FEops HEARTguide™ is indicated for patient-specific simulation of transcatheter left atrial appendage occlusion (LAAO) device implantation during procedural planning. The software performs computer simulation to predict implant frame deformation to support the evaluation for LAAO device size and placement. FEops HEARTguide™ is intended to be used by qualified clinicians in conjunction with the simulated device instructions for use, the patient's clinical history, symptoms, and other preprocedural evaluations, as well as the clinician's professional judgment. FEops HEARTguide™ is not intended to replace the simulated device instructions for use for final LAAO device selection and placement. FEops HEARTguide™ is prescription use only.
Device Story
FEops HEARTguide is a software-based simulation tool for LAAO procedural planning. Input: 2D multi-slice cardiac CT images. Process: FEops case analysts generate a 3D patient-specific anatomical model and combine it with a predefined computational model of the LAAO device; simulation predicts implant frame deformation. Output: 2D/3D/numerical data provided via PDF report and web-based viewer. Context: Operated by trained FEops analysts at FEops infrastructure; results accessed by clinicians via web browser. Usage: Clinicians use output alongside patient history and device IFU to assist in LAAO device selection and placement. Benefit: Provides pre-procedural insights into device-anatomy interaction to support clinical decision-making.
Clinical Evidence
Bench testing only. Performance validation included computational modeling credibility assessment, software verification/validation, and human factors evaluation. Performance studies for the two added LAAO devices used cohorts with sample sizes equal to or larger than the predicate. Accuracy and clinical meaningfulness were demonstrated by comparing simulation results to clinical data across the full intended operating range, using identical statistical analysis and acceptance criteria as the predicate.
Indicated for patient-specific simulation of transcatheter left atrial appendage occlusion (LAAO) device implantation during procedural planning to predict implant frame deformation and support device size/placement evaluation. Intended for use by qualified clinicians.
Regulatory Classification
Identification
An interventional cardiovascular implant simulation software device is a prescription device that provides a computer simulation of an interventional cardiovascular implant device inside a patient's cardiovascular anatomy. It performs computational modeling to predict the interaction of the interventional cardiovascular implant device with the patient-specific anatomical environment.
Special Controls
In combination with the general controls of the FD&C Act, the interventional cardiovascular implant simulation software device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Software verification, validation, and hazard analysis, with identification of appropriate mitigations, must be performed, including a full verification and validation of the software according to the predefined software specifications.
(2) Computational modeling verification and validation activities must be performed to establish the predictive capability of the device for its indications for use.
(3) Performance validation testing must be provided to demonstrate the accuracy and clinical relevance of the modeling methods for the intended implantation simulations, including the following:
(i) Computational modeling results must be compared to clinical data supporting the indications for use to demonstrate accuracy and clinical meaningfulness of the simulations;
(ii) Agreement between computational modeling results and clinical data must be assessed and demonstrated across the full intended operating range (
*e.g.,* full range of patient population, implant device sizes and patient anatomic morphologies). Any selection criteria or limitations of the samples must be described and justified;(iii) Endpoints (
*e.g.,* performance goals) and sample sizes established must be justified as to how they were determined and why they are clinically meaningful; and(iv) Validation must be performed and controls implemented to characterize and ensure consistency (
*i.e.,* repeatability and reproducibility) of modeling outputs:(A) Testing must be performed using multiple qualified operators and using the procedure that will be implemented under anticipated conditions of use; and
(B) The factors (
*e.g.,* medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data.(4) Human factors evaluation must be performed to evaluate the ability of the user interface and labeling to allow for intended users to correctly use the device and interpret the provided information.
(5) Device labeling must be provided that describes the following:
(i) Warnings that identify anatomy and image acquisition factors that may impact simulation results and provide cautionary guidance for interpretation of the provided simulation results;
(ii) Device simulation inputs and outputs, and key assumptions made in the simulation and determination of simulated outputs; and
(iii) The computational modeling performance of the device for presented simulation outputs, and the supporting evidence for this performance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2022
FEops nv % Niels Festjens RA Consultant OrthoGrow nv Davincilaan 1 Zaventem, Flemish-Brabant 1930 Belgium
Re: K214066
Trade/Device Name: FEops HEARTguide Regulation Number: 21 CFR 870.1405 Regulation Name: Interventional cardiovascular implant simulation software device Regulatory Class: Class II Product Code: OOI Dated: December 23, 2021 Received: December 27, 2021
Dear Niels Festjens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K214066
Device Name FEops HEARTguide
#### Indications for Use (Describe)
FEops HEARTguide™ is indicated for patient-specific simulation of transcatheter left atrial appendage occlusion (LAAO) device implantation during procedural planning.
The software performs computer simulation to predict implant frame deformation to support the evaluation for LAAO device size and placement.
FEops HEARTguide™ is intended to be used by qualified clinicians in conjunction with the simulated device instructions for use, the patient's clinical history, symptoms, and other preprocedural evaluations, as well as the clinician's professional judgment.
