K214076 · Spident Co., Ltd. · EMA · Dec 19, 2022 · Dental
Device Facts
Record ID
K214076
Device Name
EsCem
Applicant
Spident Co., Ltd.
Product Code
EMA · Dental
Decision Date
Dec 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
Final cementation of all-ceramic, metal-based inlays, onlays, posts, crown and bridge
Device Story
EsCem is a dual-cured, self-adhesive, self-etching resin cement used for the final cementation of indirect dental restorations (inlays, onlays, posts, crowns, bridges). The device is intended for use by dentists in a clinical setting. It functions as a luting material, providing adhesion between the tooth structure and the restoration. The cement is radiopaque, facilitating identification on radiographs. Its dual-cure mechanism allows for both light-curing and self-curing, providing flexibility in clinical application. By eliminating the need for separate pre-treatment steps, the 'one-step' system simplifies the cementation process, potentially reducing chair time and improving procedural efficiency for the clinician while ensuring secure restoration placement for the patient.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 4049:2019 standards, including film thickness (<50 μm), working time, setting time (<10 min), flexural strength (≥50 MPa), water sorption (≤40 µg/mm³), solubility (≤7.5 µg/mm³), radio-opacity (≥1 mm Al), light sensitivity, and color stability. Biocompatibility verified per ISO 10993-1, -3, -5, -6, -10, -11 and ISO 7405.
Technological Characteristics
Dual-cured resin cement. Composition: methacrylate monomers, fillers, initiators, inhibitors, and pigments. Form factor: paste. Radiopaque (≥1 mm Al). Conforms to ISO 4049:2019. Storage: 2-27°C. No software or electronic components.
Indications for Use
Indicated for final cementation of all-ceramic and metal-based inlays, onlays, posts, crowns, and bridges in dental patients.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2022
Spident Co., Ltd. Eunok Choi Manager 203 & 312, Korea Industrial Complex, 722 Gojan-Dong, Namdong-Gu, Incheon 405-821 SOUTH KOREA
Re: K214076
Trade/Device Name: EsCem Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: November 9, 2022 Received: November 18, 2022
Dear Eunok Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K214076
Device Name EsCem
Indications for Use (Describe)
Final cementation of all-ceramic, metal-based inlays, onlays, posts, crown and bridge
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K214076
Image /page/3/Picture/1 description: The image shows the word "SPIDENT" in green font. The letters are enclosed in a green box. The box is open on the top and left sides, with the letters "SP" outside of the box and the letters "IDENT" inside of the box. The background is white.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074
# Section 5. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date: December 22, 2021
## 1. Company and Correspondent making the submission:
Company Name : SPIDENT CO., LTD. Address : 203 & 312 Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074 Company Contact : J.M. Ahn / President
## 2. Device Name and Classification
Proprietary Name : EsCem Common name : Self-Adhesive Resin Cement Classification name : Dental Cement [CFR 872.3275] Product code : EMA Class : II
## 3. Predicate Devices (Legally Marketed Devices)
The predicate devices for EsCem is :
- G-CEM LinkAce (GAM-200), GC AMERICA INC., K120243
## 4. Description:
EsCem is a dual-cured resin cement that can be self-etched and self-adhesive. It is classified as type 2 and class 3 according to ISO 4049. It is easy to handle with 'one step' system that does not need pre-treatment process, and it is easy to remove excess cement, so it is possible to perform quick operation. It is also easy to identify because it has a radio-opacity(1.2mm Al). And it offers various shades for aesthetics and color stable.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Spident. The logo is green and features the word "SPIDENT" in a bold, sans-serif font. The word is enclosed in a square frame, with the top and bottom lines of the frame extending beyond the sides of the word. The logo is simple and modern, and the green color gives it a fresh and natural feel.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074
## 5. Indications for Use
Final cementation of all-ceramic, metal-based inlays, onlays, posts, crown and bridge
## 6. Performance Testing - Bench
The performance test was conducted to prove the substantially equivalence of the subject device, and the standards applied to the performance test are as follows.
| Performance | Test method | Requirement |
|----------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Film thickness | ISO 4049:2019<br>-7.5 Measurement of film thickness of luting materials | shall be no greater than 50 μm. |
| Working time | ISO 4049:2019<br>-7.7 Working time, Class 1 and Class 3 luting materials | during its formation there shall be no detectable change in its homogeneity. |
| Setting time | ISO 4049:2019<br>-7.8 Setting time, Class 1 and Class 3 materials | shall be no more than 10 min. |
| Flexural strength | ISO 4049:2019<br>-7.11 Flexural strength | shall be equal to or greater than 50 MPa |
| Water sorption | ISO 4049:2019<br>-7.12 Water sorption and solubility | shall be equal to or less than 40 µg/mm³. |
| Solubility | ISO 4049:2019<br>-7.12 Water sorption and solubility | shall be equal to or less than 7,5 µg/mm³. |
| Radio-opacity | ISO 4049:2019<br>-7.14 Radio-opacity | shall be equal to or greater than that of the same thickness of aluminium |
| Sensitivity to light | ISO 4049:2019<br>-7.9 Sensitivity to light, Class 2 materials | the material shall remain physically homogeneous. |
| Color stability | ISO 4049:2019<br>-7.13 Shade and colour stability after irradiation and water sorption | no more than a slight change in colour shall be observed |
- · ISO 4049:2019, Dentistry Polymer-based restorative materials
## 7. Biocompatibility
Biocompatibility tests and biological safety evaluation were performed and the results proved that the subject device is at least as biocompatible and biologically safe as the predicate device.
