K220050 · KLS Martin L.P. · MQN · Aug 11, 2022 · Dental
Device Facts
Record ID
K220050
Device Name
KLS Martin IPS Distraction
Applicant
KLS Martin L.P.
Product Code
MQN · Dental
Decision Date
Aug 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K220050 · Aug 11, 2022
KLS Martin IPS Distraction
KLS Martin L.P.
Published clinical literature; Retrospective clinical case data
The sponsor conducted a literature review of five clinical studies to support the safety and effectiveness of the KLS Martin IPS Distraction device in pediatric populations (infants, children, and adolescents).
Pediatric safety; Literature review; Mandibular distraction; Craniofacial reconstruction
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of 5 clinical studies (Doscher et al., Gray et al., Steinbacher, Resnick, Meena et al.); Retrospective clinical literature review
Infants (including neonates >2.5kg), children, and adolescents (26 days to 21 years of age)
Not applicable for this study
Safety and effectiveness of patient-specific implants in pediatric subpopulations
Indications for Use
KLS Martin IPS Distraction includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, and symphysis) and mid-face bones that require gradual distraction in adults, adolescents, children and infants weighing more than 2.5 kg.
Device Story
System comprises patient-specific models, guides, and distraction footplates used with previously cleared distractor bodies and metallic bone screws. Input: patient CT scan data. Process: physician provides input for model manipulation/interactive feedback; KLS Martin personnel modify digital models; verification via virtual fitting over 3D patient anatomy. Output: patient-specific titanium implants. Used in maxillofacial/midface/mandibular reconstruction. Operated by surgeons. Output allows for gradual bone distraction/stabilization. Benefits: customized fit for complex anatomical defects; enables bone lengthening/transport in pediatric and adult populations.
Clinical Evidence
No new clinical trials conducted. Safety and effectiveness for pediatric subpopulations (infants >2.5kg, children, adolescents) supported by analysis of 5 peer-reviewed clinical studies (Doscher 2014, Gray 2017, Steinbacher 2015, Resnick 2017, Meena 2017) involving patients aged 26 days to 21+ years. Studies confirm use of CT-based planning, VSP, and custom implants for mandibular distraction and craniofacial reconstruction.
Technological Characteristics
Materials: CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136). Manufacturing: Traditional (subtractive) or additive (Selective Laser Melting). Plate thickness: 0.6-10 mm. Fixation: Titanium screws (1.0-2.3 mm diameter). Sterilization: Gamma (sterile) or Steam (non-sterile). Connectivity: Uses COTS software for DICOM to .STL conversion. Design: Patient-specific based on CT imaging.
Indications for Use
Indicated for adults, adolescents, children, and infants (>2.5 kg) requiring gradual distraction for correction of congenital deficiencies or post-traumatic defects of the mandible (ramus, body, alveolar ridge, palate, symphysis) and mid-face bones. Contraindicated in active infection, insufficient bone quality/quantity, blood supply limitations, patients unable to follow post-op instructions, and foreign body sensitivity.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
KLS Martin Internal Distraction – Sterile (K161470)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
KLS Martin L.P. Susan Leander Regulatory Affairs Project Manager 11201 Saint Johns Industrial Parkway S Jacksonville, Florida 32246
8/11/22
Re: K220050
Trade/Device Name: KLS Martin IPS Distraction Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: MQN Dated: June 30, 2022 Received: July 12, 2022
Dear Susan Leander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220050
Device Name KLS Martin IPS Distraction
#### Indications for Use (Describe)
KLS Martin IPS Distraction includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, and symphysis) and mid-face bones that require gradual distraction in adults, adolescents, children and infants weighing more than 2.5 kg.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black. The logo is simple and modern.
11201 Saint Johns Industrial Pkwy S · Jacksonville, FL 32246 Phone: 904-641-7746 · 800-625-1557 · Fax: 904-641-7378 www.klsmartinnorthamerica.com
# K220050
### 510(k) Summary 21 CFR 807.92
| Submitter: | KLS-Martin L.P.<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 |
|----------------------|------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Susan Leander<br>Regulatory Affairs Project Manager<br>Phone: 800-625-1557<br>Email: susan.leander@klsmartin.com |
| Date Prepared: | August 11, 2022 |
| Trade Name: | KLS Martin IPS Distraction |
| Common Name: | Plate, Bone |
| Classification Name: | Bone Plate |
| Regulatory Class: | II |
| Product Code: | MQN |
| Primary Predicate: | KLS Martin Internal Distraction – Sterile (K161470) |
| Reference Devices: | KLS Martin Individual Patient Solutions (K191028)<br>KLS Martin Zurich Distraction System (K010139) |
### Device Description
KLS Martin IPS Distraction System is comprised of patient-specific models, guides and distraction footplates used in conjunction with previously cleared distractor bodies and metallic bone screws for internal fixation and reconstruction of maxillofacial / midface and mandibular bones. The devices are manufactured based on medical imaging (CT scan) of the patient's anatomy with input from the physician during virtual planning and prior to finalization of the device. The physician only provides input for model manipulation and interactive feedback by viewing digital models of planned outputs that are modified by trained KLS Martin personnel during the planning session. For each design iteration, verification is performed by virtually fitting the generated device model over a 3D model of the patient's anatomy to ensure its dimensional properties allow an adequate fit.
