AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213941 · Feb 24, 2022
Annalise Enterprise CXR Triage Pneumothorax
Annalise-Ai
Retrospective clinical chest X-ray cases from 4 US hospital networks; Original radiology reports
Retrospective clinical data was used to evaluate the standalone performance (accuracy) of the AI algorithm in a real-world clinical cohort, independent of the training dataset.
Standalone Performance Assessment (accuracy); Retrospective, anonymized study
Adult patients undergoing chest X-ray; Sample Size: 949 CXR cases (413 positive, 536 negative); Number of Sites: 4 US hospital network sites
Not applicable for this study
Sensitivity, specificity, and AUC for pneumothorax and tension pneumothorax detection
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Pneumothorax
Convolutional network
AUC > 0.95; Sensitivity > 80%; Specificity > 80%
AUC 0.979 (95% CI: 0.970-0.986); Sensitivity 93.9% (95% CI: 91.8-96.1); Specificity 92.2% (95% CI: 89.9-94.4) at balanced operating point
Over 1,500,000 chest x-ray images sourced from datasets across three continents.
>1 (qualified radiologists)
Retrospective, anonymized study of 949 adult CXR cases from 4 US hospital network sites.
>1 (ABR-certified and protocol-trained radiologists) + 1 (ABR-certified and protocol-trained radiologist)
Tension Pneumothorax
Convolutional network
AUC > 0.95; Sensitivity > 80%; Specificity > 80%
AUC 0.988 (95% CI: 0.981-0.993); Sensitivity 94.3% (95% CI: 90.2-98.4); Specificity 95.8% (95% CI: 94.3-97.1) at balanced operating point
Over 1,500,000 chest x-ray images sourced from datasets across three continents.
>1 (qualified radiologists)
Retrospective, anonymized study of 949 adult CXR cases from 4 US hospital network sites.
>1 (ABR-certified and protocol-trained radiologists) + 1 (ABR-certified and protocol-trained radiologist)
Indications for Use
The device is designed to aid the clinical assessment of adult chest x-ray cases with features suggestive of pneumothorax and tension pneumothorax in the medical care environment. The device analyses cases using an artificial intelligence algorithm to identify findings. It makes case-level output available to a PACS for worklist prioritization or triage. The device is intended to be used by trained clinicians who are qualified to interpret chest X-rays as part of their scope of practice. The device is not intended to direct attention to specific portions of an image or to anomalies other than pneumothorax and tension pneumothorax. Its results are not intended to be used on a standalone basis for clinical decision making nor it is intended to rule out pneumothorax or tension pneumothorax, or otherwise preclude clinical assessment of X-ray cases.
Device Story
Software workflow tool; interfaces with RIS/PACS to obtain chest X-ray images. Uses deep learning convolutional neural network to analyze images for pneumothorax and tension pneumothorax. Outputs case-level triage notifications to reporting worklists; enables prioritization of clinician queues. Used in medical care environments by trained clinicians. Does not direct attention to specific image regions; does not rule out conditions; not for standalone diagnosis. Benefits include improved workflow efficiency and reduced turnaround time for urgent findings. Does not downgrade existing worklist priorities.
Clinical Evidence
Retrospective standalone performance study (n=949; 413 positive, 536 negative) and internal bench study (n=621) for turnaround time. Standalone study used independent, multi-site data. Primary endpoints: AUC, sensitivity, specificity. Results: Pneumothorax AUC 0.979; Tension Pneumothorax AUC 0.988. Balanced operating point sensitivity/specificity: 93.9%/92.2% (pneumothorax) and 94.3%/95.8% (tension pneumothorax). Average triage turnaround time: 20.57 seconds.
Technological Characteristics
Software-only device; deep learning convolutional neural network architecture. Inputs: DICOM-compliant chest X-rays. Outputs: Triage notification to PACS/RIS. Operates via passive notification parallel to standard workflow. No hardware components. Moderate level of concern software.
Indications for Use
Indicated for adult patients undergoing chest X-ray imaging in medical care environments to aid clinicians in the triage and prioritization of cases suggestive of pneumothorax and tension pneumothorax.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Annalise-AI Pty Ltd. % Eric Qin Principal RAQA Advisor Level 21, 60 Margaret Street SYDNEY. NSW 2000 AUSTRALIA
February 24, 2022
### Re: K213941
Trade/Device Name: Annalise Enterprise CXR Triage Pneumothorax Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: December 16, 2021 Received: December 16, 2021
Dear Eric Qin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K213941
Device Name
Annalise Enterprise CXR Triage Pneumothorax
#### Indications for Use (Describe)
The device is designed to aid the clinical assessment of adult chest x-ray cases with features suggestive of pneumothorax and tension pneumothorax in the medical care environment. The device analyses cases using an artificial intelligence algorithm to identify findings. It makes case-level output available to a PACS for worklist prioritization or triage. The device is intended to be used by trained clinicians who are qualified to interpret chest X-rays as part of their scope of practice. The device is not intended to direct attention to specific portions of an image or to anomalies other than pneumothorax and tension pneumothorax. Its results are not intended to be used on a standalone basis for clinical decision making nor it is intended to rule out pneumothorax or tension pneumothorax, or otherwise preclude clinical assessment of X-ray cases.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K213941
