The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.
Device Story
RespiraSense (RS) is a wireless, body-worn, non-invasive respiratory rate (RR) monitor. Input: thoracic and abdominal displacement signals measured via piezoelectric film elements. Operation: piezoelectric sensor detects chest/abdomen movement; integrated accelerometer corrects for motion artifacts; dsPIC microcontroller processes signals to derive RR. Output: RR data transmitted via Bluetooth Low Energy (BLE) to an iPad running the RS App. Usage: hospital-type facilities; operated by medical staff for continuous monitoring (up to 4 days). RS App allows configuration, alert limit setting, and optional display of SpO2/Pulse rate from connected Nonin devices. Benefits: enables reliable remote RR monitoring for mobile patients, assisting clinical decision-making regarding patient respiratory status.
Clinical Evidence
Multi-site comparative clinical study (n=56) in intended population (adults 18+). Compared RespiraSense to gold standard EtCO2 monitor. Observers blinded to EtCO2. Primary endpoint: RR accuracy. Results demonstrated substantial equivalence to gold standard with accuracy of ±3 bpm at 95% confidence level.
Technological Characteristics
Materials: single-use biocompatible adhesive (ISO 10993-5, ISO 10993-10). Sensing: piezoelectric film elements, accelerometer. Energy: rechargeable Li-Ion battery. Form factor: body-worn Lobe (57x98x18mm) + adhesive sensor. Connectivity: Bluetooth 4.2. Standards: AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ANSI IEEE C63.27, AIM 7351731.
Indications for Use
Indicated for continuous, non-invasive, real-time respiratory rate monitoring in patients 18+ years old in hospital-type facilities, including mobile patients. Not for apnea monitoring.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
October 7, 2022
PMD Solutions % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE St. Petersburg, Florida 33704
Re: K220111
Trade/Device Name: RespiraSense Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BOZ Dated: August 30, 2022 Received: August 30, 2022
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
icies. You must comply with all the Act's
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220111
Device Name RespiraSense
Indications for Use (Describe)
The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right: 5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right: 5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# PMD Solutions RespiraSense
#### Date Summary was prepared: October 6, 2022
#### I. SPONSOR
PMD Solutions Bishoptown House Model Farm Road Cork, Ireland T12 T922 Tel - +353 (0) 212428760
| Sponsor Contact: | Myles Murray - President |
|------------------|--------------------------|
|------------------|--------------------------|
| Submission Correspondent: | Paul Dryden |
|---------------------------|--------------------------|
| | ProMedic, LLC |
| | 131 Bay Point Dr NE |
| | St. Petersburg, FL 33704 |
#### II. DEVICE
| Name of the device: | RespiraSense |
|----------------------|------------------------------|
| Common Name: | Breathing Frequency Monitor |
| Classification CFR: | 21 CFR 868.2375 |
| Classification Name: | Monitor, Breathing Frequency |
| Product Code: | BZQ |
### III. PREDICATE DEVICE AND REFERENCE DEVICE
| Predicate Device: | Philips Intellivue CL Respiration Pod (K122223) |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Breathing Frequency Monitor |
| Classification CFR: | 21 CFR 868.2375 |
| Classification Code: | BZQ / DRG / MSX |
| Classification Name: | Monitor, Breathing Frequency / Transmitters and Receivers,<br>Physiological Signal, Radio Frequency / System, Network and<br>Communication, Physiological Monitors |
| Reference Device: | Masimo Rainbow SET Acoustic Monitoring sensor (K120984) |
| Common/Usual Name: | Oximeter |
| Classification CFR: | 21 CFR 870.2700 |
| Classification Name: | Oximeter / Monitor, Breathing Frequency |
| Product Code: | DQA / BZQ |
#### IV. DEVICE DESCRIPTION
The RespiraSense (RS) monitors respiratory rate by directly measuring displacements of the chest and abdomen that occur during breathing which are analogous of respiratory rate. This is measured using piezoelectric film elements that output a varying voltage when displaced.
