← Product Code [KPR](/productcode/KPR) · K220149

# General Radiography X-ray System (Model: GR10X 40K), General Radiography X-ray System (Model: GR10X 50K) (K220149)

_Xera Medical Systems & Technology , Ltd. · KPR · Apr 19, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220149

## Device Facts

- **Applicant:** Xera Medical Systems & Technology , Ltd.
- **Product Code:** [KPR](/productcode/KPR.md)
- **Decision Date:** Apr 19, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The GR10X Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This system is not intended for mammographic applications.

## Device Story

GR10X Digital X-ray Imaging System generates radiographic images of human anatomy; used in clinical settings by healthcare professionals. System components include X-ray control, high-voltage generator, X-ray tube support, irradiation equipment, digital detectors, patient stand, and VXvue viewer software. Operation: X-ray photons pass through patient body; intercepted by detector scintillator; visible photons illuminate amorphous silicon (a-Si) array; electrical signals generated; converted to digital values. VXvue software acquires and processes data for display on monitors. Supports skeletal, respiratory, and urinary system diagnosis. Benefits include diagnostic imaging capability equivalent to predicate systems.

## Clinical Evidence

Single-blinded concurrence study conducted. Results confirmed the GR10X Digital X-ray Imaging System provides images of equivalent diagnostic capability to the predicate Ysio system, demonstrating substantial equivalence.

## Technological Characteristics

Stationary X-ray system; high-frequency inverter generator; amorphous silicon (a-Si) detectors with CsI or Gadox scintillators; 16-bit acquisition depth. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and 21 CFR 1020.30/31. Software: VXvue for image acquisition/processing.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Ysio ([K081722](/device/K081722.md))

## Reference Devices

- VIVIX-S VW ([K200418](/device/K200418.md))
- Trixell Pixium 4343RCE

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 19, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol representing the Department of Health & Human Services-USA on the left, and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue as well.

XERA Medical Systems & Technology Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 IRVINE CA 92612

Re: K220149

Trade/Device Name: GR10X-40K GR10X-50K Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR Dated: January 17, 2022 Received: January 19, 2022

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220149

Device Name GR10X-40K GR10X-50K

#### Indications for Use (Describe)

The GR10X Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This system is not intended for mammographic applications.

| Type of Use (Select one or both, as applicable)                                   |                                                                                  |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span>     ☑ Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

[As required by 21 CFR 807.92]

This 510(k) summary information is prepared in accordance with 21 CFR807.92.

#### 1. Date Prepared [21 CFR 807.92(a) (1)]

4/12/2022

#### 2. Submitter's Information [21 CFR 807.92(a) (1)]

| Name of Sponsor:      | Xera Medical Systems and Technology Ltd.    |
|-----------------------|---------------------------------------------|
| Address:              | Gulden Sk. 13/2 Kavaklidere, Ankara, Turkey |
| Contact Name:         | Oh, Kevin / Regulatory Affairs Associate    |
| Registration Number:  | 3006013411                                  |
| Name of Manufacturer: | Same as Sponsor                             |

#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

| Trade Name:                | OCTAVE                           |
|----------------------------|----------------------------------|
| Model Name:                | GR10X-40K, 50K                   |
| Common Name:               | General Radiography X-ray System |
| Classification Name:       | System, X-Ray, Stationary        |
| Classification Panel:      | Radiology                        |
| Classification Regulation: | 21 CFR 892.1680                  |
| Product Code:              | KPR                              |
| Device Class:              | 2                                |

#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

510(k) Number: K081722 Product Code: KPR Applicant: SIEMENS MEDICAL SOLUTIONS USA, INC. Trade Name: Ysio Decision Date: 03/15/2018 Type: Traditional

#### 5. Description of the Device [21 CFR 807.92(a) (4)]

This device consists of X-ray control, high-voltage generator, X-ray tube support, irradiation equipment, detectors, patient photography stand, digital imaging device, viewer SW, etc. The detectors and the viewer SW are cleared under the following 510ks.

