K220072 · Sinovation (Beijing) Medical Technology Co., Ltd. · HAW · Jun 18, 2023 · Neurology
Device Facts
Record ID
K220072
Device Name
Sinobot X1
Applicant
Sinovation (Beijing) Medical Technology Co., Ltd.
Product Code
HAW · Neurology
Decision Date
Jun 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Sinobot X1 is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
Device Story
Computer-aided, robotized image-guided surgical system for neurosurgery. Inputs: 3D MRI/CT images for pre-operative planning; 3D optical scanner (projector/camera) for patient facial surface registration. Operation: Surgeon performs pre-operative planning on medical images; system provides robotic guidance for instrument holders. Output: Stable, precise, reproducible positioning of surgical instruments (biopsy needle, stimulation/recording electrodes). Used in neurosurgical operating rooms by neurosurgeons. Benefits: Enhanced precision and reproducibility in stereotactic neurosurgical procedures.
Clinical Evidence
Bench testing only. Performance validation included positional accuracy (mean 0.613 mm) and angular accuracy (mean 0.439 degrees). Biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMD (IEC 60601-1-2), usability (IEC 62366-1), software V&V (IEC 62304), laser safety (IEC 60825-1), sterilization (SAL 10^-6), shelf life (10 years), and transportation (ASTM D4169) testing were successfully completed.
Technological Characteristics
Robotic stand with compact arm and touch screen. 3D optical scanner (projector/visible light camera). Class 2 laser (655nm, 560μW). Connectivity: DICOM compliant. Power: AC. Sterilization: Disposable sterile drapes for robotic arm/screen. Standards: IEC 60601 series, ISO 10993, IEC 62304, IEC 60825-1, ASTM D4169.
Indications for Use
Indicated for neurosurgical procedures requiring stereotactic guidance, including biopsy, stimulation, or recording electrode placement, performed by trained neurosurgeons.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
June 18, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sinovation (Beijing) Medical Technology Co., Ltd % Giselle Zhang Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K220072
Trade/Device Name: Sinobot X1 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: May 15, 2023 Received: May 18, 2023
Dear Giselle Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed Adam D . by Adam D. Pierce -5 Pierce -S Date: 2023.06.18 10:18:57 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220072
Device Name Sinobot X1
#### Indications for Use (Describe)
The Sinobot X1 is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <div> <table border="0" style="border-collapse: collapse;"><tbody><tr><td style="border: 1px solid #000000; padding: 2px;"><div> <span style="font-family: Symbol;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div></td></tr></tbody></table> </div> </span> | <div> <span style="font-family: Symbol;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-family: Symbol;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <span style="font-family: Arial, sans-serif;"> <div> <table border="0" style="border-collapse: collapse;"><tbody><tr><td style="border: 1px solid #000000; padding: 2px;"><div> Over-The-Counter Use (21 CFR 801 Subpart C) </div></td></tr></tbody></table> </div> </span> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
## 5.1 Submitter Information
| Company: | Sinovation (Beijing) Medical Technology Co.<br>Ltd<br>Rm 401, Bld 12-1, No.26 Yongwang West<br>Road, Z- park<br>Daxing, Beijing, Beijing 102600 P.R.China<br>Telephone:<br>Fax: N/A |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Manman Xu<br>RA Specialist<br>xumanman@sinovationmed.com<br>Giselle Zhang<br>Regulatory Consultant<br>Emergo Global Consulting, LLC<br>2500 Bee Cave Road, Building 1, Suite 300<br>Austin, Texas 78746 USA<br>Telephone: (512) 327-9997<br>Fax: (512)327-9998<br>LST.AUS.ProjectManagement@ul.com |
| Date Summary Prepared: | June 18, 2023 |
| 5.2 Name of the Device | |
| Trade/Device Name: | Sinobot X1 |
| Common Name: | Stereotaxic Instrument |
| Classification Name: | Neurology |
| Review Panel: | Neurology (NE) |
| Regulation: | 882.4560 |
| Class: | Class II |
| Product Code: | HAW |
## 5.3 Equivalence Claimed to Predicate Device
The Sinovation is equivalent to the ROSA ONE Brain application (K200511), manufactured by Medtech S.A.
## 5.4 Indications for Use Statement
The Sinobot X1 is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments
{4}------------------------------------------------
(biopsy needle, stimulation or recording electrode). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
## 5.5 Device Description
The Sinobot X1 device is a computer-aided, robotized image-guided interactive surgical system. The system integrates medical image processing technology with robotic surgery technology to assist the surgeon during brain surgeries. The system allows the surgeon to utilize the spatial positioning technology to pre-plan the instruments on medical images and provide stable, precise and reproducible guidance in accordance with the planning. The device consists of a robotic stand with a compact robotic arm and a touch screen. The robotic arm can attach to different types of instruments and be changed based on the intended surgical procedures. For different surgical procedures a biopsy needle, stimulation or recording electrode could be attached to the device.
