K214093 · 3D Systems · IXI · Mar 30, 2022 · Radiology
Device Facts
Record ID
K214093
Device Name
VSP Bolus
Applicant
3D Systems
Product Code
IXI · Radiology
Decision Date
Mar 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5710
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 3D Systems bolus product is placed on the skin of a patient with the intended use of helping control the dose received by that patient when undergoing radiation therapy treatment. The bolus will be manufactured according to the unique, patient-specific shape requested by a clinical customer. The 3D Systems bolus product is intended for patients of all ages requiring external beam radiotherapy, and is intended for prescription use only.
Device Story
Radiotherapy accessory; patient-specific bolus; conforms to patient anatomy to reduce airgaps. Input: DICOM medical imaging data; radiation therapy professional input (thickness, anatomical location). Transformation: Additive manufacturing (3D printing) using VisiJet M2E-BK70 elastomeric material. Output: Physical bolus placed on patient skin. Used in radiotherapy clinics; operated by radiation therapy professionals. Workflow: Professional designs bolus based on treatment plan; device manufactured; professional validates fit via CT scan prior to radiation delivery. Benefit: Improved radiation dose control for complex patient contours.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993) and simulated use testing involving clinically relevant validation study.
Technological Characteristics
Patient-specific radiotherapy bolus; additive manufacturing (3D printing). Material: VisiJet M2E-BK70 (biocompatible for intact skin). Energy source: N/A (passive accessory). Connectivity: N/A. Sterilization: Not specified. Software: N/A.
Indications for Use
Indicated for patients of all ages requiring external beam radiotherapy to help control radiation dose. Not assessed for use with protons, electrons, or orthovoltage X-rays.
Regulatory Classification
Identification
A radiation therapy beam-shaping block is a device made of a highly attenuating material (such as lead) intended for medical purposes to modify the shape of a beam from a radiation therapy source.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.
3D Systems % Benjamin Johnson Vice President, Portfolio and Regulatory 5381 S Alkire Circle LITTLETON CO 80127
March 30, 2022
Re: K214093
Trade/Device Name: VSP® Bolus Regulation Number: 21 CFR 892.5710 Regulation Name: Radiation therapy beam-shaping block Regulatory Class: Class II Product Code: IXI Dated: March 3, 2022 Received: March 4, 2022
Dear Benjamin Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K214093
Device Name
VSP Bolus
Indications for Use (Describe)
The 3D Systems VSP Bolus product is a device that will be placed on the skin of a patient as a radiotherapy accessory intended to help control the radiation dose received by the patient. VSP Bolus is generated using input from radiation therapy professionals and medical imaging data to produce a bolus that is specific to the patient being treated. The VSP Bolus product is verified and approved by the radiation therapy professional prior to use on the patient, and is intended for patients of all ages receiving radiotherapy treatment.
VSP Bolus was evaluated using 6 MV photons but has not been assessed for use with protons. electrons, or at orthovoltage X-rays.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the 3D Systems logo. The logo consists of a stylized cube with the number "3" on each visible face. To the right of the cube is the company name, "3D SYSTEMS," in a simple, sans-serif font. The overall design is clean and modern, with a focus on the company's name and a visual representation of three-dimensional technology.
# K214093
# 510(K) SUMMARY
### 1. INTRODUCTION
This document contains the 510(k) summary for VSP® Bolus. The content of this summary is based on the requirements of 21 CFR 807.92.
### 2. SUBMITTER
| | Name: | 3D Systems, Inc. |
|----|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | Address: | 5381 South Alkire Circle<br>Littleton, CO 80127, USA<br>Phone: (720) 643-1001<br>Fax: (720) 643-1009 |
| | Official Contact: | Benjamin Johnson<br>Vice President, Portfolio & Regulatory, Healthcare |
| 3. | Date Prepared: | December 20th, 2021 |
| | DEVICE<br>Trade Name:<br>Common Name:<br>Classification Name: | VSP® Bolus<br>Patient-specific Radiotherapy Bolus<br>Radiation therapy beam-shaping block |
| | Classification: | Class II, 21 CFR 892.5710 |
| 4. | Product Code: | IXI |
| | PREDICATE DEVICES | |
Predicate device: BOLUS COMPENSATOR, .DECIMAL INC. (K091911)
### 5. DESCRIPTION OF THE DEVICE
Boluses are used in external beam radiation therapy (EBRT) to change the depth of the radiation dose delivery to overcome the skin-sparing effect. The bolus is generated in accordance with a clinical treatment plan. The patient-specific bolus conforms to the patient anatomy to reduces airgaps for complex patient contours. The bolus is produced with additive manufacturing in a soft elastomeric material to conform to the patient's skin. The bolus is placed on the patient and verified for fit and accordance to the treatment plan prior to a radiation dose delivery to the patient.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the 3D Systems logo. On the left is a 3D cube with the number 3 in it. To the right of the cube is the company name, 3D Systems, in gray lettering.
