iStent infinite Trabecular Micro-Bypass System, Model iS3
Applicant
Glaukos Corporation
Product Code
KYF · Ophthalmic
Decision Date
Aug 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3920
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iStent infinite Trabecular Micro-Bypass System Model iS3 is an implantable device intended to reduce the intraocular pressure (IOP) of the eye. It is indicated for use in adult patients with primary open-angle glaucoma in whom previous medical and surgical treatment has failed.
Device Story
The iStent infinite Model iS3 is a sterile, single-use injector system pre-loaded with three titanium G2-W stents. Designed for implantation into Schlemm's canal by an ophthalmologist in a clinical setting, the device facilitates aqueous humor outflow from the anterior chamber via trabecular bypass. The stent features a monolithic design with a central lumen and four lateral outlet orifices to reduce intraocular pressure (IOP). The injector delivers the stents into the trabecular meshwork. By bypassing the trabecular meshwork, the device provides an exit route for aqueous fluid, aiming to lower IOP in patients with refractory primary open-angle glaucoma. Clinical success is measured by the proportion of patients achieving a ≥20% reduction in mean diurnal IOP from baseline at 12 months while maintaining or reducing medication usage. The device offers a surgical option for patients who have not responded to prior interventions, potentially delaying or avoiding more invasive filtering surgeries.
Clinical Evidence
Prospective, multi-center, single-arm, open-label clinical trial (N=61) in adult patients with primary open-angle, pseudoexfoliative, or pigmentary glaucoma who failed prior treatment. Primary endpoints at 12 months: 72.1% of participants achieved ≥20% mean diurnal IOP (MDIOP) reduction from baseline on same/fewer medications. Mean MDIOP change was -5.5 mmHg. Safety outcomes included adverse events (AEs); no intraoperative AEs reported. Common postoperative AEs included ocular surface disease (11.5%) and BSCVA loss ≥2 lines (11.5%). 4.9% required secondary surgical intervention.
Indicated for adult patients with primary open-angle glaucoma who have failed previous medical and surgical treatment.
Regulatory Classification
Identification
An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) “Aqueous Shunts—510(k) Submissions.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 2, 2022
Glaukos Corporation Mr. David Fernquist Vice President Regulatory Affairs 229 Avenida Fabricante San Clemente, California 92672
Re: K220032
Trade/Device Name: iStent infinite Trabecular Micro-Bypass System, Model iS3 Regulation Number: 21 CFR 886.3920 Regulation Name: Aqueous Shunt Regulatory Class: Class II Product Code: KYF Dated: June 21, 2022 Received: June 24, 2022
Dear Mr. David Fernquist:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220032
Device Name iStent infinite® Trabecular Micro-Bypass System Model iS3
#### Indications for Use (Describe)
The iStent infinite® Trabecular Micro-Bypass System Model iS3 an implantable device intended to reduce the intraocular pressure (IOP) of the eye. It is indicated for use in adult patients with primary open-angle glaucoma in whom previous medical and surgical treatment has failed.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| SUBMITTER<br>I. | Glaukos Corporation<br>229 Avenida Fabricante<br>San Clemente, CA 92672 |
|------------------------|---------------------------------------------------------------------------------------------------|
| Contact Person: | David S. Fernquist<br>Vice President, Regulatory Affairs<br>Glaukos Corporation<br>(949) 367-9600 |
| Date Summary Prepared: | August 2, 2022 |
| II.<br>DEVICE | |
| Trade Name: | iStent infinite Trabecular Micro-Bypass System Model iS3 |
| Common Name: | Glaucoma Implant |
| Classification Name: | Aqueous shunt (21 CFR 886.3920) |
| Device Class: | Class II (special controls) |
| Device Product Code: | KYF ("Implant, Eye Valve") |
#### III. PREDICATE DEVICE
- XEN Glaucoma Treatment System (K161457), 21 CFR 886.3920, Product Code KYF
#### IV. DEVICE DESCRIPTION
The iStent infinite Trabecular Micro-Bypass System Model iS3 is a sterile, single-use injector system that is pre-loaded with three G2-W stents, and is designed to deliver the stents into Schlemm's canal. The G2-W stents are manufactured from implant grade titanium (Ti6Al4V ELI per ASTM F136) and are coated with stearalkonium heparin. An area of reduced outside diameter, midway along the device, is designed to provide retention within the trabecular meshwork, while multiple outlet lateral lumens (4 outflow orifices) are designed to provide an exit route for aqueous from the anterior chamber. The stent has a single piece design, is 360 um in diameter, 360 um in height, and the central inlet and outlet lumen has a diameter of 80 um. The head of the stent has
{4}------------------------------------------------
four side outlets that each have a diameter of 50 um.
