HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Stryker Sterizone VP4 Sterilizer Cycle 1. The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Stryker Sterizone VP4 Sterilizer Cycle 1.
Device Story
Sterilization wrap system consisting of two sheets of SMS (spunbond-meltblown-spunbond) polypropylene fabric ultrasonically bonded on two edges; used by healthcare providers to wrap medical devices prior to sterilization; compatible with Stryker Sterizone VP4 Sterilizer Cycle 1 (vaporized hydrogen peroxide and ozone); allows sterilant penetration while maintaining package sterility for up to 12 months post-sterilization; facilitates aseptic opening of sterilized packages.
Clinical Evidence
Bench testing only. Evaluated sterilant penetration/efficacy (ANSI/AAMI ST79, ISO 11138-7), physical properties (ISO 11607-1, ISO 13938-2, ASTM D4966-12, CPSC 1610), package integrity (ANSI/AAMI/ISO 11607-1), and biocompatibility (ISO 10993 series). All tests passed, confirming 10^-6 sterility assurance level and 12-month shelf life.
Technological Characteristics
Three-layer SMS polypropylene fabric; antistatic treatment; ultrasonically bonded; dimensions vary by grade (H100-H600). Compatible with vaporized hydrogen peroxide and ozone sterilization. Complies with ANSI/AAMI/ISO 11607-1.
Indications for Use
Indicated for healthcare providers to enclose medical devices for sterilization using the Stryker Sterizone VP4 Sterilizer Cycle 1. Applicable to general instruments, rigid/semi-rigid channeled instruments, and flexible endoscopes.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 27, 2022
Owens & Minor (O&M) Halyard, Inc. Anureet Singh Regulatory Affairs Specialist II 9120 Lockwood Blvd Mechanicsville, Virginia 23116
# Re: K214007
Trade/Device Name: HALYARD ONE-STEP Sterilization Wrap, HALYARD QUICK CHECK Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: June 29, 2022 Received: June 30, 2022
Dear Anureet Singh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K214007
#### Device Name
HALYARD* ONE-STEP* Sterilization Wrap and HALYARD* QUICK CHECK* Sterilization Wrap
## Indications for Use (Describe)
HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Stryker Sterizone VP4 Sterilizer Cycle 1.
The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Stryker Sterizone VP4 Sterilizer Cycle 1. All models of the Halyard ONE-STEP and QUICK CHECK Sterilization Wrap have been validated for use with the Stryker Sterilizer Cycle 1 as described below:
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
| Type of Device | HALYARD ONE-STEP* and QUICK CHECK* Validation Loads Description<br>in the Stryker Sterizone VP4 Sterilizer Cycle 1 | Validated<br>Sterilization<br>Wrap GradesB |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| General<br>instruments | Load consisted of three (3) trays each containing three (3) lb of general medical<br>instruments. The load included:<br>General devices representing the following geometries:<br>Box-lock hinge<br>Pivot hinge<br>Luer-lock<br>Testing was performed with trays wrapped in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load (excluding the loading rack). The load included:<br>General devices representing the following geometries:<br>Box-lock hinge<br>Pivot hinge<br>Luer-lock<br>Testing was performed with trays wrapped in H600 sterilization wrap. | H100C<br>H200<br>H300<br>H400<br>H500<br>H600C |
| Rigid channel<br>instruments | Load #4A - Consisted of 15 lumens from rigid and semi-rigid channeled devices. The load<br>included:<br>Three (3) double channel (six (6) lumens) semi-rigid endoscopes (ureteroscope - 0.7 mm × 500 mm and 1.1 mm × 500 mm) were packaged in three (3)<br>sterilization trays including appropriate silicone brackets. Additional rigid channel instruments (nine (9) lumens) were added. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
| Single and double<br>channel flexible<br>endoscopes | Load #8A - Consisted of five (5) lumens from single and double channel flexible<br>endoscopes. The load included:<br>Two (2) double channel flexible endoscopes (ureteroscope) with inside<br>diameter of 1 mm and lengths of 850 mm and 989 mm; One (1) single channel flexible endoscope (ureteroscope) with inside diameter<br>of 1 mm and length of 850 mm; The endoscopes were individually packaged in sterilization trays, including<br>appropriate silicone brackets. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
| Multi-channel<br>flexible<br>endoscopes | Load #9A - Consisted of one (1) multichannel flexible endoscope with four (4) channels.<br>The load included:<br>One (1) multi flexible endoscope (colonoscope) with no more than four (4)<br>channels having inside diameter of 1.2 mm and lengths of 1955 mm or inside<br>diameter of 1.45 mm and lengths of 3500 mm, packaged individually in a<br>sterilization tray. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
A: Representative Sterizone® VP4 Sterilizer Cycle 1 Valifand* ONE-STEP and QUICK CHECK* Sterliization Wrap are comprised of two sheets of Halyard Sequential Sterilization Wrap ultrasonically bonded together on two sides. Therefore, these grades are applicable to HALYARD* ONE-STEP*, QUICK CHECK* and Sequential Sterilization Wrap. C: Indicates the bracketed grades for validation testing.