{4}------------------------------------------------
The RS device is a non-invasive, wireless, respiratory rate (RR) monitor that is worn on the body of the patient. It is internally-powered, and enables continuous RR monitoring from a single application for up to four days. It is designed to enhance the ability of medical staff to reliably measure respiratory rate remotely.
The RS Device is comprised of the RespiraSense Lobe (hereafter the Lobe), the reusable component that houses the RS Device's electronics, a single use adhesive RespiraSense Sensor (hereafter the Sensor) and Cradle. The Lobe and Sensor connect via a secured flat cable and are mechanically fastened together using a plastic Cradle. The Lobe and Sensor are placed on the left-hand side of a patient's torso.
The Lobe transmits wirelessly via Bluetooth to an iPad running PMDs RS App for display, control and alert functionality. The RS App allows for
- Configuring the Lobe ●
- Displaying respiration rate ●
- Display SpO2 and Pulse rate from a connected Nonin device (optional)
- Configuring alert limits .
The RS application can also connect to and display information from multiple Lobes simultaneously.
The Lobe is rechargeable. It is designed to be charged using the supplied Charging Station. Charging must be done outside of the immediate patient environment. Charging cannot occur while the device is in operation. Up to six Lobes can be charged at one time.
The Sensor is the only patient-contacting device and is secured to the patient with medical grade adhesive.
## V. INDICATIONS FOR USE
The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND REFERENCE DEVICE
The table below outlines the features of the Respiratory Rate Monitor and compares it to the predicate devices to establish substantial equivalence.
{5}------------------------------------------------
| Parameter | Subject Device RespiraSense<br>(K220111) | Predicate Device<br>Philips Intellivue CL Respiration Pod (K122223) | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984) | Comparison |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Name | Breathing Frequency Monitor | Breathing Frequency Monitor | Breathing Frequency Monitor | Similar |
| Indications for<br>Use | The RespiraSense is indicated for<br>continuous, non-invasive, and real-time<br>monitoring of respiratory rate.<br>RespiraSense is indicated for patients<br>18 years and older in hospitals,<br>hospital-type facilities and while patients<br>are mobile (e.g., walking). RespiraSense<br>is not intended to be an apnoea monitor. | The IntelliVue CL Respiration Pod is indicated for<br>use by health care professionals whenever there is a<br>need for intermittent or spot-check acquisition and<br>monitoring of physiological patient parameters<br>respiration rate and pulse rate wirelessly in specific<br>hospital areas. The IntelliVue CL Respiration Pod<br>is mainly indicated for use in general medical and<br>surgery wards and in waiting areas of emergency<br>rooms. It is not indicated for use in hospital areas in<br>which continuous patient monitoring is needed,<br>such as intensive care units or operating rooms.<br>The intended use of the IntelliVue CL Respiration<br>Pod when used together with a patient monitor is<br>for intermittent or spot-check monitoring and<br>recording of, and to generate alarms for,<br>respiration rate and pulse rate of adult patients.<br>The IntelliVue CL Respiration Pod is also<br>intended for acquisition of respiration rate and<br>pulse rate data of adult patients for a clinical<br>information management system.<br>The Intellivue CL Respiration Pod is intended for<br>use by health care professionals. It is not intended<br>for home use. It is not a therapeutic device.<br>The IntelliVue CL Respiration Pod is not intended<br>for use on patients with extremely high values for<br>respiration rate (above 60 rpm).<br>The IntelliVue CL Respiration Pod is not intended<br>for use on acutely ill cardiac patients with the<br>potential to develop life threatening arrhythmias,<br>e.g. very fast atrial fibrillation or ventricular<br>tachycardia (rapid irregular pulse rate). For<br>monitoring of these patients, a device for<br>continuous ECG monitoring is necessary. The<br>IntelliVue CL Respiration Pod is not a substitute<br>for an ECG monitor. | The Rainbow® Acoustic<br>Monitoring sensors and cables are<br>indicated for continuous,<br>noninvasive monitoring of<br>respiratory rate (RRa).