← Product Code [LCP](/productcode/LCP) · K214117

# Afinion™ HbA1c, Afinion™ 2 and Alere Afinion™ AS100 Analyzer (K214117)

_Abbott Diagnostics Technologies AS · LCP · Sep 27, 2023 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K214117

## Device Facts

- **Applicant:** Abbott Diagnostics Technologies AS
- **Product Code:** [LCP](/productcode/LCP.md)
- **Decision Date:** Sep 27, 2023
- **Decision:** SESE
- **Submission Type:** Dual Track
- **Regulation:** 21 CFR 864.7470
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

Afinion™ HbA1c is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, HbA1c) in venous and capillary human whole blood. The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus. Afinion™ 2 analyzer is a compact multi-assay analyzer for point-of-care testing, designed to analyze the Afinion™ test cartridges. Afinion™ 2 system, consisting of Afinion™ 2 analyzer and Afinion™ test cartridges is for in vitro diagnostic use only. Alere Afinion™ AS100 Analyzer with Alere Afinion™ Data Connectivity Converter (ADCC) is a compact multi-assay analyzer for point-of-care testing, designed to analyze the Afinion™ Test Cartridges. The ADCC is a small device for automatic transfer of data, including patient and control assay results, from the Alere Afinion™ Analyzer to a laboratory information system or another electronic journal system. Alere Afinion™ AS100 Analyzer System, consisting of Alere Afinion™ AS100 Analyzer with Alere Afinion™ Data Connectivity Converter (ADCC), Afinion™ Test Cartridges and Afinion™ Controls is for in vitro diagnostic use only.

## Device Story

Afinion™ HbA1c is an in vitro diagnostic test for quantitative determination of % HbA1c in human whole blood; used with Afinion™ AS100 or Afinion™ 2 analyzers. System utilizes boronate affinity assay technology; fully automated. Designed for point-of-care testing; operated by healthcare professionals. Input: venous or capillary whole blood samples in test cartridges. Output: quantitative % HbA1c result displayed on analyzer; data transferable via ADCC to laboratory information systems. Provides clinicians with objective marker for long-term metabolic control in diabetic patients; aids in clinical decision-making regarding diabetes management. Modification involves updated labeling with expanded analytical specificity data for hemoglobin variants, derivatives, and potential interfering substances.

## Clinical Evidence

Bench testing only. Analytical specificity studies evaluated interference from hemoglobin variants (HbAC, HbAD, HbAE, HbAS, HbF), hemoglobin derivatives (acetylated, carbamylated, pre-glycated), and various endogenous/exogenous substances (bilirubin, lipids, glucose, rheumatoid factor, common medications). Results demonstrated no significant interference (≤7%) up to specified concentrations. Hemolysis interference threshold established at >14%.

## Technological Characteristics

Fully automated boronate affinity immunoassay. Cartridge contains blue boronic acid conjugate, polyethersulfone membrane, and washing solution. Factory calibrated. Connectivity via ADCC for LIS integration. Software version 21.10 or higher updates result reporting for out-of-range values. ISO 14971 risk management applied.

## Regulatory Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

## Predicate Devices

- Alere Afinion™ HbA1c and ACR on Afinion™ 2 analyzer ([K171650](/device/K171650.md))

## Reference Devices

- Afinion™ HbA1c ([K050574](/device/K050574.md))
- Afinion™ Data Connectivity Converter ([K110056](/device/K110056.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K214117

B Applicant

Abbott Diagnostics Technologies AS

C Proprietary and Established Names

Afinion™ HbA1c, Afinion™ 2 and Alere Afinion™ AS100 Analyzer

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LCP | Class II | 21 CFR 864.7470 - Glycosylated Hemoglobin Assay | HE - Hematology  |
|  JQT | Class I | 21 CFR 862.2400 - Densitometer/scanner (integrating, reflectance, TLC, or radiochromatogram) for clinical use | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

This submission is a Dual 510(k) and CLIA Waiver by Application (Dual Submission) tracked as K214117 and CW210007. This 510(k) is for a modification to a cleared device.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

B Measurand:

Glycosylated Hemoglobin (HbA1c)

C Type of Test:

Quantitative colorimetric boronate affinity immunoassay

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

Afinion™ HbA1c

Afinion™ HbA1c is an in-vitro diagnostic test for quantitative determination of glycated hemoglobin (%hemoglobin A1c, % HbA1c) venous and capillary human whole blood. The measurement of % HbA1c is recommended as a marker of long term metabolic control in persons with diabetes mellitus.

Afinion™ 2

Afinion™ 2 analyzer is a compact multi-assay analyzer for point-of-care testing, designed to analyze the Afinion™ test cartridges. Afinion™ 2 system, consisting of Afinion™ 2 analyzer and Afinion™ test cartridges is for in vitro diagnostic use only.

