K220006 · A2 Medical Systems, LLC · FZP · Mar 2, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220006
Device Name
angioLOCK Polymer Ligating Clip
Applicant
A2 Medical Systems, LLC
Product Code
FZP · General, Plastic Surgery
Decision Date
Mar 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
angioLOCK® Non-Absorbable Polymer Ligation Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
Device Story
angioLOCK clips are permanent, non-absorbable, sterile, single-use surgical implants; used for vessel or tissue ligation. Clips are made of acetal homopolymer; supplied in color-coded cartridges. Surgeons use compatible manual appliers (open or laparoscopic) to load, position, and close clips. Mechanism involves haptic feedback when clip bosses seat in applier grooves; closing action pivots upper and lower curved legs with inner teeth via a hinged section to latch securely around tissue. Output is a mechanical ligation of the target structure. Device provides secure vessel occlusion; benefits include reliable hemostasis during surgery. Used in clinical settings by surgeons.
Clinical Evidence
Bench testing only. No clinical data. Performance established via dimensional analysis, biocompatibility testing, and qualitative closure testing comparing angioLOCK to the predicate Hem-o-lok. Testing confirmed identical geometry, material safety, and mechanical function (latching, applier compatibility, and release).
Technological Characteristics
Material: Tenac 5010 (polyacetal polymer). Form factor: Hinged clip with inner teeth, available in M/L, LG, and XLG sizes. Mechanism: Mechanical latching via applier. Sterilization: Ethylene Oxide (EO). Connectivity: None (mechanical).
Indications for Use
Indicated for ligation of vessels or tissue structures in surgical procedures. Contraindicated for use as a fallopian contraceptive tubal occlusion device and for ligating the renal artery during laparoscopic donor nephrectomies.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Teleflex Medical Hem-o-lok Ligating Clips (K030311)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 2, 2022
A2 Medical Systems, LLC Robert Hall Chief Operating Officer 6387 Technology Ave. Kalamazoo, Michigan 49009
Re: K220006
Trade/Device Name: angioLOCK Polymer Ligating Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP, DSS Dated: December 30, 2021 Received: January 4, 2022
Dear Robert Hall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K220006
Device Name
## angioLOCK Polymer Ligating Clip
Indications for Use (Describe)
angioLOCK® Non-Absorbable Polymer Ligation Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# SECTION 5 - 510(k) Summary
## 510(k) Number: _ Prior Formal Correspondence: None – Original Submission Prepared 12/29/2021
## Basis for the Submission: New Device
#### Name, Address, Phone and Fax Number of Applicant
Owner: Andy Ambro Phone #: 336-706-6027 Address: A2 Medical Systems, LLC 6387 Technology Ave. Suite F Kalamazoo, MI 49009
#### Contact Person
Robert Hall Chief Operating Officer Phone #: 919-538-2325 E-mail: robert@a2medicalsystems.com A2 Medical Systems, LLC Address: 6387 Technology Ave. Suite F Kalamazoo, MI 49009
#### Date Prepared
December 30 " 2021
#### A. Device Name
| Trade Name: | angioLOCK Polymer Ligating Clip |
|---------------------------|--------------------------------------|
| Common Name: | Hemostatic Clip, Ligating Clip |
| Classification Name: | Clip, Implantable |
| B. Device Classification: | Class II |
| Classification Panel: | Cardiovascular & General Surgery |
| Product Codes: | FZP per 878.4300<br>DSS per 870.3250 |
#### C. Predicate Devices
Teleflex Medical Hem-o-lok Ligating Clips (510k - K030311)
#### D. Clip Device Description
AngioLOCK ligating clips are permanent implant, non-absorbable, sterile single use, surgical clips made of implantable grade polymer and are available in multiple sizes.
ligating clips are non-absorbable, non-active implantable devices to be used for ligation of vessels and tissue structures. The clips are made of acetal homopolymer and are offered in three sizes; medium/large, large and x-large. Each clip size is compatible with a corresponding clip applier that may be designed for either general or endoscopic procedures. The clips are supplied in quantities of six according to size in color coded cartridges that are prepackaged sterile and single-use.
angioLOCK® open and Laparoscopic Ligation Clip Appliers are for use as delivery devices for angioLOCK® Non-Absorbable Polymer Ligation Clips. Other ligation clips cannot be used with these appliers. angioLOCK® Laparoscopic Polymer Ligation Clip Appliers are designed for use with specific size cannulas.
- نن Indications For Use: angioLOCK® Non-Absorbable Polymer Ligation Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
{4}------------------------------------------------
- Contraindications: angioLOCK® Non-Absorbable Polymer Ligation Clips are not intended for use as a ட் fallopian contraceptive tubal occlusion device. angioLOCK® Non-Absorbable Polymer Ligation Clips are contraindicated for use in ligating the renal artery during laparoscopic donor nephrectomies.
## G. Characteristic Comparison to Predicate Device Equivalence
The A2 Medical Systems polymer ligating clips are substantially equivalent to the predicate device Hem-o-lok in design and use.
- a. Substantially Equivalent areas are:
- i. Both are single use devices
- ii. Both will latch close over vessels
- iii. Both use the same clip Tenac 5010 material: nonabsorbable polyacetal polymer
- iv. Both have the same clinical application for indications and contraindications
- Both have the same mechanism of action cartridge and clip design allow for an applier to V. easily load a clip, haptic feedback occurs when the clip bosses locate in the applier grooves, and clips are then removed for use. When the clip is closing the upper and lower curved legs with inner teeth pivots in from a hinged section and haptic feedback occurs as clips are latched closed.
- vi. Of the four sizes provided M, M/L, LG, and XLG; they have the same geometry and beam leg height and leg thickness and round boss diameters
- vii. Both have cartridges that are color coded to match applier ring colors
- Both use traditional EO sterilization processing viiii.
- İX. Both use peel open blister packs with 6 clips each.
- X. Both use injection molding to form the clips
- b. Differences:
- i. A2 Medical Systems polymer clips are made in the USA and Hem-o-lok is made in Mexico.
#### H. Summary of Performance Data
AngioLOCK ligating clips are made of an implantable grade of polyacetal polymer with same dimensions and same material characteristics as the predicate Teleflex Medical Hem-o-lok ligating clips.
- a. Substantial Equivalence analyses preformed:
- i. A2 Medical Systems angioLOCK clip to predicate clip dimensional comparison was completed at an open and closed state within an applier that resulted in having the same function as Hem-o-lok when clips are applied and closed.
- ii. Both biocompatibility and polymer material comparisons prove that the molded and sterilized material used for the angioLOCK clip is as safe and the same as the material used in Hem-o-lok.
- iii. Qualitative - angioLOCK clip closure testing proved that the clips latch/lock, are secure in the applier, and release from applier's jaws equivalent to Hem-o-lok.
- iv. Qualitative - dimensional/geometry comparisons using precise overlays prove that angioLOCK and Hem-o-lok clips are the same size.
- Mechanism of action is the same between angioLOCK and Hem-o-lok; cartridge and clip V. design allows for an applier to easily load a clip, haptic feedback occurs when the clip bosses locate in the applier grooves, and clips are then removed for use. When the clip is closing the upper and lower curved legs with inner teeth pivots in from a hinged section and haptic feedback occurs as clips are latched closed.
#### l. Statement of Substantial Equivalence
The angioLOCK® Ligating Clips are substantially equivalent to their predicate device; the Teleflex Medical Hem-o-lok® Ligating Clips, based upon similarities in intended use, design, principles of operation and performance specifications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.