#### V. INTENDED USE
The iStent infinite Trabecular Micro-Bypass System Model iS3 has the same intended use as the predicate device and all other devices regulated within the generic type of device known as aqueous shunts in accordance with 21 CFR 886.3920. The iStent infinite is a prescription (Rx) device that is intended to be permanently implanted to reduce intraocular pressure for the management of glaucoma. Both the subject device and predicate device (XEN® Glaucoma Treatment System) have the same intended use for the treatment, namely, "to be permanently implanted to reduce intraocular pressure for the management of glaucoma", as stated above. The iStent infinite will bear the following indications for use statement:
The iStent infinite Trabecular Micro-Bypass System Model iS3 is an implantable device intended to reduce the intraocular pressure (IOP) of the eye. It is indicated for use in adult patients with primary open-angle glaucoma in whom previous medical and surgical treatment has failed.
The indications for use statement for iStent infinite is not substantially different from that of the XEN implant predicate device. The non-substantial difference in the indications for use statement for the iStent infinite vs. the predicate device is that the iStent infinite indication is narrower. Compared to the XEN indications for use statement (shown below), the iStent infinite indications for use statement does not include patients with pseudoexfolative glaucoma or pigmentary glaucoma who have failed maximally tolerated medical therapy. It should be noted that both the pivotal study for XEN and the pivotal study for iStent infinite comprised a substantial majority of primary open-angle glaucoma patients with prior failed surgical intervention and similar baseline characteristics.
The XEN® Glaucoma Treatment System is indicated for the management of refractory glaucomas, including cases where previous surgical treatment has failed, cases of primary open angle glaucoma, and pseudoexfoliative or pigmentary glaucoma with open angles that are unresponsive to maximum tolerated medical therapy.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE SUBJECT DEVICE AND PREDICATE DEVICE
Although the iStent infinite and the XEN do not share identical technological characteristics, those differences do not raise different questions of safety and effectiveness.
{5}------------------------------------------------
| Characteristics | Allergan XEN Glaucoma<br>Treatment System<br>K161457<br>PREDICATE DEVICE | iStent infinite<br>SUBJECT DEVICE |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | To be permanently implanted to<br>reduce intraocular pressure for the<br>management of glaucoma | To be permanently implanted to<br>reduce intraocular pressure for the<br>management of glaucoma |
| Regulation<br>Number/Product<br>Code | 886.3920, KYF | 886.3920, KYF |
| Indication | The XEN Glaucoma Treatment<br>System is indicated for the<br>management of refractory<br>glaucomas, including cases where<br>previous surgical treatment has<br>failed, cases of primary open angle<br>glaucoma, and pseudoexfoliative or<br>pigmentary glaucoma with open<br>angles that are unresponsive to<br>maximum tolerated medical therapy. | The iStent infinite Trabecular Micro-<br>Bypass System Model iS3 is an<br>implantable device intended to<br>reduce the intraocular pressure (IOP)<br>of the eye. It is indicated for use in<br>adult patients with primary open-<br>angle glaucoma in whom previous<br>medical and surgical treatment has<br>failed. |
| Rx or OTC | Rx | Rx |
| Permanent Implant | Yes | Yes |
| Design | Monolithic, round tube design with a<br>central inlet and outlet lumen | Monolithic round tube design with a<br>central inlet and outlet lumen |
| Material | Gelatin derived from porcine dermis,<br>formed into a tube, and then cross-<br>linked with glutaraldehyde; no<br>coating | Implant grade titanium (Ti6Al4V<br>ELI) with a stearalkonium heparin<br>coating |
| Size (nominal<br>dimensions) | Dry dimensions:<br>6 mm length<br>0.045 mm inner diameter<br>0.15 mm outside diameter | 0.36 mm length<br>0.36 mm flange diameter<br>0.23 mm head diameter<br>0.08 mm central outlet diameter<br>0.05 mm flow outlet diameter |
| Sterilization | Terminal gamma ray radiation | Terminal gamma ray radiation |
| Single-Use | Yes | Yes |
| Anatomical site | Angle-based implant | Angle-based implant |
| Mechanism of action | Outflow of aqueous fluid via a<br>subconjunctival bleb | Outflow of aqueous fluid via<br>trabecular bypass |
| Method of Insertion | Via a preloaded XEN injector | Via a preloaded injector |
