JJGC Indústria e Comércio de Materiais Dentários S.A.
Product Code
KCT · General Hospital
Decision Date
Apr 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Helix NGM Compact Surgical Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre-vacuum)- Exposure at 132 °C for 4 minutes, 20-minute dry time. Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time. Neodent Instrument Kit Cases are intended for sterilization of non-porous loads. The combined weight of the GM Surgical Kit Case and the associated instruments is 268.0 g. The weight of the empty Kit Case is 208.15 grams. Neodent Instrument Kit Cases should not be stacked during sterilization.
Device Story
Reusable rigid container system; comprises base, removable inner tray, and lid. Manufactured from autoclavable polysulfone and polyphenylsulfone; includes medical-grade silicone grommets for instrument retention. Used by healthcare providers to organize, protect, and sterilize dental surgical/prosthetic instruments. Input: instruments placed in tray; device acts as a protective, perforated barrier during steam sterilization cycles (fractionated vacuum or gravity displacement). Output: sterilized, organized instrument kit. Facilitates safe transport and storage post-sterilization. Requires FDA-cleared sterilization wrap. Benefits: ensures instrument organization and sterility maintenance for surgical procedures.
Clinical Evidence
Bench testing only. Validated via manual cleaning (blood soil removal), sterilization efficacy (ANSI/AAMI/ISO 17665-1/2), life-cycle testing (100 cycles of cleaning/sterilization), and biocompatibility (ANSI/AAMI/ISO 10993-5). All tests passed acceptance criteria.
Technological Characteristics
Rigid container system; materials: polysulfone (base), polyphenylsulfone (lid), medical-grade silicone (grommets). Perforated design for steam penetration. Dimensions: 195 x 90 x 54 mm. Reusable (up to 100 cycles). Sterilization: moist heat (steam) via fractionated vacuum or gravity displacement. No software or electronics.
Indications for Use
Indicated for organization, storage, transport, and steam sterilization of surgical and prosthetic instruments in healthcare settings. Intended for non-porous loads. Contraindicated for stacking during sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
April 13, 2022
JJGC Indústria e Comércio de Materiais Dentários S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K214051
Trade/Device Name: Neodent Implant System-Helix NGM Compact Surgical Kit Cases Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: Class II Product Code: KCT Dated: March 18, 2022 Received: March 23, 2022
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214051
#### Device Name
Neodent Implant System - Helix NGM Compact Surgical Kit Cases
Indications for Use (Describe)
Helix NGM Compact Surgical Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures.
Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:
Fractionated vacuum (pre-vacuum)- Exposure at 132 °C for 4 minutes, 20-minute dry time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.
Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.
The combined weight of the GM Surgical Kit Case and the associated instruments is 268.0 g. The weight of the empty Kit Case is 208.15 grams. Neodent Instrument Kit Cases should not be stacked during sterilization.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
## ADMINISTRATIVE INFORMATION
| Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Parana, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs,<br>Straumann USA<br>E-mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509 |
| Date Prepared | 5/Apr/2022 |
| Preparer / Alternate Contact | Mariana Hartmann<br>Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA<br>E-mail: mariana.hartmann@neodent.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/ Proprietary Name | Neodent Implant System – Helix NGM Compact Surgical Kit Cases |
|------------------------------|-------------------------------------------------------------------------------------------------|
| Common Name | Instrument Sterilization Trays |
| Classification Name | Sterilization Wrap Containers, Trays, Cassettes & Other |
| Classification Regulations | 21 CFR 880.6850, Class II |
| Product Code | KCT |
| Classification Panel | General Hospital |
| Reviewing Branch | Infection Control Devices Branch |
| PREDICATE DEVICE INFORMATION | |
| Primary Predicate Device | K192670 - Neodent Instrument Kit Cases, JJGC Indústria e<br>Comércio de Materiais Dentários S.A |
{4}------------------------------------------------
#### INDICATIONS FOR USE
#### Indications for Use for Helix NGM Compact Surgical Kit Case and Pre-Mounted Helix NGM Compact Surgical Kit Case:
Helix NGM Compact Surgical Kit Cases are indicated for organization of surgical and/or prosthetic instruments during sterilization, storage and transport. The use of this product facilitates storage and organization of instruments during and after surgical procedures.
Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:
Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20-minute dry time.
