← Product Code [MQC](/productcode/MQC) · K214102

# Glidewell Appliance Resin, Hard/Soft (K214102)

_Prismatik Dentalcraft, Inc. · MQC · Feb 23, 2022 · DE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K214102

## Device Facts

- **Applicant:** Prismatik Dentalcraft, Inc.
- **Product Code:** [MQC](/productcode/MQC.md)
- **Decision Date:** Feb 23, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE

## Indications for Use

Glidewell™ Appliance Resin. Hard/Soft is indication of orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners.

## Device Story

Glidewell Appliance Resin (Hard/Soft) is a light-cured, methacrylate-based resin used for 3D printing orthodontic and dental appliances. The system utilizes DLP printers (385nm wavelength) to convert liquid resin into solid devices. It supports single-layer 3D printed hard appliances or dual-layer devices (3D printed hard outer layer combined with injection-molded soft inner layer). Used in dental clinics or laboratories; operated by dental professionals. The device enables custom fabrication of splints and guards to assist in patient dental/orthodontic treatment. Safety and performance are verified through physical property testing and biocompatibility assessments.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by non-clinical bench testing: flexural strength/modulus (ISO 20795-2), water sorption/solubility (ISO 20795-2), tensile strength/modulus/elongation (ASTM D638), printing accuracy/orientation validation, packaging validation (ASTM D4169-14), and biocompatibility (ISO 10993-1).

## Technological Characteristics

Light-cured methacrylate-based resin. Compatible with 385nm DLP additive manufacturing systems. Non-sterile. Physical properties tested per ISO 20795-2 and ASTM D638. Biocompatible per ISO 10993-1.

## Predicate Devices

- KeyPrint KeySplint Soft Resin ([K183598](/device/K183598.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 23, 2022

Prismatik Dentalcraft, Inc. Jiahe Li Regulatory Affairs Specialist 2144 Michelson Drive Irvine, California 92612

Re: K214102

Trade/Device Name: Glidewell Appliance Resin, Hard/Soft Regulatory Class: Unclassified Product Code: MQC, KMY Dated: December 27, 2021 Received: December 29, 2021

Dear Jiahe Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K214102

Device Name Glidewell™ Appliance Resin, Hard/Soft

Indications for Use (Describe)

Glidewell™ Appliance Resin. Hard/Soft is indication of orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 SubpartC)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

«An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonnation unless it displays a currently valid 0MB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a color gradient that starts with pink at the top and transitions to blue at the bottom. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." in a sans-serif font.

## K214102

### 510(k) Summary

#### I. SUBMITTER

Prismatik Dentalcraft, Inc. 2144 Michelson Drive, Irvine, CA 92612, USA

Primary Contact Person: Jiahe Li, Regulatory Affairs Specialist Email: Jiahe.Li(a)glidewelldental.com Phone: (949) 222-3516

Secondary Contact Person: Herbert Crane, VP RA/QA Email: Herbert.Crane(@glidewelldental.com Phone: (949) 222-3531

Date Prepared: December 27, 2021

#### II. DEVICE

Name of Device: Glidewell™ Appliance Resin, Hard/Soft Classification Product Code: MQC Regulatory Class: Unclassified Common Name: Mouthguard, Prescription; Classification Name: N/A, Pre-Amendment Subsequent Product Code: KMY Regulatory Class: Class I Common Name: Positioner, Tooth, Preformed Classification Name: Preformed tooth positioner

#### III. PREDICATE DEVICE

KeyPrint KeySplint Soft Resin (K183598)

#### IV. DEVICE DESCRIPTION

Glidewell™ Appliance Resin, Hard/Soft is a light-cured resin for fabrication of orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners. The Glidewell™ Appliance Resin. Hard and the Glidewell Appliance Resin, Soft can be used in combination to create a dual layer device consisting of a 3D printed hard layer, made from Glidewell™ Appliance Resin, Hard, and a soft layer that is molded with Glidewell Appliance Resin, Soft. The Glidewell™ Appliance Resin, Hard can also be used alone to create a 3D printed single layer device. The Glidewell™ Appliance Resin, Hard is compatible with DLP printers utilizing a wavelength of 385nm.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a gradient of colors, including pink, purple, blue, and yellow. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC." in blue text.