<br>The RAS- 125 is indicated for adult<br>patients in hospitals, hospital-type<br>facilities, mobile and home<br>environments.<br>The RAS- 125c is indicated for<br>adult and pediatric patients in<br>hospitals, hospital-type facilities,<br>mobile and home environments. | Similar to both<br>predicate and<br>reference for<br>measuring respiratory<br>rate in the intended<br>population and<br>environment. |
| Parameter | Subject Device RespiraSense<br>(K220111) | Predicate Device<br>Philips Intellivue CL Respiration Pod<br>(K122223) | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984) | Comparison |
| Measurement<br>Principle | Thoracic and abdominal movements<br>from respiratory effort. Bending of the<br>piezo sensor caused by chest and<br>abdomen movements during breathing<br>produce a voltage signal, from which<br>respiration rate is derived. Integrated<br>accelerometer provides means to<br>correct artefacts. | Thoracic movements from respiratory effort -<br>Inclination changes of the incorporated<br>accelerometer sensor, caused by chest and<br>abdomen movements during breathing and heart<br>contraction, produce a voltage signal, from which<br>respiration and pulse rate signals are derived. | Use of piezoelectric sensor element<br>for the purpose of non-invasive<br>respiratory rate monitoring. As air<br>flows, the acoustic vibrations are<br>measured by the sensors and<br>processed to determined respiratory<br>rate. | Similar principle of<br>operations, sensor<br>technology of subject<br>device is different<br>from predicate but<br>similar to reference<br>device. |
| Patient Interface | Lobe attached to sensor. Sensor<br>adhesively attached to patient's chest. | Pod attached to strap. Strap adhesively attached to<br>patient. | Sensor adhered to the neck of the<br>patient, close to the windpipe. A<br>cable connects the sensor to<br>a nearby monitor. | Similar |
| Patient<br>Population | Patients aged 18 or over | Adult | Adult and pediatric patients | Similar |
| Prescriptive | Yes | Yes | Yes | Similar |
| Environment | Hospital, hospital-type facilities, and<br>mobile | Hospital | Hospitals, hospital-type facilities,<br>mobile and home environments. | Similar |
| Hardware | Small body-worn device with adhesive<br>attachment, data is sent to the<br>RespiraSense Software Application.<br>Lobe hardware consists of a battery and<br>an electronic module with the embedded<br>processor, accelerometer, dsPIC<br>microcontroller, Bluetooth Low Energy<br>(BLE) transceiver, Sounder and LED.<br>Encased in a plastic housing. | Small body-worn pod with adhesive<br>attachment, data is sent to an assigned<br>compatible IntelliVue Patient Monitor or<br>IntelliVue Guardian Software.<br>Pod hardware consists of a battery and an<br>electronic module with the embedded processor,<br>accelerometer, SRR transceiver and speaker.<br>Encased in a plastic housing. | A small piezoelectric sensor strip<br>adhered to the neck of the patient<br>and connected via a Masimo RD<br>cables from the sensor to a<br>peripheral bedside patient monitor. | Similar |
| Software | The RespiraSense Lobe provides and<br>communicates measurement values<br>and other information (e.g., battery<br>state) wirelessly as encrypted data via<br>an encrypted BLE wireless protocol to<br>the RespiraSense Software Application<br>on a iPad. The RespiraSense<br>Application manages bidirectional<br>communication between Lobe and the<br>RespiraSense Application installed on<br>a COTS Tablet Computer | The CL Respiration Pod provides and<br>communicates measurements values and other<br>information (e.g., battery state) wirelessly via Short<br>Range Radio (SRR) to an assigned compatible<br>IntelliVue Patient Monitor or IntelliVue Guardian<br>Software.<br>The CL Respiration Pod can be controlled from<br>the assigned IntelliVue Patient Monitor or<br>IntelliVue Guardian Software run on PC or<br>server. | All software is managed by the<br>peripheral bedside patient monitor<br>to which the sensor is connected to.<br>The Bedside patient monitor<br>provides power to the sensor. | Similar |
