Multicenter observational clinical study (Apollo, NCT04690569) specimen repository
A subset of 285 clinical samples from the Apollo study was used to compare the LIAISON MeMed BV assay performance against the predicate device and physician adjudication results.
Apollo Study (subset); Multicenter, observational, blinded study
Hospital admitted, Emergency Department (ED), and Urgent Care patients with suspected acute bacterial or viral infections; Sample Size: 285; Number of Sites: 11
Predicate device (MeMed BV) and physician adjudication
Diagnostic performance (bacterial vs. viral differentiation) and assay score agreement
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Bacterial vs. Viral Infection Differentiation
Weighted multinomial logistic regression model
Bias < 12.5 score units
Bias at cut-offs 10, 35, 65, 90 were 1.1, 1.7, 2.5, and 3.2 score units respectively
—
—
Apollo clinical study (NCT04690569): 285 samples from 11 medical centers
—
Indications for Use
The DiaSorin LIAISON MeMed BV is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoassay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON MeMed BV assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON MeMed BV assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. The assay has to be performed on the automated LIAISON XL Analyzer. The DiaSorin LIAISON MeMed BV Control Set is intended for use as assayed quality control to monitor the performance of the DiaSorin LIAISON MeMed BV assay. The performance characteristics of the LIAISON controls have not been established for any other assays or instrument platforms different from the automated LIAISON Analyzer family. The control set is intended for in vitro diagnostic use in a professional laboratory only.
Device Story
Automated in vitro diagnostic assay; measures host proteins TRAIL, IP-10, and CRP in serum using chemiluminescent immunoassay (CLIA) technology. Operates on LIAISON XL Analyzer; requires three reagent packs per test. Computational integration of protein measurements produces a numeric score (0-100) indicating likelihood of bacterial vs. viral infection. Used in emergency departments/urgent care by professional laboratory staff. Results aid clinicians in differentiating infection etiology alongside clinical assessments. Benefits include improved diagnostic accuracy for suspected acute infections, potentially guiding appropriate antibiotic use.
Clinical Evidence
Clinical validation utilized 285 samples from a multicenter observational study (Apollo, NCT04690569). Performance compared to forced expert adjudication and consensus adjudication. Deming regression showed high correlation with predicate (Score R²=0.94). Results demonstrate significant trend between score bins and bacterial infection likelihood, with equivalent performance to predicate device.
Technological Characteristics
Chemiluminescent immunoassay (CLIA) using paramagnetic particles coated with monoclonal antibodies. Measures TRAIL, IP-10, and CRP. Automated on LIAISON XL Analyzer. Uses isoluminol-conjugated antibodies. Requires 200 μL serum sample. Connectivity via LIAISON XL platform. Software performs computational integration of protein concentrations to generate a final score.
Indications for Use
Indicated for adult and pediatric patients presenting to emergency departments or urgent care centers with suspected acute bacterial or viral infection and symptoms duration of seven days or less; samples collected at hospital admission.
Regulatory Classification
Identification
A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.
Special Controls
A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis must comply with the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
*e.g.,* age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
*e.g.,* collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K213936
B Applicant
DiaSorin Inc.
C Proprietary and Established Names
LIAISON MeMed BV, LIAISON MeMed BV Control Set
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QPS | Class II | 21 CFR 866.3215 - Device To Detect And Measure Non-Microbial Analyte(S) In Human Clinical Specimens To Aid In Assessment Of Patients With Suspected Sepsis | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the LIAISON MeMed BV device and the LIAISON MeMed BV Control Set
B Measurand:
Three host immune protein biomarkers: TRAIL, IP-10, and CRP.
C Type of Test:
Chemiluminescent immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The DiaSorin LIAISON MeMed BV is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoassay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON MeMed BV assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON MeMed BV assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. The assay has to be performed on the automated LIAISON XL Analyzer.
The DiaSorin LIAISON MeMed BV Control Set is intended for use as assayed quality control to monitor the performance of the DiaSorin LIAISON MeMed BV assay. The performance characteristics of the LIAISON controls have not been established for any other assays or instrument platforms different from the automated LIAISON Analyzer family. The control set is intended for in vitro diagnostic use in a professional laboratory only.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
The LIAISON MeMed BV Assay is performed on the LIAISON XL Analyzer.
IV Device/System Characteristics:
A Device Description:
The LIAISON MeMed BV assay is an in vitro diagnostic device that measures in parallel the blood concentrations of TRAIL, IP-10, and CRP. The test consists of an automated analyzer that conducts chemiluminescence-based analyte measurements of patient serum samples and their computational integration. Reagents are provided in individual compartments within a plastic container called the Reagent Integral. The assembled kit includes one each of the TRAIL Reagent Integral, IP-10 Reagent Integral, CRP Reagent Integral and two levels of calibrators for each specific assay. The assay configuration allows for the performance of 100 tests.
The LIAISON MeMed BV Control Set is quality control material that is available for purchase separately and that consists of two vials each of two control levels (viral and bacterial) provided lyophilized. The volume of each vial is sufficient for at least 10 determinations. The LIAISON
K213936 - Page 2 of 22
{2}
MeMed BV Control Set is intended to monitor for substantial reagent failure. The controls should be evaluated on each day of testing or as required by guidelines or requirements of local regulations or accredited organizations.
## B Principle of Operation:
The LIAISON MeMed BV assay consists of three individual chemiluminescence immunoassays (CLIAs) for quantitative determination of TRAIL, IP-10, and CRP. All three reagent packs must be the same lot and present at the same time on the same instrument used for sample testing. All three reagent packs are individually calibrated and quality controlled.
Each individual reagent pack for TRAIL, IP-10, and CRP utilizes monoclonal antibodies for capture and antibody-conjugates for the detection of the specific assay target. The assay incubates sample, calibrator, or control with assay buffer and paramagnetic particles coated with a monoclonal antibody that specifically recognizes the target of interest. Following the incubation, an isoluminol conjugated antibody specific for each analyte is then added to the reaction and incubated. The unbound conjugate is removed with a wash step. Starter reagents are then added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is proportional to the concentration of analyte present in the calibrators, controls or samples.
