The FLOW COUPLER Device and System is intended to be used in the anastomosis of veins and arteries normally encountered in microvascular and vascular reconstructive procedures and in the detection of blood flow and confirmation of vessel patency following end-to-end anastomosis of vessels.
Device Story
System comprises single-use implantable rings and removable 20MHz ultrasonic Doppler probe; used with portable monitor. Probe press-fits onto rings to secure anastomosis; monitor provides audible pulsed Doppler ultrasound signal indicating blood flow. Used in hospital OR and PACU by surgeons and nurses. Monitor processes Doppler phase shift via analog front end; includes FFT filtering and noise reduction to produce audible output. Healthcare providers use audio feedback to confirm vessel patency and monitor for flow changes. Subject device removes remote monitoring capability present in predicate; includes hardware/firmware updates for power management, power button ergonomics, and audio system clarity.
Clinical Evidence
Bench testing only. Testing included cleaning validation (IEC 60601-1), design inspection/functional demonstration of power switch, audio quality testing (SNR > 20, THD within specs), and EMC testing (IEC 60601-1-2 Ed 4.1). All tests met predefined acceptance criteria.
Technological Characteristics
20MHz pulsed wave Doppler ultrasound; 156.25 KHz PRF. Powered by 12VDC (AC/DC converter) or internal Li-ion battery. Dimensions: 6.17" x 8.18" x 3.20". Audio system: 4W speaker, convex grill. Connectivity: Standalone (remote monitoring disabled). Sterilization: Not specified (implantable rings). Software: Firmware-based signal processing (analog front end, phase shift detection, FFT filtering).
Indications for Use
Indicated for use in end-to-end anastomosis of veins and arteries in microsurgical and vascular reconstructive procedures. Used with FLOW COUPLER Monitor to detect blood flow and confirm vessel patency intra-operatively and post-operatively (up to 7 days). Doppler probe is not a permanent implant; must be removed 3-14 days post-operatively.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 20, 2022
Synovis Micro Companies Alliance, Inc. Julie Carlston Senior Manager, Regulatory Affairs (a Subsidiary of Baxter International Inc.) 439 Industrial Lane Birmingham, Alabama 35211
Re: K213974
Trade/Device Name: GEM FLOW COUPLER Monitor (GEM1020M-2) Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW, MVR Dated: March 11, 2022 Received: March 14, 2022
Dear Julie Carlston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Deborah Fellhauer, RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213974
Device Name GEM FLOW COUPLER Monitor (GEM1020M-2)
Indications for Use (Describe)
The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
-Submitter Synovis Micro Companies Alliance, Inc. (A Subsidiary of Baxter International, Inc.) 439 Industrial Lane Birmingham, AL 35211-4464 Establishment Registration Number: 1062741
Julie Carlston Senior Regulatory Affairs Manager at Synovis Micro Companies Alliance, Inc. (a subsidiary of Baxter International Inc.) Telephone: 651.796.7511 Fax: 651.642.9018 Email: Julie Carlston@baxter.com
Date Prepared: April 14, 2022
### II. Subject Device
| Device Trade Name | GEM FLOW COUPLER Monitor (GEM1020M-2) |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Cardiovascular Blood Flowmeter |
| Classification Name | Flowmeter, Blood, Cardiovascular<br>Regulation Number: 21 CFR 870.2100<br>Classification: Class II<br><br>Implantable Clip<br>Regulation Number: 21 CFR 878.4300<br>Classification: Class II |
| Product Code | MVR, DPW |
# III. Predicate Device
| Device Trade Name | GEM FLOW COUPLER System; K191252 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Cardiovascular Blood Flowmeter |
| Classification Name | Flowmeter, Blood, Cardiovascular<br>Regulation Number: 21 CFR 870.2100<br>Classification: Class II<br>Implantable Clip<br>Regulation Number: 21 CFR 878.4300<br>Classification: Class II |
| Product Code | MVR, DPW |
{4}------------------------------------------------
# IV. Device Description
The GEM FLOW COUPLER System is designed for the detection of blood flow in vessels. The GEM FLOW COUPLER System consists of components as described below:
1. FLOW COUPLER Device (Product Code MVR) - single-use implantable rings used to secure anastomosis of small vessels. This includes a 20MHz ultrasonic Doppler transducer (probe) which attaches to one of the FLOW COUPLER rings, and an external lead.
