Percent Oxygen Sensors are intended to replace the original oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures.
Device Story
Electrochemical galvanic oxygen sensors; function as replacement components for medical oxygen analyzers. Sensors generate electrical current proportional to oxygen concentration in breathing gas mixtures; output signal used by host equipment to measure and display oxygen levels. Used in hospital, pre-hospital, industrial, and home settings; operated by clinicians or patients depending on host device. Provides real-time oxygen concentration data to healthcare providers or users to monitor inspired oxygen levels; facilitates clinical decision-making regarding respiratory therapy and anesthesia gas management. Benefits include continued accurate monitoring of oxygen delivery.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 80601-2-55, including accuracy, precision, specificity, sensitivity, linearity, range, and response time.
Technological Characteristics
Electrochemical galvanic oxygen sensors. Principle of operation based on Faraday's law of electrolysis; current generation proportional to oxygen volume and lead anode consumption. Variations in physical housing, signal connection, and output to match OEM host equipment. No software or connectivity features.
Indications for Use
Indicated for use as replacement oxygen-sensing components in oxygen analyzers measuring breathing gas concentrations. Patient population follows the equipment to which the sensor is attached; intended for hospital, institutional, pre-hospital, industrial, and home settings.
Regulatory Classification
Identification
An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
August 19, 2022
CareOx, LLC % Paul Dryden Consultant ProMedic Consulting, LLC 131 Bav Point Dr NE Saint Petersburg, Florida 33704
Re: K213933
Trade/Device Name: Percent Oxygen Sensors Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: Class II Product Code: CCL Dated: July 20, 2022 Received: July 21, 2022
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K213933
Device Name Percent Oxygen Sensors
Indications for Use (Describe)
Percent Oxygen Sensors are intended to replace the original oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instruction of infing data sources, gather and maintain the data needed and complete and review the collection, includion. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
## PDF Page 14 of 70
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
| Date: | August 16, 2022 |
|----------------------------|---------------------------------------------------------------------|
| Sponsor: | CAREOX®, LLC<br>103 Carnegie Center Ste. 300<br>Princeton, NJ 08540 |
| | Tel - 609.524.2540 |
| Sponsor Contact: | Young Kim – CEO |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | Percent Oxygen Sensors |
| Common/Usual Name: | Oxygen Sensors |
| Regulation Number: | 21CFR 868.1720 |
| Regulation Code: | Analyzer, Gas, Oxygen, Gaseous-Phase |
| Product Code: | CCL |
| Regulatory Class: | II |
| Predicate Device: | K952736 Analytic Industries Percent Oxygen Sensors |
#### Device Description:
The subject device is a family of Medical Oxygen Sensors which may be used as industry replacement types with various medical inspired-oxygen measuring devices.
The oxygen sensors are all electrochemical galvanic type devices. The difference between models is merely the physical shape of the external housing (the basic oxygen sensor is often time encapsulated in a secondary housing), signal connection, signal output and the response time.
The family of oxygen sensors concept extends to OEM manufacturers of anesthesia and respiratory therapy equipment. Again, the difference between models is merely the physical shape of the external housing (the basic oxygen sensor is often time encapsulated in a secondary housing), signal connection, signal output and the response time in order for the OEM to exercise a degree of control over the recurring replacement of oxygen sensors.
#### Indications for Use:
Percent Oxygen Sensors are intended to replace the original oxygen-sensing components of an oxygen analyzer that measures oxygen concentration in breathing gases.
| Patient Population: | The sensor follows the population of the equipment for which it is attached. |
|---------------------|------------------------------------------------------------------------------|
|---------------------|------------------------------------------------------------------------------|
Environment of Use: Hospital/institutional, pre-hospital, industrial and home settings.
{4}------------------------------------------------
#### Substantial Equivalence Discussion:
The subject devices are manufactured by Analytical Industries Inc. and were cleared under K952736. CareOx, the sponsor of this submission is purchasing final, finished oxygen sensors from Analytical Industries Inc. with the private label of CareOx.
Indications - Equivalent to the predicate
Technology - The technology is identical
Principal of Operation - The principal of operation is identical
Operating specifications – Equivalent
Environment of Use - Similar
Patient Population – Similar
Non-clinical Testing
We have performed the applicable test from ISO 80601-2-55.
- These tests includes: Accuracy Precision Specificity Sensitivity Linearity Range Response time
#### Differences:
There are no differences between the proposed device and the predicate device.
{5}------------------------------------------------
## Table of Comparison to Predicate
| Attribute | Predicate<br>Analytic Industries<br>Percent Oxygen Sensors | Subject Device<br>CareOx | Equivalency to at least one or<br>both devices |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| K# | K952736 | TBD | Similar |
| Product Code<br>Classification Name | Analyzer, Gas, Oxygen, Gaseous-Phase<br>CCL | Analyzer, Gas, Oxygen, Gaseous-Phase<br>CCL | Similar |
| Indications for Use | Percent Oxygen Sensors are intended to replace the<br>original oxygen-sensing components of an oxygen<br>analyzer that measures oxygen concentration in<br>breathing gases. | Percent Oxygen Sensors are intended to replace the<br>original oxygen-sensing components of an oxygen<br>analyzer that measures oxygen concentration in<br>breathing gases. | Similar |
| Environments of<br>use | Replacement sensor for equipment that provides<br>inspired oxygen concentrations | Replacement sensor for equipment that provides<br>inspired oxygen concentrations | Similar |
| Principle of<br>operation | Electrochemical galvanic | Electrochemical galvanic | Similar |
| Prescriptive | Yes | Yes | Similar |
| Service Life | The life of sensor is based on the Faraday's law of<br>electrolysis, i.e., the current generated by the sensor<br>per unit volume of oxygen and the amount of the<br>lead anode in the sensor determines the life of the<br>sensor. Only 70% of theoretically calculated sensor<br>life is used to claim the useful life of the sensor. For<br>example, if theoretical life is calculated as 90<br>months, the claimed life of the sensor would be 63<br>months. | The life of sensor is based on the Faraday's law of<br>electrolysis, i.e., the current generated by the sensor<br>per unit volume of oxygen and the amount of the<br>lead anode in the sensor determines the life of the<br>sensor. Only 70% of theoretically calculated sensor<br>life is used to claim the useful life of the sensor. For<br>example, if theoretical life is calculated as 90<br>months, the claimed life of the sensor would be 63<br>months. | Similar |
### Substantial Equivalence Conclusion:
The sponsor has demonstrated through performance testing, design and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.