The LIAISON MeMed BV test result is a score between 0 and 100 derived from computational integration of the measurements of the three proteins TRAIL, IP-10, and CRP, where low scores are indicative of viral infection and high score of bacterial infection.
- 0 ≤ score ≤10: High likelihood of viral infection (or other non-bacterial etiology)
- 10 < score <35: Moderate likelihood of viral infection (or other non-bacterial etiology)
- 35 ≤ score ≤65: Equivocal
- 65 < score <90: Moderate likelihood of bacterial infection (or co-infection)
- 90 ≤ score ≤100: High likelihood of bacterial infection (or co-infection)
## C Instrument Description Information:
1. Instrument Name:
LIAISON XL Analyzer
2. Specimen Identification:
The LIAISON MeMed BV test is intended to be performed on human serum specimens. To prevent confusion, a distinct identification number must be assigned to every sample in the LIAISON XL software. This sample ID can be entered either by scanning the bar-code using the barcode scanner located in the loading bay for sample racks or typing it manually.
3. Specimen Sampling and Handling:
K213936 - Page 3 of 22
{3}
The loading bay for sample racks allows continuous loading of the LIAISON XL system with samples. The samples in the tubes are placed in special racks and loaded in one of the 10 lanes afterwards. As support, a LED below each lane indicates the relevant usage status.
## 4. Calibration:
Two point kit calibrators specific to each Reagent Integral lot are used to establish specific working curves for each of the three assays based on assay master curves stored on the analyzer. In order to correctly run the test, the TRAIL, IP-10, and CRP Reagent Integrals must be calibrated.
Calibration is required every 21 days, or
- With each new lot of reagents (Reagent Integral or Starter Reagents)
- If Quality Control results are out of the acceptable range
- After each servicing of the LIAISON XL Analyzer
## 5. Quality Control:
The LIAISON MeMed BV Controls are representative of patient specimens intended for use with the LIAISON MeMed BV assay. Control 1 represents a low viral score and Control 2 represents a high bacterial score.
## V Substantial Equivalence Information:
A Predicate Device Name(s): MeMed BV
B Predicate 510(k) Number(s): K210254
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K213936 | K210254 |
| --- | --- | --- |
| Device Trade Name | LIAISON MeMed BV | MeMed BV |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The DiaSorin LIAISON MeMed BV is an automated in vitro diagnostic semi-quantitative assay that uses chemiluminescent immunoassay (CLIA) technology to measure three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use | The MeMed BV test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use |
K213936 - Page 4 of 22
{4}
K213936 - Page 5 of 22
| | IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The LIAISON MeMed BV assay is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for seven days or less. The LIAISON MeMed BV assay generates a numeric score that falls within discrete interpretation ranges based on the increasing likelihood of bacterial infection. | in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. The MeMed BV is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection. |
| --- | --- | --- |
| Results | Same | Semi-quantitative |
| Measurand | Same | Three non-microbial host proteins (TRAIL, IP-10, and CRP) |
| Sample Type | Same | Human serum |
| Measurement System | Same | Photomultiplier |
| Output | Same | SCORE result |
| General Device Characteristic Differences | | |
| Antibody Conjugate | Isoluminol Derivative | Alkaline Phosphatase |
| Sample volume | 200 μL | 100 μL |
| Instrument System | LIAISON XL Analyzer | MeMed Key Analyzer |
| Calibration | Low and High calibrators for each | High, Medium, and Low calibrators for |
{5}
| | analyte | each analyte |
| --- | --- | --- |
| Calibration Interval | 3 weeks | 2 weeks |
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3: 2019 – Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, 3rd Edition
CLSI EP15-A3: 2019 – User Verification of Precision and Estimation of Bias; Approved Guideline, 3rd Edition
CLSI EP06-Ed2: 2020 – Evaluation of Linearity of Quantitative Measurement Procedures
CLSI EP07-3rd Edition: 2018 – Interference Testing in Clinical Chemistry
CLSI EP37-1st Edition: 2018 – Supplemental Tables for Interference Testing in Clinical Chemistry
CLSI EP17-A2: 2012 – Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures
CLSI EP28-A3c: 2010 – Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory
ISO 14971: 2019 – Medical devices. Application of risk management to medical devices
IEC 62304:2006/A1: 2016 – Medical Device Software – Software life cycle processes
CLSI LIS01 – A2; Standard Spec for Low-Level Protocol to Transfer Messages Between Clinical Lab Instruments and Computer Systems
CLSI LIS02-A2; Standard Specification for Transferring Information Between Clinical Instruments and Computer Systems
AAMI TIR57: 2016 – Principles for medical device security – Risk Management
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision Study
A 20-day within-laboratory precision study was conducted at DiaSorin Inc. The study was performed for 20 days with two assay runs per day and two replicates per run by multiple
K213936 - Page 6 of 22
{6}
technicians for a total of 80 replicates per panel member. The test days spanned two calibration cycles. The test panel included the LIAISON MeMed BV Control set (Control 1 and Control 2 with two levels containing three analytes) and four serum specimens. Three of the serum specimens were contrived by spiking neat serum to generate target SCORE values that correspond to one viral (Low Score), one bacterial (High Score), and one equivocal assay (Medium Score) result. The final serum specimen was a native patient sample (Native) that corresponded to a low (viral/non-infectious) score.
