K220012 · Bresotec, Inc. · MNR · Dec 21, 2022 · Anesthesiology
Device Facts
Record ID
K220012
Device Name
BresoDX1
Applicant
Bresotec, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Dec 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
BresoDX1 is indicated for use as an aid in the diagnosis of moderate to severe sleep apnea in adult patients. BresoDX1 records a patient's physiological signals during sleep and scores apneas and hypopneas. BresoDX1 is intended to be used in the home and clinic settings.
Device Story
BresoDX1 is a wearable data acquisition system for home or clinic sleep apnea diagnosis. Inputs include tracheal breathing sounds, tracheo-sternal movement, and body position (via BresoSensor on suprasternal notch) plus SpO2 and heart rate (via wrist-worn Nonin WristOx2 pulse oximeter). Data transmits wirelessly to a bedside BresoHub for storage and cloud upload. Proprietary software processes signals to estimate airflow and respiratory effort, generating waveforms and sleep study reports. Healthcare professionals review reports, edit automated scoring, and re-generate final interpretations. System aids diagnosis of moderate to severe sleep apnea by providing objective respiratory event data.
Clinical Evidence
Prospective, single-arm, multi-site, blinded study (n=164) comparing BresoDX1 to in-lab polysomnography (PSG). Results for REIb-based diagnosis (AHI threshold 15): Sensitivity 88.5% (95% CI: 79.2%-94.6%), Specificity 84.9% (95% CI: 75.5%-91.7%), PPV 84.1%, NPV 89.0%. Correlation with PSG: REIb vs AHI(PSG) r=0.93; ODI(BresoDX1) vs ODI(PSG) r=0.95.
Technological Characteristics
Wearable system comprising BresoSensor (microphone, 3D accelerometer), BresoHub, and Nonin WristOx2 oximeter. Materials: plastic housing and adhesive tape (ISO 10993-5/10 compliant). Connectivity: Wi-Fi/Bluetooth. Power: Rechargeable Li-ion (sensor), wall adapter (hub), alkaline batteries (oximeter). Software: Cloud-based analysis, proprietary algorithm for respiratory event scoring. Sterilization: Nonsterile.
Indications for Use
Indicated for adult patients with suspected moderate to severe sleep apnea. Used as an aid in diagnosis by recording physiological signals during sleep to score apneas and hypopneas.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 21, 2022
Bresotec Inc Mehrnaz Tabibi Regulatory and Quality Director 55 York Street Unit 200 Toronto, Ontario M5J 1R7 Canada
Re: K220012
Trade/Device Name: BresoDX1 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: November 21, 2022 Received: November 21, 2022
Dear Mehrnaz Tabibi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K220012
Device Name
#### BresoDX1
Indications for Use (Describe)
BresoDX1 is indicated for use as an aid in the diagnosis of moderate to severe sleep apnea in adult patients. BresoDX1 records a patient's physiological signals during sleep and sores and hypopneas. BresoDX1 is intended to be used in the home and clinic settings.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. '
{3}------------------------------------------------
### K220012 - 510(k) SUMMARY
Submitter
Bresotec, Inc. 55 York Street Suite 200 Toronto, Ontario, M5J 1R7 Canada
Phone: +1 (647)930-1777 Facsimile: +1 (437)800-1139 Contact Person: Mehrnaz Tabibi Date Submitted: January 4, 2022
Name of Device: BresoDX1
Common or Usual Name: Ventilatory Effort Recorder
Classification Name: 21 C.F.R. 868.2375, Breathing Frequency Monitor
Regulatory Class: II
Product Code: MNR
Predicate Device: ACCUSOM manufactured by NovaSom, Inc. (K110486)
Referenced Devices: WatchPat200U, manufactured by Itamar Medical, Ltd. (K203839), and AcuPebble SA100 manufactured by Acurable Limited (K210480)
#### Device Description
BresoDX1 is a lightweight data acquisition system intended for use by Health Care Professional (HCP) as an aid in the diagnosis of moderate to severe sleep apnea for adults. It is suitable for independent use in unattended settings, like in the home, or with minimal assistance in clinic settings.
