K220124 · Pollogen, Ltd. · GEI · Feb 17, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220124
Device Name
YandR System, VoluDerm Handpiece, VoluDerm Tips
Applicant
Pollogen, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 17, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The YandR System is intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Handpiece.
Device Story
YandR System is a computer-controlled RF electrosurgical device for skin ablation and resurfacing. It consists of a tabletop console, a VoluDerm handpiece, and single-use sterile tips. The device generates a 1MHz sinusoidal RF signal delivered via bi-polar electrode pins on the tips (up to 62 Joules/pin). Operated by clinicians in a professional setting, the system features an integrated handpiece trigger for activation, replacing traditional footswitches. The device is a modernized, compact version of the Legend Pro RF family. By delivering RF energy to the skin, it facilitates controlled ablation and resurfacing, which can improve skin texture and appearance. The clinician monitors the procedure via the system interface; output is controlled by the user to achieve desired dermatological outcomes.
Clinical Evidence
Bench testing only. Evidence includes risk analysis per ISO 14971, electrical/RF/EMC safety testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2, and software verification/validation per IEC 62304. System testing included energy measurements, safety controls, and emission indicators. No clinical data provided.
Technological Characteristics
Electrosurgical RF device; 1MHz sinusoidal signal; bi-polar electrode pins; tabletop console form factor; hand-switch activation. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Software validated per IEC 62304.
Indications for Use
Indicated for dermatological procedures requiring ablation and resurfacing of the skin. Patient population not specified.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 17, 2022
Pollogen Ltd. % Kathy Maynor Regulatory Consultant Kathy Maynor 26 Rebecca Ct Homosassa, Florida 34446
Re: K220124
Trade/Device Name: YandR System, VoluDerm Handpiece and VoluDerm Tips Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 17, 2022 Received: January 18, 2022
Dear Kathy Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220124
#### Device Name
YandR System and VoluDerm Handpiece and VoluDerm Tips
Indications for Use (Describe)
The YandR System is intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Handpiece.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logos for Pollogen and Lumenis. The Pollogen logo is on the left and consists of the word "Pollogen" inside of a series of overlapping circles. To the right of the Pollogen logo is the text "a company of" above the Lumenis logo. The Lumenis logo is a blue circle with a white "L" inside of it, with the word "Lumenis" to the right of the circle and the tagline "Energy to Beauty" below it.
# 510(k) SUMMARY
The YandR System (member of the Legend Pro Family of Radio Frequency (RF) Systems)
| Applicant Name:<br>(21 CFR 807.92(a)(1) | Pollogen Ltd.<br>6 Kaufman Street<br>Tel Aviv, Tel-Aviv, ISRAEL 6801298<br>Phone: +972 3 5104110<br>Fax: +972 3 5104112 | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:<br>(21 CFR 807.92(a)(1) | Kathy Maynor<br>Regulatory Consultant<br>26 Rebecca Ct<br>Homosassa Florida<br>34446<br>Phone: 352-586-3113 | and<br>Amaya De Levie<br>Lumenis Be Ltd.<br>6 Hakidma Street PO Box 240,<br>Yokneam Industrial Park<br>Yokneam 2069204<br>Israel<br>Phone: +972-549599073<br>E-mail: Amaya.de-levie@lumenis.com |
| Date Prepared: | 17 January 2022 | |
| Trade Name:<br>(21 CFR 807.92(a)(2) | YandR System and VoluDerm Handpiece and VoluDerm Tips | |
| Classification Name: | Powered laser surgical instrument | |
| Product Code: | GEI | |
| Device Class: | Class II | |
| Regulation Number: | 21 CFR Part 878.4400 | |
| Panel: | General & Plastic Surgery | |
| Predicate Devices:<br>(21 CFR 807.92(a)(3) | Pollogen Legend+ system (K173503) | |
## Intended Use/ Indications for Use (21 CFR 807.92(a)):
The YandR System is intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Handpiece.
> The YandR System 510K Summary 510(k) Summary Page 1 of 3
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image contains two logos. The first logo is for "Pollogen" and it is encased in a circular design. The second logo is for "Lumenis" and it includes the tagline "Energy to Beauty".
