 x 910 mm) and are 37.4" x 22.6" x 36.2" (960 x 580 x 930 mm) for the predicate device, and the differences between two devices are so small and are validated by the compliance testing of RESNA WC-1 & WC-2 on the subject device. It is demonstrated there are no any new safety and effectiveness concerns raised by the unfolded dimensional differences for the subject device.
Folded dimensions (L x W x H) are 43.8" x 22.8" x 12.5" (1115 x 580 x 320 mm) for the subject device and 22.6" 12.6" 30.4" (580 * 325 *780 mm ) for the predicate device. The differences are not small and the subject device has larger folded dimensions, and the smaller dimensions can bring about more convenient wheelchair
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for DK City. The logo has the letters DK in large, bold, dark blue font. To the right of the letters DK are a series of arrows in shades of blue. To the right of the arrows is the word City in a cursive black font.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
transfer than the subject device. The differences between two devices are validated by the compliance testing of RESNA WC-1 & WC-2 on the subject device. It is demonstrated there are no any new safety and effectiveness concerns raised by the unfolded dimensional differences for the subject device.
Travel range per full charge for the subject device is 9.3 miles (15 km) and 11.2 miles (18 km) for the predicate device. The difference of traveling range is due to the use of different batteries, different weights of user and other driving situations on the two devices. There are no any new safety and effectiveness concerns raised by this difference of travel range for the subject device.
Max Speed Backward is 2.4 mph (3.84 km/h) for subject device and 1.86 mph (3.0 km/h) for the predicate device. The difference of the max speed backward between two devices is small and there are no any new safety and effectiveness concerns raised by the difference of the max speed backward for the subject device.
Maximum obstacle climbing for two devices is 1.96" (50 mm) for subject device and 1.36" (34.5 mm) for predicate device. The data were measured per the ANSI RESNA WC-2:2019 Section 10 and are shown in the use manual to mitigate the risk due to the different obstacle climbing. Subject device has larger obstacle climbing and it is safer than the predicate device when climbing the obstacle. There are no any new safety and effectiveness concerns raised by the small difference of the maximum obstacle climbing for the subject device.
Electronic controller installed in the subject device is PG VR2 BLDC, Brushless double drive joystick controller and it is Changzhou Billon Electronic, WS-1 for the predicate device. The controlling software was installed on the electronic controller, which is referred to the MAF-2116 for Agency review. Since the control systems passed the requirements of ISO 7176-14 and RESNA WC-2:2008 section 14, their performances are surely validated. The differences of the electronic controller used by the predicate and subject devices will not raise any new safety and effectiveness concerns for the subject device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DK City. The letters "DK" are in a bold, dark blue font, followed by a series of blue chevrons that get progressively lighter in color. The word "City" is written in a black, italicized font to the right of the chevrons.
No. 1, Lane 160, Sec. 2, Tan-Fu Road, Tan-Tzu Dist., Taichung City, Taiwan Tel: +886-4-25321000 Fax: +886-4-25325000 Website: www.dkcity.com
#### CONCLUSIONS .
The subject device, Joy Rider, are as safe and effective as, and function in a manner equivalent to the predicate device, Nanjing Jin Bai He Medical Apparatus Co., Ltd., Powered Wheelchair, DYW30A(D09) K170787. The conclusions drawn from the non-clinical tests demonstrate that the subject device performs as well as the legally marketed device identified in the submission. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.