FLOW COUPLER Monitor (Product Code DPW) - portable monitor that provides the 2. audible output of the FLOW COUPLER Device's pulsed Doppler ultrasound signal. A varying audible signal is produced when the FLOW COUPLER Device (probe) detects flow.
The predicate device (K191252) submission included an additional remote monitoring capability for the remote retrieval of FLOW COUPLER Monitor data via the GEMflow App (which is part of the GEM Cloud System). This optional remote monitoring capability was not extensively utilized and has been excluded from the subject device. The remote monitoring capability is not available with the subject devic.
# V. Intended Use / Indications for Use
The Intended Use and Indications for Use remain the same as the predicate device (K191252).
#### Intended Use
The FLOW COUPLER Device and System is intended to be used in the anastomosis of veins and arteries normally encountered in microvascular and vascular reconstructive procedures and in the detection of blood flow and confirmation of vessel patency following end-to-end anastomosis of vessels.
#### Indications for Use
The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intraoperatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days postoperatively.
# VI. Comparison of Technological Characteristics with the Predicate Device
The overall functionality and features necessary to support the device's intended use remain unchanged from the previous device clearance (K191252). Minor modifications to the product hardware, firmware and labeling have been made in this submission:
> . Power management update (monitor firmware update): revised logic for handling a battery-protection shutdown scenario.
{5}------------------------------------------------
Power button update (monitor hardware update): recessed the button to be . flush with the device bezel and increased the contact force.
Audio system update (monitor hardware update): updated speaker assembly ● and capacitors on Printed Circuit Board Assembly to reduce background noise and enhance overall audio clarity.
List of approved cleaning solutions update (labeling update): updated . Instructions for Use with an expanded list of approved cleaning solutions.
. Disabling of the optional remote monitoring capability: updated monitor firmware to disable the connection to the cloud backend server (GEM Cloud System) to disable the optional remote monitoring capability.
The technological characteristics and specifications of the FLOW COUPLER Monitor have been evaluated against its predicate device (previous version of the subject device) to determine equivalence.. A summary of the comparison between the predicate and subject devices is provided in Table 1 below.
| Attribute | Predicate Device | Subject Device |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | GEM FLOW COUPLER System | Same |
| 510(k)-holder | Synovis Micro Companies Alliance, Inc.<br>(A Subsidiary of Baxter International, Inc.) | Same |
| 510(k) Number | K191252 | K213974 |
| Product Code | MVR, DPW | Same |
| Regulation | 21 CFR 878.4300<br>21 CFR 870.2100 | Same |
| Classification | Class II | Same |
| Intended Use | The FLOW COUPLER Device and System is intended to be used<br>in the anastomosis of veins and arteries normally encountered<br>in microvascular and vascular reconstructive procedures and<br>in the detection of blood flow and confirmation of vessel<br>patency following end-to-end anastomosis of vessels. | Same |
| Indications for Use | The FLOW COUPLER Device is a single use, implantable device<br>that is intended to be used in the end-to-end anastomosis of<br>veins and arteries normally encountered in microsurgical and<br>vascular reconstructive procedures. The FLOW COUPLER<br>Device includes a pair of permanently implanted rings which<br>secure the anastomosis and a removable Doppler probe that is<br>press-fit onto one of the rings. When the FLOW COUPLER<br>Device is used in conjunction with the FLOW COUPLER<br>Monitor, the FLOW COUPLER System is intended to detect<br>blood flow and confirm vessel patency intra-operatively and<br>post-operatively at the anastomotic site. Post-operatively,<br>blood flow can be detected on an as needed basis for up to 7<br>days. The FLOW COUPLER Doppler probe is not intended to be<br>a permanent implant and should be removed 3 to 14 days<br>postoperatively. | Same |
| Product Use<br>Environment | Hospital Operating Room and Post-Anesthesia Care Unit | Same |
| End Users of Product | Nurses and surgeons | Same |
| Dimensions | 6.17" D x 8.18" W x 3.20" H | Same |