# Table 1. Comparison of the iStent infinite System and XEN Glaucoma Treatment System (Predicate Device)
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
{6}------------------------------------------------
#### Bench Testing A.
The nonclinical bench testing conducted on the iStent infinite Trabecular Micro-Bypass System Model iS3 included design verification and functional product testing, sterilization validation, packaging and shelf life testing, and biocompatibility testing. Results of the nonclinical testing demonstrate that theiStent infinite Trabecular Micro-Bypass System Model iS3 functions as intended.
## Design Verification and Functional Product Testing:
The iStent infinite Trabecular Micro-Bypass System Model iS3 Stent and injector were evaluated to verify that the design output met the original design input and intent. This testing was based on tests described in ANSI Z80.27, Implantable Glaucoma Devices, the FDA guidance document "Aqueous Shunts - 510(k) Submissions", and the FDA guidance document "Premarket Studies of Implantable Minimally Invasive Glaucoma Surgical (MIGS) Devices". All physical and mechanical testing demonstrated that the stent and injector function as intended.
{7}------------------------------------------------
| Table 2. Physical & Mechanical Testing | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Results |
| Surface & Edge<br>Quality | High magnification SEM photos of the G2-W stent demonstrated that the stent<br>had smooth edges and was free from surface defects. |
| Dimensions | Glaukos has validated that stent production meets tolerances to appropriate<br>statistical levels. |
| Physical Stability | An in-vivo test to evaluate the physical stability of the titanium stent was<br>performed per ANSI Z80.27 Section 5.5. A total of 10 etched and coated<br>stents were pulled from a production lot for the validation, inspected, and were<br>placed into BSS for 14 days at a temperature 35 $\pm$ 2 °C. Visual inspection (at<br>least 10x) and dimensional measurements were performed at baseline and after<br>14 days. The results of the dimensional inspection showed that the four<br>critical dimension measurements remained the same before and after<br>incubation. The results of the visual inspection demonstrated that the surface<br>finish on the incubated stents maintained the same quality as prior to<br>incubation. The data also show that the coating on the stents remains intact<br>after incubation. |
| Pressure/Flow<br>Characteristics | Numerical modeling, including computational fluid dynamics, was used to<br>evaluate the flow through the stents over physiologically relevant boundary<br>conditions. The stents were found to have negligible flow resistance. |
| Structural Integrity | A study was undertaken to evaluate the stress levels during the highest<br>anticipated load conditions for the stent by Finite Element Analysis (FEA).<br>Based on the modeling data, it was determined that the safety factors at the<br>lowest and highest implant velocities were 41x and 14x, respectively. The<br>results confirm that the stent will maintain its structural integrity after<br>implantation with the velocity range seen clinically. |
| Insertion Testing | As part of the shelf life testing for the iS3 injector, functional testing was<br>performed at baseline and after 1 year of aging to demonstrate that all specified<br>requirements were met e.g. stent delivery, stent singulation, stent implantation<br>and trocar penetration in synthetic tissue. All tested injectors successfully<br>passed all predetermined acceptance criteria for stent delivery. |
| Stability of Coating | Stability of the stearalkonium heparin coating on the stent was demonstrated<br>for the shelf life period of the finished, sterile device. |
| MRI Compatibility | Non-clinical testing has demonstrated that the iStent infinite Trabecular Micro-<br>Bypass System (Model iS3) is MR Conditional. A patient with this device can<br>be safely scanned in an MR system meeting conditions specified in the IFU<br>and patient implant card. |
| Corrosion Resistance | Glaukos submitted samples that were representative of the finished, sterile<br>titanium stent to a contract laboratory for electrochemical evaluation in<br>accordance with ASTM (American Society For Testing and Materials) F2129-<br>15, Standard Test Method for Conducting Cyclic Potentiodynamic Polarization<br>Measurements to Determine the Corrosion Susceptibility of Small Implant<br>Devices". The test lab concluded that test samples displayed acceptable<br>corrosion resistance to pitting and crevice corrosion in the ASTM F2129 test in<br>the received condition. |
Sterilization Validation:
The gamma irradiation sterilization method was validated using the VDmx 35 method described in ISO 11137-1:2015 and ISO 11137-2:2015. Validation results demonstrate that a minimum
{8}------------------------------------------------
exposure dose of 25 kGy has been substantiated for the routine sterilization of the iStent infinite Trabecular Micro-Bypass System Model iS3 to provide a 10-6 sterility assurance level (SAL).