| Parameter | Subject Device RespiraSense<br>(K220111) | Predicate Device<br>Philips Intellivue CL Respiration Pod<br>(K122223) | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984) | Comparison |
| Performance<br>Specification | Range: 6 - 60 breaths per minute<br>Resolution: 1 breath per minute<br>Accuracy: $\pm$ 3 breaths per minute | Range: 5-60 breaths per minute Resolution: 1 breath per minute Accuracy: $\pm$ 1 breath per minute | Range: 4-70 breaths per minute<br>Resolution: 1 breath per minute<br>Accuracy $\pm$ 1 breaths per minute | Difference on the<br>lower range of<br>respiration has no<br>impact on safety or<br>efficacy. For accuracy<br>there is reference<br>Biobeat K212153<br>which has BPM<br>accuracy of $\pm$ 3 bpm<br>to support<br>equivalence. |
| Wireless<br>Connectivity | Bluetooth 4.2 | Short Range Radio | N/A | Similar principle<br>of operation |
| Power Source | Rechargeable Lithium-Ion battery | Rechargeable Lithium-Ion battery | Peripheral Bedside Monitor | Similar |
| Patient<br>contacting parts | Adhesive attaching device to<br>patient's chest. | Adhesive attaching device to patient's chest. | Adhesive attaching device to the<br>patient's neck. | Similar |
| Safety and EMC<br>Standards | AAMI ANSI ES 60601-1 :2015<br>+A1:2012<br>IEC 60601-1-2:2014<br>IEC 60601-1-8:2012<br>ANSI 63.27-2017<br>AIM 735173 | AAMI ANSI ES 60601-1:2005<br>IEC 60601-1-2:2007<br>IEC 60601-1-8:2006 | IEC 60601-1<br>IEC 60601-1-2 | Subject device<br>complies with the<br>current versions of<br>the applicable<br>standards. |
| Dimensions | 57 mm x 98 mm x 18 mm (without<br>sensor) | Pod: 53.5 mm x 27 mm x 65 mm ±5% (2.1 in x<br>1.1 in x 2.6 in ±5%) (without cradle and sensor) | Sensor: 135 mm x 22mm x 3 mm<br>accompanied by a 3-3.5m cable | Similar |
| Weight | 57 g | 80 g ±10% (2.8 oz ±10%) | <20g (excluding cable) | Similar |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# VII. PERFORMANCE DATA:
Bench testing - Bench testing was performed that includes:
- Shelf-life / Aging ●
- Software Verification and Validation Guidance for the Content of Premarket Submissions . for Software Contained in Medical Devices
- AAMI ANSI ES 60601-1:2015+A1:2012 Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Collateral standard: Electromagnetic Disturbances Requirements ● and Tests
- IEC 60601-1-8 2012 Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- ANSI IEEE C63.27-2017 American National Standard for Evaluation of Wireless ● Coexistence
- AIM 7351731 Rev. 2.00 Medical Electrical Equipment and System Electromagnetic ● Immunity Test for Exposure to Radio Frequency Identification Readers
- Cybersecurity Content of Premarket Submissions for Management of Cybersecurity in Medical devices
Discussion - The subject device meets the applicable standards similar to the predicate device.
Biocompatibility - The adhesive of the sensor is applied to intact skin with a prolonged duration (up to 4 days). Testing was performed per:
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- . ISO 10993-10:2010 - Biological evaluation of medical devices - Part 10: Tests for skin sensitization
Discussion - Both devices use single-use disposable biocompatible adhesive compliant with applicable parts of the ISO 10993-1 Biological Evaluation standard to adhere to intact skin.
# Clinical Testing
We performed a multi-site comparative clinical testing in the intended population and the environment comparing the subject device to the gold standard EtCO2. A total of 56 subjects were evaluated by manual counting and / or comparison to a FDA cleared end-tidal CO2 monitor with the observer blinded to the EtCO2 monitor. The subject covered the age range of > 18 years old and split by gender. The statistical analysis showed that the subject device are substantially equivalent to the gold standard of + 3 bpm at a 95% confidence level. The results supported the claims of substantial equivalence for the intended use.
# VII. CONCLUSION:
The testing performed and comparison to the predicate demonstrate that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.