#### V. INDICATIONS FOR USE

Glidewell™ Appliance Resin, Hard/Soft is indicated for the fabrication of orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

| Technological<br>Characteristics        | Subject Device<br>Glidewell™ Appliance<br>Resin, Hard/Soft                                                                                                                                                    | Predicate Device<br>KeyPrint KeySplint Soft<br>Resin (K183598)                                                                                                                                                                    | Comparison                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                            | Prismatik Dentalcraft,<br>Inc.                                                                                                                                                                                | Keystone Industries                                                                                                                                                                                                               | N/A                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product Code                            | MQC, KMY                                                                                                                                                                                                      | MQC, KMY                                                                                                                                                                                                                          | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Prescription Device                     | Yes                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                               | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use                            | Resin for orthodontic and<br>dental appliances                                                                                                                                                                | Resin for orthodontic and<br>dental appliances                                                                                                                                                                                    | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use                     | Glidewell™ Appliance<br>Resin, Hard/Soft is<br>indicated for the<br>fabrication of<br>orthodontic and dental<br>appliances such as bite<br>planes, mouthguards,<br>nightguards, splints and<br>repositioners. | The KeyPrint KeySplint<br>Soft device is indicated for<br>the fabrication of<br>orthodontic and dental<br>appliances such as bite<br>planes, mouthguards,<br>nightguards, splints and<br>repositioners.                           | The indications for use<br>are the same except for<br>the device trade name.                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Design<br>Characteristics               | General Design                                                                                                                                                                                                | Light-cured resin using 3D<br>printing technology<br>or combination of 3D<br>printing and injection<br>molding to convert the<br>liquid resin to solid<br>orthodontic and dental<br>appliances.                                   | Light-cured resin using 3D<br>printing technology to<br>convert the liquid resin to<br>solid orthodontic and<br>dental appliances. | The subject device uses<br>3D printing technology<br>to convert the liquid<br>resin to solid<br>orthodontic and dental<br>appliances, same as the<br>predicate device.<br>The only difference is<br>that the subject device<br>has the additional<br>option of fabricating a<br>dual-layered device<br>through 3D printing of<br>the hard outer layer<br>with the hard resin and<br>injection molding of<br>the soft inner layer with<br>the soft resin. |
| Technological<br>Characteristics        | Subject Device<br>Glidewell™ Appliance<br>Resin, Hard/Soft                                                                                                                                                    | Predicate Device<br>KeyPrint KeySplint Soft<br>Resin (K183598)                                                                                                                                                                    | Comparison                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Material<br>composition                 | Light cured methacrylate<br>based-resin.                                                                                                                                                                      | Light cured methacrylate<br>based-resin.                                                                                                                                                                                          | Substantially<br>equivalent                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Additive<br>Manufacturing<br>System     | Glidewell™ Appliance<br>Resin, Hard/Soft is<br>intended to be used in<br>conjunction with an<br>additive Computer-<br>Aided Manufacturing<br>(CAM) and curing<br>system.                                      | The KeyPrint KeySplint<br>Soft device is intended to<br>be used in conjunction<br>with an additive<br>Computer- Aided<br>Manufacturing (CAM) and<br>curing system.                                                                | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Bench Testing<br>(physical<br>property) | Physical properties<br>according to:<br>ISO 20795-2:<br>Flexural Strength,<br>Flexural<br>Modulus, Water<br>Sorption/SolubilityASTM D638:<br>Tensile Strength,<br>Tensile Modulus,<br>Elongation              | Physical properties<br>according to:<br>ASTM D790:<br>Flexural PropertiesISO 20795-2:<br>Flexural Strength,<br>Flexural Modulus,<br>Water<br>Sorption/SolubilityASTM D638:<br>Tensile Strength,<br>Tensile Modulus,<br>Elongation | The predicate device<br>was tested for flexural<br>properties according to<br>ASTM D790 in<br>addition to ISO 20795-<br>2.         |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Biocompatibility                        | Biocompatible per<br>testing results according<br>to ISO 10993-1                                                                                                                                              | Biocompatible per testing<br>results according to ISO<br>10993-1                                                                                                                                                                  | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterility                               | Non-Sterile                                                                                                                                                                                                   | Non-Sterile                                                                                                                                                                                                                       | Same                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections with different colors, including pink, purple, blue, and yellow. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

## DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The subject device, Glidewell™ Appliance Resin, Hard/Soft, is substantially equivalent to the primary predicate device, KeyPrint KeySplint Soft Resin (K183598) in intended use, indications for use and technological characteristics, including technical specifications/features, material and principles of operation.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in a sans-serif font, with the words "PRISMATIK" and "DENTALCRAFT, INC." stacked on top of each other.

The subject device, Glidewell™ Appliance Resin, Hard/Soft, has the same intended use as the predicate device, KeyPrint KeySplint Soft Resin (K183598) as material for fabricating prescription mouthguards and preformed tooth positioners. The subject device, Glidewell™ Appliance Resin, Hard/Soft, has the same Indications for Use Statement (IFUS) as the predicate device, KeyPrint KeySplint Soft Resin (K183598), except for the device trade name. Both the subject device, Glidewell™ Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598) are dental resins indicated for the fabrication of orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners.