Table 1. Precision Study Summary Results
| TRAIL | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | Mean (pg/mL) | Repeatability | | Between-Run | | Between Days | | Total | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 131 | 2.6 | 2.0 | 1.4 | 1.1 | 4.8 | 3.7 | 5.6 | 4.3 |
| Control 2 | 24.7 | 0.5 | 2.1 | 0.2 | 2.2 | 1.1 | 4.5 | 1.3 | 5.4 |
| Low Score | 182 | 3.7 | 2.0 | 6.4 | 3.5 | 2.3 | 1.3 | 7.7 | 4.2 |
| High Score | 28.1 | 0.99 | 3.5 | 0.6 | 1.9 | 1.41 | 5.0 | 1.8 | 6.4 |
| Medium Score | 59.8 | 1.1 | 1.8 | 1.1 | 1.8 | 2.3 | 3.8 | 2.7 | 4.6 |
| Native | 68.2 | 1.7 | 2.4 | 0 | 0.0 | 2.6 | 3.8 | 3.02 | 4.4 |
| IP-10 | | | | | | | | | |
| Sample ID | Mean (pg/mL) | Repeatability | | Between-Run | | Between Days | | Total | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 1207 | 20.7 | 1.7 | 14.4 | 1.2 | 25.3 | 2.1 | 35.8 | 3.0 |
| Control 2 | 191 | 2.8 | 1.5 | 2.8 | 1.4 | 9.4 | 4.9 | 10.2 | 5.3 |
| Low Score | 782 | 19.7 | 2.5 | 11 | 1.4 | 14.0 | 1.8 | 26.5 | 3.4 |
| High Score | 1244 | 26.1 | 2.1 | 0 | 0.0 | 22.8 | 1.8 | 33.5 | 2.7 |
| Medium Score | 537 | 10.4 | 1.9 | 8.0 | 1.5 | 16.8 | 3.1 | 21.4 | 4.0 |
| Native | 79.0 | 2.2 | 2.8 | 1.2 | 1.5 | 5.0 | 6.3 | 5.6 | 7.0 |
| CRP | | | | | | | | | |
| Sample ID | Mean (mg/L) | Repeatability | | Between-Run | | Between Days | | Total | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 15.9 | 0.3 | 2.0 | 0.2 | 1.1 | 0.6 | 3.5 | 0.67 | 4.1 |
| Control 2 | 139 | 4.1 | 2.9 | 2.0 | 1.4 | 5.2 | 3.7 | 6.9 | 4.9 |
| Low Score | 13.7 | 0.4 | 2.7 | 0.2 | 1.1 | 0.7 | 4.9 | 0.8 | 5.7 |
| High Score | 188 | 2.8 | 1.5 | 3.9 | 2.1 | 6.9 | 3.7 | 8.5 | 4.5 |
| Medium Score | 47 | 1.2 | 2.5 | 1.7 | 3.6 | 1.2 | 2.6 | 2.4 | 5.1 |
| Native | 2.4 | 0.09 | 3.9 | 0.06 | 2.8 | 0.2 | 10.4 | 0.3 | 11.4 |
K213936 - Page 7 of 22
{7}
| Score | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | Mean | Repeatability | | Between-Run | | Between Days | | Total | |
| | | SD | % CV1 | SD | % CV1 | SD | % CV1 | SD | % CV1 |
| Control 1 | 2.9 | 0.4 | N/A | 0.0 | N/A | 0.3 | N/A | 0.5 | N/A |
| Control 2 | 99 | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A |
| Low Score | 1 | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A |
| High Score | 99 | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A |
| Medium Score | 54.7 | 1.9 | N/A | 1.8 | N/A | 2.6 | N/A | 3.6 | N/A |
| Native | 7.9 | 0.8 | N/A | 0.0 | N/A | 0.6 | N/A | 0.9 | N/A |
$^{1}$ CV analysis was not considered for the logistic scale of the LIAISON MeMed BV Score.
## Reproducibility Study
A five-day reproducibility study was conducted at DiaSorin Inc. and two external laboratories. Testing was performed over five days with one run per day and six replicates per run, for a total of 90 replicates per panel member. IP-10 values below the established measuring range were observed for the Native panel member in the reproducibility study; therefore, IP-10 Nateve panel member data are presented both as concentration and Relative Light Unit (RLU) values.
Table 2. Reproducibility Study Summary Results
| TRAIL | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | Mean (pg/mL) | Repeatability | | Between-Days | | Within-Laboratory | | Between-Sites | | Reproducibility | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 134 | 2.5 | 1.9 | 8.0 | 6.0 | 8.5 | 6.3 | 2.0 | 1.5 | 8.7 | 6.5 |
| Control 2 | 27.8 | 0.6 | 2.1 | 1.7 | 6.0 | 1.8 | 6.3 | 0.0 | 0.0 | 1.6 | 5.8 |
| Low Score | 189 | 5.0 | 2.6 | 3.1 | 1.7 | 5.9 | 3.1 | 1.2 | 0.7 | 6.0 | 3.2 |
| High Score | 27.9 | 0.6 | 2.1 | 0.7 | 2.4 | 0.9 | 3.2 | 0.7 | 2.5 | 1.1 | 4.0 |
| Medium Score | 60.4 | 1.1 | 1.8 | 0.9 | 1.5 | 1.4 | 2.3 | 0.0 | 0.0 | 1.4 | 2.2 |
| Native | 70.1 | 1.6 | 2.3 | 0.8 | 1.2 | 1.8 | 2.6 | 1.1 | 1.5 | 2.1 | 3.0 |
| IP-10 | | | | | | | | | | | |
| Sample ID | Mean (pg/mL) | Repeatability | | Between-Days | | Within-Laboratory | | Between-Sites | | Reproducibility | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 1368 | 27.0 | 2.0 | 57.9 | 4.2 | 63.9 | 4.7 | 191.8 | 14.0 | 202.2 | 14.8 |
| Control 2 | 186 | 4.5 | 2.4 | 8.9 | 4.8 | 10.0 | 5.4 | 26.1 | 14.0 | 28.0 | 15.0 |
| Low Score | 815 | 24.3 | 3.0 | 11.6 | 1.4 | 26.9 | 3.3 | 81.2 | 10.0 | 85.6 | 10.5 |
| High Score | 1348 | 33.6 | 2.5 | 31.7 | 2.4 | 46.2 | 3.4 | 151.5 | 11.2 | 158.4 | 11.7 |
| Medium Score | 567 | 13.6 | 2.4 | 11.6 | 2.1 | 17.9 | 3.2 | 52.5 | 9.3 | 55.5 | 9.8 |
K213936 - Page 8 of 22
{8}
| Native (concentration)1 | 75.4 | 2.5 | 3.4 | 1.1 | 1.4 | 2.7 | 3.6 | 8.2 | 10.8 | 8.6 | 11.4 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Native (RLU) | 9698 | 255.2 | 2.6 | 117.7 | 1.2 | 281.1 | 2.9 | 429.2 | 4.4 | 513.1 | 5.3 |
| CRP | | | | | | | | | | | |
| Sample ID | Mean (mg/L) | Repeatability | | Between-Days | | Within-Laboratory | | Between-Sites | | Reproducibility | |
| | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 | 13.7 | 0.4 | 2.9 | 0.6 | 4.4 | 0.7 | 5.2 | 0.6 | 4.6 | 1.0 | 7.0 |
| Control 2 | 129 | 4.3 | 3.3 | 4.8 | 3.8 | 6.5 | 5.0 | 3.9 | 3.0 | 7.6 | 5.9 |
| Low Score | 14.7 | 0.3 | 1.9 | 0.4 | 2.6 | 0.5 | 3.2 | 0.8 | 5.4 | 0.9 | 6.3 |
| High Score | 201 | 4.2 | 2.1 | 5.1 | 2.5 | 6.6 | 3.3 | 6.2 | 3..1 | 9.1 | 4.5 |
| Medium Score | 50.0 | 2.4 | 4.8 | 1.6 | 3.2 | 2.9 | 5.8 | 2.7 | 5.4 | 4.0 | 7.9 |
| Native | 2.6 | 0.1 | 3.6 | 0.1 | 4.9 | 0.2 | 6.1 | 0.2 | 9.3 | 0.3 | 11.1 |
| Score | | | | | | | | | | | |
| Sample ID | Mean | Repeatability | | Between-Days | | Within-Laboratory | | Between-Sites | | Reproducibility | |
| | | SD | % CV1 | SD | % CV1 | SD | % CV1 | SD | % CV1 | SD | % CV1 |
| Control 1 | 2.2 | 0.2 | N/A | 0.7 | N/A | 0.0 | N/A | 0.0 | N/A | 0.7 | N/A |