BresoDX1 is to be worn while sleeping and consists of: (1) a BresoSensor which is placed on the patient's suprasternal notch with a disposable BresoSensor Frame (2) a BresoHub, which is located at the bedside; (3) a Nonin WristOx2® pulse oximeter, which is worn on the patient's wrist, where the oximeter sensor is to be placed on the patient's finger; and (4) BresoDX1 software.
The BresoSensor and the Nonin WristOx2® pulse oximeter sensor will record patient's tracheal breathing sounds, neck/body position and movement, tracheo-sternal movement, arterial oxyhemoglobin saturation and heart rate overnight during sleep. Recorded data on the BresoSensor and oximeter will be sent to the BresoHub wirelessly via Wi-Fi and Bluetooth for pre-processing and storing. When the patient returns the BresoDX1 system to the health care professional, stored data that had been uploaded will be processed for analysis and sleep study report generation. The generated report will show the following channels: heart rate, SpO2, airflow, respiratory effort, and body position. The health care professional will be able to use the software to review data, edit automated scoring, and re-generate a sleep study report based on the edited scoring for further interpretation. The health care professional will reprocess the device after each use.
{4}------------------------------------------------
## Intended Use / Indications for Use
BresoDX I is indicated for use as an aid in the diagnosis of moderate to severe sleep apnea in adult patients. BresoDX1 records a patient's physiological signals during sleep and scores apneas and hypopneas. BresoDX1 is intended to be used in the home and clinic settings.
## Comparison of Technological Characteristics
| Comparison of technological characteristics with predicate and reference devices | | |
|----------------------------------------------------------------------------------|--|--|
|----------------------------------------------------------------------------------|--|--|
| | Bresotec's<br>BresoDX1 | NovaSom's<br>ACCUSOM<br>(Predicate) | Itamar' s<br>WatchPat200U<br>(Reference Device) | AcuPebble SA100<br>(Reference Device) | Comparison |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory<br>Class,<br>Product<br>Code | Class II, MNR | Class II, MNR | Class II, MNR | Class II, MNR | |
| 510k # | K220012 | K110486 | K203839 | K210480 | |
| Intended<br>Use /<br>Indications<br>for Use | The BresoDX1 is<br>indicated for use as<br>an aid in the<br>diagnosis<br>of moderate to severe<br>sleep apnea in adult<br>patients. The<br>BresoDX1 records a<br>patient's<br>physiological signals<br>during sleep and<br>scores apneas and<br>hypopneas. The<br>BresoDX1 is<br>intended to be used in<br>the home and clinic<br>settings. | The ACCUSOM device<br>is indicated for use in the<br>diagnostic evaluation of<br>adults with possible<br>sleep apnea. The<br>ACCUSOM can score<br>obstructive apneas,<br>which includes mixed<br>apneas.<br>The ACCUSOM device<br>is intended for use in the<br>home and clinic setting. | The Watch-PAT200U<br>(WP200U) device is a<br>non-invasive home care<br>device for use with<br>patients suspected to have<br>sleep related breathing<br>disorders. The WP200U is<br>a diagnostic<br>aid for the detection of<br>sleep related breathing<br>disorders, sleep staging<br>(Rapid Eye<br>Movement (REM) Sleep,<br>Light Sleep, Deep Sleep<br>and Wake), snoring level<br>and body<br>position. The WP200U<br>generates a peripheral<br>arterial tonometry<br>("PAT") Respiratory<br>Disturbance index<br>("PRDI"), Apnea-<br>Hypopnea index<br>("PAHI"), Central<br>Apnea-<br>Hypopnea index<br>("PAHIc"), PAT sleep<br>staging identification<br>(PSTAGES) and optional<br>snoring level and body<br>position discrete states<br>from an external<br>integrated snoring and<br>body position sensor. The<br>WP200U's PSTAGES and<br>snoring level and body<br>position<br>provide supplemental<br>information to its<br>PRDI/PAHI/PAHIc. The<br>WP200U's PSTAGES<br>and snoring level and<br>body position are not | AcuPebble SA100 is<br>indicated to sense, record,<br>and interpret a patient's<br>physiological signals<br>(including respiratory<br>pattern) during sleep for the<br>purpose of pre-screening<br>patients for obstructive<br>sleep apnea (OSA)<br>syndrome. The device is<br>designed for use in home-<br>screening of adults with<br>suspected possible sleep<br>breathing disorders<br>(although it can also be<br>used in clinic). Results are<br>used to assist the healthcare<br>professional in the patient's<br>evaluation.