Alere Afinion™ AS100 Analyzer

Alere Afinion™ AS100 Analyzer with Alere Afinion™ Data Connectivity Converter (ADCC) is a compact multi-assay analyzer for point-of-care testing, designed to analyze the Afinion™ Test Cartridges. The ADCC is a small device for automatic transfer of data, including patient and control assay results, from the Alere Afinion™ Analyzer to a laboratory information system or another electronic journal system.

Alere Afinion™ AS100 Analyzer System, consisting of Alere Afinion™ AS100 Analyzer with Alere Afinion™ Data Connectivity Converter (ADCC), Afinion™ Test Cartridges and Afinion™ Controls is for in vitro diagnostic use only.

K214117 - Page 2 of 8

{2}

C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use
For Point of Care use.

D Special Instrument Requirements:
For use with the Afinion™ 2 and the Alere Afinion™ AS100 Analyzer

IV Device/System Characteristics:

A Device Description:
The Afinion HbA1c is a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human venous and capillary whole blood. Each assay cartridge contains blue boronic acid conjugate, a tube with a polyethersulfone membrane, and washing solution (morpholine buffered sodium chloride solution with detergents and preservative). Each cartridge is labeled with a unique barcode.

The Afinion HbA1c is factory calibrated and intended for use with the previously cleared Afinion AS100 Analyzer (K050574, K110056, K151809), the Afinion 2 (K171650) and the Afinion HbA1c Controls (K050574).

B Principle of Operation:
Afinion™ HbA1c is a fully automated boronate affinity assay. The test Cartridge contains all of the necessary reagents. The sample is collected with the integrated sampling device and the Test Cartridge is placed in the cartridge chamber of the analyzer. The blood sample is then automatically diluted and mixed with a solution that releases hemoglobin from the erythrocytes. The hemoglobin precipitates. This sample mixture is transferred to a blue boronic acid conjugate, which binds to the cis-diols of glycated hemoglobin. This reaction mixture is soaked through a filter membrane and all precipitated hemoglobin, conjugate bound and unbound (i.e., glycated and non-glycated hemoglobin) remains on the membrane. Any excess of conjugate is removed with washing reagent.

The analyzer evaluates the precipitate on the membrane. By measuring the reflectance, the blue (glycated hemoglobin) and the red (total hemoglobin) color intensities are evaluated. The ratio between them is proportionate to the percentage of HbA1c in the sample. The % HbA1c is displayed on the Afinion™ AS100 Analyzer and the Afinion™ 2.

K214117 - Page 3 of 8

{3}

V Substantial Equivalence Information:

A Predicate Device Name(s):

Alere Afinion™ HbA1c and ACR on Afinion™ 2 analyzer

B Predicate 510(k) Number(s):

K171650

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K214117 | K171650  |
| --- | --- | --- |
|  Device Trade Name | Afinion™ HbA1c
Afinion™ 2
Afinion™ AS100 Analyzer | Afinion™ HbA1c
Afinion™ 2 analyzer  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Afinion HbA1c is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, % HbA1c). The measurement of % HbA1c is recommended as a marker of long term metabolic control in persons with diabetes mellitus. | Same  |
|  Assay principle | Quantitative colorometric boronate affinity immunoassay | Same  |
|  General Device Characteristic Differences |  |   |
|  Analyzer | Alere Afinion™ AS100 Analyzer and Afinion™ 2 | Afinion™ 2 analyzer  |
|  Sampling from Control Vial | Mix the control well by thoroughly shaking the vial for 30 seconds. You may also use a vortex mixer for 30 seconds. | Mix the control material thoroughly by shaking the vial for 30 seconds.  |

K214117 - Page 4 of 8

{4}

|  Device & Predicate Device(s): | K214117 |   |   | K171650  |   |
| --- | --- | --- | --- | --- | --- |
|  Results outside reportable range | Code | Cause | Comment | Code | Cause  |
|   |  105 | HbA1c below 4.0% | If your analyzer has software version 21.10 or higher, this code will not be displayed. The results will be shown as «HbA1c <4.0% » | 105 | HbA1c below 4.0%  |
|   | 106 | HbA1c above 15% | If your analyzer has software version 21.10 or higher, this code will not be displayed. The results will be shown as «HbA1c <15.0% » | 106 | HbA1c above 15%  |

VI Standards/Guidance Documents Referenced:

ISO 14971 Medical devices - Application of risk management to medical devices

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Previously established in K050574

2. Linearity/ assay reportable range:

Linearity was previously established for the range of 4.00 – 15.00 % HbA1c in K050574.