### Bacterial Endotoxin
Endotoxin Limulus amebocyte lysate (LAL) testing has been performed as recommended in the FDA guidance document "Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices."
### Packaging and Shelf Life Testing:
The iStent infinite Trabecular Micro-Bypass System Model iS3 is labeled with an expiration date of 1 year. The shelf life study evaluated the functional performance of the G2-W Stent and the iS3 injector, as well as the packaging integrity of the tray sealed with the Tyvek lid. Additional testing was completed to evaluate the impact of environmental conditioning and distribution factors. Test results confirm that the G2-W Stent and the iS3 injector meet their functional requirements and the sterile barrier (package integrity) remains intact after simulated distributionand aging. This testing provides the justification for the 1-year shelf life and the maintenance of the sterile barrier.
### Biocompatibility Testing
The biocompatibility testing outlined in the tables below (Table 4) was performed on the stent (or representative samples of the finished device) and the patient-contacting portion of the injector in accordance with the relevant parts of International Organization for Standardization (ISO) standard 10993. All testing demonstrated that the device materials have an acceptable biocompatibility profile. With respect to physico-chemical testing of the stent, there is an extensive history of titanium use in medical devices. Therefore, tests for extraction in aqueous and organic solvents and for hydrolytic stability were not performed on the device, since they were considered unnecessary. In addition, the device's titanium material contains no monomers and is not subject to hydrolytic degradation.
{9}------------------------------------------------
| Test | Purpose | Acceptance<br>Criteria | Results |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|---------|
| Cytotoxicity: | | | |
| ISO Inhibition Of Cell<br>Growth | To determine the potential biological reactivity of a<br>mammalian cell culture (L929) in response to the test<br>article extract | Cell growth<br>inhibition <<br>30% | Pass |
| ISO L929 MEM<br>Elution Test | To determine the biological reactivity of a<br>mammalian cell culture (L929) in response to the test<br>article extract | No cell lysis or<br>toxicity | Pass |
| Agar Diffusion Test | To determine the biological reactivity of a<br>mammalian monolayer cell culture (L929) in<br>response to the test article | No cell lysis or<br>toxicity | Pass |
| Genotoxicity: | | | |
| Bacterial Reverse<br>Mutation Study | To evaluate the potential of the test article to induce<br>reverse mutations in histidine and tryptophan genes<br>in <i>S. typhimurium</i> and <i>E. coli</i> respectively | No mutagenic<br>changes | Pass |
| Mouse Bone Marrow<br>Micronucleus Study | To determine the ability of the test article and/or its<br>metabolites to induce micronuclei in maturing<br>erythrocytes of mice | No toxicity or<br>mutagenic<br>effects | Pass |
| In Vitro Chromosomal<br>Aberration Study | To determine the ability of the test article to induce<br>chromosome aberrations, structural or numerical, in<br>CHO cells in the presence or absence of an<br>exogenous mammalian activation system | No<br>chromosomal<br>aberrations<br>induced | Pass |
| Other: | | | |
| Intraocular Irritation<br>Study in the Rabbit | To evaluate the potential of the test article extract to<br>cause intraocular irritation or toxicity following an<br>intracameral injection in rabbits | No evidence of<br>irritation | Pass |
| Guinea Pig Kligman<br>Maximization Test | To evaluate the allergenic potential or sensitizing<br>capacity of the test article | No evidence of<br>delayed dermal<br>contact<br>sensitization | Pass |
| Muscle Implantation<br>in the Rabbit (2, 4, and<br>13 Weeks) | To evaluate the test article for local tissue responses<br>and the potential to induce local toxic effects after<br>implantation | No significant<br>reaction | Pass |
| Acute Systemic<br>Toxicity in the Mouse | To evaluate the test article extracts for potential toxic<br>effects following a single-dose systemic injection in<br>mice | No evidence of<br>systemic<br>toxicity | Pass |
| USP Material-<br>Mediated Rabbit<br>Pyrogen Study | To determine the potential presence of chemical<br>pyrogens in extracts of the test article | Non-pyrogenic | Pass |