Gravity displacement – Exposure at 132 °C for 15 minutes, 20-minute dry time.
Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.
The combined weight of the GM Surgical Kit Case and the associated instruments is 268.0 g. The weight of the empty Kit Case is 208.15 grams. Neodent Instrument Kit Cases should not be stacked during sterilization.
#### SUBJECT DEVICE DESCRIPTION
The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject kit cases are manufactured in autoclavable polymer. The design of the subject devices include grommets manufactured from medical grade silicone that retain the instruments within the tray. They also presents markings that guide instrument use during procedures. The subject device kit cases are provided nonsterile to the end-user.
The dimensions for each part of the model and the overall dimensions are presented in the table below:
| Assembled<br>Kit Case | Description | Assembled<br>Kit Case<br>Dimension<br>(L x W x H) | Component<br>Number | Component Dimension<br>(L x W x H) |
|-----------------------|----------------------------------------------------|---------------------------------------------------|---------------------|------------------------------------|
| 110.315 | Helix NGM Compact<br>Surgical Kit Case | 195 x 90 x<br>54 mm | 212.293 (Lid) | 195 x 90 x 36 mm |
| | | | 704346 (Tray) | 180 x 76 x 16.5 mm |
| | | | 704345 (Base) | 188 x 84 x 32 mm |
| 110.316 | Pre-Mounted Helix NGM<br>Compact Surgical Kit Case | 195 x 90 x<br>54 mm | 212.293 (Lid) | 195 x 90 x 36 mm |
| | | | 704346 (Tray) | 180 x 76 x 16.5 mm |
| | | | 704345 (Base) | 188 x 84 x 32 mm |
Note: The instrument and accessory devices that are sterilized and stored within the subject Kit Cases are not themselves subject devices of this submission.
{5}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Helix NGM Compact Surgical Kit Cases<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K192670<br>Neodent Instrument Kit Cases<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | COMPARISON |
| Indications for<br>Use Statement | Indications for Use for Helix NGM Compact<br>Surgical Kit Case<br>Helix NGM Compact Surgical Kit Cases are<br>indicated for organization of surgical and/or<br>prosthetic instruments during sterilization,<br>storage and transport. The use of this<br>product facilitates storage and organization<br>of instruments during and after surgical<br>procedures.<br>Neodent Instrument Kit Cases are intended<br>to allow sterilization of the enclosed medical<br>devices. Neodent Instrument Kit Cases<br>require the use of FDA cleared wrap to<br>maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization<br>wrap that is FDA-cleared for the indicated<br>cycles, and moist heat (steam) sterilized<br>using one of the following cycles:<br>Fractionated vacuum (pre-vacuum)<br>Exposure at 132 °C for 4 minutes, 20-minute<br>dry time.<br>Gravity displacement — Exposure at 132 °C<br>for 15 minutes, 20-minute dry time.<br>Neodent Instrument Kit Cases are intended<br>for sterilization of non-porous loads.<br>The combined weight of the GM Surgical Kit<br>Case and the associated instruments is 268.0<br>g. The weight of the empty Kit Case is 208.15<br>grams. Neodent Instrument Kit Cases should<br>not be stacked during sterilization.<br>Indications for Use for Pre-Mounted Helix<br>NGM Compact Surgical Kit Case<br>Product indicated for conditioning and safe<br>fastening surgical and/or prosthetic<br>instruments during their use and<br>sterilization. The use of this product<br>facilitates storage and organization of<br>instruments during and after surgical<br>procedures.<br>Neodent Instrument Kit Cases are intended<br>to allow sterilization of the enclosed medical<br>devices. Neodent Instrument Kit Cases<br>require the use of FDA cleared wrap to<br>maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization<br>wrap that is FDA-cleared for the indicated<br>cycles, and moist heat (steam) sterilized<br>using one of the following cycles:<br>Fractionated vacuum (pre-vacuum)<br>Exposure at 132 °C for 4 minutes, 20-minute<br>dry time.<br>Gravity displacement — Exposure at 132 °C<br>for 15 minutes, 20-minute dry time.<br>Neodent Instrument Kit Cases are intended<br>for sterilization of non-porous loads.