Type of Use (Select one or both, as applicable)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K214007
| Submitter: | O&M Halyard, Inc.<br>9120 Lockwood Boulevard<br>Mechanicsville, VA 23116<br>Phone: 804-723-7000/800-488-8850<br>Fax: 804-723-7100 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Contact: | Anureet Singh<br>Regulatory Affairs Specialist II<br>O&M Halyard, Inc.<br>1 Edison Drive<br>Alpharetta, GA 30005 |
| Date of Summary: | 25 July 2022 |
| Device Trade Name: | HALYARD* ONE-STEP* Sterilization Wrap and HALYARD* QUICK CHECK*<br>Sterilization Wrap |
| Common Name: | Sterilization Wrap |
| Classification Name: | Sterilization wrap (21 CFR 880.6850, Product Code FRG) |
| Predicate Device: | Halyard ONE-STEP Sterilization Wrap, K192147 |
| Device Description: | Halyard Sterilization Wrap is supplied to the customer as bulk packages of single<br>sheets, where in accordance with standard hospital practices, two sheets are then<br>used to wrap a medical device or a collection of medical devices for sterilization.<br>HALYARD ONE-STEP and HALYARD QUICK CHECK Sterilization Wraps<br>are comprised of two sheets of HALYARD* Sequential Sterilization Wrap<br>ultrasonically seamed on two edges. This allows for convenient wrapping with<br>two sheets simultaneously.<br><br>Sterilization wrap is a square or rectangular sheet made of three-layer SMS<br>(spunbond-meltblown-spunbond) polypropylene fabric treated with an antistatic<br>treatment. The wrap allows a sterilized package to be opened aseptically.<br><br>The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are<br>intended to enclose another medical device that is to be sterilized by a healthcare<br>provider using the Stryker Sterizone VP4 Sterilizer Cycle 1 which is a dual-<br>sterilant, low temperature sterilizer that uses vaporized hydrogen peroxide (H2O2)<br>and ozone (O3) in a single cycle. All models of the Halyard ONE-STEP and<br>QUICK CHECK Sterilization Wrap have been validated for use with the Stryker<br>Sterizone VP4 Sterilizer Cycle 1 as described below. |
| Indication for Use: | HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended<br>to enclose another medical device that is to be sterilized by a healthcare provider<br>using the single cycle Stryker Sterizone VP4 Sterilizer Cycle 1.<br><br>The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are<br>intended to allow sterilization of the enclosed devices by the Stryker Sterizone |
{5}------------------------------------------------
VP4 Sterilizer Cycle 1. All models of the Halyard ONE-STEP and QUICK CHECK Sterilization Wrap have been validated for use with the Stryker Sterizone VP4 Sterilizer Cycle 1 as described below:
| Type of Device | HALYARD ONE-STEP* and QUICK CHECK* Validation Loads Description<br>in the Stryker Sterizone VP4 Sterilizer Cycle 1 | Validated<br>Sterilization<br>Wrap GradesB |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| General<br>instruments | Load consisted of three (3) trays each containing three (3) lb of general medical<br>instruments. The load included:<br>General devices representing the following geometries: Box-lock hinge Pivot hinge Luer-lock Testing was performed with trays wrapped in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load (excluding the loading rack). The load included: General devices representing the following geometries: Box-lock hinge Pivot hinge Luer-lock Testing was performed with trays wrapped in H600 sterilization wrap. | H100C<br>H200<br>H300<br>H400<br>H500<br>H600C |
| Rigid channel<br>instruments | Load #4A - Consisted of 15 lumens from rigid and semi-rigid channeled devices. The load<br>included: Three (3) double channel (six (6) lumens) semi-rigid endoscopes (ureteroscope<br>- 0.7 mm × 500 mm and 1.1 mm × 500 mm) were packaged in three (3)<br>sterilization trays including appropriate silicone brackets. Additional rigid channel instruments (nine (9) lumens) were added. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