| Test | Purpose | Acceptance<br>Criteria | Results |
| Cytotoxicity: | | | |
| ISO Medium Eluate<br>Method Test (1x CMEM<br>Extract) | To determine the biological reactivity of a mammalian<br>cell culture (L929) in response to the test article extract | No cell lysis or<br>toxicity | Pass |
| ISO Agar Diffusion Test<br>(Solid Sample) | To determine the biological reactivity of a mammalian<br>monolayer cell culture (L929) in response to the test<br>article | No cell lysis or<br>toxicity | Pass |
| Other: | | | |
| Intraocular Irritation<br>Test (Phosphate<br>Balanced Saline Extract) | To evaluate the potential of the test article extract to<br>cause intraocular irritation or toxicity following an<br>intracameral injection in rabbits | No evidence of<br>irritation | Pass |
| Guinea Pig Kligman<br>Maximization Test<br>(Saline & Vegetable Oil<br>Extracts) | To evaluate the allergenic potential or sensitizing<br>capacity of the test article | No evidence of<br>delayed dermal<br>contact<br>sensitization | Pass |
| Rabbit Intracutaneous<br>Reactivity/Irritation Test<br>(Saline & Vegetable Oil<br>Extracts) | To evaluate the test article for potential irritation effects<br>as a result of an intracutaneous injection in New Zealand<br>White rabbits | Non-irritating | Pass |
| Acute Systemic Toxicity<br>in the Mouse (Saline &<br>Vegetable Oil Extracts) | To evaluate the test article extracts for potential toxic<br>effects following a single-dose systemic injection in mice | No evidence of<br>systemic toxicity | Pass |
| Rabbit (Material-<br>Mediated) Pyrogen Test<br>(Normal Saline Extract) | To determine the potential presence of chemical pyrogens<br>in extracts of the injector test article using the ISO Rabbit<br>Pyrogen Test (Material Mediated) procedure. | Non-pyrogenic | Pass |
Table 3. Biocompatibility Testing - Stent
{10}------------------------------------------------
| | Table 4. Biocompatibility Testing - Injector | | |
|--|----------------------------------------------|--|--|
|--|----------------------------------------------|--|--|
#### B. Clinical Performance Testing
A prospective, multi-center, single-arm, open-label, clinical trial was conducted at 14 sites in the US and one OUS site to evaluate the safety and effectiveness of the iStent infinite in adult patients with primary open-angle glaucoma, pseudoexfoliative or pigmentary glaucoma with open angles, and in whom previous medical and surgical treatment has failed. 61 participants were implanted with the iStent infinite and followed post-operatively for 12 months. The two effectiveness endpoints were 1) the proportion of participants achieving a 20% or greater mean diurnal IOP (MDIOP) reduction from baseline at 12 months on the same or fewer number of medication classes and 2) the change from baseline in MDIOP at 12 months. No washout of IOP-lowering medications was performed before implantation. Safety outcomes included adverse events (AEs), surgical complications, best spectacle corrected visual acuity (BSCVA), and ocular findings from slit-lamp biomicroscopy, fundus examination, gonioscopy, pachymetry, and visual field testing.
# Demographics and Preoperative Characteristics
61 participants were implanted. The mean age of participants was 71.7 years (median 71.0, range 49 to 88) and there were 28 men (28/61 or 45.9%) and 33 women (33/61 or 54.1%). 37 of 61 (60.7%) participants were White, 15 of 61 (24.6%) were Black, six of 61 (9.8%) were Asian: race
{11}------------------------------------------------
was not reported for three of 61 (4.9%) subjects. 11 of 61 participants (18.0%) had ethnicity reported as Hispanic or Latino. 55 participants (90.1%) were diagnosed with primary open angle glaucoma (POAG), three (4.9%) had pseudoexfoliative glaucoma, and three (4.9%) had pigmentary glaucoma. All 61 participants had undergone prior filtering or cilioablative glaucoma procedures. Preoperatively, the mean visual field mean deviation (MD) score was -15.1 (SD 8.56) dB (median -13.7 dB, range -31.82 to -1.79 dB). The mean baseline medicated MDIOP was 23.5±2.82 mm Hg (median 22.7 mm Hg, range 20-35 mm Hg). At baseline, participants were using a mean of 3.0 (± 0.9) ocular hypotensive medications, with 19 of 61 (31.1%) on two or fewer medications and 42 of 61 (68.9%) on three or more medications.