Implants are provided sterile or non-sterile, range in thickness from 0.6 - 10 mm, and are manufactured using additive or traditional (subtractive) methods from either CP Titanium (ASTM F67) or Ti-6A1-4V (ASTM F136) materials. These patient-specific devices are fixated with previously cleared KLS Martin screws.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black. The logo is simple and modern.
### Indications for Use
KLS Martin IPS Distraction includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, and symphysis) and mid-face bones that require gradual distraction in adults, adolescents, children, and infants weighing more than 2.5 kg.
### Technological Characteristics & Substantial Equivalence Discussion
The intended use and technological characteristics of the subject device, KLS Martin IPS Distraction are the same as those of the predicate device. KLS Martin Internal Sterile Distraction, cleared under K161470. Both are distraction devices made of titanium. Both use standard distractors with attached footplates. The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing.
### Similarities to Predicate and Reference Devices
The primary predicate device, KLS Martin Internal Sterile Distraction, K161470, and the subject device, KLS Martin IPS Distraction are intended for gradual distraction in the facial skeleton and share the same fundamental principle of operation - metallic bone plates used in conjunction with metallic bone screws for bone stabilization and lengthening of bones. The indications for use for the subject device are the same as the predicate device. The skeletal regions for the subject and predicate devices are similar.
Both the primary predicate device and the KLS Martin Zurich Distraction System reference device, cleared under K010139, present evidence of substantially equivalent use of similar devices. The method to attach the patient-specific footplates to the distraction bodies is the same for the subject, predicate and reference devices.
The KLS Martin IPS Distraction devices are provided sterile (same as K161470) or non-sterile (same as K010139 and K191028). The non-sterile devices require the end-user to process them using validated cleaning and sterilization methods prior to use as recommended in the labeling.
The KLS Martin Individual Patient Solutions is included as a reference device to support the use of patient-specific metallic implants. Both the subject and the KLS Martin Individual Patient Solutions (K191028) reference devices use image data obtained from medical scanners, such as a CT scan. They both use validated commercial-off-the-shelf (COTS) software applications to transfer patient imaging from a DICOM format to an .STL format and manipulate the images to produce a final design file.
The manufacturing materials used for the implants in the subject device are similar to those cleared in their own prior clearance. K191028. The design and dimensions of the plates will be based on patientspecific data. using similar methods described in K191028. made from either CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136) using traditional (subtractive) or additive manufacturing methods. The bone plates are similar in final finished form to those presented in the K191028 reference device, and therefore performance and biocompatibility are adequately addressed.
IPS Distractor bone plates will be attached to maxillofacial, midface or mandibular bones and to distractor bodies for distraction osteogenesis. Bone plates in the subject device may be implanted permanently in skeletally mature patients or they may be removed on completion of the procedure, as is common for K161470 and K010139 predicate and reference devices. Permanent implants are not recommended for use in skeletally immature patients (16 years of age or younger).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KLS Martin L.P. The text "KLS" is in red, while "martin" is in gray. "L.P." is in black and has a unique font style.
### Differences from Predicate and Reference Devices
The indications for use statements for the subject device and predicate devices are the same, except the subject device statement specifically lists the following targeted patient populations:
- · Infants that weigh more than 2.5 kg (29 days to < 2 years of age) *
- · Children (2 years of age to < 12 years of age)
- · Adolescents (12 years of age 21 years of age)
- Adults
### *Premature neonates are not included in the proposed target population. The lowest patient weight for the subject device is 2.5 kg (5 pounds, 8 ounces), which is considered normal weight at delivery.
A risk assessment has been performed based on FDA guidance, "Premarket Assessment of Pediatric Medical Devices, March 24, 2014" for these pediatric subpopulations with supporting peer-reviewed clinical literature to demonstrate the safety and effectiveness of the subject device implants for use in the target pediatric subpopulations.