#### SUBMITTER I.
| Company Name | Annalise-AI Pty Ltd |
|----------------|---------------------------------------------------------------|
| Address | Level 21, 60 Margaret Street<br>Sydney, NSW 2000<br>Australia |
| Phone Number | +61 2 7204 0817 |
| Contact Person | Michele Houldsworth |
| Date Prepared | December 16, 2021 |
#### II. DEVICE
| Device Name | Annalise Enterprise CXR Triage Pneumothorax |
|---------------------|---------------------------------------------------------------------------------|
| Classification Name | Radiological computer aided triage and notification software<br>(21CFR892.2080) |
| Regulatory Class | Class II |
| Product Code | QFM |
## III. PREDICATE DEVICE
| Manufacturer Name | Behold.AI Technologies Limited |
|---------------------|------------------------------------------------------------------------------|
| Device Name | Red Dot |
| 510(k) reference | K191556 |
| Classification Name | Radiological computer aided triage and notification software (21CFR892.2080) |
| Regulatory Class | II |
| Product Code | QFM |
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Annalise Enterprise CXR Triage Pneumothorax is a software workflow tool that interfaces with RIS and PACS to obtain chest x-ray images to process. The artificial intelligence algorithm within the device uses deep learning techniques to identify the presence of pneumothorax and tension pneumothorax.
The AI algorithm used in the device is a convolutional network trained using deep learning techniques. The training dataset included over 1,500,000 chest x-ray images sourced from datasets across three continents from different x-ray manufacturers, machines and a range of patient demographics. Cases in the training dataset were labeled by at least three qualified radiologists for the presence or absence of radiographic features suggestive of pneumothorax or tension pneumothorax.
The AI results output from the device are sent to the reporting worklist software to enable AI assisted triage of the reporting worklist. The exact functionality available depends on the worklist software being used (RIS, PACS).
This triage functionality uses the findings detected in each study by the AI model to provide information into the worklist software enabling the prioritization of the reporting worklist. Each organization can specify which findings will result in triage and the priority of each finding. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded by the AI software.
#### INDICATIONS FOR USE V.
The Indications For Use statement is as follows;
The device is designed to aid the clinical assessment of adult chest x-ray cases with features suggestive of pneumothorax and tension pneumothorax in the medical care environment. The device analyses cases using an artificial intelligence algorithm to identify findings. It makes caselevel output available to a PACS or RIS for worklist prioritization or triage.
The device is intended to be used by trained clinicians who are qualified to interpret chest X-rays as part of their scope of practice.
The device is not intended to direct attention to specific portions of an image or to anomalies other than pneumothorax and tension pneumothorax. Its results are not intended to be used on a standalone basis for clinical decision making nor it is intended to rule out pneumothorax or tension pneumothorax, or otherwise preclude clinical assessment of X-ray cases.
The Indications for Use statement for the subject device is not identical to the predicate device, in that the subject device can identify and triage Pneumothorax and Tension Pneumothorax. The differences, however, do not alter the intended use of the device nor do they affect the safety and effectiveness of the device, as compared to the predicate. Both devices have the same intended use to assist with worklist triage by providing notification of pneumothorax findings.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL VI. CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject and predicate devices are both Radiological Computer-Assisted Prioritization Software intended to assist with worklist triage by providing notification of pneumothorax findings in adults to trained clinicians. The technologies use AI techniques to analyze radiological images. The devices establish effective triage within a clinician's queue based on high sensitivity and specificity.
At a high level, the subject and predicate devices share similar technological characteristics as follows:
- Target population ●
- Technical method for notification and prioritization
- Anatomical site and modality ●
- Intended user and use environment ●
- Image protocol (input)
- Means of notification to user (output)
- System components
- Prioritization relationship to the standard of care workflow
- Standalone performance level and associated study methods ●
- Triage effectiveness (turnaround time) .
The subject and predicate devices also share equivalent technological characteristics. These characteristics exhibit minor differences, however, do not raise new questions of safety and effectiveness. The equivalent technological characteristics include:
- Capability of the subject device to identify and triage tension pneumothorax
- Results are not delivered to an Electronic Patient Record (EPR) ●
- Use of multiple operating points
Therefore, by examination of the device intended use and technological attributes, substantial equivalence is supported.
{6}------------------------------------------------
## VII. PERFORMANCE DATA
The following performance data have been provided to support evaluation of substantial equivalence.
#### Software Verification and Validation Testing A.
Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005.
The software for this device was considered as posing moderate level of concern, since prior to mitigation of hazards, a failure of the software device could result in minor injury.