### Effectiveness Results
Tables 5 and 6 summarize the primary effectiveness analyses based on 12-month diurnal IOP data.
| Analysis Population/Imputation Method for Missing Data | iStent infinite<br>n/N (%)<br>(95% CI)<br>N= 61 |
|--------------------------------------------------------------------------------|-------------------------------------------------|
| ITT Population/Worst Postoperative IOP & Last Available<br>Medication Classes1 | 44/61 (72.1%)<br>(59.2%, 82.9%) |
| ITT Population/Failure Assumption2 | 44/61 (72.1%)<br>(59.2%, 82.9%) |
| PP Population | 43/59 (72.9%)<br>(59.7%, 83.6%) |
| ITT Population/Exclusion from Cohort3 | 44/60 (73.3%)<br>(60.3%, 83.9%) |
| ITT Population/Multiple Imputation4 | 73.4%<br>(62.2%, 84.6%) |
# Table 5. Analyses of Responder Effectiveness Endpoint Proportion of Responders at Month 12
Participants with hypotony (IOP < 6 mmHg) associated with clinically significant findings, loss of light perception, IOP-related SSIs, cyclodialysis cleft, and/or no stents visible were treated as non-responders.
- 1. Responder status for the single participant with missing data at Month 12 was determined using the worst postoperative IOP and the last available number of medication classes.
- 2. The single participant who missed the 12-month evaluation was treated as a non-responder.
- 3. The single participant who missed the 12-month evaluation was excluded.
- 4. Multiple imputation was used for the single participant with missing data at Month 12.
{12}------------------------------------------------
| Analysis Population/Imputation Method for Missing Data | iStent infinite<br>N<br>Mean ± SD<br>(95% CI) |
|--------------------------------------------------------|-----------------------------------------------|
| ITT Population/Worst Postoperative IOP1 | 61<br>$-5.5 \pm 5.24$<br>(-6.9, -4.2) |
| PP Population | 59<br>$-5.5 \pm 5.29$<br>(-6.9, -4.1) |
| ITT Population/Exclusion from Cohort2 | 60<br>$-5.6 \pm 5.27$<br>(-6.9, -4.2) |
| ITT Population/Multiple Imputation3 | 61<br>$-5.5 \pm 0.67$ 4<br>(-6.9, -4.2) |
## Table 6. Analyses of MDIOP Change from Baseline Effectiveness Endpoint 12-Month Diurnal IOP Change from Baseline
1. The worst postoperative IOP was used as the 12-month MDIOP for the single participant who missed the 12-month evaluation.
2. The single participant who missed the 12-month evaluation was excluded.
3. Multiple imputation was used for the single participant with missing data at Month 12.
4. Mean ± SE for this value.
### Safety Results
Ocular adverse events (AEs) are summarized in Table 7. There were no intraoperative adverse events (AEs). Five of 61 participants (8.2%) needed the use of a second injector due to nondeployment of the second or third stent, and there was also head movement in one of these 5 participants. There were no reports of corneal decompensation, choroidal effusion, choroidal hemorrhage, hypotony maculopathy, deep stents ("buried" in the trabecular meshwork) that were not visible at the last three scheduled visits of the study, stent explantation, stent dislocation, or stent repositioning. The most common AEs reported were ocular surface disease, substantial increase in IOP vs. baseline, and loss of BSCVA ≥ 2 lines. Stent obstruction occurred in two of 61 participants (3.3%). Two instances of stent migration occurred in one participant (1.6%). Three of 61 participants (4.9%) required secondary surgical intervention (implantation of aqueous shunt) to lower IOP.