<br>The system is not intended<br>as a substitute for full<br>polysomnography when<br>additional parameters such<br>as sleep stages, limb<br>movements or EEG activity<br>are required. | Substantially equivalent to<br>predicate device.<br>BresoDX1 does not<br>differentiate between<br>obstructive and mixed<br>apneas. This difference<br>does not impact the ability<br>of BresoDX1 to provide<br>information to the user as<br>an aid in the diagnosis of<br>moderate to severe sleep<br>apnea. |
| | | | intended to be used as the<br>sole or primary<br>basis for diagnosing any<br>sleep related breathing<br>disorder, prescribing<br>treatment, or<br>determining whether<br>additional diagnostic<br>assessment is warranted.<br>PAHiC is indicated for use<br>in patients 17 years and<br>older. All other<br>parameters are indicated<br>for 12 years and older. | | |
| User<br>Population | Home and clinic<br>setting | Home and clinic setting | A non-invasive home care<br>device | For use in home-screening<br>(although it can be used in<br>clinic) | Substantially equivalent to<br>predicate device. |
| Sensors | -Microphone in<br>BresoSensor placed<br>on the suprasternal<br>notch<br>-Oximeter sensor<br>placed on finger -<br>Three-dimensional<br>accelerometer in<br>BresoSensor placed<br>on suprasternal notch | -Breath Sensor (sensing<br>microphone and ambient<br>noise microphone)<br>placed under the nose<br>-Oximeter finger sensor<br>placed on the finger<br>- Chest Sensor located<br>on the chest. | -uPAT finger<br>probe placed on the<br>finger<br>-Actigraphy on the wrist<br>- (optional) External<br>SBP/RESBP chest sensor<br>(microphone, three-<br>dimensional<br>accelerometer) placed on<br>the chest right under the<br>sternal notch | -Microphone placed on the<br>neck above the suprasternal<br>notch | Substantially Equivalent to<br>predicate device.<br>In addition, BresoDX1 and<br>reference device uses three-<br>dimensional accelerometer. |
| Channels | Airflow, respiratory<br>effort, tracheal breath<br>sounds tracheo-<br>sternal movement,<br>pulse rate, SpO2,<br>body position | Airflow, respiratory<br>sounds, pulse rate, SpO2,<br>snoring level | PAT, pulse rate, oximetry,<br>actigraphy, snoring, body<br>position, chest movement | Sounds (Cardiac features,<br>respiratory features,<br>movement features) | Substantially Equivalent to<br>predicate device.<br>BresoDX1 does not provide<br>the output of snoring levels.<br>Snoring is optional for<br>HSAT and is not essential<br>for the diagnosis of sleep<br>apnea.<br>The difference of system<br>output between BresoDX1<br>with its predicates does not<br>raise any uncertainty of<br>safety, effectiveness, and<br>efficacy. |
| Portability | Yes (wearable) | Yes (wearable) | Yes (wearable) | Yes (wearable) | Identical to predicate<br>device. |
| Data<br>Download | Transfer through<br>wireless connection | Transfer through<br>wireless connection | Transfer through USB<br>connection | Transfer through wireless<br>connection | Substantially equivalent to<br>predicate device. |
| Size and<br>Weight of<br>sensor | BresoSensor:<br>17.4×38.7×35.5 mm<br>17g | Patient Module :<br>100.3×71.9×29.5 mm<br>116.2 g | Chest sensor:<br>1.3inch (32mm) diameter<br>12g | 29.5mm diameter × 16mm<br>height<br>7g | Different but difference in<br>size and weight do not raise<br>safety or effectiveness<br>question. |