The reference device. K191028, was previously cleared only for the adult population, but the primary predicate, K161470, supports the use of implants in pediations infant, children, and adolescents.
The KLS Martin IPS Distraction alloy footplates are attached to titanium alloy distraction bodies, whereas the predicate and reference device CP titanium footplates are attached to the same style titanium alloy distraction bodies. The different material combinations were addressed through performance testing.
Previously cleared specifications for the K191028 reference device plate thicknesses range from 0.3 mm - 10 mm and are fixated with previously cleared KLS Martin titanium screws ranging in diameter from 1.0 mm - 3.2 mm in lengths from 3.5 mm - 22 mm. The subject device plates range in thickness from 0.6 mm - 10 mm and are fixated with previously cleared KLS Martin titanium screws ranging in diameter from 1.0 mm - 2.3 mm in lengths from 3 mm - 22 mm. The screw dimensions for the subject device are similar to those previously evaluated in K161470. The predicate and reference device dimensions encompass the entire range of dimensions for the subject device.
A device comparison table of the subject, predicate, and reference devices is presented, below.
{6}------------------------------------------------
| Device Comparison Table | | | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KLS Martin IPS Distraction | KLS Martin Internal Distraction - Sterile<br>K161470 | KLS Martin Individual Patient Solutions<br>K191028 | KLS Martin Zurich Distraction System<br>K010139 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) |
| Indications for Use | KLS Martin IPS Distraction includes devices<br>intended as bone stabilizers and lengthening<br>(and or transport) devices for correction of<br>congenital deficiencies or post traumatic defects<br>of the mandible (including ramus, body,<br>alveolar ridge, palate, and symphysis) and mid-<br>face bones that require gradual distraction in<br>adults, adolescents, children and infants<br>weighing more than 2.5 kg. | Internal Distraction - Sterile includes devices<br>intended as bone stabilizers and lengthening<br>(and or transport) devices for correction of<br>congenital deficiencies or post traumatic defects<br>of the mandible (including ramus, body,<br>alveolar ridge, palate, and symphysis) and mid-<br>face bones that require gradual distraction. | KLS Martin Individual Patient Solutions implant<br>devices are intended for use in the stabilization,<br>fixation, and reconstruction of the maxillofacial /<br>midface and mandibular skeletal regions. | The Zurich Distraction System includes devices<br>intended as a bone stabilizer and lengthening<br>(and or transport) device when correction of<br>congenital deficiencies or post traumatic<br>defects of the mandible (including ramus, body,<br>alveolar ridge, palate, symphysis), mid-face,<br>and cranial bones require gradual distraction. |
| Contraindications | 1. Active infection<br>2. Patient conditions including: blood supply<br>limitations, insufficient quantity or quality<br>of bone or latent infections<br>3. Patients with mental or neuralgic conditions<br>who are unwilling or incapable of following<br>postoperative care instructions<br>4. Foreign body sensitivity- where material<br>sensitivity is suspected, tests are to be made<br>prior to implantation | 1. Active infection<br>2. Patient conditions including: blood supply<br>limitations, insufficient quantity or quality<br>of bone or latent infections<br>3. Patients with mental or neuralgic<br>conditions who are unwilling or incapable<br>of following postoperative care instructions<br>4. Foreign body sensitivity- where material<br>sensitivity is suspected, tests are to be<br>made prior to implantation | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Suspected sensitivity to the implant material.<br>4. Circulatory problems, systemic diseases, and<br>metabolic disorders.<br>5. Insufficient or inadequate bone tissue.<br>6. Secondary diseases such as degenerative<br>processes that may negatively influence the<br>healing process.<br>7. Interventions carried out in a non-sterile<br>environment (e.g. paranasal sinuses).<br>8. Regions exposed to inappropriate forces or<br>excessive weight loads.<br>9. Patients unwilling or unable to follow<br>instructions during the postoperative phase<br>due to their mental, neurological, or physical<br>condition.