The 3D Optical scanner is an optical device combined by a projector and a visible light camera. Which provide a high-resolution point cloud scanning of patients' facial skin surface in order to perform patient registration.
### 5.6 Substantial Equivalence Discussion
The following table compares the Sinovation to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
| Attribute | Sinovation Sinobot X1 | ROSA ONE 3.1.3.2<br>Brain Application<br>(K200511) | Comparison |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Sinovation (Beijing)<br>Medical Technology<br>Co., Ltd | Medtech S.A | N/A |
| Product Code | HAW | HAW | Same |
| Regulation<br>Number | 882.4560 | 882.4560 | Same |
| Indications for Use | The Sinobot X1 is<br>intended for the spatial<br>positioning and<br>orientation of<br>instruments holders or<br>tool guides to be used by<br>trained neurosurgeons to guide standard<br>neurosurgical<br>instruments (biopsy<br>needle, stimulation or | The device is intended for<br>the spatial positioning<br>and orientation of<br>instruments holders or<br>tool guides to be used<br>by trained<br>neurosurgeons to guide<br>standard neurosurgical<br>instruments (biopsy<br>needle, stimulation or<br>recording electrode,<br>recording electrode). The device is | Same |
{5}------------------------------------------------
| | device is indicated for<br>any neurosurgical<br>procedure in which the<br>use of stereotactic<br>neurosurgery<br>may be appropriate. | indicated for any<br>neurosurgical procedure<br>in which the use of<br>stereotactic neurosurgery<br>may be appropriate. | |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Pre-Operative Planning<br>Intra-Operative<br>Registration<br>Intra-Operative<br>Guidance | Pre & intraoperative<br>images<br>Surgical planning Patient<br>Registration Guidance of<br>instruments | Same |
| Mechanism of<br>Action | | | Only the description is<br>different. |
| Where Used | Neurosurgical operating<br>room | Neurosurgical operating<br>room | Same |
| User | Neurosurgeon | Neurosurgeon | Same |
| Anatomical Site | Head | Head | Same |
| Images Type | 3D MRI / CT | 3D MRI / CT | Same |
| DICOM<br>Compliance | Yes | Yes | Same |
| Integrated<br>Planning Software | Stereotactic Planning<br>System Software | ROSANNA BRAIN<br>(Medtech) | Different<br>All software functions<br>and features are similar. |
| Trajectory<br>Planning<br>Parameters | Parameters for planning<br>trajectories: entry point,<br>target point, length of the<br>instrument, diameter,<br>name | Parameters for planning<br>trajectories: entry point,<br>target point, length of the<br>instrument, diameter,<br>name, security radius<br>(10mm by default),<br>security aperture (10° by<br>default) | Different |
| Trajectory<br>Definition<br>(Endoscopy<br>Module) | No endoscopic module | Parameters for planning<br>trajectories: entry point,<br>target point, length of the<br>instrument, diameter,<br>name, security radius<br>(10mm by default),<br>security aperture (10° by<br>default) | Different<br>The subject device does<br>not contain endoscopic<br>module, the difference of<br>the two devices will not<br>raise concerns related to<br>the safety and<br>effectiveness. |
| Save/Load<br>Planning | Yes | Yes | Same |
| Patient<br>Registration<br>Methods | Fiducial markers<br>(bone)<br>Optical registration<br>device | Fiducial markers (skin,<br>bone)<br>Optical registration<br>device | Same |
{6}------------------------------------------------
| | | Stereotactic<br>frame<br>(fiducials mounted on the<br>frame) | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fiducial<br>Markers<br>Registration<br>with<br>Pointer Probe | Yes | Yes | Same |
| Surface Matching<br>Registration with<br>Optical Distance<br>Sensor | Yes | Yes | Same |
| | | | Different |
| Laser Class for<br>Optical<br>Registration | Class 2 laser Wavelength<br>- 655nm<br>Maximum Output –<br>560μW<br>(complies with 21 CFR<br>1040.10) | Class 2 laser Wavelength<br>- 658 nm<br>Maximum output – 1 mW<br>(complies with 21 CFR<br>1040.10) | Even though the<br>specifications are<br>different, the subject<br>device passed the<br>standards requirements<br>and is safe to use, hence<br>the difference will not<br>raise concerns in safety<br>and effectiveness. |
| Image-Guided | Yes | Yes | Same |
| Real Time Display<br>of the Instrument<br>Position | Yes | Yes | Same |
| Provide Guidance<br>for Surgical<br>Instruments | Yes | Yes | Same |
| Instrument Guide<br>Position<br>Adjustment | Automatic (robotized) | Automatic (robotized) | Same |
