Breast BI Coil Set 0.55T; Breast BI 7 MR Coil Set 1.5T; Breast BI 7 MR Coil Set 3T
Applicant
Noras Mri Products GmbH
Product Code
MOS · Radiology
Decision Date
Apr 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the Breast Coil is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and placement of localization-wire by a trained physician. For use with Siemens 0.55T, 1.5T, and 3T systems
Device Story
Specialty MRI coil sets for female breast imaging; used in conjunction with Siemens 0.55T, 1.5T, or 3T MRI scanners. Device consists of rigid coil frame, immobilization system, and biopsy unit. Functions as pure receiving coil; detects high-frequency signals from hydrogen nuclei induced by MRI scanner's transmitter. Signals digitized and processed by MRI system to produce transverse, sagittal, coronal, and oblique images. Used in clinical settings by trained physicians to assist in diagnosis and guide breast biopsies or wire localizations. Benefits include high-resolution imaging and precise MR-guided interventional procedures.
Clinical Evidence
No clinical testing performed. Bench testing included SNR (NEMA MS 1-2008), uniformity (NEMA MS 6-2008), surface coil heating, and usability testing (IEC 62366-1/2, IEC 60601-1-6). Biocompatibility testing per ISO 10993. Sample test images demonstrated good coverage, quality, uniformity, and SNR.
Technological Characteristics
7-channel phased array coil configuration. Materials: fiberglass-reinforced polycarbonate frame, PE breast cushion/headrest, Alexit biocompatible coating. Sensing: pure receiving coil for hydrogen nuclei signals. Connectivity: interfaces with Siemens MRI systems. Standards: IEC 60601-1, IEC 60601-1-2, NEMA MS-1, NEMA MS-6, ISO 10993.
Indications for Use
Indicated for female patients undergoing MR examination of the breast for diagnostic imaging and MR-guided biopsy or localization-wire placement. For use with Siemens 0.55T, 1.5T, and 3T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 8, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized image of an eagle. The FDA acronym is in a blue square, and the full name, "U.S. Food & Drug Administration," is in blue text.
NORAS MRI Products GmbH % Daniel Kamm Principal Engineer Kamm & Assosciates 8870 Ravello Ct NAPLES FL 34114
## Re: K220144
Trade/Device Name: Breast BI Coil Set 0.55T; Breast BI 7 MR Coil Set 1.5T; Breast BI 7 MR Coil Set 3T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: February 17, 2022 Received: February 18, 2022
## Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K220144
Device Name
Breast BI Coil Set 0.55T; Breast BI 7 MR Coil Set 1.5T; Breast BI 7 MR Coil Set 3T
Indications for Use (Describe)
The intended use of the Breast Coil is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and placement of localization-wire by a trained physician. For use with Siemens 0.55T, 1.5T, and 3T systems
Type of Use (Select one or both, as applicable)
| <span style="font-size: 1em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size: 1em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K220144
Image /page/3/Picture/1 description: The image is a logo for NORAS MRI products. The logo is in blue and white, with the word "NORAS" in large, bold letters. Underneath the word "NORAS" is a horizontal line that extends to the right and forms a heartbeat-like line. To the right of the heartbeat line are the words "MRI products" in smaller letters.
NORAS MRI Products GmbH Leibnizstrasse 4 Hoechberg Germany 97204 Telephone: +49 931/2 99 27-0 Telefax: +49 931/2 99 27-20 E-Mail: mri@noras.de www.noras.de Date Prepared: April 8, 2022
- A) SUBMITTED BY: NORAS MRI Products GmbH, Inc
- B) CONTACT: Manuel Noras, Chief Executive Officer
- C) Device Identification
Trade/Device Names: Breast Bl Coil Set 0.55T; Breast Bl 7 MR Coil Set 1.5T; Breast Bl 7 MR Coil Set 3T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: ll Product Code: MOS
| D) Predicate Device: | K162651 |
|----------------------|--------------------------------------|
| Manufacturer: | NORAS MRI Products GmbH |
| Trade/Device Name: | Breast BI 7 MR Coil Mammavention 3T |
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Name: | Magnetic resonance diagnostic device |
| Regulatory Class: | II |
| Product Code: | MOS |
Additional Predicate Device: K180123 Manufacturer: NORAS MRI Products GmbH Trade/Device Name: Breast Bl 7 MR Coil 1.5T Mammavention Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: ll Product Code: MOS
{4}------------------------------------------------
D) DEVICE DESCRIPTION: These three models are known as specialty MRI coils for use in generating MRI images of the female breast. MRI procedures that include the evaluated device are conducted for clinical purposes at the discretion of the ordering physician.