<br>10. Osteoporosis or osteomalacia or other<br>structural bone damage preventing the stable<br>fixation of implant components.<br>11. Bone tumors located in the implant base<br>region.<br>12. Obvious drug or alcohol abuse. | 1. Active infection<br>2. Patient conditions including: blood supply<br>limitations, insufficient quantity or quality<br>of bone or latent infections<br>3. Patients with mental or neuralgic<br>conditions who are unwilling or incapable<br>of following postoperative care<br>instructions<br>4. Foreign body sensitivity- where material<br>sensitivity is suspected, tests are to be<br>made prior to implantation<br>5. For Symphysis: In those patients where<br>there is inadequate volume or quality of<br>bone to place the distractor securely.<br>6. For Ramus: In these cases where is an in-<br>adequate volume or quality of bone to<br>place the distractor securely. |
| Device Comparison Table | | | | |
| | KLS Martin IPS Distraction | KLS Martin Internal Distraction - Sterile<br>K161470 | KLS Martin Individual Patient Solutions<br>K191028 | KLS Martin Zurich Distraction System<br>K010139 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) |
| Patient-specific? | Yes. Devices are manufactured patient-specific,<br>based on a CT scan of the patient. | No. Standard distraction plates are available in a<br>variety of sizes in thicknesses of 0.6 mm – 2.0<br>mm. | Yes. Devices are manufactured patient-specific,<br>based on a CT scan of the patient. | No. Standard distraction plates are available in<br>a variety of sizes in thicknesses of 0.6 mm –<br>1.0 mm. |
| Classification | 21 CFR 872.4760, Class II | 21 CFR 872.4760, Class II | 21 CFR 872.4760, Class II | 21 CFR 872.4760, Class II |
| Product Code | MQN | MQN | JEY | MQN |
| Material | Anatomical Models: Epoxy/Acrylic Resins<br>Implants: CP Titanium or Ti-6Al-4V | Implants: CP Titanium or Ti-6Al-4V | Anatomical Models: Epoxy/Acrylic Resins<br>Implants: CP Titanium or Ti-6Al-4V | Traditional (Subtractive) and Ti-6Al-4V:<br>Traditional (Subtractive) |
| Manufacturing Method | Epoxy/Acrylic Resins: Stereolithography (SLA)<br>CP Titanium or Ti-6Al-4V: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive; Selective Laser<br>Melting) | Traditional (Subtractive) and Ti-6Al-4V:<br>Traditional (Subtractive) | Epoxy/Acrylic Resins: Stereolithography (SLA)<br>CP Titanium: Traditional (Subtractive)<br>Ti-6Al-4V: 3D (Additive; Selective Laser Melting) | Traditional (Subtractive) and Ti-6Al-4V:<br>Traditional (Subtractive) |
| Sterilization | Sterile (Gamma)<br>Non-sterile (Steam) | Sterile (Gamma Radiation) | Non-sterile (Steam) | Non-sterile (Steam) |
| Anatomical Sites | Maxillofacial / Midface & Mandible | Maxillofacial / Mandible | Maxillofacial / Midface & Mandible | Maxillofacial / Mandible |
| Patient Population | • Infants that weigh at least 2.5kg (29 days to<br><2 years of age)<br>• Children (2 years of age to < 12 years of age)<br>• Adolescents (12 years of age - 21 years of<br>age)<br>• Adults | • Infants (29 days to < 2 years of age)<br>• Children (2 years of age to < 12 years of age)<br>• Adolescents (12 years of age - 21 years of<br>age)<br>• Adults | Adult | Pediatric & Adult |
| Distraction<br>Length | Max 60 mm | Max 60 mm | Not specified | 15mm - 50 mm |
| Device Comparison Table | | | | |
| | KLS Martin IPS Distraction | KLS Martin Internal Distraction - Sterile<br>K161470 | KLS Martin Individual Patient Solutions<br>K191028 | KLS Martin Zurich Distraction System<br>K010139 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) |
| Plate Specifications | | | | |
| Thickness | • 0.6mm-10mm | • 0.6mm-2.0mm | • Orbital floor only: 0.3mm–1.0mm<br>• Maxillofacial / midface reconstruction: 0.6mm–10 mm<br>• Mandibular reconstruction: 1.0mm–3.0mm | • 0.6mm-1.0mm |
| Plate Style | Locking & non-locking | Locking & non-locking | Locking & non-locking | Locking & non-locking |
| Plate Width | Maxillofacial / midface /mandibular:<br>Min: ≥ 3 mm<br>Max: Dependent on screw-hole | Not specified | Maxillofacial / midface:<br>Min: ≥ 3 mm<br>Max: Dependent on screw-hole<br>Mandibular:<br>Min: ≥ 7 mm<br>Max: 8.5 mm | Not specified |
| Plate Length | Maxillofacial / midface /mandibular:<br>Min: 5 mm<br>Max: 350 mm | Not specified | Maxillofacial / midface:<br>Min: 18 mm<br>Max: 350 mm<br>Mandibular:<br>Min: 31 mm<br>Max: 320 mm | Not specified |
| Degree of Curvature | In-plane<br>Min: 30°<br>Max: 180°<br>Out-of-plane<br>Min: 15°<br>Max: 180°…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.