| Instrument<br>Fixation | Instruments are<br>mounted on robotic arm | Instruments are<br>mounted on robotic arm | Same |
| | | | Different |
| Instruments | Instrument holder,<br>holder and adaptors,<br>optical sensor | Instrument<br>endoscope<br>holder and<br>adaptors,<br>optical sensor | The instruments of both<br>devices are the same<br>except the endoscopic<br>module, which the<br>subject device does not<br>have. The difference will<br>not affect the use of the<br>device and will not affect |
{7}------------------------------------------------
| | | | and raise new risks<br>related to safety and<br>effectiveness of the<br>device. |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | device. |
| Instrument<br>Calibration<br>Method | Factory calibration | Factory calibration | Same |
| Associated<br>Equipment | Navigation Probe<br>(Pointer Probe)<br>Standard Tool Holder<br>(Guider)<br>Microdrive Holder<br>Optical Sensor (Optical<br>Distance Sensor and<br>Optical 3D scanner)<br>Fiducial markers Head<br>holder<br>Frame Registration<br>Plates Frame Adapter<br>Head Holder Adapter | Navigation probe<br>Standard tool holder<br>Endoscope holder<br>Microdrive holder<br>Optical sensor Fiducial<br>markers Head holder<br>Leksell frame registration<br>plates<br>CRW Frame<br>Head Holder Adapter | Different<br>Both the subject device<br>and predicate device<br>shares similar<br>components or associated<br>equipment, only the<br>predicate device contains<br>the endoscopic module,<br>and the difference will<br>not affect the use of the<br>device and will not affect<br>and raise new risks<br>related to safety and<br>effectiveness of the<br>device. |
| Patient<br>Immobilization | Yes - The device is<br>attached to the head<br>holder or the frame via<br>an<br>adaptor | Yes - The device is<br>attached to the head<br>holder or the frame via an<br>adaptor | Same |
| Device Mobility | Yes - Mobile stands<br>with wheels; Robotic<br>stand immobilized<br>with stabilization feet | Yes - Mobile stands<br>with wheels; Robotic<br>stand immobilized<br>with stabilization feet | Same |
| Vigilance System | Yes - foot pedal | Yes - foot pedal | Same |
| Sterile | Non-sterile and sterile<br>instruments<br>Disposable sterile<br>drapes for the robotic<br>arm and touch screen | Non-sterile and sterile<br>instruments<br>Disposable sterile<br>drapes for the robotic<br>arm and touch screen | Same |
| Single-Use | No | No | Same |
| Ac Powered | Yes | Yes | Same |
| Electrical Safety<br>Testing | IEC 60601 Series<br>of<br>Standards | IEC 60601 Series<br>of<br>Standards | Same |
{8}------------------------------------------------
## 5.7 Non-Clinical Performance Data
To demonstrate safety and effectiveness of Sinovation and to show substantial equivalence to the predicate device, Sinovation completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The Sinovation passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Risk Assessment of the Biocompatibility per ISO 10993-1 Requirements Met ●
- Cytotoxicity testing per ISO 10993-5 Passed ●
- Sensitization testing per ISO 10993-10 Passed ●
- Electrical safety testing per IEC 60601-1 Passed ●
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 Passed ●
- Usability per IEC 60601-1-6 Demonstrates the usability was considered during design ●
- Usability per IEC 62366-1 Demonstrates the usability was considered during design ●
- Software verification and validation per IEC 62304/FDA Guidance results ● /conclusion
- Safety of Laser Products per IEC 60825-1 Passed ●
- Sterilization validation - demonstrates SAL of 10-6
- Shelf Life Testing Supports shelf life of 10 years ●
- Transportation Testing per ASTM D4169 Demonstrates package integrity maintained ● Sinovation also conducted the accuracy tests and the data as reported in Table 5.2 below:
| Device Performance | Mean | Standard<br>deviation | Confidence interval at<br>99% confidence level |
|---------------------------|-------|-----------------------|------------------------------------------------|
| Positional accuracy [mm] | 0.613 | 0.197 | 1.205 |
| Angular accuracy [degree] | 0.439 | 0.241 | 1.163 |
#### Table 5.2 Accuracy Testing Results
### 5.8 Statement of Substantial Equivalence
The Sinobot X1 has the same intended use as the predicate device, and the same or similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated the Sinobot X1 is as safe and effective as the predicate device. Therefore, the Sinobot X1 is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.