Self-curing temporary crown resin that is used temporarily before repairing a permanent restoration
Device Story
Hexa-Temp is a self-curing temporary crown and bridge resin used by dentists in clinical settings. The device consists of a methacrylate matrix (Bis-GMA, TEGDMA) and inorganic fillers (size 16nm to 3.22µm). It is applied to the tooth to fabricate temporary restorations. The material cures via a self-curing process. The output is a temporary dental restoration that provides esthetic coverage and protection for the tooth while a permanent restoration is prepared. The device benefits patients by maintaining tooth function and esthetics during the interim period between procedures.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 10477:2018 (Dentistry Polymer-based crown and veneering materials). Mechanical properties compared to predicate include water sorption, solubility, polymerization temperature, polymerization shrinkage, compressive strength, and flexural strength. Biocompatibility evaluated per ISO 10993 series (parts 1, 3, 5, 6, 10, 11, 18) and ISO 7405:2018. Results demonstrate performance comparable or superior to the predicate device.
Technological Characteristics
Self-curing methacrylate-based resin (Bis-GMA, TEGDMA). Contains ~38% inorganic filler (16nm to 3.22µm). Conforms to ISO 10477:2018. Mechanical properties: Compressive strength 330.38 MPa, Flexural strength 69.93 MPa. Storage 2-27°C. No software or electronic components.
Indications for Use
Indicated for dental patients requiring temporary crowns, bridges, inlays, or onlays prior to permanent restoration.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 10, 2022
Spident Co., Ltd. J.M. Ahn President 203 & 312, Korea Industrial Complex, 722 Gojan-Dong, Namdong-Gu, Incheon 405-821 Korea, South
Re: K214071
Trade/Device Name: Hexa-Temp Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: December 21, 2021 Received: December 27, 2021
Dear J.M. Ahn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214071
Device Name Hexa-Temp
Indications for Use (Describe)
Self-curing temporary crown resin that is used temporarily before repairing a permanent restoration
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SPIDENT. The logo is green and features the word "SPIDENT" in a sans-serif font. The letters are all capitalized and are a bright green color. The logo is enclosed in a green box that is open on the top and left sides.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82-32-821-0071 Fax : +82-32-821-0074
## Section 5. 510(k) Summary
K214071
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date: December 21, 2021
- 1. Company and Correspondent making the submission: Company Name : SPIDENT CO., LTD. Address : 203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea, 405-821 Tel : +82-32-821-0071 Fax : + 82-32-821-0074 Company Contact : J.M. Ahn / President
## 2. Device Name and Classification
Proprietary Name : Hexa-Temp Common name : Temporary crown and bridge resin Classification name : Temporary crown and bridge resin [CFR 872.3770] Product code : EBG Class : II
### 3. Predicate Devices (Legally Marketed Devices)
The predicate devices for Hexa-Temp is :
- · STRUCTUR 2 SC, VOCO GMBH, K040769
### 4. Description:
Hexa-Temp is temporary crown and bridge resin that is used temporarily before repairing a permanent restoration. It can be self-cured resin. Filler size ranges from 0.7 to 1.2 µm, resulting in good radio-opacity and high flexural strength. Various shades of TW, A1, A2, A3, B1, C2, AO3 and BL for esthetic crown are provided. Hexa-Temp contains about 38% inorganic filler. Inorganic filler has a size of 16nm ~ 3.22µm.
#### 5. Indications for Use
Self-curing temporary crown resin that is used temporarily before repairing a permanent restoration
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for SPIDENT. The logo is green and features the word "SPIDENT" in a sans-serif font. The word is enclosed in a square with a break in the upper left corner. The logo is simple and modern.
#### 6. Performance Testing - Bench
The performance test was conducted to prove the substantially equivalence of the subject device, and the standards applied to the performance test are as follows.
- · ISO 10477:2018, Dentistry Polymer-based crown and veneering materials.
#### 7. Biocompatibility
Biocompatibility tests and biological safety evaluation were performed and the results proved that the subject device is at least as biocompatible and biologically safe as the predicate device.
The standards applied are as follows.
- · ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- · ISO 10993-3:2014, Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- · ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- · ISO 10993-6:2016, Biological evaluation of medical devices Part 6: Tests for local effects after implantation
- · ISO 10993-10:2010, Biological evaluation of medical devices. Tests for irritation and skin sensitization
- · ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- · ISO 10993-18:2020, Biological evaluation of medical devices Part 18: Chemical characterization of medical device materials within a risk management process
- · ISO 7405:2018, Dentistry Evaluation of biocompatibility of medical devices used in dentistry
| Product Name | Hexa-Temp | Structur 2 SC |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | N/A | K040769 |
| Manufacturer | SPIDENT CO., LTD. | VOCO GMBH |
| Product<br>description | Hexa-Temp is temporary crown and<br>bridge resin that is used temporarily<br>before repairing a permanent<br>restoration. It can be self-cured<br>resin. Filler size ranges from 0.7 to<br>1.2 µm, resulting in good radio-<br>opacity and high flexural strength.<br>Various shades of TW, A1, A2, A3,<br>B1, C2, AO3 and BL for esthetic<br>crown are provided. Hexa-Temp<br>contains about 38% inorganic filler.<br>Inorganic filler has a size of 16nm<br>~ 3.22µm. | Structur 2 SC is a fluorescent self-<br>curing paste-paste system for the<br>production of temporary crowns,<br>bridges, inlays and onlays. Structur<br>2 SC consists of base paste and<br>catalyst paste. |
| Chemical<br>composition of<br>Resin | Methacrylate matrix(Bis-GMA,<br>TEGDMA) | Methacrylates, amines, terpenes,<br>benzoyl peroxide and BHT |
| Indications for use | Self-curing temporary crown resin<br>that is used temporarily before<br>repairing a permanent restoration | Structur 2SC is indicated for the<br>fabrication of temporary crowns,<br>bridges, inlays and onlays. |
| Principle of<br>operation | Self-cured | Self-cured |
| Mechanical<br>Properties<br>•Water sorption<br>(µg/mm³) | 24.89 | 14.59 |
| •Solubility<br>(µg/mm³) | 3.85 | 3.15 |
| •Color and color<br>consistency | There is Color consistency and<br>Color stability. | There is Color consistency and<br>Color stability. |
| •Polymerization<br>Temperature | Did not exceed 45°C | Did not exceed 45°C |
| •Polymerization<br>Shrinkage (%) | 2.63 | 2.50 |
| •Compressive<br>Strength<br>(MPa) | 330.38 | 294.82 |
| •Flexural Strength<br>(MPa) | 69.93 | 86.42 |
| •Surface finish | It's smooth and glossy. | It's smooth and glossy. |
| Standard<br>conformed | ISO 10477 | ISO 10477 |
| Biocompatibility | Yes | Yes |
| Application Area | Tooth | Tooth |
| Target Population | Dental patient | Dental patient |
| Intended<br>Operator | Dentist | Dentist |
| Storage condition | 2-27°C | 4-23°C |
| Shade | TW, A1, A2, A3, B1, C2, AO3 and<br>BL | A1, A2, A3, A3.5, B1, B3, C2 and<br>BL |
#### 8. Substantial Equivalence Discussion
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Spident. The logo is green and features the word "SPIDENT" in a sans-serif font. The letters are enclosed in a square border that is open at the top and bottom. The logo is simple and modern.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82-32-821-0071 Fax : +82-32-821-0074
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "SPIDENT" in green font. The word is partially enclosed in a green square. The square is open on the top and left sides, and the word is positioned inside the square.
# SPIDENT CO., LTD.
203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82-32-821-0071 Fax : +82-32-821-0074
#### 9. Similarities and Differences with Marketed Devices:
The subject device, Hexa-Temp and the predicate device, Structur 2 SC was compared as above. Both products have the similar product description as a temporary crown and bridge resin.
In case of composition, two products contain slightly different raw materials, but the subject device and the predicate device show the same performance as methacrylic monomer and inhibitor, initiator, pigment, filler. Filler information of predicate device is unknown, but it will be similar as the general characteristics of the temporary crown and bridge resin.
Indication for use of the two products are almost same and both products can be used as self-cure.
As the mechanical properties, clinically important items of the temporary crown resin were selected: Water sorption, Solubility, Color and Color stability, Polymerization temperature, Polymerization shrinkage, Compressive strength, and Flexural strength. Among them,
Water sorption, Solubility, Color and Color stability, and Flexural strength belong to ISO 10477 items.
The performance results of the two products are not the same, but both products meet the requirements of ISO 10477 and it can be seen that the main clinical performance is comparable or superior.
For this reason, despite difference in material, both the subject device and the equivalent device have no problem in terms of the clinical performance.
Biocompatibility, Application Area, Target Population, Intended Operator of both products are the same.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SPIDENT. The logo is green and features the word "SPIDENT" in a sans-serif font. The letters are enclosed in a square with a break in the upper left corner. The logo is simple and modern.
SPIDENT CO., LTD. 203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82-32-821-0071 Fax : +82-32-821-0074
In case of Storage condition and Shade, there is a slight difference but performance and biocompatibility test showed that these differences would not raise any new questions of safety and effectiveness. Therefore, Hexa-Temp is substantially equivalent with predicate device, Structur 2 SC, and at least as safe and effective as the predicate device.
#### 10. Conclusion
Based on a comparison of subject device, Hexa-Temp (SPIDENT CO., LTD) and predicate device, Structur 2 SC (VOCO GMBH), it is confirmed that the subject device is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.