{4}------------------------------------------------

- VIVIX-S VW(K200418) . Model Names: FXRD-2530VW, FXRD-2530VW PLUS, FXRD-3643VW, FXRD-3643VW PLUS, FXRD-4343VW, FXRD-4343VW PLUS, with imaging areas of 25cm x 30cm, 36cm x 43cm, 43cm x 43cm, respectively.
A high frequency inverter (Inverter) X-sensor voltage device designed to generate X-rays by combination of tube voltage, tube current, irradiation time, etc. so that it can be filmed at various angles for diagnosis of the patient's skeletal, respiratory, and urinary systems. The digital imaging system is used to obtain the images taken by the X-ray unit as radiation from the X-ray unit passes through the human body and is transmitted by the detector intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the Software, VXvue, acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors.

# 6. Indications for Use [21 CFR 807.92(a)(5)]

The GR10X Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This system is not intended for mammographic applications.

# 7. Technological Characteristics [21 CFR 807.92(a) (6)]

Comparisons with the predicate, the system shows the technological characteristics of the proposed GR10X-40K, 50K device to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.

# 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

When compared to the predicate devices (K081722), the GR10X-40K has the subsequently equivalent:

- Intended Use
- Technological characteristics
- · Operating principle
- · Design features
- Communication Method

There is no significant difference between the GR10X Digital X-ray Imaging System and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.

{5}------------------------------------------------

|                     | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                        |           |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Feature             | Ysio                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | GR10X-40K                                                                                                                                                                                                                                                                             | GR10X-50K |
| Indications for Use | The Ysio (New RAD -FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen ,and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio system is not meant for mammography. The Ysio uses an integrated or portable digital detector for generating diagnostic images by converting x-rays into electronic signals. Ysio is also designed to be used with conventional film/screen or Computed Radiography(CR) cassettes. | The GR10X Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This system is not intended for mammographic applications. |           |

Table 1: Comparison of Subject Device GR10X-40K, 50K to the Predicate

# Table 2: Comparison of Subject Device GR10X-40K, 50K to the Predicate

|                      | Predicate Device                                     | Subject Device                                       |             |                             |
|----------------------|------------------------------------------------------|------------------------------------------------------|-------------|-----------------------------|
| Feature              | Ysio                                                 | GR10X-40K                                            | GR10X-50K   | Comparison<br>Results       |
| Technical Comparison |                                                      |                                                      |             |                             |
| X-ray                |                                                      |                                                      |             |                             |
| Generator            | Polydoros<br>65/80 kW                                | DMG-40KAX A                                          | DMG-50KAX B |                             |
| X-ray tube           | OPTITOP<br>150/40/80/HC-100                          | E7239X                                               | E7252X      |                             |
| Collimator           | Collimator N                                         | M38                                                  |             | Substantially<br>Equivalent |
| X-ray techniques     | Radiography                                          | Radiography                                          |             |                             |
| Organ programs       | X-ray parameters<br>Imaging processing<br>parameters | X-ray parameters<br>Imaging processing<br>parameters |             |                             |
| Testing              |                                                      |                                                      |             |                             |

{6}------------------------------------------------

| IEC Compliance       | IEC 4th edition for<br>EMC testing                                                     | IEC 4th edition for<br>EMC testing                                     | Testing according to<br>current IEC test<br>scope |
|----------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------|
| Digital Imaging      |                                                                                        |                                                                        |                                                   |
| SSXI for Rad imaging | Trixell Pixium<br>MAX wi-D =<br>3543EZh<br>MAX mini = 2430EZ<br>MAX Static =<br>4343RC | VIVIX-S 4343VW,<br>VIVIX-S 3643VW,<br>VIVIX-S 2530VW,<br>VIVIX-S 1717V | Substantially<br>Equivalent                       |

