Diathermic Slitter (FlushKnife) DK2620JI and DK2623JI
Applicant
Fujifilm Corporation
Product Code
KGE · Gastroenterology, Urology
Decision Date
Feb 4, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These products are used in combination with an applicable endoscope to perform the endoscopic treatments for the target site in the gastrointestinal tract, such as ablation, incision, cauterization, coagulation and avulsion of tissue and arrest of bleeding, and to inject saline or submucosal injection agent for endoscopic surgery into the submucosal layer in the gastrointestinal tract, which is exposed after dissection, under the management of physicians in medical facilities. Never use this product for any other purpose.
Device Story
FlushKnife DK2620JI/DK2623JI is a monopolar electrosurgical instrument used with an endoscope for gastrointestinal procedures. Device features needle or ball tip slitter to deliver high-frequency electrical current for tissue cutting, coagulation, and hemostasis. Integrated water feed function allows injection of saline or submucosal agents. Operated manually via handle slider by physicians in medical facilities. Rotatable water supply connector improves usability compared to predicate. Device is single-use and sterile. Output is thermal energy for tissue modification; clinical benefit includes precise tissue dissection and bleeding control during endoscopic surgery.
Clinical Evidence
Bench testing only. Evaluated electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-2/18), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993 series), and sterility (ISO 11135). Device-specific bench testing validated compatibility with submucosal injection agents and confirmed performance specifications for insertion diameter, working length, slitter length, conductivity, and liquid delivery. No clinical data provided.
Technological Characteristics
Monopolar electrosurgical instrument; needle or ball tip; manual operation via handle slider. Features rotatable water supply connector. Single-use; sterile. Complies with ISO 10993 (biocompatibility), ISO 11135 (sterility), and IEC 60601 series (electrical/EMC).
Indications for Use
Indicated for patients requiring endoscopic gastrointestinal surgery for tissue ablation, incision, dissection, cauterization, coagulation, avulsion, hemostasis, and submucosal injection of saline or agents.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 4, 2022
FUJIFILM Corporation % Dhara Buch Regulatory Affairs Specialist FUJIFILM Healthcare Americas Corporation 81 Hartwell Avenue, Suite 300 Lexington, MA 02421
Re: K220053
> Trade/Device Name: Diathermic Slitter (FlushKnife) DK2620JI and DK2623JI Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KGE Dated: January 5, 2022 Received: January 6, 2022
Dear Dhara Buch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220053
Device Name
FUJIFILM Diathermic Slitter (FlushKnife) DK2620JI and DK2623JI
Indications for Use (Describe)
These products are used in combination with an applicable endoscope to perform the endoscopic treatments for the target site in the gastrointestinal tract, such as ablation, incision, cauterization, coagulation and avulsion of tissue and arrest of bleeding, and to inject saline or submucosal injection agent for endoscopic surgery into the submucosal layer in the gastrointestinal tract, which is exposed after dissection, under the management of physicians in medical facilities. Never use this product for any other purpose.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
### FUJIFILM Corporation
### FUJIFILM Diathermic Slitter (FlushKnife) DK2620JI and DK2623JI
Date: January 05, 2021
# Submitter's Information:
FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan
# Contact Person:
Dhara Buch Requlatory Affairs Specialist Phone: 781-824-2708 E-Mail: dhara.buch(@fujifilm.com
# Identification of the Proposed Device:
| Device Name: | FUJIFILM Diathermic Slitter (FlushKnife) DK2620J<br>and DK2623JI |
|-----------------|------------------------------------------------------------------|
| Common Name: | Electrosurgical Instruments |
| Product Code: | KGE |
| Device Class: | Class II |
| Classification: | Endoscopic electrosurgical unit and accessories |
| | 21 C.F.R. § 876.4300 |
| Review Panel: | Gastroenterology/Urology |
#### Predicate Device:
- FUJIFILM Diathermic Slitter FlushKnife (K171096)
#### Intended Use / Indications for Use:
These products are used in combination with an applicable endoscope to perform the endoscopic treatments for the target site in the gastrointestinal tract, such as ablation, incision, dissection, cauterization, coagulation and avulsion of tissue and arrest of bleeding, and to inject saline or submucosal injection agent for endoscopic surgery into the submucosal layer in the gastrointestinal tract, which is exposed after dissection, under the management of physicians in medical facilities. Never use this product for any other purpose.
# Device Description:
FUJIFILM Diathermic Slitter (FlushKnife) is an electrosurgical instrument that removes tissue and controls bleeding by use of high-frequency ("HF") electrical current. The device is available with either a needle or ball tip slitter and comes in a variety of sizes.