The standards applied are as follows.
- · ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- · ISO 10993-3:2014, Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- · ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SPIDENT. The logo is green and features the word "SPIDENT" in a sans-serif font. The letters are enclosed in a square-shaped border that is open on the left and right sides. The logo is simple and modern.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074
vitro cytotoxicity
- ISO 10993-6:2016, Biological evaluation of medical devices Part 6: Tests for local effects after implantation
- · ISO 10993-10:2010, Biological evaluation of medical devices. Tests for irritation and skin sensitization
- ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- ISO 7405:2018, Dentistry Evaluation of biocompatibility of medical devices used in dentistry
| Product Name | EsCem | G-CEM LinkAce |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | N/A | K120243 |
| Manufacturer | SPIDENT CO., LTD. | GC Corporation |
| Product description | EsCem is a dual-cured resin cement<br>that can be self-etched and self-<br>adhesive and is classified as type 2<br>and class 3 according to ISO 4049.<br>It is easy to handle with 'one step'<br>system that does not need pre-<br>treatment process, and it is easy to<br>remove excess cement, so it is<br>possible to perform quick<br>operation. It is also easy to identify<br>because it has a radiopaque. And it<br>offers various shades for aesthetics. | GC G-CEM LinkAce is a dual-cure<br>self-adhesive universal resin luting<br>cement delivered in double barrel<br>automix syringe, designed for the<br>adhesive luting of all-ceramic,<br>metal or composite indirect<br>restorations. |
| Chemical<br>composition of<br>Resin | Methacrylate monomer<br>Filler<br>Initiator<br>Inhibitor<br>pigment | Methacrylate monomer<br>Filler<br>Initiator<br>Inhibitor<br>pigment |
| Indications for use | Final cementation of all-ceramic,<br>metal-based inlays, onlays, posts,<br>crown and bridge | - Cementation of all types of all<br>ceramic, resin and metal-based<br>inlays, onlays, crowns and bridges.<br>- Cementation of metal, ceramic,<br>fibre posts, and cast post and cores. |
| Principle of<br>operation | Dual cured | Dual cured |
## 8. Substantial Equivalence Discussion
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SPIDENT. The logo is green and features the word "SPIDENT" in a bold, sans-serif font. The word is enclosed in a square bracket that is missing the top right corner. The logo is simple and modern.
SPIDENT CO., JLTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, NamdongGu, Incheon, Korea 405-821
Tel : +82-32-821-0071 Fax : + 82-32-821-0074
| Mechanical<br>Properties | | |
|------------------------------|-------------------|---------------------------|
| • Setting time (sec) | Less than 600 sec | Less than 600 sec |
| • Flexural strength<br>(MPa) | More than 50 | More than 50 |
| • Film thickness<br>(μm) | Less than 50 | Less than 50 |
| • Water sorption<br>(µg/mm³) | Less than 40 | Less than 40 |
| • Solubility<br>(µg/mm³) | Less than 7.5 | Less than 7.5 |
| • Radio-opacity<br>(mmAl) | More than 1 mm | More than 1 mm |
| Standard<br>conformed | ISO 4049 | ISO 4049 |
| Biocompatibility | Yes | Yes |
| Light Curing Time | 20 Seconds | 20 Seconds |
| Application Area | Tooth | Tooth |
| Target Population | Dental patient | Dental patient |
| Intended Operator | Dentist | Dentist |
| Storage condition | 2-27°C | 4-25°C |
| Shade | A2, A3, TW | A2, AO3, BO1, Translucent |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Spident. The logo is green and features the word "SPIDENT" in a sans-serif font. The letters are all capitalized and are a light shade of green. The word is enclosed in a square border that is also green.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074
## 9. Similarities and Differences with Marketed Devices:
The subject device, EsCem and the predicate device, G-CEM LinkAce (GAM-200) was compared as above. Both products have the similar product description as a dental cement. In case of composition, two products contain slightly different raw materials, but the subject device and the predicate device show the same performance as monomer, filler, photo inhibitor (stabilizer), self-cure initiator, photo initiator and pigment. For this reason, despite difference in material, both the subject device and the equivalent device have no problem in terms of the clinical performance.
Indication for use of the two products are almost same and both products can be used as dual cure. The performance results of the two products are not the same, but both products meet the requirements of ISO 4049. Biocompatibility, Light Curing Time, Application Area, Target Population, Intended Operator of both products are the same. In case of Storage condition and Shade, there is a slight difference but performance and biocompatibility test showed that these differences would not raise any new questions of safety and effectiveness. Therefore, EsCem is substantially equivalent with predicate device, G-CEM LinkAce, and at least as safe and effective as the predicate device.
#### 10. Conclusion
Based on a comparison of subject device, EsCem (SPIDENT CO., LTD) and predicate device, G-CEM LinkAce (GC America INC.), it is confirmed that the subject device is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.