| Single and double<br>channel flexible<br>endoscopes | Load #8A - Consisted of five (5) lumens from single and double channel flexible<br>endoscopes. The load included: Two (2) double channel flexible endoscopes (ureteroscope) with inside<br>diameter of 1 mm and lengths of 850 mm and 989 mm; One (1) single channel flexible endoscope (ureteroscope) with inside diameter<br>of 1 mm and length of 850 mm; The endoscopes were individually packaged in sterilization trays, including<br>appropriate silicone brackets. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
| Multi-channel<br>flexible<br>endoscopes | Load #9A - Consisted of one (1) multichannel flexible endoscope with four (4) channels.<br>The load included: One (1) multi flexible endoscope (colonoscope) with no more than four (4)<br>channels having inside diameter of 1.2 mm and lengths of 1955 mm or inside<br>diameter of 1.45 mm and lengths of 3500 mm, packaged individually in a<br>sterilization tray. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C |
A: Representative Sterizone® VP4 Sterilizer Cycle 1 Validation Load from K172191. B'HALYARD* ONE-STEP* and QUICK CHECK* Sterilization Wrap are comprised of two sheets of Halyard Sequential Sterilization Wrap ultrasonically bonded together on two sides are applicable to HALYARD* ONE-STEP*, QUICK CHECK* and Sequential Sterilization Wrap. C: Indicates the bracketed grades for validation testing.
{6}------------------------------------------------
| | Proposed<br>HALYARD ONE-STEP and<br>QUICK CHECK Sterilization Wrap<br>(K214007) | Predicate<br>HALYARD ONE-STEP<br>Sterilization Wrap<br>(K192147) | Same<br>Similar, or<br>Different |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Manufacturer | O&M Halyard, Inc. | O&M Halyard, Inc. | Same |
| Device Model<br>Numbers | H100<br>H200<br>H300<br>H400<br>H500<br>H600 | H100<br>H200<br>H300<br>H400<br>H500<br>H600 | Same |
| Common or<br>Usual Name | Sterilization Wrap | Sterilization Wrap | Same |
| Classification | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| Class | II | II | Same |
| Product Code | FRG | FRG | Same |
| Indication for<br>Use | HALYARD ONE-STEP and QUICK<br>CHECK Sterilization Wraps are<br>intended to enclose another medical<br>device that is to be sterilized by a<br>healthcare provider using the Stryker<br>Sterizone VP4 Sterilizer Cycle 1.<br>The HALYARD ONE-STEP and<br>QUICK CHECK Sterilization Wraps<br>are intended to allow sterilization of<br>the enclosed devices by the Stryker<br>Sterizone VP4 Sterilizer Cycle 1. All<br>models of the Halyard ONE-STEP and<br>QUICK CHECK Sterilization Wrap<br>have been validated for use with the<br>Stryker Sterizone VP4 Sterilizer Cycle<br>1 as described below:<br><br>Type of Device: General instruments<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load consisted of three (3) trays each<br>containing three (3) lb of general medical<br>instruments. The load included general<br>devices representing the following<br>geometries: Box-lock hinge, Pivot hinge,<br>Luer-lock.<br>Testing was performed with trays wrapped<br>in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays<br>each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load<br>(excluding the loading rack). The load<br>included general devices representing the<br>following geometries: Box-lock hinge,<br>Pivot hinge, Luer-lock.<br>Testing was performed with trays wrapped<br>in H600 sterilization wrap. | The Halyard ONE-STEP Sterilization<br>Wraps are intended to allow<br>sterilization of the enclosed devices by<br>the Sterilucent HC 80TT Hydrogen<br>Peroxide Sterilizer (ie, both the Lumen<br>and Flexible Cycles) Additionally, the<br>Halyard ONE-STEP Sterilization<br>Wrap was validated to allow effective<br>aeration under the pre-programmed<br>HC 80TT Sterilization Cycles. All<br>models of the Halyard ONE-STEP<br>Sterilization Wrap have been validated<br>for use with the Sterilucent HC 80TT<br>Hydrogen Peroxide Sterilizer cycles as<br>described below.