| Power<br>Source | BresoHub: wall<br>mount power adaptor<br>BresoSensor:<br>Lithium-ion<br>rechargeable battery<br>Oximeter: 2 × 1.5V<br>Alkaline AAA<br>battery | Patient Module:<br>Rechargeable battery<br>operated | One OTS 1.5V Alkaline<br>AAA battery OR<br>One rechargeable AAA<br>1.2V Nickel-metal<br>hydride battery<br>rechargeable (NiMH)<br>battery | Rechargeable lithium<br>polymer battery | Different but difference in<br>power source do not raise<br>safety or effectiveness<br>question. |
| Indicators | Visual indicators<br>LEDs on<br>BresoSensor and<br>BresoHub.<br>Indicators on<br>BresoHub Web<br>Interface | Audio and visual<br>indicators | Visual indicator – error<br>messages | N/A | Different. BresoDX1 does<br>not have audio indicators,<br>but it has visual indicators.<br>This difference does not<br>raise safety or effectiveness<br>question. |
| Sterilization | Nonsterile | Nonsterile | Nonsterile | Nonsterile | Identical. |
{5}------------------------------------------------
{6}------------------------------------------------
BresoDX1 and its predicate device, ACCUSOM, utilize the same technology to monitor key signals. Both devices use a microphone to detect a patient's breathing pattern, and a pulse oximeter to determine a patient's arterial oxyhemoglobin saturation and pulse rate. BresoDX1 processes recorded sound data and accelerometer data using a proprietary software algorithm to estimate a subject's airflow changes and respiratory effort and to generate the corresponding waveforms.
## Software
BresoDX I Software collects and stores the data during a test. Data will be uploaded to the cloud. BresoDX1 Software will process and analyze the data to generate a test report. The test report and the data will be reviewed and interpreted by a Health Care Professional.
Some functionalities require connectivity with the cloud system. Those functionalities are as follows:
- i. Data upload
- ii. Data Storage
- iii. Data Analysis
- User and Study Management iv.
- V. Study Editor
- vi. Study Report Generator
The BresoDX1 software is developed according to Medical Device Software/ Software lifecycle processes: IEC 62304:2006/A12016. Also, BresoDX1 software is developed according to Information security, cybersecurity and privacy protection, Information security controls: ISO/ IEC 27002: 2022.
## Performance Data
A series of safety and functional testing was performed to confirm that BresoDX1 meets the requirements of the relevant standards. These tests include electrical safety testing according to IEC 60601-1:2005+A1:2012, CAN/CSA-C22.2 No. 60601-1:2014, ANSI/AAMI ES60601-1:2005/A1:2012, home healthcare environment testing according to IEC 60601-1-11:2015, usability according to IEC60601-1-6:2010+A1:2013, application of usability engineering to medical devices according to IEC 62366-1:2015 and IEC 62366-1:2015/COR1:2016, EMC testing according to IEC 60601-1-2:2014, FCC part 15 tests.
Biocompatibility of the patient contacting components (BresoDX1 plastic housing, tape on the BresoSensor Frame) in direct contact with patient's healthy closed skin for up to 8 hours were tested in accordance with ISO10993-1 and FDA's corresponding guidance document, Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Biocompatibility tests for these patient contacting components of the subject device complaint with the following standards:
{7}------------------------------------------------
- · Cytotoxicity per ISO 10993-5 (2009)
- · Irritation per ISO 10993-10 (2010)
- · Skin sensitization per ISO 10993-10 (2010)
Bench performance tests were performed to demonstrate that the system meets its performance requirements by verifying the device precision and evaluating the inter-device and intra-device repeatability and reproducibility in detecting sleep apnea events and evaluating the impact of environmental noises on the device performance. Oximetry input integration validation test was performed to evaluate the accuracy of the integration of Nonin WristOx2® Model 3150 BLE pulse oximeter with the BresoDX1 system.