#### B. Performance Testing
Performance of the subject device was assessed in two studies to satisfy requirements set forth in Special Controls per 21CFR892.2080. These included:
- Standalone Performance Assessment (accuracy); and .
- Triage Effectiveness Assessment (turn-around time). .
Standalone performance was assessed via a retrospective, anonymized study of adult patient, DICOM-compliant CXR cases. The test dataset analyzed included a total of 949 CXR cases (positive pneumothorax n=413 (including tension pneumothorax n=123 as a subset) and negative n=536) from unique patients. To obtain this cohort, eligible chest x-ray cases were collected consecutively from 4 US hospital network sites by manual review (for pneumothorax) or natural language processing analysis (for tension pneumothorax) of the original radiology reports associated with each case. No data had previously been collected from these sites for training or testing of the device's AI algorithm. That is, the test dataset applied during the standalone performance evaluation was newly acquired and independent from the training dataset.
The cohort collected for this evaluation included chest x-ray cases taken using a range of radiographic imaging protocols, including different patient positions (erect and supine), view positions (anteroposterior, posteroanterior and lateral) and x-ray imaging equipment, including different equipment manufacturers (n=10) and equipment models (n=22). Within this cohort, 427 patients were female (45%) and 522 patients were male (55%), with the mean patient age of 62.3 years (SD=17.5, range 22-99 years old). A range of clinical confounders were also present across the cohort, including various clinical conditions and other radiographic findings. The ethnicity breakdown for the cohort was 7,7% Hispanic. 86,7% Not Hispanic (the remaining proportion of the cohort was either unavailable or declined to declare). The race breakdown for the cohort was 0.3% Two or more races, 3.3% Asian, 4.8% Other, 5.5% African American, 83% White (the remaining proportion of the cohort was either unavailable or declined to declare).
To determine the ground truth, each deidentified CXR case was annotated in a blinded fashion by at least two American Board of Radiology (ABR)-certified and protocol-trained radiologists (ground truthers), with consensus determined by two ground truthers and a third ground truther in the event of disagreement.
{7}------------------------------------------------
The results included an AUC of 0.979 (95% CI: 0.970-0.986) and 0.988 (95% CI: 0.981-0.993) for pneumothorax and tension pneumothorax respectively, thus meeting the AUC>0.95 requirement for product code OFM. The sensitivity and specificity of the device was reported for three operating points and shown to meet >80% requirement for product code OFM. The default "balanced sensitivity and specificity" operating point demonstrated sensitivity of 93.9% (95% CI: 91.8. 96.1) and specificity of 92.2% (95% CI: 89.9. 94.4) for pneumothorax and sensitivity of 94.3% (95% CI: 90.2, 98.4) and specificity of 95.8% (95% CI: 94.3, 97.1) for tension pneumothorax. The "optimized for sensitivity" operating point demonstrated sensitivity of 96.6% (95% CI: 94.7, 98.3) and specificity of 84.1% (95% CI: 82.1, 87.1) for pneumothorax and sensitivity of 95.9% (95% CI: 91.9, 99.2) and specificity of 94.9% (95% CI: 93.3, 96.4) for tension pneumothorax. The "optimized for specificity" operating point demonstrated sensitivity of 89.1% (95% CI: 86.2, 92.0) and specificity of 95.7% (95% CI: 94.0, 97.4) for pneumothorax and sensitivity of 83.7% (95% CI: 76.4, 90.2) and specificity of 97.8% (95% CI: 96.7, 98.7) for tension pneumothorax.
The results demonstrate the subject device establishes effective triage within a clinician's queue based on high sensitivity and specificity. Further, these results are substantially equivalent to those of the predicate device.
Triage effectiveness (turn-around time) was assessed by an internal bench study using a dataset of n=621 cases positive for pneumothorax and/or tension pneumothorax eligible for prioritization. These cases were collected from multiple data sources spanning a variety of geographical locations, patient demographics and technical characteristics. The results demonstrated an average triage turn-around time of 20.57 seconds, (95% CI: 19.90-21.24), which is substantially equivalent to the total performance time published for the predicate device.
Therefore, the subject device has been shown to satisfy the performance requirements per 21 CFR 892.2080, for radiological triage and notification software, by providing clinically effective triage for chest x-ray studies containing features suggestive of pneumothorax and tension pneumothorax. This data demonstrates the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
## VIII. CONCLUSIONS
The subject and predicate devices are both software-only devices intended to assist with worklist triage by providing notification of pneumothorax findings in adults to trained clinicians. The labelling of both devices is limited to worklist triage and are not to be used in-lieu of full patient evaluation or relied upon to provide direct diagnosis. The devices both operate by passive notification of suspected cases parallel to the standard of care workflow. The evaluation demonstrates that minor differences between the subject and predicate devices do not raise new questions of safety and effectiveness. In addition, the performance testing conducted demonstrates the subject device performs as intended, that it is as safe, as effective and performs comparably to the predicate device. The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.