| Control 2 | 98.3 | 0.2 | N/A | 0.3 | N/A | 0.3 | N/A | 0.3 | N/A | 0.5 | N/A |
| Low Score | 1.0 | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A |
| High Score | 99.0 | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A | 0.0 | N/A |
| Medium Score | 55.0 | 1.9 | N/A | 1.3 | N/A | 1.5 | N/A | 1.5 | N/A | 2.8 | N/A |
| Native | 7.8 | 0.5 | N/A | 0.4 | N/A | 0.9 | N/A | 0.9 | N/A | 1.1 | N/A |
$^{1}$ IP-10 values for the Native Panel Member were below the assay measuring range.
2. Linearity:
A study was performed to assess the linearity of measurement for each of the three measurands (TRAIL/IP-10/CRP) with acceptance criteria for bias due to non-linearity of less than $15\%$ . Linearity is not applicable to the test result (Score), as it is calculated using a predefined weighted multinomial logistic regression model. Serial dilutions of specimens containing high levels of each analyte were evaluated on the LIAISON MeMed BV Test. Specifically, two serum specimens (CRP and IP-10) and a high-concentration buffer spiked sample (TRAIL), were diluted to prepare a dilution series that spanned the relevant measuring range for each analyte (15-300 pg/mL for TRAIL, 100-2000 pg/mL for IP-10, and 1-250 mg/L for CRP). Linearity for each of the MeMed BV measurands fell within the acceptance criteria with no deviations above $5.5\%$ observed throughout the relevant range for all analytes.
3. Analytical Specificity/Interference:
An interference study was performed to evaluate the impact of select interferents and cross-reactants on the test score. Performance in the presence of potentially interfering substances was evaluated using two serum samples, one that corresponded to a low score value (5) and one that corresponded to a high score value (95). Each sample was divided into equal
K213936 - Page 9 of 22
{9}
aliquots and spiked with the known amount of potential interferant or with the same volume of vehicle control. Matched spiked and vehicle control aliquots were evaluated in five replicates in the same run.
Table 3. Interference Study Results
| Substance | Conc. Tested | Sample ID | Mean Vehicle Score | Mean Test Score | Absolute Difference |
| --- | --- | --- | --- | --- | --- |
| Bilirubin (conjugated) | 0.4 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Bilirubin (unconjugated) | 0.4 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Cholesterol | 4 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Hemoglobin | 10 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Human serum Albumin | 60 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Triglycerides | 15 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Acetaminophen | 0.156 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Acetyl Salicylic Acid | 0.03 mg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Amoxicillin | 54 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Ampicillin | 75 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Azithromycin | 11.1 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Biotin | 3600 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Caffeine | 108 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Cetirizine HCL | 4.35 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Dextramethorphan | 15.6 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Doxycycline | 18 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Ethanol | 0.5% v/v | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Heparin | 3300 U/L | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Ibuprofen | 219 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Levofloxacin | 36 μg/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
K213936 - Page 10 of 22
{10}
These results demonstrate that the presence of the evaluated exogenous and endogenous interferants do not significantly impact the LIAISON MeMed BV Score values.
Additional interference testing was performed to establish whether the presence of human anti-mouse antibodies (HAMA) interfere with LIAISON MeMed BV assay results. Potential HAMA interference was assessed by evaluating three independent clinical samples that each contained HAMA levels varying from 316 to >600 ng/mL as determined by a sample vendor certificate of analysis. Samples were tested both before and after processing with heterophilic blocking tubes to remove HAMA. Results from testing are displayed below.
Table 4. HAMA Interference Testing Results
| Substance | Conc. Tested | Sample ID | Mean Vehicle Score | Mean Test Score | Absolute Difference |
| --- | --- | --- | --- | --- | --- |
| HAMA | >600 ng/mL | 1 | 39 | 51 | 12 |
| | >600 ng/mL | 2 | 75 | 77 | 2 |
| | 316 ng/mL | 3 | 9 | 14 | 5 |
The observed difference in Score for each HAMA sample met acceptance criteria of less than 12.5 score units and indicate a low likelihood of specimens shifting to a non-adjacent bin.
Additional interference testing was performed to determine whether the presence of rheumatoid factor interferes with the LIAISON MeMed BV assay results. Again, two serum samples including one low Score (5) and one high score (95) were divided into equal aliquots and spiked with high RF Concentrations. Results are displayed in the table below.
Table 5. Rheumatoid Factor Interference Testing Results
| Substance | Conc. Tested | Sample ID | Mean Vehicle Score | Mean Test Score | Absolute Difference |
| --- | --- | --- | --- | --- | --- |
| Rheumatoid Factor | 500 IU/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
These results demonstrate that the presence of the evaluated RF concentrations do not significantly impact the LIAISON MeMed BV Score Values.