FEops HEARTguide™ is not intended to replace the simulated device instructions for use for final LAAO device selection and placement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/3 description: The image shows the logo for FEOPS. The logo consists of the word "FEOPS" in large, bold, black letters, followed by the tagline "insights for excellence" in a smaller, lighter font. To the right of the text is a vertical line, followed by a black speech bubble containing a white icon that resembles an open book or a stylized pair of brackets.
# 510(K) SUMMARY (21CFR807.92)
## SUBMITTER
Company Name: Establishment registration number: Address:
Phone number: Principal contact person: Principal contact e-mail address Additional contact person: Additional contact e-mail address Summary date:
### DEVICE
Name & trade name: Classification name:
FEops nv 3020703662 Technologiepark 122, 9052 Gent - Zwijnaarde, Belgium +32486435671 Niels Festjens niels@ortho-grow.com Peter Mortier peter.mortier@feops.com
FEops HEARTguide™ Interventional Cardiovascular Implant Simulation Software Device
## PREDICATE DEVICE
The predicate device to which substantial equivalence is claimed:
| Trade or proprietary or model name | FEops HEARTguide™ |
|------------------------------------|-------------------|
| De Novo class II petition: | DEN200030 |
| Decision date | 08 SEP 2021 |
| Classification product code | QQI |
| Regulation Number | 870.1405 |
| Manufacturer | FEops NV |
{4}------------------------------------------------
# DESCRIPTION AND FUNCTIONING OF THE DEVICE
FEops HEARTguide™ predicts implant frame deformation after percutaneous LAAO device implantation through computer simulation. The predicted deformation provides additional information during LAAO procedural planning.
The simulation is based on a 3D model of the patient anatomy which is generated from 2D medical images of the patient anatomy (multi-slice Cardiac Computed Tomography). The simulation is executed by FEops Case Analysts and run on FEops infrastructure.
The simulation report is created by combining a predefined device model with a patient-specific model of the patient anatomy. This is performed by trained operators at FEops using an internal software platform through an established workflow. The purposely qualified case analysts and quality control analysts process the received medical images of the patient to produce the simulation results.
The simulation results are provided as 2D and numerical data shown in a PDF report and 3D, 2D and numerical data shown in a web-based Viewer application accessible through a standard web browser.
# INTENDED USE
FEops HEARTguide™ is indicated for patient-specific simulation of transcatheter left atrial appendage occlusion (LAAO) device implantation during procedural planning.
The software performs computer simulation to predict implant frame deformation to support the evaluation for LAAO device size and placement.
FEops HEARTguide™ is intended to be used by qualified clinicians in conjunction with the simulated device instructions for use, the patient's clinical history, symptoms, and other preprocedural evaluations, as well as the clinician's professional judgment.
FEops HEARTguide™ is not intended to replace the simulated device instructions for use for final LAAO device selection and placement.
FEops HEARTguide™ is prescription use only.
Image /page/4/Picture/13 description: The image is a logo for FEOPS, which stands for "insights for excellence". The logo consists of the word "FEOPS" in large, bold, black letters, followed by the phrase "insights for excellence" in smaller, gray letters. To the right of the text is a vertical line, followed by a black speech bubble containing a white icon that resembles an open book.
{5}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS
The subject device shares the same code base as the predicate device and differs in the addition of computer models for two additional LAAO devices, the Abbott Amplatzer Amulet (AMU) and the Boston Scientific Watchman FLX (FLX).
| | Predicate device | Subject Device | Comparison |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | FEops HEARTguide™ | FEops HEARTguide™ | / |
| 510(k)/De<br>Novo Number | DEN200030 | K214066 | / |
| Manufacturer | FEops nv | FEops nv | Same |
| Regulation<br>Number | 21 CFR 870.1405 | 21 CFR 870.1405 | Same |
| Device<br>Classification<br>Name | Interventional<br>cardiovascular<br>implant<br>simulation software device | Interventional<br>cardiovascular<br>implant<br>simulation software device | Same |
| Common name | FEops HEARTguide™ | FEops HEARTguide™ | Same |
| Product Code | QQI | QQI | Same |
| Intended<br>Use/Indications<br>for Use | FEops HEARTguide is<br>indicated for patient-specific<br>simulation of transcatheter<br>left atrial appendage<br>occlusion (LAAO) device<br>implantation during<br>procedural planning.<br><br>The software performs<br>computer simulation to<br>predict implant frame<br>deformation to support the<br>evaluation for LAAO device<br>size and placement.<br><br>FEops HEARTguide is<br>intended to be used by<br>qualified clinicians in<br>conjunction with the<br>simulated device<br>instructions-for-use, the<br>patient's clinical history,<br>symptoms, and other | FEops HEARTguide is<br>indicated for patient-specific<br>simulation of transcatheter<br>left atrial appendage<br>occlusion (LAAO) device<br>implantation during<br>procedural planning.<br><br>The software performs<br>computer simulation to<br>predict implant frame<br>deformation to support the<br>evaluation for LAAO device<br>size and placement.<br><br>FEops HEARTguide is<br>intended to be used by<br>qualified clinicians in<br>conjunction with the<br>simulated device<br>instructions-for-use, the<br>patient's clinical history,<br>symptoms, and other | Same |
| | | | |
| | preprocedural evaluations, as well as the clinician's professional judgment.<br><br>FEops HEARTguide is not intended to replace the simulated device's instructions for use for final LAAO device selection and placement.<br><br>FEops HEARTguide is prescription use only. | preprocedural evaluations, as well as the clinician's professional judgment.<br><br>FEops HEARTguide is not intended to replace the simulated device's instructions for use for final LAAO device selection and placement.<br><br>FEops HEARTguide is prescription use only. | |
