← Product Code [FXX](/productcode/FXX) · K213920

# Kangbeier Surgical Mask (Model: KBR-1001) (K213920)

_Xinxiang Kangbeier Medical Technology Co., Ltd. · FXX · Jan 16, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213920

## Device Facts

- **Applicant:** Xinxiang Kangbeier Medical Technology Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Jan 16, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The disposable surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

## Device Story

Kangbeier Surgical Mask (Model: KBR-1001) is a 3-ply, flat-pleated, single-use, non-sterile surgical mask. Construction includes inner/outer spun-bond polypropylene layers and a middle melt-blown polypropylene filter layer; nose piece is galvanized wire coated with polyethylene; ear loops are nylon and polyurethane. Device is worn by healthcare personnel and patients in clinical settings to provide a physical barrier against microorganisms, body fluids, and particulates. Performance meets ASTM F2100 Level 3 standards for fluid resistance, particulate filtration, and bacterial filtration. Device is intended for infection control to reduce exposure to blood and body fluids.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration efficiency ≥ 98%), ASTM F2101 (bacterial filtration efficiency ≥ 98%), ASTM F2100 (differential pressure < 6.0 mmH2O/cm2), and 16 CFR 1610 (flammability Class 1). Biocompatibility testing per ISO 10993-1:2018 confirmed the device is non-cytotoxic, non-sensitizing, and non-irritating.

## Technological Characteristics

3-ply surgical mask; materials: spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter); nose piece: galvanized wire with polyethylene coating; ear loops: nylon/polyurethane. Dimensions: 17.8 cm x 9.2 cm. Performance: ASTM F2100 Level 3. Non-sterile, single-use. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- DemeMASK ([K201479](/device/K201479.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 16, 2022

Xinxiang Kangbeier Medical Technology Co., Ltd. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K213920

Trade/Device Name: Kangbeier Surgical Mask (Model: KBR-1001) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 14, 2021 Received: December 15, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K213920

Device Name Kangbeier Surgical Mask (Model: KBR-1001)

Indications for Use (Describe)

The disposable surgical face masks are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

|  | <span></span> |
|--|---------------|
|  | <span></span> |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY K213920

This summary of 510(k) is submitted in accordance with the requirements of 21 CFR §807.92:

#### l. SUBMITTER

Xinxiang Kangbeier Medical Technology Co., Ltd. East Mancun Industrial District, Changyuan, Henan 48105 China Tel: +86-13341254319 Fax: N/A

| Contact Person: | Shaoju Tian      |
|-----------------|------------------|
| Date Prepared:  | December 8, 2021 |

#### II. SUBJECT DEVICE

| Device Trade Name:           | Kangbeier Surgical Mask (Model: KBR-1001) |
|------------------------------|-------------------------------------------|
| Classification Name:         | Surgical Mask                             |
| Regulation:                  | 21 CFR §878.4040                          |
| Regulatory Class:            | Class II                                  |
| Device Panel:                | General Hospital                          |
| Product Classification Code: | FXX                                       |

#### lll. PREDICATE DEVICE

| Predicate Manufacturer: | DemeTECH Corporation |
|-------------------------|----------------------|
| Predicate Trade Name:   | DemeMASK             |
| Predicate 510(k):       | K201479              |

No reference devices were used in this submission.

#### IV. DEVICE DESCRIPTION

The Kangbeier Surgical Mask (Model: KBR-1001) is non-sterile, disposable, single-use surgical mask. It is manufactured with three layers, the inner and outer layers are made of non-woven, spun-bond polypropylene, while the middle filter layer is made of a non-woven, melt-blown polypropylene.

#### V. INDICATIONS FOR USE

The disposable surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVCE

The following characteristics were compared between the subject device and the predicate device in Table 1 below.

