K213958 · Embody, Inc. · MAI · Apr 8, 2022 · Orthopedic
Device Facts
Record ID
K213958
Device Name
Bioabsorbable Anchor
Applicant
Embody, Inc.
Product Code
MAI · Orthopedic
Decision Date
Apr 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.
Device Story
Bioabsorbable Anchor; absorbable implant for soft tissue/bone fixation. Composed of polydioxanone (PDO) dyed with D&C Violet No. 2. Provided sterile, preloaded in disposable Dual Anchor Inserter instrument. Used by surgeons in orthopedic shoulder procedures. Implant provides mechanical fixation of prosthetic material to tissue/bone. Device undergoes hydrolytic degradation in vivo; degradation products absorbed and excreted. Benefits include secure graft fixation during healing without need for permanent hardware removal.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity, bacterial endotoxin, genotoxicity, implantation). Chemical characterization per ISO 10993-18 and toxicological risk assessment per ISO 10993-17. Non-clinical performance testing included insertion, pullout, and fatigue testing. 16-week in vitro hydrolytic degradation study per ASTM F1635-16. In vivo degradation profile evaluated via intramuscular and femoral condyle implantation with histological observations.
Technological Characteristics
Material: Polydioxanone (PDO) dyed with D&C Violet No. 2. Form factor: Absorbable implant preloaded in disposable inserter. Principle: Hydrolytically degradable polymer for soft tissue/bone fixation. Standards: ISO 10993-1, ISO 10993-17, ISO 10993-18, ASTM F1635-16. Sterilization: Sterile, single-use.
Indications for Use
Indicated for fixation of soft tissue grafts in minimally invasive and open orthopedic shoulder procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Rotation Medical Bone Staple (RMB Staple) (K131635)
{0}------------------------------------------------
April 8, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Embody, Inc. Brianna Schehr Director of Quality, Regulatory & Clinical Affairs 4211 Monarch Way, Suite 500 Norfolk, Virginia 23508
## Re: K213958
Trade/Device Name: Bioabsorbable Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: March 10, 2022 Received: March 10, 2022
Dear Brianna Schehr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213958
Device Name Bioabsorbable Anchor
Indications for Use (Describe)
The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Embody. The logo is written in a dark blue font. The "O" in embody is an orange square with white lines inside.
## K213958 - 510(k) Summary
#### SUBMITTER l.
| Owner/Submitter: | Embody, Inc.<br>4211 Monarch Way<br>Suite 500<br>Norfolk, VA 23508 |
|------------------|----------------------------------------------------------------------|
| Phone: | 236-994-7912 |
| Contact Person: | Brianna Schehr<br>Director of Quality, Regulatory & Clinical Affairs |
| Date Prepared: | April 8, 2022 |
#### ll. DEVICE
| Name of Device: | Bioabsorbable Anchor |
|-----------------------|------------------------------------------------|
| Common or Usual Name: | Bone Anchor |
| Classification Name: | Fastener, fixation, biodegradable, soft tissue |
| Regulation Number: | 21 CFR 888.3030 |
| Regulatory Class: | Class II |
| Product Code: | MAI |
| Medical Specialty: | Orthopedic |
#### PREDICATE DEVICE lll.
Rotation Medical Bone Staple (RMB Staple), K131635. This predicate has not been subject to a design-related recall.
ORTHOSORB Resorbable Pins, K111077, is being referenced as a device made of polydioxanone. Therefore, the use of polydioxanone as a fixation material is not new.
#### DEVICE DESCRIPTION IV.
The Bioabsorbable Anchor is an absorbable implant device that is intended to provide fixation of a prosthetic material to soft tissue and/or bone. The Bioabsorbable Anchor is composed of polydioxanone (PDO) dyed with D&C Violet No. 2. Two Bioabsorbable Anchors are provided sterile for single use only, packaged preloaded in a disposable anchor inserter instrument, the Dual Anchor Inserter. The preloaded Dual Anchor Inserter is provided in a dual sterile pouch confiquration.
#### V. INDICATION FOR USE
The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "EMBODY" in a stylized font. The letters are dark blue, except for the "O", which is inside an orange square with rounded corners. Inside the orange square, there are several diagonal white lines.
Both the subject and predicate devices have the same intended use for the fixation of a prosthetic material.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Bioabsorbable Anchor is substantially equivalent in terms of indications for use, technological characteristics, and performance characteristics to the predicate device. Fixation is the technological principle for both the subject and predicate devices and is based on the use of a permanent implant in a U-shape polymer strap design with barbed ends, to directly affix a prosthetic material. The Bioabsorbable Anchor is substantially equivalent to the reference device in terms of material composition of an absorbable fixation device.
The following technological differences exist between the subject and predicate devices:
- The subject device is composed of an absorbable polymer (polydioxanone), whereas the . predicate device is a nonabsorbable polymer (PEEK), and
- The subject device is optimized for fixation to both soft tissue and bone, whereas the . predicate device is optimized only for fixation to bone.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
### Biocompatibility Testing
The biocompatibility evaluation for the Bioabsorbable Anchor device was conducted in accordance with the FDA quidance document "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" dated September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation Testing Within a Risk Management Process," as recognized by FDA. The finished device met the requirements of the standards for the following battery of tests; cytotoxicity, sensitization, acute systemic toxicity, systemic toxicity (material-mediated pyrogenicity), bacterial endotoxin testing, genotoxicity, and implantation. Chemical characterization in accordance with ISO 10993-18 and a toxicological risk assessment in accordance with ISO 10993-17 confirmed the Bioabsorbable Anchor is considered safe for its intended use.
The Bioabsorbable Anchor is considered a permanent implant device in contact with tissue/bone, while the anchor inserter instrument is considered tissue/bone contacting for a duration of less than 24 hours.
### Non-Clinical Performance Testing
The Bioabsorbable Anchor was evaluated in comparison to the predicate device when applicable in accordance with the FDA guidance document "Bone Anchors—Premarket Notification (510(k)) Submissions" and included the following tests:
- Insertion Testing .
- Pullout Testing .
- Fatigue Testing o
Both the subject and predicate devices had suitable characteristics for their indication for use.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "EMBODY" in a stylized font. The letters are primarily dark blue, except for the "BOD" portion, where the "O" is replaced by an orange square with diagonal lines inside. The font is bold and modern, with rounded edges, giving the word a sleek and professional appearance. The overall design is clean and eye-catching.
A sixteen-week in vitro hydrolytic degradation study guided by ASTM F1635-16 "Standard Test Method for in vitro Degradation Testing of Hydrolytically Degradable Polymer Resins and Fabricated Forms for Surgical Implants" was also performed. This study provided characterization of the degradation profile via mass loss.
### Absorption Profile
The Bioabsorbable Anchor is hydrolytically degraded in tissues and demonstrated expected local tissue response, histological observations, and degradation after 13 weeks in both intramuscular and femoral condyle implantation. The degradation products are subsequently absorbed by tissues and excreted from the body.
### VIII. CONCLUSIONS
Results of the biocompatibility testing and non-clinical performance testing demonstrate that the Bioabsorbable Anchor is as safe and as effective, and substantially equivalent to the predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.