# Table 3: Comparison of VIVIX-S VW Series and VIVIX-S 1717V to the predicate Trixell Pixium 4343RCE to the predicate Trixell Pixium 4343RCE

| Technical<br>Specifications | Trixell Pixium 4343RCE<br>detector | VIVIX-S VW Series, VIVIX-<br>S 1717V<br>(Subject)                                                                                                   | Comparison<br>Results    |
|-----------------------------|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Dimensions                  | 423.3mm x 425.4 mm                 | VIVIX-S 4343VW: 460 mm x 460 mm<br>VIVIX-S 3643VW: 384mm × 460mm<br>VIVIX-S 2530VW: 287mm x 350mm<br>VIVIX-S 1717V: 460mm x 460mm                   | Substantially Equivalent |
| Resolution                  | 2860 x 2874 pixels                 | VIVIX-S 4343VW: 3072 x 3072 pixels<br>VIVIX-S 3643VW: 2560 x 3072 pixels<br>VIVIX-S 2530VW: 2048 x 2560 pixels<br>VIVIX-S 1717V: 3072 x 3072 pixels | Substantially Equivalent |
| Pixel size                  | 148 µm                             | VIVIX-S 4343VW: 140µm<br>VIVIX-S 3643VW: 140µm<br>VIVIX-S 2530VW: 124µm<br>VIVIX-S 1717V: 140µm                                                     | Substantially Equivalent |
| Semiconductor<br>Material   | Amorphous silicon, a-Si            | Amorphous silicon, a-Si                                                                                                                             | Substantially Equivalent |
| Scintillator                | Cesium iodide (CsI)                | Cesium iodide (CsI),<br>Gadolinium Oxide(Gadox)                                                                                                     | Substantially Equivalent |
| Acquisition<br>depth        | 16 bit                             | 16 bit                                                                                                                                              | Substantially Equivalent |

{7}------------------------------------------------

| DQE (Detective Quantum Efficiency)  | DQE @ 0.05 lp/mm (2 µGy), 67% | FXRD-4343VAW:                                                                                                                                                                                        | Substantially Equivalent |
|-------------------------------------|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
|                                     |                               | 47%<br>FXRD-4343VAW PLUS:<br>62%<br>FXRD-3643VAW:<br>45.5%<br>FXRD-3643VAW PLUS:<br>61%<br>FXRD-2530VAW:<br>49%<br>FXRD-2530VAW PLUS:<br>61%<br>FXRD-1717NAW:<br>50%<br>FXRD-1717NBW:<br>27%         |                          |
| MTF (Modulations transfer function) | MTF @ 1 lp/mm, 62%            | FXRD-4343VAW:<br>76%<br>FXRD-4343VAW PLUS:<br>60%<br>FXRD-3643VAW:<br>74%<br>FXRD-3643VAW PLUS:<br>61%<br>FXRD-2530VAW:<br>76%<br>FXRD-2530VAW PLUS:<br>60%<br>FXRD-1717NAW:<br>66%<br>FXRD-1717NBW: | Substantially Equivalent |

# 9. Summary of Non-Clinical Data

A comparison test was conducted between the subject devices and the predicate device (K081722) on the items such as DQE, MTF and spatial resolution.

These detectors comply with the following international and FDA-recognized consensus standards:

- 21CFR1020.30, Diagnostic X-ray Systems, and their major components
- 21CFR1020.31, Radiographic equipment
- IEC 60601-1 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and Essential Performance.
- CAN/CSA-C22.2 No. 60601-1 (2008) (Medical Equipment –Part 1 : General Requirements for Basic Safety and Essential Performance) (includes National Differences for Canada)

{8}------------------------------------------------

- ANSI/AAMI ES60601-1 (2005+ C1:09+A2:10) (Medical Electrical Equipment Part 1
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2 : General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests

We also referenced the following FDA guidance:

- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software contained in Medical Devices, Document issued on: May 11, 2005 Medical Devices, Document issued on: May 11, 2005
- Guidance for Industry and Food and Drug Administration Staff Pediatric Information for X-ray Imaging Device Premarket Notifications, November 2017

#### 10. Summary of Clinical Data

A single-blinded concurrence study was conducted, and the study confirmed that the GR10X Digital X-ray Imaging System provides images of equivalent diagnostic capability to the predicate devices, the Yiso and its results demonstrate substantial equivalence.

# 11. Conclusion [21 CFR 807.92(b) (3)]

The GR10X Digital X-ray Imaging System is substantially equivalent to the currently marketed and predicate devices (K081722) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, 3 and the clinical test. The results of these tests demonstrate that GR10X Digital X-ray Imaging System meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K220149](https://fda-staging.innolitics.com/device/K220149)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