<br>Lumen Cycle<br>Reusable metal and nonmetal devices<br>including devices with diffusion-<br>restricted spaces such as the hinged<br>portion of forceps and scissors and up<br>to fifteen (15) stainless steel lumens<br>per load with the following<br>dimensions:<br>Single or dual channeled rigid and<br>semi-rigid endoscopes, with stainless<br>steel lumens that are<br>$\ge$ 0.77 mm internal diameter (ID) and<br>$\le$ 410 mm long, or $\ge$ 1.33 mm ID and $\le$<br>430 mm long;<br>and, Triple channeled rigid and semi-<br>rigid endoscopes, with stainless steel<br>lumens that are $\ge$ 1.00 mm ID and $\le$<br>310 mm long (Refer to the HC 80TT<br>User Manual for complete instructions<br>on load(s) and cycle(s), including<br>chamber loading instructions (i.e. 20.1<br>lb per load)<br>Flexible Cycle | Different |
Technological Characteristics Comparison Table:
{7}------------------------------------------------
| Validated Sterilization Wrap GradesB:<br>H100C, H200, H300, H400, H500, H600C<br><br>Type of Device: Rigid channel<br>instruments<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #4A - Consisted of 15 lumens from<br>rigid and semi-rigid channeled devices.<br>The load included:<br>Three (3) double channel (six (6) lumens)<br>semi-rigid endoscopes (ureteroscope - 0.7<br>mm × 500 mm and 1.1 mm × 500 mm)<br>were packaged in three (3) sterilization<br>trays including appropriate silicone<br>brackets.<br>Additional rigid channel instruments (nine<br>(9) lumens) were added.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C | Reusable rigid or semi-rigid non-<br>lumen medical devices including non-<br>lumen devices with metallic diffusion-<br>restricted spaces such or mated<br>surfaces such as the hinged portion of<br>forceps or scissors;<br>Single channel flexible endoscopes<br>with flexible lumens that are ≥ 1.00<br>mm ID and ≤ 1280 mm long; and Dual<br>channel flexible endoscopes with<br>flexible lumens that are ≥ 0.80 mm ID<br>and ≤ 1000 mm long.<br>(Refer to the HC 80TT User Manual<br>for complete instructions on load(s)<br>and cycle(s), including chamber<br>loading instructions (i.e. 25 lb per<br>load). |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Device: Single and double<br>channel flexible endoscopes<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #8A - Consisted of five (5) lumens<br>from single and double channel flexible<br>endoscopes. The load included: Two (2)<br>double channel flexible endoscopes<br>(ureteroscope) with inside diameter of 1<br>mm and lengths of 850 mm and 989 mm;<br>One (1) single channel flexible endoscope<br>(ureteroscope) with inside diameter of 1<br>mm and length of 850 mm; The<br>endoscopes were individually packaged in<br>sterilization trays, including appropriate<br>silicone brackets.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C | |
| Type of Device: Multi-channel flexible<br>endoscopes<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #9A - Consisted of one (1)<br>multichannel flexible endoscope with four<br>(4) channels. The load included: One (1)<br>multi flexible endoscope (colonoscope)<br>with no more than four (4) channels having<br>inside diameter of 1.2 mm and lengths of<br>1955 mm or inside diameter of 1.45 mm | |
{8}------------------------------------------------
| | and lengths of 3500 mm, packaged<br>individually in a sterilization tray.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C<br><br>A: Representative Sterizone® VP4<br>Sterilizer Cycle 1 Validation Load from<br>K172191. B: HALYARD* ONE-STEP*<br>and QUICK CHECK* Sterilization Wrap<br>are comprised of two sheets of Halyard<br>Sequential Sterilization Wrap<br>ultrasonically bonded together on two<br>sides. Therefore, these grades are<br>applicable to HALYARD* ONE-STEP*,<br>QUICK CHECK* and Sequential<br>Sterilization Wrap. C: Indicates the<br>bracketed grades for validation testing. | | |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Sterilization<br>Parameters | Stryker Sterizone VP4 Sterilizer Cycle<br>1 which is a dual-sterilant, low<br>temperature sterilizer that uses<br>vaporized hydrogen peroxide (H2O2)<br>and ozone (O3) in a single cycle | Sterilucent PSD-85 Hydrogen<br>Peroxide Sterilizer that includes<br>● Lumen Cycle<br>● Flexible Cycle | Different |