{4}------------------------------------------------

| Device              | Subject Device -<br>Kangbeier Surgical Mask<br>(Model: KBR-1001)                                                                                                                                                                                                                                                                                                                                                            | Predicate Device -<br>DemeMASK (K201479)                                                                                                                                                                                                                                                                                                                                                                               | Result                                    |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Manufacturer        | Xinxiang Kangbeier<br>Medical Technology Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                       | DemeTECH Corporation                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                       |
| 510K Number         | TBD                                                                                                                                                                                                                                                                                                                                                                                                                         | K201479                                                                                                                                                                                                                                                                                                                                                                                                                | N/A                                       |
| Product Common Name | SURGICAL FACE MASK                                                                                                                                                                                                                                                                                                                                                                                                          | SURGICAL FACE MASK                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                 |
| Trade Name          | Kangbeier Surgical Mask<br>(Model: KBR-1001)                                                                                                                                                                                                                                                                                                                                                                                | DemeMASK Surgical Mask                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                                       |
| Product Code        | FXX                                                                                                                                                                                                                                                                                                                                                                                                                         | FXX                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                 |
| Classification      | Class II (21 CFR 878.4040)                                                                                                                                                                                                                                                                                                                                                                                                  | Class II (21 CFR 878.4040)                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                 |
| Intended Use        | The disposable surgical face<br>masks are intended to be worn<br>to protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material.<br>These face masks are intended<br>for use in infection control<br>practices to reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device<br>provided non-sterile. | The Disposable Surgical Face Masks<br>are intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material.<br>These face masks are intended for<br>use in infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single use, disposable device<br>provided non- sterile. | Identical                                 |
| Model               | 3 Ply, Ear Loops, Flat-Pleated<br>Style                                                                                                                                                                                                                                                                                                                                                                                     | 3 Ply, Ear Loops or Tie-On                                                                                                                                                                                                                                                                                                                                                                                             | No tie-on<br>option for<br>subject device |

Table 1 is continued on page 2

{5}------------------------------------------------

| Materials             |                                             |                                         |                                                |
|-----------------------|---------------------------------------------|-----------------------------------------|------------------------------------------------|
| Outer Facing<br>Layer | Spun-Bond Polypropylene                     | Spun-Bond Polypropylene                 | Similar                                        |
| Middle Layer          | Melt-Blown Polypropylene                    | Melt-Blown Polypropylene Filter         | Similar                                        |
| Inner Facing Layer    | Spun-Bond Polypropylene                     | Spun-Bond Polypropylene                 | Similar                                        |
| Nose Piece            | Galvanized wire coated with<br>polyethylene | Single Galvanized Wire, Coated<br>By PE | Similar                                        |
| Ear Loops             | Nylon & polyurethane                        | Not made with natural rubber latex      | Similar                                        |
| Color                 | Blue                                        | White                                   | Subject device<br>includes blue<br>colorant    |
| Dimension (Width)     | $9.2 cm \pm 0.2 cm$                         | $9.5 cm \pm 1.0 cm$                     | More narrow<br>tolerance for subject<br>device |
| Dimension<br>(Length) | $17.8 cm \pm 0.2 cm$                        | $17.5 cm \pm 1.0 cm$                    | More narrow<br>tolerance for subject<br>device |
| OTC Use               | Yes                                         | Yes                                     | Same                                           |
| Sterility             | Non-Sterile                                 | Non-Sterile                             | Same                                           |
| Use                   | Single Use                                  | Single Use                              | Same                                           |
| ASTM F2100 Level      | Level 3                                     | Level 3                                 | Same                                           |

{6}------------------------------------------------

## VII. NON-CLINICAL DATA

The subject device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004. A summary of the benchtop performance testing results is provided below in Table 2 along with the acceptance criteria.