The Breast BI Coil Sets described in this document has been designed, depending upon model type, for use with an MRI system with field strength of 0.55T 1.5T or 3.0T.
The coil system consists of pure receiving coils for the reception of high frequency signals from the hydrogen-('H)-nuclei. The hydrogen nuclei are induced into precession by the transmitting coil of the MRI device.
The precise magnetization induces potential differences in the Breast Bl Coil Sets which are digitized and further processed in the MRI system.
The Breast BI Coil Sets consist of a rigid Coil Frame with Immobilization, Biopsy and Breast Coil System. Imaging is performed with a 7-Channel "phased array" Coil developed and manufactured by the NORAS company. The coils are mounted in the rigid Coil Frame. Interconnection is handled by the software of the MRI.
The Breast BI Coil Set 0.55T is an innovative concept designed to offer flexible use to meet the needs of the procedure to be performed in your clinic. It is removable and enables the biopsy device to be used with the respective corresponding coil with the different field strengths. The rigid Coil Frame forms a high resolution, 7-channel "phased array" configuration.
The Coil Frame is made of fiberglass-reinforced polycarbonate. The surface has been finished with Alexit biocompatible coating. The breast cushion and the headrest are made of PE
E) INDICATIONS FOR USE The intended use of the Breast Coil is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and placement of localization-wire by a trained physician. For use with Siemens 0.55T, 1.5T, and 3T systems.
F) SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION: Except for the new MR field strength of 0.55 T the devices are functionally identical. All models conform to NEMA Standards for the measurement of performance and safety parameters and the IEC standards for safety issues with the Magnetic Resonance Imaging Devices, IEC 60601-2-33:2002. All device testing have been completed successfully. The indications for use statements are functionally identical. Please refer to the detailed comparison table presented below. The added models do not raise any new questions about safety or effectiveness.
{5}------------------------------------------------
| | K162651: Breast BI 7 MR Coil Mammavention 3T<br>K180123: Breast BI 7 MR Coil 1.5T Mammavention | Breast BI Coil Set 0.55T;<br>Breast BI 7 MR Coil Set 1.5T;<br>Breast BI 7 MR Coil Set 3T | Comparison |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indication for Use | K162651: The intended use of Breast BI 7 MR Coil 3T<br>Mammavention is, in conjunction with a Magnetic<br>Resonance Scanner, the MR examination of the female<br>breast. It is indicated for use as a diagnostic imaging device<br>to produce transverse, sagittal, coronal ad oblique images<br>of the internal structures of the female breast. When<br>interpreted by a trained physician, these images provide<br>information that can be useful in determining diagnosis.<br>The included Breast Biopsy Unit permits MR guided breast<br>biopsy and wire localization of lesions can be performed by<br>a trained physician. The coil system Breast BI 7 MR Coil<br>Mammavention 3T can be used with the following MRI<br>systems: 3T: Siemens 3T: Skyra, Skyra fit, Prisma, Prisma<br>fit, Spectra<br>K180123: The intended use of Breast BI 7 MR Coil 1.5T<br>Mammavention is, in conjunction with a Magnetic<br>Resonance Scanner, the MR examination of the female<br>breast. It is indicated for use as a diagnostic imaging device<br>to produce transverse, sagittal, coronal and oblique images<br>of the internal structures of the female breast. When<br>interpreted by a trained physician, these images provide<br>information that can be useful in determining diagnosis.<br>The included Breast Biopsy Unit permits MR<br>guided breast biopsy and wire localization of lesions can<br>be performed by a trained physician.