# Device Description (21 CFR 807.92(a)(4):
The YandR System, subject of this submission as well as the cleared systems of the Legend Pro RF Family of Systems (Pollogen Legend+ system (listed as Legend Pro); see K122200, K131758, K171359 and K173503) are computer-controlled RF devices connected to a treatment Handpiece/Applicators and disposable tips. The system is comprised of:
- System Console
- . VoluDerm Handpiece/Applicator and single-use sterile removable tips (present only in the Surgen, Legend Pro and YandR Systems)
The YandR System with its VoluDerm Handpiece is a new model of the modified Legend Pro Family of RF Systems The YandR System as the cleared Pollogen Legend+ system can generate a 1MHz sinusoidal signal applied by the VoluDerm Handpiece and the same VoluDerm Tips with bi-polar electrode pins at up to 62 Joules/per pin. The proposed system relies on the same fundamental underlying technology of the cleared systems with some hardware and software modifications to meet the marketing requirements for a more compact and simple device (tabletop console instead of a floor-mounted, rolling console) with a modern interface and the support of only the VoluDerm Handpiece (removing the TriPollar Applicators), similarly to its predecessor, the Pollogen Ltd.'s Surgen U (K131758). Also, to simplify the system, the YandR system has a hand switch, integrated into the VoluDerm Handpiece (called the Handpiece Trigger), instead of a separated Footswitch.
The new member of the of the cleared Legend Pro Family of Radio Frequency (RF) systems (Apollo, Surgen and Pollogen Legend+ systems) is a class II electrosurgical cutting and coagulation device (21 C.F.R. §878.4400; Product Code GEI). The subject Y&R System with its VoluDerm Handpiece and VoluDerm Tips can be regarded as a modification of the Pollogen Legend+ System, a member of the Legend Pro Family of RF systems and accessories. As such, it shares with the cleared family members the same subset of intended use when using VoluDerm Handpiece and underlying technology. The subject modified YandR System with the VoluDerm Handpiece is claimed to be equivalent, regarding both its intended use and its technological characteristics to its cleared Pollogen Legend+ system, member of Legend Pro Family of RF Systems.
## Substantial Equivalence (21 CFR 807.92(a)(6)
The YandR System, subject of this submission, is a modification of the Pollogen Legend+ system, recently cleared under. It is regarded as a modified member to the Legend Pro Family of RF Systems (cleared under K122200, K131758, K171359 and K173503). As such it shares with the cleared family members, the same intended use and underlying technology. The modifications introduced to the subject YandR System as compared to its cleared predicate are designed and intended mainly for system modernization and increased user convenience in accordance with market/design inputs.
Design verification and validation processes were performed as a result of this risk analysis assessment to verify that no different questions of safety and effectiveness have been raised due
> The YandR System 510K Summary 510(k) Summary Page 2 of 3
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logos of two companies, Pollogen and Lumenis. The Pollogen logo is on the left and consists of the word "Pollogen" inside a series of overlapping circles. The Lumenis logo is on the right and consists of a blue circle with a white "U" inside, followed by the word "Lumenis" and the tagline "Energy to Beauty".
to the modifications to this system. The test methods are essentially the same as those used to support to the clearance of the Pollogen Legend Pro+ (see K171359 and K173503 for the latest clearances).
The following activities were performed (21 CFR 807.92(b)(1):
- Risk analysis per ISO 14971 .
- . Electrical, Radio Frequency and electromagnetic compatibility safety testing to IEC 60601-1, IEC 60601-1-2, IEC and 60601-2-2
- Software verification and validation according to IEC 62304 and FDA Guidance "Principles . of Software Validation Guidance for Industry and FDA Staff, January 2002".
- . System testing (e.g., energy measurements, safety controls, emission indicator, scanners, aiming beam).
Test results indicated that the YandR System performs in accordance with its requirements and specifications similarly to its predicate Pollogen Legend+ system member of the Legend Pro Family of RF Systems (cleared under K122200, K131758, K171359 and K173503). Consequently, Pollogen Ltd. believes that the YandR System is substantially equivalent to the cleared predicate and it does not raise any different questions of safety and/or effectiveness (21 CFR 807.92(b)(2).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.