| Weight | 1.8 lb. (0.8 kg) | Same |
| Acoustical Output | Acoustical Power will be less than 94 mW/cm2 | Same |
| Depth | Not selectable | Same |
| Transmission<br>Frequency | 20 MHz | Same |
| Transmission<br>Characteristic | Pulsed wave transmission, continuous reception | Same |
| Pulse Repetition<br>Frequency | 156.25 KHz | Same |
| Energy Used and/or<br>Delivered | 12 Volts direct current supplied by A/C to D/C converter<br>Rechargeable internal lithium ion battery pack | Same |
| Power Input | Input Voltage: 90-264 VAC<br>Frequency: 47-63 Hz | Same |
| Power Output | Total Output Power: 30 watts<br>Voltage: 12VDC<br>Current: 2.5A<br>Output Regulation ± 5% | Same |
| Power management update | | |
| Monitor firmware | Analog Front End, Doppler phase shift detection, plus FFT<br>filtering, noise reduction, and background noise reduction | Revised logic for handling a<br>proper battery-protection<br>shutdown scenario to prevent<br>complete battery drainage and<br>inoperable monitor |
| Power button update | | |
| Power button | Location: rear of unit<br>Elevation: raised (3.50 mm) | Location: rear of unit<br>Elevation: flush with device<br>bezel<br>Activation force: 400gf |
| Audio system update | | |
| Printed Circuit Board<br>Assembly (PCBA) | PCBA per K191252 | PCBA components have been<br>changed to improve audio<br>quality |
| Speaker Assembly | 57.5mm wide, 3W Max. Power speaker<br>Concave speaker grill | 60mm wide, 4W Max. Power<br>speaker<br>Convex speaker grill |
| List of approved cleaning solutions update | | |
| Cleaning instructions | Wipe the Monitor with a dry or water-moistened soft cloth.<br>Ensure any residual organic material is removed.<br>Do not pour or spray liquid directly on the Monitor. Allow to<br>air dry before use. | Wipe the Monitor with a dry or<br>water-moistened soft cloth,<br>Isopropyl Alcohol, Ammonium<br>Hydroxide based surface<br>cleanser, Ammonium Chloride<br>based surface cleanser or 2%<br>bleach solution. Ensure any<br>residual organic material is<br>removed. Do not pour or spray<br>liquid directly on the Monitor.<br>Allow to air dry before use. |
| Disabling of the optional remote monitoring capability | | |
| Remote Access<br>Function (Optional) | Android and iOS App to access audio files remotely | Disabled optional remote<br>monitoring capability |
Table 1 - FLOW COUPLER Monitor Predicate and Subject Devices - Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
#### VII.Performance Data
Synovis conducted performance testing to evaluate safety and effectiveness of the FLOW COUPLER Monitor to support substantial equivalence to the predicate device, as well as to evaluate that product requirements are met. The following testing of the FLOW COUPLER Monitor has been performed.
| Performance Testing | Acceptance Criteria | Result |
|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| FLOW COUPLER Monitor cleaning per<br>IEC 60601-1:2005+AMD1:2012 section<br>11.6.6<br>FLOW COUPLER Monitor Cleaning<br>Conditioning | Passing inspection for the damage and post-test leakage and dielectric test. Functioning monitor after the cleaning. | Pass |
| FLOW COUPLER Monitor Design<br>Inspection and Functional<br>Demonstration | The power switch meets design and functional specifications | Pass |
| FLOW COUPLER Monitor Audio Quality<br>Testing | Audio level greater than 70dB Signal to Noise ratio of the output electrical signal to the speaker shall be greater than 20 at all tested frequencies Total Harmonic Distortion (THD) within specification at varying frequencies | Pass |
| FLOW COUPLER Monitor EMC Testing | IEC 60601-1-2 Edition 4.1: 2020 Class A for Emissions, Immunity for Professional Healthcare Facility Environment And 47 CFR, Part 15:2022. §15.107 and §15.109. Class A | Pass |
Applicable verification and validation testing was completed with passing results per the predefined acceptance criteria for each test case. These results support that the modified device performs as intended, and meets all functional requirements and performance specifications.
#### VIII.Conclusions
The FLOW COUPLER System indications for use and principles of operation remain unchanged (are the same as the predicate device). Performance testing has been completed to evaluate safety and effectiveness of the subject device as it compares to its predicate. The conclusion drawn from the riskbenefit assessment and nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device. No new issues of safety and effectiveness have been identified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.