Cross-reactivity Testing
K213936 - Page 11 of 22
{11}
Evaluation of potential cross-reactants was performed similarly to the interference testing described previously. Briefly, tesiting included two serum samples, one high score and one low score, in the presence and absence of the potential cross-reactant.
Table 6. Cross-reactivity Testing Summary Results
| Substance | Conc. Tested | Sample ID | Mean Vehicle Score | Mean Test Score | Absolute Difference |
| --- | --- | --- | --- | --- | --- |
| 4-1BB Ligand | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| CD40 Ligand (carrier) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| CD40 Ligand (carrier-free) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| LT α1/β2 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| LT α2/β1 (recombinant mouse) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| LT α2/β1 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| TNF-α | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| TNF-β | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| BLC/BCA-1 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| ENA-78 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| GCP-2 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| GRO α | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| GRO γ | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| IFN γ | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| IL-8 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| I-TAC | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| NAP-2 | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| MIG | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| SDF-1α (recombinant mouse) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
K213936 - Page 12 of 22
{12}
| SDF-1α | 50 ng/mL | 1 | 99 | 99 | 0 |
| --- | --- | --- | --- | --- | --- |
| | | 2 | 1 | 1 | 0 |
| SDF-1β
(aa 19-93) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| SDF-1β
(aa 22-93) | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Pentraxin 2/SAP | 500 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Pentraxin 3/TSG-14 | 500 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
| Adiponectin | 50 ng/mL | 1 | 99 | 99 | 0 |
| | | 2 | 1 | 1 | 0 |
No cross-reactivity or significant impact of LIAISON MeMed BV test score was observed.
## Hook Effect Study
A study was performed to evaluate whether a high dose hook effect occurs in samples containing high amounts of analyte. Samples containing analyte levels above the upper end of the analytical measuring range were evaluated using one reagent lot. All samples demonstrated an increasing amount of chemiluminescence with increasing concentration of analyte. No hook effect was observed for TRAil up to 1,000 pg/mL, IP-10 up to 10,000 pg/mL, and CRP up to 500 mg/L.
4. **Assay Reportable Range:**
Not applicable.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
### Calibration
The LIAISON MeMed BV assay consists of three separate reagent integrals, TRAIL, IP-10, and CRP that generate a continuous response (relative light units, RLU) which is used in sample grading to provide a reportable quantitative result. Sample grading is based on the use of a calibration curve referenced to 'In-house' TRAIL, IP-10, and CRP reference standards (master standards). Each calibration curve (master curve) is controlled by the use of two lyophilized calibrators (Calibrator 1 and Calibrator 2) provided with the Reagent Integral. Calibrator 1 and 2 are assayed by the user to transform the master curve into a working curve and further used to calculate sample results. The LIAISON XL Analyzer working curve is obtained by the user during assay calibration by assigning a curve to the two point kit calibrators based upon the master curve. This working curve is used to calculate the patient sample results.
Recalibration in triplicate is mandatory whenever at least one of the following conditions occurs: new lot of reagents are being used, previous calibration was performed more than 21 days prior, quality control results are out of the acceptable range, or the analyzer has been serviced.
K213936 - Page 13 of 22
{13}
# Controls
The LIAISON MeMed BV Controls are representative of patient specimens intended for use with the LIAISON MeMed BV assay. Control 1 represents a low viral score specimen and control 2 represents a high bacterial score specimen. The range of concentrations for each analyte present in each control is reported on the certificate of analysis and indicates the limits established by DiaSorin for control values that can be obtained in reliable assay runs.
# Specimen Stability
Several studies were performed to assess stability of serum samples exposed to different storage conditions.
To demonstrate room temperature sample storage stability, five serum samples that spanned the entire LIAISON MeMed BV score range were aliquoted and stored at 22°C across several timepoints before being tested in duplicate on the LIAISON MeMed BV assay. Mean values were calculated for each sample and mean percent differences from the baseline were evaluated by linear regression to demonstrate a deviation of less than 10% from samples at baseline. The study data support a room temperature sample stability claim of three hours.
To demonstrate refrigerated sample storage stability, five serum samples spanning the entire LIAISON MeMed BV score range were aliquoted and stored at 2-8°C. Mean values were plotted and evaluated over several timepoints to demonstrate a deviation of less than 10% from baseline. These study data support a refrigerated sample storage claim of eight hours.
To demonstrate frozen sample storage stability, five serum samples that spanned the entire range of the LIAISON MeMed BV assay were aliquoted and stored at both -20°C and -70°C before being tested in duplicate. Mean values were calculated for each sample and mean percent differences from the baseline were evaluated by linear regression to demonstrate a deviation of less than 10% from samples at baseline. Data from these studies support a frozen sample storage claim of four weeks at -20°C or -70°C.
# Freeze/Thaw Stability
To evaluate the effect of freeze-thawing serum samples on the LIAISON MeMed BV assay results, five serum samples that spanned the entire LIAISON MeMed BV score range were aliquoted, frozen, and thawed for up to five freeze/thaw cycles before being tested in triplicate on the LIAISON MeMed BV assay. Mean values for each sample were evaluated at each time point and analyzed via linear regression to determine the point at which a 10% or greater deviation from baseline occurs. Results from these studies demonstrate sample stability for up to three freeze/thaw cycles.
# Stability of LIAISON MeMed BV Calibrators
Stability of reconstituted calibrators for the LIAISON MeMed BV assay were evaluated across multiple storage conditions: room temperature on board the LIAISON XL instrument, refrigerated at 2-8°C, and frozen at -20°C. Performance was evaluated by re-calibrating the assay with the reconstituted calibrators at designated intervals. Samples tested included unexpired kit controls and four internal controls that spanned the LIAISON MeMed BV
K213936 - Page 14 of 22
{14}
Score range. Results from these stability studies support storage of controls at room temperature for 48 hours, refrigerated for up to four weeks, and frozen for up to 8 weeks. Additionally, calibrator stability was demonstrated for up to 4 freeze/thaw cycles.