<br>The coil system Breast BI 7 MR Coil Mammavention 1.5T<br>can be used with the following MRI systems: 1.5T: Siemens<br>1.5T: Avanto fit, Aera, Amira | The intended use of the Breast Coil is, in conjunction<br>with a Magnetic Resonance Scanner, the MR<br>examination of the female breast. It is indicated for<br>use as a diagnostic imaging device to produce<br>transverse, sagittal, coronal and oblique images of the<br>internal structures of the female breast. When<br>interpreted by a trained physician, these images<br>provide information that can be useful in determining<br>diagnosis. The included Breast Biopsy Unit permits<br>MR guided breast biopsy and placement of<br>localization-wire by a trained physician. For use with<br>Siemens 0.55T, 1.5T, and 3T systems | Functionally the<br>same. An 0.55 T<br>version has been<br>added. |
| | K162651: Breast BI 7 MR Coil Mammavention 3T<br>K180123: Breast BI 7 MR Coil 1.5T Mammavention | Breast BI Coil Set 0.55T;<br>Breast BI 7 MR Coil Set 1.5T;<br>Breast BI 7 MR Coil Set 3T | Comparison |
| Appearance | Image: Breast BI 7 MR Coil Mammavention 3T and Breast BI 7 MR Coil 1.5T Mammavention | Image: Breast BI Coil Set 0.55T, Breast BI 7 MR Coil Set 1.5T, and Breast BI 7 MR Coil Set 3T | SIMILAR |
| MR Compatible<br>Systems | 1.5 T or 3 T | 0.55 T, 1.5 T, 3 T depending on model chosen. | Wider<br>compatibility |
| Standards | IEC 60601-1:2005, IEC 60601-1:2005 /AMD1:2012, Medical<br>Electrical Equipment<br>IEC 60601-1-2:2014/ Electromagnetic disturbances -<br>Requirements and tests<br>NEMA MS-1-2008 (R2020) Determination of Signal-to-Noise Ratio<br>(SNR) in Diagnostic Magnetic Resonance Imaging<br>ISO 10993-5:2009 Biological evaluation of medical devices<br>Cytotoxicity<br>ISO 10993-10:2009 Biological evaluation of medical devices<br>Irritation and Sensitization<br>60601-1-6 Edition 3.1 2013-10 Collateral standard: Usability | Current versions used | SAME |
| Function | Employs a coil system which serves solely as a receiving coil<br>for the reception of high frequency signals from the<br>hydrogen -(1-H) nuclei. The hydrogen nuclei are induced<br>into precession by the transmitting coil of the MRT device. | Identical mode of operation | SAME |
| Biocompatibility | Patient contact parts tested to<br>ISO 10993 | SAME | SAME |
{6}------------------------------------------------
{7}------------------------------------------------
## G) SUMMARY OF PERFORMANCE TESTING.
Patient contact materials were subjected to ISO 10993 tests for: Cytoxicity, Irritation, and sensitization. The devices were subjected to IEC60601-1 Safety Testing and IEC60601-1-2 ESD Testing.
Testing was performed to in accordance with NEMA MS 1-2008 (R2014, R2020), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Resonance Imaging in order to assure good S/N performance. Additionally, uniformity (NEMA MS 6-2008) and surface coil heating testing was successfully performed.
Usability testing was performed in accordance with: IEC 62366-1.2015 Medical Devices - Part 1: Application of Usability Engineering To Medical Devices IEC/TR 62366-2:2016 Medical Devices - Part 2: Guidance On The Application Of Usability Engineering To Medical Devices, and IEC 60601-1-6:2016 Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
Clinical sample test images were obtained for the new model listed above. The images exhibited good coverage, quality, uniformity and SNR.
H) CLINICAL TESTING: Not required.
## I) CONCLUSION:
In all material aspects, the new models of breast MR coils performed equivalently to the predicate device, had similar construction methods, and have not raised any new issues of safety or effectiveness. They have an indications for use statement that is virtually identical to the predicate, thus rendering the new models of MRI coils substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.