← Product Code [EIH](/productcode/EIH) · K220017

# Coloring Liquid (K220017)

_Dongguan Xiangtong Co., Ltd. · EIH · Jun 29, 2022 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220017

## Device Facts

- **Applicant:** Dongguan Xiangtong Co., Ltd.
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Jun 29, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Coloring Liquid is a liquid used for the complete or partial coloration of milled zirconia substructure and anatomy before sintering.

## Device Story

Coloring Liquid is a water-based, acidic metal salt solution used in dental laboratories. It is applied to milled zirconia dental frameworks or restorations via immersion or brushing prior to the final sintering process. The liquid contains inorganic pigments that penetrate the zirconia to achieve specific shades, allowing the final restoration to match the patient's natural tooth color. The device is intended for professional use by dental technicians. By enabling precise color matching before the high-temperature sintering phase, the device helps ensure the aesthetic quality of the final dental prosthesis.

## Clinical Evidence

No clinical data. Bench testing and biocompatibility testing were performed. Biocompatibility testing included cytotoxicity (ISO 10993-5), oral mucosa irritation (ISO 10993-10), sensitization (ISO 10993-10), acute/subchronic systemic toxicity (ISO 10993-11), and genotoxicity (ISO 10993-3). All tests met acceptance criteria.

## Technological Characteristics

Water-based solution containing inorganic pigments, polyethylene glycol, and inorganic salts. Non-sterile. Applied via brush or immersion. Designed for use with zirconia ceramic materials prior to high-temperature sintering.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Upcera Coloring Liquid (I and II) ([K141723](/device/K141723.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 29, 2022

Dongguan Xiangtong Co., Ltd. Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland, New Jersey 07436

Re: K220017

Trade/Device Name: Coloring Liquid Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: May 28, 2022 Received: June 3, 2022

Dear Charles Shen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K220017

Device Name Coloring Liguid

Indications for Use (Describe)

Coloring Liquid is a liquid used for the complete or partial coloration of milled zirconia substructure and anatomy before sintering.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### K220017 Section 5: 510(k) Summary:

This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92

#### Submitter & Foreign Manufacture Identification 5.1

DONGGUAN XIANGTONG CO., LTD NO.4. Tech 9th Rd, Hi-Tech Industrial Development Zone, Songshan Lake, Dongguan, Guangdong, China., Zipcode 523000 Tel: (086)-0769-22895688 Submitter's FDA Registration Number: N/A

#### Contact Person 5.2

Image /page/3/Picture/6 description: The image shows a logo for a company that appears to be in the medical field. The logo features a stylized letter 'M' in blue, with a crescent shape above it in light blue. The company name, "曼度医械", is written in orange above the 'M'. The overall design is simple and modern, with a focus on the company's initials.

Charles Shen Manton Business and Technology Services 37 Winding Ridge, Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com

#### 5.3 Date of Summary: December 19, 2021

| 5.4 | Device Name: |
|-----|--------------|
|-----|--------------|

| Proprietary Name:      | Coloring Liquid   |
|------------------------|-------------------|
| Common Name:           | Coloring Liquid   |
| Classification Name:   | Powder, Porcelain |
| Device Classification: | II                |
| Regulation Number:     | 21 CFR 872.6660   |
| Panel:                 | Dental            |
| Product Code:          | EIH               |

#### 5.5 Predicate Device Information:

- K141723, "Upcera Coloring Liquid (I and II)", manufactured by "Liaoning (1) Upcera Co., Ltd." located in Benxi, China
#### 5.6 Device Description:

Coloring Liquid is water based, which consist of watery, acidic metal salt solutions. They are used for the individual staining of dental zirconia frameworks and restorations prior to the final sintering of the restoration, enabling the user to adjust the restoration to match the natural color of the patient's teeth.

{4}------------------------------------------------

Coloring Liquid is provided in 45 different shades, which are corresponding to patient's closest tooth color. For staining, the zirconia materials have to be immerged into the liquids or to be brushed with the liquids, prior to sintering at high temperatures

#### 5.7 Indications for Use:

Coloring Liquid is a liquid used for the complete or partial coloration of milled zirconia substructure and anatomy before sintering.