## Stability of LIAISON MeMed BV Controls
Stability of reconstituted LIAISON MeMed BV controls was evaluated using two lots of control materials across multiple storage conditions: room temperature on board the LIAISON XL instrument, refrigerated at 2-8°C, and frozen at -20°C. Controls were evaluated in triplicate at designated intervals after storage. Mean values were analyzed via linear regression to identify the point where deviation greater than 10% occurs. Results from these studies demonstrate control stability for up to 72 hours at room temperature, eight weeks when refrigerated, and eight weeks when stored frozen. Additionally, stability of the LIAISON MeMed BC controls were also demonstrated for up to four freeze/thaw cycles.
## Fresh vs. Frozen Study
A study was conducted to establish that fresh and frozen serum samples are equivalent and can be used interchangeably in the assay. Forty-three fresh serum samples were collected from patients and spiked with recombinant TRAIL, IP-10, and/or CRP to span the entire range of the LIAISON MeMed BV Score. Samples were tested fresh in triplicate and then frozen, thawed, and tested again in triplicate. Sample results were analyzed by Passing Bablock regression with overall acceptance criteria for the study requiring a slope 0.9-1.1 and R² ≥ 0.9. Results for all individual analytes and Score values met the acceptance criteria and establish equivalent assay results for both fresh and frozen specimens.
## 6. Detection Limit:
### Limit of Blank
Five blank samples were tested on the LIAISON MeMed BV assay on one analyzer, with two reagent lots and two technicians over six runs. Testing spanned three days and included 2 replicates per run for a total of 60 results per reagent lot. LoB was calculated as described in CLSI EP17-A2. Results are summarized in the table below.
Table 7. Limit of Blank Test Results
| | Limit of Blank | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | CRP (mg/L) | | IP-10 (pg/mL) | | TRAIL (pg/mL) | |
| | Lot 1 | Lot 2 | Lot 1 | Lot 2 | Lot 1 | Lot 2 |
| Mean | 0.000 | 0.002 | 0.000 | 0.067 | 4.73 | 3.75 |
| SD | 0.000 | 0.014 | 0.000 | 0.309 | 0.366 | 0.35 |
| Concentration | 0.000 | 0.024 | 0.000 | 0.578 | 5.33 | 4.33 |
The assay's LoB was defined as the higher of the two kit lots (0.024 mg/L for CRP, 0.578 pg/mL for IP-10, and 5.33 pg/mL for TRAIL).
### Limit of Detection
K213936 - Page 15 of 22
{15}
Four samples containing low levels of CRP, IP-10, and TRAIL were evaluated on the LIAISON MeMed BC assay on one analyzer, with two reagent lots and two technicians over six runs. Testing spanned three days and included 2 replicates per run for a minimum of 96 dose results. LoD was calculated as described in CLSI EP17-A2. Results are summarized in the table below.
Table 8. Limit of Detection Study Results
| | Limit of Detection | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | CRP (mg/L) | | IP-10 (pg/mL) | | TRAIL (pg/mL) | |
| | Lot 1 | Lot 2 | Lot 1 | Lot 2 | Lot 1 | Lot 2 |
| Mean | 0.392 | 0.375 | 30.95 | 48.57 | 8.21 | 7.13 |
| SD | 0.022 | 0.026 | 2.296 | 2.254 | 1.025 | 0.437 |
| Concentration | 0.037 | 0.067 | 3.8 | 4.31 | 7.03 | 5.05 |
The overall assay LoD for each analyte was defined as the higher of the two kit lots (0.067 mg/L for CRP, 4.31 pg/mL for IP-10, and 7.03 pg/mL for TRAIL).
## Limit of Quantitation
Seven serum samples for CRP, seven serum and/or spiked matrix samples for TRAIL and eight serum samples for IP-10 containing low concentrations of each analyte were evaluated to determine the assay limit of quantitation as described in CLSI EP17-A2. The % CV for each individual sample was plotted against the mean dose result and a best fit curve was applied to determine when the regression line crossed the 20% CV threshold. The assay LoQ for each analyte was defined as the higher of the two kit lots. LoQ for the LIAISON MeMed BV test was identified as 1.0 mg/L for CRP, 100 pg/mL for IP-10, and 15 pg/mL for TRAIL. When one of the three proteins (CRP, IP-10 and TRAIL) is detected below the respective assay LoQ or above the assay specific maximum measurable value, the LoQ value or the maximum is used as the input into the formula.
## 7. Assay Cut-Off:
See clinical cut-off.
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
Clinical validation for the LIAISON MeMed BV assay was performed utilizing a subset of samples used to support clearance of the predicate MeMed BV assay (K210254). Briefly, samples were collected in a multicenter, observational, blinded study (Apollo, Clinicaltrials.gov identifier: NCT04690569) across 11 medical centers (9 in the US and 2 in Israel). The study population included hospital admitted, Emergency Department (ED) and Urgent Care patients with suspected acute bacterial or viral infections.
K213936 - Page 16 of 22
{16}
The primary analysis of the complete clinical study cohort established the diagnostic performance for differentiating bacterial from viral infection in patients with suspected acute bacterial or viral infection using forced expert adjudication as the comparator method (in which physicians were forced to make a bacterial, viral, or noninfectious diagnosis with categorization of patients as indeterminate not allowed). In this analysis, experts were blinded to both C-reactive protein (CRP) and procalcitonin (PCT) values. A secondary analysis was also performed using consensus expert adjudication as a comparator method (in which indeterminate cases were removed from analysis and with the experts given CRP and PCT values). Data from the secondary objective is also presented below and should be considered supplementary information.
A subset of the original Apollo study cohort consisting of 300 samples (delinked and de-identified specimen aliquots) was utilized to demonstrate equivalent clinical performance of the LIAISON MeMed BV test. Fifteen specimens were not evaluated due to sample integrity errors on the LIAISON XL Anlayzer due to low serum volume which precluded the possibility of repeat testing. Of the remaining 285 subjects considered, 51 resided in Israel and 234 in the U.S. Age, gender, ethnicity and racial diversity are summarized in Table 9 below.