# Comparison of Technological Characteristics:
{4}------------------------------------------------
A comparison of technological characteristics between subject device and the predicate device is provided table below:
| Device Details | Predicate Device | Subject Device |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | FUJIFILM Diathermic Slitter Knife | FUJIFILM Diathermic Slitter<br>(FlushKnife) DK2620JI and<br>DK2623JI |
| Model | MODEL DK2618J -N10-, DK2618J -<br>N15-, DK2618J -N20-, DK2618J -N25-,<br>DK2618J -N30-, DK2618J -B15-,<br>DK2618J -B20-, DK2618J -B25-,<br>DK2618J -B30-, DK2623J -N15-,<br>DK2623J -N20-, DK2623J -B15-,<br>DK2623J -B20- | MODEL DK2620JI N10-,<br>DK2620JI -N15, DK2620JI -N20-<br>, DK2620JI -N25-, DK2620JI -<br>N30-, DK2620JI -B15-,<br>DK2620JI -B20-, DK2620JI -<br>B25-, DK2620JI -B30-,<br>DK2623JI -N15-, DK2623JI -<br>N20-, DK2623JI -B15-,<br>DK2623JI -B20- |
| 510(k) number | K171096 | To be assigned |
| Product code | KGE | KGE |
| Manufacturer | FUJIFILM Corporation | FUJIFILM Corporation |
| Intended Use | This instrument is designed to be used<br>with specified endoscopes to cut tissue<br>within the digestive tract and using high-<br>frequency current. Both of the ball tip<br>type and the needle type instruments<br>are indicated for ablation, incision,<br>dissection, avulsion, cauterization,<br>coagulation and hemostasis of tissue<br>within the digestive tract. | These products are used in<br>combination with an applicable<br>endoscope to perform the<br>endoscopic treatments for the<br>target site in the gastrointestinal<br>tract, such as ablation, incision,<br>dissection, cauterization,<br>coagulation and avulsion of<br>tissue and arrest of bleeding,<br>and to inject saline or<br>submucosal injection agent for<br>endoscopic surgery into the<br>submucosal layer in the<br>gastrointestinal tract, which is<br>exposed after dissection, under<br>the management of physicians in<br>medical facilities. |
| Slitter Length | 1.0/1.5/2.0/2.5/3.0mm (for DK2618J<br>series)<br>1.5/2.0mm (for DK2623J series) | Same as K171096 |
| Slitter Diameter | 0.5mm | Same as K171096 |
| Slitter Shape | Needle Type<br>(With Ball Tip:-BXX-)<br>(Without Ball Tip:-NXX-) | Same as K171096 |
| Maximum Diameter<br>of Insertion Portion | 2.7mm | Same as K171096 |
| Water Feed Function | Yes | Same as K171096 |
| Method of Operation | Manually (handle slider) | Same as K171096 |
| Water supply<br>connector | Yes, fixed | Yes, rotatable |
| Energy | Energy delivered from an<br>electrosurgical generator | Same as K171096 |
| Monopolar /<br>Bipolar | Monopolar | Same as K171096 |
| Sterilization | Yes (Single Use Device) | Same as K171096 |
| Combination<br>Tools | Endoscope, Electrosurgical generator,<br>A Cord, Water supply unit | Endoscope, Electrosurgical<br>generator, A Cord |
| Injection agent | Saline and other sterile liquids | Saline, Eleview (K150852),<br>ORISE Gel(K180068) |
{5}------------------------------------------------
#### Performance Data:
Electrical safety and electromagnetic compatibility of the proposed device was evaluated using the following standards: ANSI/AAMI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014, IEC 60601-1-6:2013, IEC 60601-2-2:2017, and IEC 60601-2-18:2009.
Biocompatibility of the proposed device was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-7:2008, ISO 10993-10:2010, and ISO 10993-11:2017. Biocompatibility testing was performed in accordance with FDA's guidance, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process,"" published September 4, 2020.
Sterility of the proposed device was evaluated according to ISO 11135:2014 and the FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
Device specific testing was conducted using the following consensus standard: ISO 8600-1:2015. Bench testing was conducted to validate the compatibility of subject device with submucosal injection agent.
The proposed device met performance specifications in the following additional testing:
- Maximum diameter of insertion portion
- . Working length
- Slitter length
- Conductivity
- Liquid delivery amount ●
In all cases, the device met the pre-defined acceptance criteria for the test.
#### Conclusions:
The subject device shares the same intended use and similar indications as the predicate device (K171096).
Bench testing demonstrates that the subject device is as safe and effective as the predicate device. Thus, subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.