| Prescription<br>Use only | Yes | Yes | Same |
| Software build | 1.4.1 | 1.4.2 | Changes limited to extension of device library and optimization of FLX and AMU modelling strategy |
| Modelling<br>strategy | Device specific computational model applied to patient-specific geometry | Device specific computational model applied to patient-specific geometry | Same |
| Software<br>architecture | Simulation results prepared by FEops Case Analysts and made available in Web based Viewer | Simulation results prepared by FEops Case Analysts and made available in Web based Viewer | Same |
| Simulated<br>devices | 1. Boston Scientific WATCHMAN (P130013) | 1. Boston Scientific WATCHMAN (P130013)<br>2. Boston Scientific WATCHMAN FLX (P130013/S035)<br>3. Abbott Amplatzer | The subject device library has 2 additional LAA occluders to select. All 3 included LAA occluders are approved for use on the US market. |
Image /page/5/Picture/5 description: The image shows the logo for FEOPS, which is written in a bold, sans-serif font. Below the logo, the words "insights for excellence" are written in a smaller, lighter font. The logo is simple and modern, and the words below it suggest that the company is focused on providing insights that help its clients achieve excellence.
Image /page/5/Picture/6 description: The image shows a black and white logo. The logo is a speech bubble with a bracket symbol inside. The bracket symbol is white and the speech bubble is black.
{6}------------------------------------------------
The provided detailed comparison demonstrates the subject device is substantially equivalent in intended use, design, operating principles, materials and performance characteristics to the primary predicate device. The difference in technological characteristics do not raise new questions of safety and effectiveness.
Image /page/6/Picture/4 description: The image shows the logo for FEOPS, which is a company that provides insights for excellence. The logo consists of the word "FEOPS" in large, bold, black letters, with the words "insights for excellence" in smaller, gray letters below. To the right of the word "FEOPS" is a vertical line, and to the right of the vertical line is a black speech bubble with a white icon inside. The icon inside the speech bubble looks like a stylized open book.
{7}------------------------------------------------
# PERFORMANCE DATA
The 510(k)-submission includes data to demonstrate the Special Controls defined in 21CFR870.1405 are implemented, including:
- -Software verification, validation, and hazard analysis, with identification of appropriate mitigations, including full verification and validation of the software according to the predefined software specifications with detail expected for a moderate level of concern software,
- A detailed credibility assessment of the device models demonstrating computational modeling verification and validation activities have been performed to establish the predictive capability of the device for its indications for use. The applied methods and results are similar to the activities performed for the predicate device,
- The provided performance validation testing data shows the applied methods are similar to the methods applied for the design validation of the predicate device and a similar performance level has been demonstrated. The data includes a comparison of the results to clinical data supporting the indications for use to demonstrate accuracy and clinical meaningfulness of the simulations, an assessment of the agreement between the computational model results and clinical data across the full intended operating range, a justification of the endpoints and sample sizes determination and clinical meaningfulness
- To ensure consistency of modeling outputs, the validation was performed with multiple qualified operators using the procedure that will be implemented under anticipated conditions of use and the factors which were held constant where identified.
- -A detailed description of the computations and statistical analyses used to evaluate the data is included in the 510(k) submission,
- A Human factors evaluation report was provided demonstrating the ability of the user interface and labeling to allow for intended and qualified users to correctly use the device and interpret the provided information
- Device labeling is included in the submission which includes warnings that identify anatomy and image acquisition factors that may impact simulation results, provide cautionary guidance for interpretation of the provided simulation results, describes the device simulation inputs and outputs, and key assumptions made in the simulation and determination of simulated outputs as well as the computational modeling performance of the device for presented simulation outputs, and the supporting evidence for this performance
For both added LAAO devices, the performance study was performed on a cohort with a sample size equal to or larger than the predicate device. Acceptance criteria were defined using the same method as for the predicate device demonstrating the same clinical meaningfulness. The same type of measurements and an identical statistical analysis was performed. This demonstrates the performance of the subject device is equivalent to the performance of the predicate device.
{8}------------------------------------------------
## SUMMARY
The characteristics that determine the functionality and performance of FEops HEARTguide™, the subject device, are substantially equivalent to the predicate device cleared under DEN200030. The testing indicates that the subject device is as safe, as effective, and performs as well as the predicate.
Image /page/8/Picture/4 description: The image shows the logo for FEOPS. The logo consists of the word "FEOPS" in large, bold, black letters, with the tagline "insights for excellence" in smaller, gray letters underneath. To the right of the wordmark is a vertical gray line, followed by a black speech bubble containing a white icon that resembles an open book.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.