Table 9. Demographics of Clinical Samples Evaluated
| Demographic/Clinical Category | | Study Population (N = 285) |
| --- | --- | --- |
| Gender | Female | 156 (54.7%) |
| | Male | 129 (45.3%) |
| Age | 5 months to ≤ 2 years | 14 (4.9%) |
| | > 2 to ≤ 12 years | 44 (15.4%) |
| | > 12 to ≤ 18 years | 17 (6.0%) |
| | > 18 years | 210 (73.7%) |
| Ethnicity | Hispanic/Latino | 38 (13.3%) |
| Race | Native American | 0 (0.0%) |
| | Asian | 16 (5.6%) |
| | Black or African American | 60 (21.1%) |
| | Native Hawaiian | 0 (0.0%, 0.0%) |
| | White | 143 (50.2%) |
| | Other | 15 (5.3%) |
| | Non-US | 51 (17.9%) |
| Comorbidity | Diabetes | 18 (6.3%) |
| | Hypertension | 16 (5.6%) |
| | COPD | 5 (1.8%) |
| | Hyperlipidemia | 13 (4.6%) |
| | None Reported | 233 (81.8%) |
| Clinical Syndrome | LRTI | 29 (10.2%) |
| | URTI | 149 (52.3%) |
| | LRTI/URTI | 12 (4.2%) |
K213936 - Page 17 of 22
{17}
| | LRTI, UTI | 1 (0.4%) |
| --- | --- | --- |
| | UTI | 7 (2.5%) |
| | UTI; URTI | 2 (0.7%) |
| | Other | 85 (29.8%) |
All specimens were shipped frozen on dry ice to DiaSorin Inc. Specimen aliquots were randomized and shipped frozen to two additional external sites for testing. Results of the LIAISON MeMed BV assay for internal assay concentration values for individual analytes and score values were compared directly to the predicate results utilizing Deming Regression Analysis. Results are summarized in the table below.
Table 10. Deming Regression Analysis Results – LIAISON MeMed BV Test Results Compared to the Predicate Device.
| Analyte | N | Range | Slope (95% CI) | Y-intercept (95% CI) | R² (Spearman) | R (Spearman) |
| --- | --- | --- | --- | --- | --- | --- |
| TRAIL | 238 | 15.6 - 300 pg/mL | 0.98 (0.91 – 1.05) | -2.47 (-8.03 to 3.09) | 0.86 | 0.93 |
| IP-10 | 211 | 100 – 2000 pg/mL | 1.13 (1.05 – 1.21) | -13.67 (-45.08 to 17.7) | 0.94 | 0.97 |
| CRP | 259 | 1.0 – 229.3 mg/L | 1.15 (1.10 – 1.20) | -0.92 (-2.08 to 0.23) | 0.98 | 0.99 |
| Score | 285 | 0.0 - 100 | 1.03 (1.0 – 1.05) | 0.84 (0.02 to 1.65) | 0.94 | 0.97 |
Additional analyses were performed to calculate the bias at each relevant cut-off across the LIAISON MeMed BV Score range as detailed below.
Table 11. Deming Regression Bias Analysis – Bias estimates at relevant cutoffs
| | LIAISON MeMed BV Score Bias | 95% CI | |
| --- | --- | --- | --- |
| 10 | 1.1 | 0.3 | 2.1 |
| 35 | 1.7 | 0.5 | 3.1 |
| 65 | 2.5 | 0.6 | 4.7 |
| 90 | 3.2 | 0.7 | 6.0 |
The bias at each bin cut-off met acceptance criteria of less than 12.5 score units, and ensure a low likelihood of specimens shifting to a non-adjacent bin. A comparison of the LIAISON MeMed BV Bin result to that obtained with the predicate device on the same clinical samples is summarized below.
Table 12. Clinical Sample Bin Results – LIAISON MeMed BV Comparison to the Predicate
K213936 - Page 18 of 22
{18}
| LIAISON MeMed BV Bin Results | Predicate Bin Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Bin 1 | Bin 2 | Bin 3 | Bin 4 | Bin 5 |
| Bin 1 | 146 | 7 | 0 | 0 | 0 |
| Bin 2 | 13 | 28 | 7 | 0 | 0 |
| Bin 3 | 0 | 12 | 8 | 5 | 0 |
| Bin 4 | 0 | 1 | 9 | 15 | 1 |
| Bin 5 | 0 | 0 | 0 | 4 | 29 |
| Agreement | 91.8% (146/159) | 58.3% (28/48) | 33.3% (8/24) | 62.5% (15/24) | 96.7% (29/30) |
Out of all clinical samples evaluated, only 1 sample exhibited a shift to a non-adjacent bin when tested with the LIAISON MeMed BV test.
Additional analyses of the LIAISON MeMed BV Results were performed by comparing results from the LIAISON test to forced expert adjudication result utilized as the comparator in the Apollo clinical study.
Table 13. LIAISON MeMed BV Comparison to Physician Adjudication Result – Forced Adjudication
| LIAISON MeMed BV Score Bin | N | Forced Bacterial Diagnosis | Forced Viral/ Noninfectious Diagnosis | % Total Patients | % Patients Bacterial | % Patients Viral/ Noninfectious | LR (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 90 ≤ score ≤ 100 | 33 | 22 | 11 | 11.6 | 66.7 | 33.3 | 13.0 (7.09-25.83) |
| 65 < score < 90 | 26 | 8 | 18 | 9.1 | 30.8 | 69.2 | 2.89 (1.27-5.95) |
| 35 ≤ score ≤ 65 | 25 | 6 | 19 | 8.8 | 24.0 | 76.0 | 2.05 (0.79-4.51) |
| 10 < score < 35 | 48 | 1 | 47 | 16.8 | 2.1 | 94.7 | 0.14 (0.01-0.6) |
| 0 ≤ score ≤ 10 | 153 | 1 | 152 | 53.7 | 0.7 | 99.3 | 0.043 (0.002-0.18) |
| Total | 285 | 38 | 247 | 100 | | | |
A significant trend was demonstrated between the LIAISON MeMed BV Score and increasing likelihood of bacterial infection across Score bins. Additionally, individuals with Scores in the lowest bin have a high likelihood of viral or non-bacterial etiology and the individuals with scores in the highest bin have a high likelihood of bacterial infection. To further demonstrate equivalent performance to the predicate device, a similar analysis was performed for the predicate device results on the same clinical samples evaluated as part of this study.