#### న్ని శిశు సంరక్షణ కేంద్రం, మూలాలు గ్రామం నుండి 10 కి.మీ. దూరంలో ఉంది. ప్రాథమిక ఆరోగ్యామ్ నిర్మాణ ప్రాథమిక ఆరోగ్యామ్ కూడా ఉంది. మాతా శిశు సంరక్షణ కేంద్రం, మూలాలు విశాఖపట్న Summary of Device Testing:

Bench testing was performed to ensure that the Coloring Liquid met its specifications. All tests were verified to meet acceptance criteria. Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.

#### 5.9 Comparison With Predicate Device

Coloring Liquid is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

- (1) K141723, "Upcera Coloring Liquid (I and II)", manufactured by "Liaoning Upcera Co., Ltd." located in Benxi, China
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.

| Description            | Subject Device                                                                                                                                 | Predicate Device (K141723)                                                                                                                                                 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use  | Coloring Liquid is a liquid used for the<br>complete or partial coloration of milled<br>zirconia substructure and anatomy before<br>sintering. | Upcera Coloring Liquid (I and II) is a<br>liquid used for the complete or partial<br>coloration of milled Upcera zirconia<br>substructure and anatomy before<br>sintering. |
| Technology             | Water based with inorganic pigments                                                                                                            | Water based with inorganic pigments                                                                                                                                        |
| Operating<br>Principle | Brush or immerse zirconia ceramic materials<br>with coloring liquid before sintering                                                           | Brush or immerse zirconia ceramic<br>materials with coloring liquid before<br>sintering                                                                                    |
| Ingredient             | Water, Polyethylene glycol, inorganic salts                                                                                                    | Water, Polyethylene glycol, inorganic<br>salts                                                                                                                             |
| Bottle Size            | Various                                                                                                                                        | Various                                                                                                                                                                    |
| Shade                  | Various                                                                                                                                        | Various                                                                                                                                                                    |
| Prescription           | Yes                                                                                                                                            | Yes                                                                                                                                                                        |

Table 5.1: Comparison of Intended Use, Design, Material, and Processing

{5}------------------------------------------------

| Sterile | Non-sterile | Non-sterile |
|---------|-------------|-------------|
|---------|-------------|-------------|

The subject device is similar to the predicate device in terms of indications for use, technology, and principle of operation.

The following table shows similarities and differences of the safety tesing between our device and the predicate devices.

## Table 5.2: Comparison of Safety Testing

| Description                                                                | Subject Device                                                             | Predicate Device (K141723)                                                 |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Cytotoxicity<br>(ISO 10993-5:2009)                                         | No cytotoxicity effect                                                     | No cytotoxicity effect                                                     |
| Irritation Oral Mucosa<br>Irritation<br>(ISO 10993-10: 2010)               | Not a primary oral mucosa<br>irritant under the conditions of<br>the study | Not a primary oral mucosa<br>irritant under the conditions of<br>the study |
| Sensitization (ISO<br>10993-10: 2010)                                      | Not a sensitizer under the<br>conditions of the study                      | Not a sensitizer under the<br>conditions of the study                      |
| Acute Systemic Toxicity<br>and Subchronic Toxicity<br>(ISO 10993-11: 2006) | No Acute Systemic Toxicity and<br>subchronic toxic effects<br>observed     | No acute and subchronic toxic<br>effects observed                          |
| Genotoxicity (ISO<br>10993-3:2003)                                         | No genotoxic effects observed                                              | No genotoxic effects observed                                              |

#### 5.10 Non-clinical Testing

Non-clinical Testing Bench testing was performed to ensure the Coloring Liquid met its specifications. All tests were verified to meet acceptance criteria. Biocompatibility testing was performed to verify the substantially equivalent safety of the materials that are used.

#### 5.11 Conclusions

It has been shown in this 510(k) submission that "Coloring Liquid" and its predicate devices have the identical indications for use, similar composition and biocompatibility, similar manufacturing process, and similar performance.

The difference between the "Coloring Liquid" and their predicate device do not raise any question regarding its equivalence.

"Coloring Liquid", as designed and manufactured, is substantially equivalent as its predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K220017](https://fda-staging.innolitics.com/device/K220017)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