The subject device. Glidewell™ Appliance Resin. Hard/Soft, is substantially equivalent to the predicate device, KeyPrint KeySplint Soft Resin (K183598) in technical specifications/features. The product specifications of the subject device, Glidewell™ Appliance Resin, Hard/Soft, is based on benchmarking comparable devices with similar indications for use on the market, including the predicate device, KeyPrint KeySplint Soft Resin (K183598). The same property testing according to ISO 20795-2 and ASTM D638 that applied to the predicate device, KeyPrint KeySplint Soft Resin (K183598), were performed on the subject device, Glidewell™ Appliance Resin, Hard/Soft, including flexural strength, flexural modulus, water sorption, water solubility, tensile strength, tensile modulus and elongation. The physical property of the subject device, Glidewell™ Appliance Resin, Hard/Soft, passed the threshold of performance criteria in ISO 20795-2 when applicable to the design, and met the same performance criteria used by the predicate device, KeyPrint KeySplint Soft Resin (K183598).

The subject device, Glidewell™ Appliance Resin, Hard/Soft and the predicate device, KeyPrint KeySplint Soft Resin (K183598) are similar in material composition. Both the subject device. Glidewell™ Appliance Resin. Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598) are light cured methacrylate-based resin. Despite the actual difference that might exist between the formulations, the difference does not affect the safety and effectiveness for the indended use, as verified by the safety and performance testing.

The subject device, Glidewell™ Appliance Resin, Hard/Soft, is substantially equivalent to the predicate device, KeyPrint KeySplint Soft Resin (K183598) in terms of principle of operation. Both the subject device, Glidewell™ Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598) are light-cured resin in liquid form that through light-mediated conversion can be fabricated into a solid finished device. The finished device has the desired physical properties suitable for orthodontic and dental appliances such as bite planes, mouthguards, nightguards, splints and repositioners.

#### VII. PERFORMANCE DATA

Non-clinical data submitted to demonstrate substantial equivalence include:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for PRISMATIK DENTALCRAFT, INC. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

- Flexural strength and flexural modulus, according to ISO 20795-2
- Water sorption and water solubility, according to ISO 20795-2 ●
- Tensile Strength, Tensile Modulus and Elongation, according to ASTM D638 ●
- Printing Accuracy and Printing Orientation Validation ●
- Packaging validation ●
- . Biocompatibility

No clinical data is included in this submission.

### Flexural Strength and Flexural Modulus

The subject device, Glidewell™ Appliance Resin, Hard/Soft, was tested for flexural strength and flexural modulus in accordance with the test methods outlined in ISO 20795-2. All the testing results met the acceptance criteria. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device. Glidewell™ Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598).

### Water Sorption and Water Solubility

The subject device, Glidewell™ Appliance Resin, Hard/Soft, was tested for water sorption and water solubility in accordance with ISO 20795-2. All the testing results met the acceptance criteria. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Glidewell™ Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598).

### Tensile Strength, Tensile Modulus and Elongation

The subject device, Glidewell™ Appliance Resin, Hard/Soft, was tested for tensile strength, tensile modulus and elongation in accordance with ASTM D638. All the testing results met the acceptance criteria. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, GlidewellTM Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598).

### Printing Accuracy and Printing Orientation Validation

Printing accuracy test was performed to validate that the physical output of the additive manufacturing system and procedure for Glidewell™ Appliance Resin. Hard are able to meet design input dimensions within the pre-specified tolerance. Printing orientation test was performed to validate that the hard resin printed at different print direction within the build space relative to the device orientation and at different build plate locations are able to meet the same performance criteria. The results met the pre-specified acceptance criteria and demonstrated that the subject device. Glidewell™ Appliance Resin. Hard/Soft, can be reliably fabricated at different print directions within the build space and at different build plate locations using the additive manufacturing system and procedure.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangular prism on the left and the company name on the right. The company name is written in a sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

### Packaging Validation

Packaging validation was performed for the subject device, Glidewell™ Appliance Resin, Hard/Soft. Per ASTM D4169-14, the shipping unit was tested for manual handling, compressive loads, repetitive shocks from vibration, vertical vibration environments, concentrated impacts and secondary manual handling drops. It was determined that Glidewell™ Appliance Resin, Hard/Soft with the respective packaging, is suitable for use. The results of the testing were used to address questions related to substantial equivalence based on differences in product packaging between the subject device. Glidewell™ Appliance Resin. Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598).

### Biocompatibility

The subject device, Glidewell™ Appliance Resin, Hard/Soft, was tested in accordance with ISO 10993-1. Based on the biocompatibility testing results, it was determined that there is no biocompatibility concern for the subject device. The results of the testing were used to address questions related to substantial equivalence based on differences in chemical composition between the subject device, Glidewell™ Appliance Resin, Hard/Soft, and the predicate device, KeyPrint KeySplint Soft Resin (K183598).

#### VIII. CONCLUSION

Based on technological characteristics and non-clinical test data included in this submission, the subject device, Glidewell™ Appliance Resin, Hard/Soft, has been shown to be substantially equivalent to the predicate device. KeyPrint KeySplint Soft Resin (K183598).

---

**Source:** [https://fda-staging.innolitics.com/device/K214102](https://fda-staging.innolitics.com/device/K214102)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).