K214117 - Page 5 of 8

{5}

K214117 - Page 6 of 8

3. Analytical Specificity/Interference:

Interference was evaluated and found to support the sponsor's claims, included in the candidate labeling, that there is no significant interference (≤7%) up to the following concentrations:

|  Substance | Highest concentration tested with no significant interference  |
| --- | --- |
|  Acetaminophen | 20 mg/dL  |
|  Acetylcysteine | 166 mg/dL  |
|  Acetylsalicylic acid | 100 mg/dL  |
|  Ampicillin | 100 mg/dL  |
|  Ascorbic acid | 30 mg/dL  |
|  Bilirubin -conjugated | 60 mg/dL  |
|  Bilirubin - unconjugated | 60 mg/dL  |
|  Cefoxitin | 250 mg/dL  |
|  Cholesterol | 351 mg/dL  |
|  Cyclosporine A | 0.5 mg/dL  |
|  Cyclosporine C | 0.5 mg/dL  |
|  Doxycyclin | 5 mg/dL  |
|  Fructosamine | 680 μmol/L  |
|  Glucose | 1000 mg/dL  |
|  Glyburide | 0.19 mg/dL  |
|  Heparin | 5000 U/L  |
|  Ibuprofen | 50 mg/dL  |
|  Intralipid | 1000 mg/dL  |
|  Levodopa | 2 mg/dL  |
|  Metformin | 4 mg/dL  |
|  Methyldopa | 2 mg/dL  |
|  Metronidazole | 20 mg/dL  |
|  Phenylbutazone | 40 mg/dL  |
|  Rheumatoid factor | 780 000 IU/L  |
|  Rifampicin | 6.4 mg/dL  |
|  Salicylic acid | 59 mg/dL  |
|  Theophylline | 10 mg/dL  |
|  Total protein | 150 g/L  |
|  Triglycerides | 1389 mg/dL  |

{6}

K214117 - Page 7 of 8

# Cross reactivity with Hemoglobin Derivatives

Potential cross reactivity with acetylated hemoglobin, carbamylated hemoglobin, labile HbA1c, and glycated albumin was evaluated and found to support the sponsor's claims, included in the candidate labeling, that there is no cross reactivity with acetylated hemoglobin (4.6 mg/mL), carbamylated hemoglobin (13.8 mg/mL), labile HbA1c (11.4 mg/mL), or glycated albumin (7.7 mg/mL).

# Hemoglobin Variant Interference

Hemoglobin variant testing demonstrated no significant interference for Hemoglobin A2 (≤ 5.7%), Hemoglobin S (≤ 42%), Hemoglobin C (≤ 36%), Hemoglobin E (≤ 26%), Hemoglobin D (≤ 42%) and Hemoglobin F (≤10.4%). The sponsor's definition of non-significant interference is ≤7% mean relative deviation between the candidate method and the comparator method.

The labeling of the candidate HbA1c assay was updated to reflect these results as well as the following: "The highest HbF concentration where no significant interference (= 7%) is observed is 10.4% HbF. Above 10.4% HbF, a negative interference is observed".

4. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

The Afinion HbA1c assay is traceable to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Reference Method for Measurement of HbA1c. The Afinion HbA1c assay is certified with the National Glycohemoglobin Standardization Program (NGSP).

5. Detection Limit:

The Afinion HbA1c assay reportable range is 4.0-15.0% HbA1c as established in K050574.

6. Assay Cut-Off:

Not applicable.

# B Comparison Studies:

1. Method Comparison with Predicate Device:

Performance was established in K050574

{7}

2. **Matrix Comparison:**
Matrix comparison was evaluated in K050574

**C Clinical Studies:**

1. **Clinical Sensitivity:**
Not applicable.

2. **Clinical Specificity:**
Not applicable.

3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not applicable.

**D Clinical Cut-Off:**
Not applicable.

**E Expected Values/Reference Range:**

The following is included in the labeling:

**Recommendations from American Diabetes Association (ADA):**
A reasonable goal for many nonpregnant adults with diabetes is HbA1c &lt; 7.0% (53 mmol/mol). More or less stringent glycemic goals may be appropriate for individual patients. Goals should be individualized based on duration of diabetes, age/life expectancy, comorbid conditions, known CVD or advanced microvascular complications, hypoglycemia unawareness, and individual patient considerations.

American Diabetes Association. Glycemic Target. Diabetes Care 2019;42(Suppl. 1):S61-S70

**VIII Proposed Labeling:**

The labeling supports the finding of substantial equivalence for this device.

**IX Conclusion:**

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K214117 - Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/K214117](https://fda-staging.innolitics.com/device/K214117)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
