K214070 · Tela Bio · FTM · Mar 21, 2023 · General, Plastic Surgery
Device Facts
Record ID
K214070
Device Name
OviTex PRS (Long Term Resorbable)
Applicant
Tela Bio
Product Code
FTM · General, Plastic Surgery
Decision Date
Mar 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OviTex PRS (Long-Term Resorbable) is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one time use.
Device Story
Surgical mesh composed of multi-layer ovine forestomach matrix sheets sewn with resorbable PLGA suture; used for soft tissue reinforcement in plastic and reconstructive surgery. Device implanted by surgeons to provide structural support where tissue weakness exists. Benefits include reinforcement of surgical sites; resorption over time. Available in 2-8 ply configurations and various shapes/sizes up to 434 cm².
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing (mechanical strength, suture retention, compliance), biocompatibility (ISO 10993-1), toxicological risk assessment (ISO 10993-18), and a 78-week animal implantation study (ISO 10993-6) demonstrating complete resorption and non-irritant local tissue response.
Indicated for patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery where soft tissue weakness exists.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 21, 2023
TELA Bio John Urtz Senior Manager, Regulatory and Quality 1 Great Valley Parkway, Suite 24 Malvern, Pennsylvania 19355
Re: K214070
Trade/Device Name: OviTex PRS (Long Term Resorbable) Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM, FTL, FTM, FTL Dated: December 23, 2021 Received: December 27, 2021
Dear John Urtz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Deborah A. Fellhauer -S
Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214070
Device Name OviTex PRS (Long-Term Resorbable)
Indications for Use (Describe)
OviTex PRS (Long-Term Resorbable) is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one time use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Disposition (Part 1 CFR 301 Subpart A) |
| Sale or Transfer (19 CFR 301 Subpart C) |
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
| Device trade name: | OviTex PRS (Long-Term Resorbable) |
|---------------------|---------------------------------------------------------------------------------------------------------|
| Device common name: | Mesh, surgical |
| Regulation number: | §878.3300, Surgical Mesh |
| Product code(s): | FTM / FTL |
| Predicate device: | Endoform Restella (K183398) |
| Reference device: | Endoform Reconstructive Template (Non-Absorbable)<br>(K181935) |
| Owner / Submitter: | TELA Bio, Inc.<br>1 Great Valley Parkway<br>Suite 24<br>Malvern, PA 19355 |
| Contact Person: | John Urtz<br>Sr. Manager, Quality and Regulatory<br>TELA Bio, Inc.<br>jurtz@telabio.com<br>484-320-2884 |
| Date prepared: | 16 March 2023 |
## Device Description
OviTex PRS (Long-Term Resorbable) is a surqical mesh manufactured by layering sheets of ovine forestomach matrix to create multi-layer configurations of devices sewn together with resorbable Poly(lactic-co-glycolic Acid) ("PLGA") suture for use in plastic and reconstructive surgery. The 2-8 ply devices are available in surface areas up to 434 cm² in various shapes.
## Intended Use / Indications for Use
OviTex PRS (Long-Term Resorbable) is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one time use.
## Summary/Comparison of Technological Characteristics
OviTex PRS (Long-Term Resorbable) has the same indications for use as the predicate device, Endoform Restella. The principle of operation is also the same.
The subject device, OviTex PRS (Long-Term Resorbable) has similar technological characteristics as the predicate device. Both devices are composed of sheets of ovine forestomach matrix and are sewn together with a polymeric suture. The main differences between the subject and predicate device are the polymeric suture material (PLGA in the subject device vs. Polyglycolic Acid ("PGA") or Polypropylene ("PP") in the predicate device) and the directionality of the ovine tissue fenestrations (bi-directional for the subject device and uni-directional for the predicate device). The subject device is offered in a broader range of shapes, sizes, and thicknesses compared to the predicate, but within the range of
{4}------------------------------------------------
sizes/thicknesses of the reference device. OviTex PRS (Long-Term Resorbable) maintains the same fundamental technological characteristics as the predicate device with respect to material types, biocompatibility, device specifications, and sterilization.
| | | | Reference Device |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | Endoform |
| | | | Reconstructive |
| | Subject Device | Predicate Device | Template (Non- |
| Characteristic | OviTex PRS (LTR) | Endoform Restella | absorbable) |
| 510(k) Number | K214070 | K183398 | K181935 |
| Product Code | FTM / FTL | FTM / FTL | FTL / FTM |
| Regulation<br>Number | §878.3300, Surgical<br>Mesh | §878.3300, Surgical<br>Mesh | §878.3300, Surgical<br>Mesh |
| Common Name | Mesh, surgical | Mesh, surgical | Mesh, surgical |
| Indications for<br>Use | OviTex PRS (Long-Term Resorbable) is intended for use in the reinforcement of soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use. | Endoform Restella is intended for use in the reinforcement of soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use. | Endoform Reconstructive Template - Non-Absorbable is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing or bridging material to obtain the desired surgical outcome. |
| Animal Derived<br>Tissue | Ovine forestomach<br>matrix | Ovine forestomach<br>matrix | Ovine forestomach<br>matrix |
| Polymeric Suture | Poly(lactic-co-glycolic Acid) | Polyglycolic Acid OR<br>Polypropylene | Polypropylene |
| Layering<br>Configurations | 2-8 layers | 3 layers | 1-10 layers |
| Maximum Device<br>Size | 434cm² | 399cm² | 1000cm² |
| Shapes Offered | Rectangle, Contour,<br>Oval | Rectangle, Contour | Rectangle |
| Tissue<br>Fenestrations | Bi-directional | Uni-directional | None |
| Biocompatibility | Established | Established | Established |
| Characteristic | Subject Device<br>OviTex PRS (LTR) | Predicate Device<br>Endoform Restella | Reference Device<br>Endoform<br>Reconstructive<br>Template (Non-absorbable) |
| Endotoxin<br>Content | < 20 EU/device | < 20 EU/device | < 20 EU/device |
| Sterilization<br>Method and<br>Sterility<br>Assurance Level<br>(SAL) | Ethylene Oxide, 10-6 | Ethylene Oxide, 10-6 | Ethylene Oxide, 10-6 |
| Usage | Single use only | Single use only | Single use only |
{5}------------------------------------------------
## Non-clinical Performance Data
Preclinical testing was conducted on OviTex PRS (Long-Term Resorbable) devices to demonstrate substantial equivalence to the predicate device. New bench testing included mechanical strength, suture retention, endotoxin, and compliance testing. Results of the testing confirms that the proposed device meets all product specifications.
Biocompatibility assessment and testing was performed in accordance with ISO 10993-1 on the final, finished device. Additionally, extractable and Leachable (E&L) testing was conducted on the final finished PLGA polymeric suture. A comprehensive assessment of each of the characterized chemicals obtained from this testing was incorporated into a toxicological risk assessment (TRA) regarding the finished subject device in accordance with ISO 10993-18. The results of all performed biocompatibility testing and the accompanying toxicological risk assessment establish the biocompatibility of the OviTex PRS (LTR) device family.
An implantation study was conducted in accordance with ISO 10993-6 for a permanent implant contacting tissue which compared the local tissue response of the subject device to that of the predicate device. The results of this study showed that all devices had advanced to complete resorption by 78 weeks and that the subject device was substantially equivalent to the predicate device, classified as a non-irritant in terms of the local tissue response.
### Clinical Performance Data
Substantial equivalence was not based on an assessment of clinical performance data.
### Conclusion
The results of preclinical, mechanical, and biocompatibility testing demonstrate that OviTex PRS (Long-Term Resorbable) is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.