| Technology | Tortuous sheet material used to<br>enclose medical devices that are to be<br>sterilized by a healthcare provider to<br>allow sterilization of the enclosed<br>medical device(s) and maintain<br>sterility of the enclosed device(s) until<br>used | Tortuous sheet material used to<br>enclose medical devices that are to be<br>sterilized by a healthcare provider to<br>allow sterilization of the enclosed<br>medical device(s) and maintain<br>sterility of the enclosed device(s) until<br>used | Same |
| Device Design | Two sheets of nonwoven<br>polypropylene fabric. Each sheet is<br>composed of three thermally-bonded<br>layers consisting of a meltblown<br>polypropylene layer surrounded by<br>spunbond polypropylene layers (SMS) | Two sheets of nonwoven<br>polypropylene fabric. Each sheet is<br>composed of three thermally-bonded<br>layers consisting of a meltblown<br>polypropylene layer surrounded by<br>spunbond polypropylene layers (SMS) | Same |
| Method for<br>Bonding SMS<br>Layers | Thermal bonding with round pin,<br>hexagonal, triangle bond pattern<br>( | | |
## Summary of Performance Testing
Performance Testing
(Bench):
Performance testing of HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap was evaluated and the results showed that acceptance criteria were met demonstrating that the HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap allows sterilization of its contents using the Stryker Sterizone VP4 Sterilizer Cycle 1 and that sterility is maintained for the testing period of 12 months.
{9}------------------------------------------------
| Purpose | Test | Acceptance Criteria | Results |
|-----------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------|
| Sterilant<br>Penetration/Efficacy | ANSI/AAMI ST79<br>ANSI/AAMI/ISO 11138-7 | Achieving a 10-6 sterility assurance<br>level following processing in a worst-case half-cycle | Passed |
| Performance<br>Testing (Non-sterile<br>and Sterile) | ANSI/AAMI/ISO 11607-1 Annex B<br>ISO 13938-2<br>ASTM D4966-12<br>CPSC 1610 | Complies with the selected physical<br>properties | Passed |
| Maintenance of<br>Package Integrity | ANSI/AAMI/ISO 11607-1<br>ANSI/AAMI ST79 | Maintain sterility for up to 12 months | Passed |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-7 | Non-cytotoxic<br>Non-irritating<br>H2O2 ≤ 0.56 μg/cm2 | Passed |
## Summary of Non-Clinical Testing Performed
## Performance Testing (Clinical): Clinical evaluations were not required and therefore are not submitted with this 510(k). Discussion: The HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap in this submission and the predicate device submission are intended to enclose another medical device that is to be sterilized by a healthcare provider, to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). The ONE-STEP and QUICK CHECK Sterilization Wrap in this submission and the predicate device submission have identical intended use, design, materials, specifications, and composition, and are manufactured using identical production methods. The different technological characteristics, that is, the Indication for Use and the Sterilization Parameters, do not affect the safety and effectiveness of the device as evidenced by the results of the nonclinical testing. Overall Performance Conclusions: The conclusions drawn from the nonclinical tests demonstrate that the HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap (K214007) is as safe, as effective, and performs as well as or better than the legally marketed HALYARD ONE-STEP Sterilization Wrap (K192147).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.