<br>The combined weight of the GM Surgical Kit<br>Case and the associated instruments is 268.0<br>g. The weight of the empty Kit Case is 208.15<br>grams. Neodent Instrument Kit Cases should | Indications for Use for GM Surgical Kit Case:<br>Neodent Instrument Kit Cases are intended to<br>be used to enclose other medical devices that<br>are to be sterilized by a health care provider.<br>Neodent Instrument Kit Cases are intended to<br>allow sterilization of the enclosed medical<br>devices. Neodent Instrument Kit Cases require<br>the use of FDA cleared wrap to maintain the<br>sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization<br>wrap that is FDA-cleared for the indicated<br>cycles, and moist heat (steam) sterilized using<br>one of the following cycles:<br>Fractionated vacuum (pre-vacuum) — Exposure<br>at 132 °C for 4 minutes, 20-minute dry time.<br>Gravity displacement — Exposure at 132 °C for<br>15 minutes, 40-minute dry time.<br>Neodent Instrument Kit Cases are intended for<br>sterilization of non-porous loads.<br>The combined weight of the GM Surgical Kit<br>Case and the associated instruments is 613.1 g.<br>The weight of the empty Kit Case is 510 grams.<br>Neodent Instrument Kit Cases should not to be<br>stacked during sterilization.<br>Indications for Use for GM Helix Compact<br>Surgical Kit Case:<br>Neodent Instrument Kit Cases are intended to<br>be used to enclose other medical devices that<br>are to be sterilized by a health care provider.<br>Neodent Instrument Kit Cases are intended to<br>allow sterilization of the enclosed medical<br>devices. Neodent Instrument Kit Cases require<br>the use of FDA cleared wrap to maintain the<br>sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization<br>wrap that is FDA-cleared for the indicated<br>cycles, and moist heat (steam) sterilized using<br>one of the following cycles:<br>Fractionated vacuum (pre-vacuum) — Exposure<br>at 132 °C for 4 minutes, 20-minute dry time<br>Gravity displacement — Exposure at 132 °C for<br>15 minutes, 20-minute dry time.<br>Neodent Instrument Kit Cases are intended for<br>sterilization of non-porous loads.<br>The combined weight of the GM Helix Compact<br>Surgical Kit Case and the associated<br>instruments is 308.2 g. The weight of the empty<br>Kit Case is 231 grams. Neodent Instrument Kit<br>Cases should not to be stacked during<br>sterilization. | Similar |
| | not be stacked during sterilization. | | |
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | |
| | Helix NGM Compact Surgical Kit Cases<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K192670<br>Neodent Instrument Kit Cases<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | COMPARISON |
| Intended Use | The Helix Narrow GM Compact Surgical<br>Kit Cases is suitable for safe storage of<br>surgical and prosthetic instruments,<br>organizing the instruments in the<br>sequence in which they will be used, as<br>well as for sterilization of such<br>instruments. | The GM Helix Compact Surgical Kit Case is<br>suitable for safe storage of surgical and<br>prosthetic instruments, organizing the<br>instruments in the sequence in which they<br>will be used, as well as for sterilization of<br>such instruments. | Same |
| Design | Rigid polysulfone polymer base and<br>removable inner tray with a<br>polyphenylsulfone lid.<br>Retention grommets of medical grade<br>silicone. | Rigid polysulfone polymer base and<br>removable inner tray with a<br>polyphenylsulfone lid.<br>Retention grommets of medical grade<br>silicone. | Identical |
| Perforated | Yes; allows moist heat (steam)<br>penetration to achieve sterilization | Yes; allows moist heat (steam)<br>penetration to achieve sterilization | Identical |
| Reusable | Yes | Yes | Identical |
| Overall<br>dimensions | 110.315: 195 L x 90 W x 54 H, mm<br>110.316: 195 L x 90 W x 54 H, mm | 110.288: 264 L x 163 W x 54 H, mm<br>110.297: 195 L x 90 W x 64 H, mm | Similar |
| Volume to Vent<br>Ratio | 110.315: 52.3 cm³/cm² (20.59 in³/in²)<br>110.316: 52.3 cm³/cm² (20.59 in³/in²) | 110.288: 98.04 cm³/cm² (38.6 in³/in²)<br>110.297: 63.5 cm³/cm² (25.0 in³/in²) | Similar |
| Useful Life | Yes, reusable up to 100 cycles | Yes, reusable up to 100 cycles…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.