{13}------------------------------------------------
| | iStent infinite N = 61<br>Number (Percent) of<br>Subjects with Event |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Postoperative Adverse Event | |
| A significant increase in crystalline lens opacity from baseline defined as a<br>change of ARLNS grade of three half-step increments of 0.5 per increment or<br>greater for nuclear opalescence, cortical or posterior subcapsular opacities (as<br>applicable to phakic eyes) | 0 (0.0%) |
| Age-related macular degeneration | 0 (0.0%) |
| Allergic reaction | 0 (0.0%) |
| Anterior chamber fill | 0 (0.0%) |
| Anterior chamber tap | 0 (0.0%) |
| An increase of three half-step increments of 0.5 per increment or greater in<br>anterior subcapsular opacities or a clinically significant cataract eligible for<br>phacoemulsification with BCVA loss (ETDRS) of greater than 10 letters from<br>baseline (as applicable to phakic eyes) | 0 (0.0%) |
| Aqueous misdirection | 0 (0.0%) |
| Atrophy/phthisis | 0 (0.0%) |
| Bleb leak | 0 (0.0%) |
| Blepharitis | 3 (4.9%) |
| Chalazion | 0 (0.0%) |
| Choroidal effusion | 0 (0.0%) |
| Choroidal hemorrhage | 0 (0.0%) |
| Chronic pain in the study eye present greater than 3 months postoperative | 0 (0.0%) |
| Clinically significant cystoid macular edema | 0 (0.0%) |
| Conjunctival erosion due to tube shunt | 1 (1.6%) |
| Conjunctivitis | 1 (1.6%) |
| Corneal abrasion | 0 (0.0%) |
| Deep stents ("buried" in the trabecular meshwork) that are not visible at the last<br>three scheduled visits of the study | 0 (0.0%) |
| Dellen | 0 (0.0%) |
| Disc hemorrhage | 1 (1.6%) |
| Dysesthetic bleb | NA |
| Elevated IOP¹ | 1 (1.6%) |
| Endophthalmitis | 0 (0.0%) |
| Fixed dilated pupil | 0 (0.0%) |
| Flat or shallow anterior chamber (e.g., shallowing of the anterior chamber that<br>causes any amount of iris-cornea touch) | 0 (0.0%) |
| Hyperemia | 2 (3.3%) |
| Hypotony (IOP < 6 mmHg) associated with clinically significant findings | 1 (1.6%) |
| Implant exposure | 0 (0.0%) |
| Implant migration | 0 (0.0%) |
| IOP increase >= 10 mmHg vs. baseline IOP¹ | 5 (8.2%) |
| IOP increase requiring oral medication¹ | 2 (3.3%) |
| IOP increase requiring surgical intervention¹ | 3 (4.9%) |
| Increase in C/D ratio of > 0.3 units on ophthalmoscopic examination | 0 (0.0%) |
| Intraocular inflammation arising after the protocol's specified medication regimen<br>is complete | 1 (1.6%) |
| Intraocular inflammation following tube shunt surgery | 2 (3.3%) |
| Iridodialysis | 0 (0.0%) |
| Lens/IOL dislocation | 0 (0.0%) |
| Loss of best spectacle corrected visual acuity (BSCVA) of 2 lines or more | 7 (11.5%) |
| <= 30 days | 1 (1.6%) |
| > 30 days² | 6 (9.8%) |
| Loss of eye | 0 (0.0%) |
| Macular edema | 2 (3.3%) |
| Macular puckering | 0 (0.0%) |
| Nd:YAG capsulotomy | 0 (0.0%) |
| Needling procedure | NA |
| iStent infinite N = 61 | |
| | Number (Percent) of<br>Subjects with Event |
| Postoperative Adverse Event | |
| Ocular hypotensive medication intolerance | 3 (4.9%) |
| Ocular pain | 1 (1.6%) |
| Ocular surface disease | 7 (11.5%) |
| Perioperative inflammation | 4 (6.6%) |
| Posterior vitreous detachment | 1 (1.6%) |
| Proliferative vitreoretinopathy | 0 (0.0%) |
| Ptosis | 0 (0.0%) |
| Pupillary block | 0 (0.0%) |
| Retinal detachment | 0 (0.0%) |
| Retinal dialysis | 0 (0.0%) |
| Retinal flap tears | 0 (0.0%) |
| Secondary surgical intervention | 3 (4.9%) |
| Significant corneal complications including opacification and decompensation | 0 (0.0%) |
| Significant corneal edema | 0 (0.0%) |
| Significant corneal injury | 0 (0.0%) |
| Significant damage to trabecular meshwork | 0 (0.0%) |
| Significant hyphema (i.e, >= 10% of anterior chamber) | 2 (3.3%) |
| Significant iris damage | 0 (0.0%) |
| Stent dislocation | 0 (0.0%) |
| Stent explant | 0 (0.0%) |
| Stent migration3 | 1 (1.6%) |
| Stent obstruction4 | 2 (3.3%) |
| Stent-cornea touch | 0 (0.0%) |
| Stye | 1 (1.6%) |
| Subconjunctival hemorrhage | 1 (1.6%) |
| Toxic Anterior Segment Syndrome (TASS) | 0 (0.0%) |
| Transient hypotony…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.