All tests confirmed that the subject device meets the pre-determined acceptance criteria and the requirements of the relevant standards.
Clinical performance testing was performed in compliance with the recognized consensus standard ISO 14155:2020 (Clinical investigation of medical devices for human subjects-Good clinical practice to evaluate the capability of BresoDX1. A prospective, single-arm, multiple-site, blinded study was performed to validate the BresoDX1 in comparison with in-laboratory polysomnography (PSG). The study involved overnight recording of the BresoDX1 and clinical PSG in the sleep laboratory, simultaneously. One hundred and sixty-four (164) subjects were evaluated. Based on the clinical performance as documented in the pivotal clinical study, the BresoDX1 has a safety and effectiveness profile that is similar to the predicate device.
The BresoDX1 diagnosis accuracy from the clinical study is summarized as follows:
| | REIb based diagnosis*<br>n/Total, (95% Confidence Interval) |
|---------------------------|-------------------------------------------------------------|
| Sensitivity | 69/78, 88.5% (79.2%, 94.6%) |
| Specificity | 73/86, 84.9% (75.5%, 91.7%) |
| Positive Predictive Value | 69/82, 84.1% (74.4%, 91.3%) |
| Negative Predictive Value | 73/82, 89.0% (80.2%, 94.9%) |
* REIb is the total number of respiratory events scored × 60 divided by monitoring time (in minutes). The diagnostic criteria were based on the established threshold, AHI cut-off 15.
The REIn accuracy from the clinical study is summarized as follows:
Correlation between REIb and AHI(PSG) is 0.93 AHI(PSG) = $1.033 × REI_b - 0.0705$
REIb is Respiratory Event Index calculated by BresoDX1 software and AHI(PSG) is Apnea and Hypopnea Index calculated by clinical PSG.
The Oxygen Desaturation Index (ODI) accuracy from the clinical study is summarized as follows:
Correlation between ODI(BresoDX1) and ODI(PSG) is 0.95 ODI(PSG) = $1.334 × ODI_{(BresoDX1)} + 5.317$
ODI(BresoDX1) is Oxygen Desaturation Index calculated by BresoDX1 and ODI(PSG) is Oxygen Desaturation Index calculated by clinical PSG.
{8}------------------------------------------------
### SpO2 and Pulse Rate Measurement Accuracy
The accuracy of the Nonin 3150BLE SpO2 and pulse rate values in comparison to the co-oximeter samples measured over the SpO2 range of 70% - 100% are validated in accordance with Medical Electrical Equipment- ISO 80601-2-61:2017 Sub-clauses 201.12.1.103 & 201.12.1.104 Pulse Rate and SpO2 accuracy. Accuracy data is calculated using the root-mean-square ($A_{rms}$ value) for all data points using the following RMS figure of merit calculation:
$$A_{rms} = \frac{\sqrt{\Sigma_{i=1}^n (X_i - X_{Ri})^2}}{n}$$
$X_i$ is the oximeter 4 beat average reading and $X_{Ri}$ is the Oxitest setting at the corresponding sample time.
The accuracy of SpO2 ($A_{rms}$ value) is ±2% (non-motion, low perfusion) ±3% (motion) within the SpO2 range of 70% - 100%. The accuracy of pulse rate ($A_{rms}$ value) is ±3% (motion, non-motion, low perfusion) within the pulse rate range of 18-300 BPM (beats per minutes).
## Conclusions
BresoDX1 has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate device (K110486). Performance data demonstrated substantially equivalent performance as the ACCUSOM. Thus, BresoDX1 is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.