| Item                                                     | Purpose                              | Acceptance Criteria              | Result                                                                                                                                                |
|----------------------------------------------------------|--------------------------------------|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Level 3 Fluid<br>Resistance<br>Performance ASTM<br>F1862 | Resistance to Fluid                  | 29 Out of 32 pass at 160<br>mmHg | Pass<br>30-31 Out of 32 pass at 160 mmHg<br>LOT #1, 30/32 passed, @ 160 mmHg<br>LOT #2 30/32 passed, @ 160 mmHg<br>LOT #3 31/32 passed, @ 160 mmHg    |
| Particulate Filtration<br>Efficiency ASTM<br>F2299       | Particulate Filtration<br>Efficiency | ≥ 98%                            | Pass<br>98.62-99.34%<br>LOT #1 AVE Filtration Efficiency ≥ 98%<br>LOT #2 AVE Filtration Efficiency ≥ 98%<br>LOT #3 AVE Filtration Efficiency ≥ 98%    |
| Bacterial Filtration<br>Efficiency ASTM<br>F2101         | Bacterial Filtration Efficiency      | ≥ 98%                            | Pass<br>99.5 - 99.1%<br>LOT #1 AVE Filtration Efficiency ≥ 98%<br>LOT #2 AVE Filtration Efficiency ≥ 98%<br>LOT #3 AVE Filtration Efficiency ≥ 98%    |
| Differential Pressure<br>ASTM F2100                      | Differential Pressure                | < 6.0 mmH20/cm2                  | Pass<br>3.92 - 4.78 mmH20/cm2<br>LOT #1 AVE Differential Pressure 5.0<br>LOT #2 AVE Differential Pressure 5.0<br>LOT #3 AVE Differential Pressure 5.0 |
| Class 1 Flammability<br>16 CFR 1610                      | Flammability                         | < 3.5 second burn time           | Pass<br>Class 1, Did Not Ignite (DNI)<br>LOT #1 Did not ignite<br>LOT #2 Did not ignite<br>LOT #3 Did not ignite                                      |

|  | Table 2 - Benchtop Performance Testing |  |
|--|----------------------------------------|--|
|  |                                        |  |

{7}------------------------------------------------

## BIOCOMPATIBILITY TESTING

Biocompatibility testing was performed in accordance with ISO 10993-1:2018. Specifically, the following testing endpoints were evaluated.

|                                      |                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                 | Result                                                                      |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Standards                            | Purpose / Method                                                                                                                | Acceptance<br>Criteria                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                             |
| Cytotoxicity -<br>ISO 10993-5        | Purpose: Cytotoxic<br>potential<br>Method: Evaluation of<br>test Articles by the<br>MEM Extraction Test                         | If viability is reduced to<br><70% of the reagent<br>control extract, a<br>cytotoxic potential<br>exists.                                                                                                                                                                                                                                                                                                                       | Pass<br>Under the conditions of the study the device is<br>non- cytotoxic   |
| Skin Sensitization<br>- ISO 10993-10 | Purpose: Sensitizing<br>potential<br>Method: Magnusson-<br>Kligman Sensitization /<br>Guinea Pig<br>Maximization Test<br>(GPMT) | Magnusson and Kligman<br>grades of less than 1                                                                                                                                                                                                                                                                                                                                                                                  | Pass<br>Under the conditions of the study the device is<br>non- sensitizing |
| Skin Irritation -<br>ISO 10993-10    | Purpose: Potential for<br>irritation<br>Method: Intracutaneous<br>Injection Test                                                | Erythema and Eschar<br>Formation Value (total<br>possible =4)<br>No erythema = 0<br>Very slight erythema<br>(barely perceptible) =1<br>Well defined<br>erythema=2<br>Moderate erythema = 3<br>Severe erythema (beet<br>redness) to eschar<br>Formation (preventing<br>grading of erythema) =4<br>Total possible erythema<br>score = 4<br>Edema Formation Value<br>No edema = 0<br>Very slight edema<br>(barely perceptible) = 1 | Pass<br>Under the conditions of the study the device is<br>non- irritating  |

Table 3 - Biocompatibility Testing

## VIII. CONCLUSIONS

The conclusions drawn from the nonclinical and clinical tests that the device is as safe, as effective, and performs as well as or better than the legally marketed device(K201479), manufactured by DemeTECH Corporation.

---

**Source:** [https://fda-staging.innolitics.com/device/K213920](https://fda-staging.innolitics.com/device/K213920)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).