Table 14. Predicate Performance Comparison to Physician Adjudication Result – Forced Adjudication
K213936 - Page 19 of 22
{19}
| Predicate Score Bin | N | Forced Bacterial Diagnosis | Forced Viral/ Noninfectious Diagnosis | % Total Patients | % Patients Bacterial | % Patients Viral/ Noninfectious | LR (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 90 ≤ score ≤ 100 | 30 | 20 | 10 | 10.5 | 66.7 | 33.3 | 13.0 (6.83-27.01) |
| 65 < score < 90 | 24 | 10 | 14 | 8.4 | 41.7 | 58.3 | 4.64 (2.16-9.63) |
| 35 ≤ score ≤ 65 | 24 | 5 | 19 | 8.4 | 20.8 | 79.2 | 1.71 (0.6-3.96) |
| 10 < score < 35 | 48 | 2 | 46 | 16.8 | 4.2 | 95.8 | 0.28 (0.05-0.85) |
| 0 ≤ score ≤ 10 | 159 | 1 | 158 | 55.8 | 0.6 | 99.4 | 0.041 (0.002-0.173) |
| Total | 285 | 38 | 247 | 100 | | | |
Overall, these results demonstrate equivalent performance between the LIAISON MeMed BV test and the predicate device.
Additional analyses of the LIAISON MeMed BV test was performed according to the secondary analysis in which the adjudicators were un-blinded to PCT and CRP results. Individuals for whom adjudicators could not agree were labeled indeterminate and excluded from this analysis. Twenty eight of the 285 samples were excluded.
Table 15. LIAISON MeMed BV Comparison to Physician Adjudication Result – Consensus Adjudication
| LIAISON MeMed BV Score Bin | N | Consensus Bacterial Diagnosis | Consensus Viral/ Noninfectious Diagnosis | % Total Patients | % Patients Bacterial | % Patients Viral/ Noninfectious | LR (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 90 ≤ score ≤ 100 | 24 | 21 | 3 | 9.3 | 87.5 | 12.5 | 52.97 (19.9-214.87) |
| 65 < score < 90 | 20 | 5 | 15 | 7.8 | 25.0 | 75.0 | 2.52 (0.87-5.99) |
| 35 ≤ score ≤ 65 | 23 | 3 | 20 | 9.0 | 13.0 | 87.0 | 1.14 (0.28-3.07) |
| 10 < score < 35 | 42 | 0 | 42 | 16.3 | 0.0 | 100.0 | 0.0 (0.0-0.341) |
| 0 ≤ score ≤ 10 | 148 | 1 | 147 | 57.6 | 0.7 | 99.3 | 0.051 (0.003-0.214) |
| Total | 257 | 30 | 227 | 100 | | | |
Similar analyses were performed for the predicate device in these samples.
K213936 - Page 20 of 22
{20}
Table 16. Predicate Performance Comparison to Physician Adjudication Result – Consensus Adjudication
| Predicate Score Bin | N | Consensus Bacterial Diagnosis | Consensus Viral/ Noninfectious Diagnosis | % Total Patients | % Patients Bacterial | % Patients Viral/ Noninfectious | LR (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 90 ≤ score ≤ 100 | 21 | 18 | 3 | 8.2 | 85.7 | 14.3 | 45.4 (16.75-185.3) |
| 65 < score < 90 | 20 | 9 | 11 | 7.8 | 45 | 55.0 | 6.19 (2.72-13.77) |
| 35 ≤ score ≤ 65 | 19 | 2 | 17 | 7.4 | 10.5 | 89.5 | 0.89 (0.15-2.89) |
| 10 < score < 35 | 44 | 0 | 44 | 17.1 | 0.0 | 100.0 | 0.0 (0.0-0.326) |
| 0 ≤ score ≤ 10 | 153 | 1 | 152 | 59.5 | 0.7 | 99.3 | 0.050 (0.003-0.207) |
| Total | 257 | 30 | 227 | 100 | | | |
Cumulatively, the results from clinical sample testing demonstrate equivalent performance between the LIAISON MeMed BV test and the predicate device.
2. **Matrix Comparison:**
Not applicable.
C **Clinical Studies:**
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
See section B.1 above.
D **Clinical Cut-Off:**
The Score algorithm of the LIAISON MeMed BV test is identical to that of the predicate device (K210254). See the K210254 decision summary for additional details regarding establishment of clinical cut-offs.
K213936 - Page 21 of 22
{21}
E Expected Values/Reference Range:
Predictive values depend on the likelihood ratios and the prevalence of disease. Laboratories and other users should establish their own reference intervals for their patient populations using the LIAISON MeMed BV test to reflect potential sources of variability, such as patient gender, race, or age.
To assess the expected reference range for the LIAISON MeMed BV assay, serum samples from 150 apparently healthy asymptomatic adults were collected and evaluated on a single lot of LIAISON MeMed BV reagents. Study subjects were 65% female and 35% male and ranged in age from 21-71.
Table 14. Reference Interval for the LIAISON MeMed BV Score in Healthy Individuals
| Race | LIAISON MeMed BV Score Band | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | 1
(0 ≤ score ≤ 10) | 2
(10 < score < 35) | 3
(35 ≤ score ≤ 65) | 4
(65 < score < 90) | 5
(90 ≤ score ≤ 100) |
| American Indian | N | 5 | 0 | 0 | 0 | 0 |
| | % | 100% | 0 | 0 | 0 | 0 |
| Asian | N | 8 | 2 | 0 | 0 | 0 |
| | % | 80% | 20% | 0 | 0 | 0 |
| Black | N | 6 | 1 | 0 | 0 | 0 |
| | % | 85.7% | 14.3% | 0 | 0 | 0 |
| Hispanic | N | 49 | 7 | 2 | 1 | 0 |
| | % | 83.1% | 11.9% | 3.4% | 1.7% | 0 |
| White | N | 94 | 14 | 1 | 1 | 0 |
| | % | 85.5% | 12.7% | 0.9% | 0.9% | 0 |
The Mean Score value in the healthy population was 6.9, the median score was 4, and the central 95% interval spanned the range 1-33.
F Other Supportive Instrument Performance Characteristics Data:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K213936 - Page 22 of 22
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.