### 6. INTENDED USE
The 3D Systems bolus product is placed on the skin of a patient with the intended use of helping control the dose received by that patient when undergoing radiation therapy treatment. The bolus will be manufactured according to the unique, patient-specific shape requested by a clinical customer. The 3D Systems bolus product is intended for patients of all ages requiring external beam radiotherapy, and is intended for prescription use only.
#### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The intended use and technological characteristics of the subject device are either identical or substantially equivalent to the predicate device, differing only in the materials and manufacturing methods to produce the device. The potential impact on substantial equivalence of each technological difference was addressed by risk analysis and clinically oriented testing.
A comparison of the technical characteristics between VSP® Bolus and predicate device is provided in the following Comparison Table.
| Predicate Comparison Table | | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | VSP® Bolus<br>Subject Device | Bolus Compensator<br>Predicate Device |
| | | |
| Indications for Use | The 3D Systems VSP® Bolus product is a device<br>that will be placed on the skin of a patient as a<br>radiotherapy accessory intended to help control<br>the radiation dose received by the patient. VSP<br>Bolus is generated using input from radiation<br>therapy professionals and medical imaging data to<br>produce a bolus that is specific to the patient<br>being treated. The VSP Bolus product is verified<br>and approved by the radiation therapy<br>professional prior to use on the patient, and is<br>intended for patients of all ages receiving<br>radiotherapy treatment.<br><br>VSP Bolus was evaluated using 6 MV photons but<br>has not been assessed for use with protons,<br>electrons, or at orthovoltage X-rays. | The Bolus Compensators are used by<br>radiation therapy professionals for the<br>treatment of cancer patients. They are<br>designed by radiation therapy professionals<br>for a unique patient and are intended to<br>modify the shape of abeam from a radiation<br>therapy source. Each Bolus Compensator<br>must be validated and approved by the<br>radiation therapy professional prior to use on<br>a patient. |
| Device Input<br>Parameters | 1. DICOM<br>2. Radiotherapy professional input (i.e.<br>thickness, anatomical location) | 1. DICOM<br>2. The device is designed by the radiation<br>therapy professional. |
| Manufacturing<br>Method | Additive Manufacturing (3D Printing) | Subtractive Manufacturing (Machining) |
| Device Material | VisiJet® M2E-BK70 | Deep Blue Wax |
| Patient Contact<br>Materials | Biocompatible for intact skin contact. | "negligible irritation to skin at ambient<br>temperatures"1 |
| Device Clinical<br>Acceptance | Each VSP® Bolus must be validated and approved<br>by the radiation therapy professional through a CT<br>scan prior to the first treatment fraction. | Each Bolus Compensator must be validated<br>and approved by the radiation therapy<br>professional to use on a patient. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the 3D Systems logo. On the left is a cube-like shape with the number 3D on each of the three visible sides. To the right of the cube is the company name, "3D SYSTEMS," in a sans-serif font. The logo is in grayscale.
| Predicate Comparison Table | | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | VSP® Bolus<br>Subject Device | Bolus Compensator<br>Predicate Device |
| Target Population | Cancer patients requiring external beam<br>radiotherapy | Cancer Patients requiring external beam<br>radiation therapy |
| Target Population<br>Anatomical Sites | Various | Various |
| Use Environment | Radiotherapy Clinic | Radiotherapy Clinic |
| Performance | Clinically oriented validation test cases were<br>written and executed. VSP Bolus was deemed fit<br>for clinical use by radiation therapy professionals. | Clinically oriented validation test cases were<br>written and executed in house .decimal<br>personnel including Board Certified Medical<br>Physicists where Bolus Compensators was<br>deemed fit for clinical use. |
Deep Blue Wax MSDS, QAF-312 Rev 071814 1)
#### 8. SUMMARY OF CLINICAL TESTING
Clinical testing was not performed as part of the development of this product. Clinical testing is not advantageous in demonstrating substantial equivalence, safety, or effectiveness of the device since testing can be performed such that no human subjects are exposed to risk.
#### 9. SUMMARY OF PERFROMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The biocompatibility evaluation for the VSP Bolus was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medica Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- . Irritation
#### Simulated Use Testing
Clinically relevant validation study was performed.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the 3D Systems logo. On the left is a 3D cube with the number 3 in it. To the right of the cube is the text "3D SYSTEMS" in a sans-serif font. The text is gray and the cube is silver.
## 10. CONCLUSION
The VSP® Bolus has the same intended use and similar or substantially equivalent technological characteristics as the predicate. Minor differences in the technological characteristics do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.