K213907 · Care Essentials Pty, Ltd. · FRG · Mar 14, 2022 · General Hospital
Device Facts
Record ID
K213907
Device Name
CareGUARD Sterilization Wrap
Applicant
Care Essentials Pty, Ltd.
Product Code
FRG · General Hospital
Decision Date
Mar 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
CareGUARD* Sterilization Wrap is a single-use, non-woven sterilization wrap intended to be sterilized by a health care provider using pre-vacuum steam at 273ºF / 134ºC for 4 minutes. The CareGUARD* Sterilization Wrap was validated for a dry time of 30 minutes. The CareGUARD* Sterilization wrap is recommended to be used with trays weighing up to 13.7 lbs. The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used.
Device Story
CareGUARD* Sterilization Wrap is a single-use, non-woven fabric wrap used in clinical settings to enclose medical devices prior to sterilization. The device is composed of 100% polypropylene spunbond-meltblown-spunbond (SMS) layers. It is used by healthcare personnel to wrap surgical instrument trays (up to 13.7 lbs) using sequential or simultaneous double-wrapping techniques. The wrap allows for steam penetration during pre-vacuum sterilization cycles and maintains the sterility of the enclosed instruments post-sterilization. The device benefits patients by ensuring the sterility of surgical instruments until the moment of use in a procedure.
Clinical Evidence
No clinical data. Bench testing only. Testing included physical and chemical property verification, biocompatibility (ISO 10993-5, 10993-10), bacterial filtration efficiency (ASTM F2101-14), and validation of pre-vacuum steam sterilization (SAL 10^-6), sterilant penetration, dry time, and sterility maintenance (30 days).
Technological Characteristics
Material: 100% polypropylene spunbond-meltblown-spunbond (SMS) non-woven fabric. Construction: Single or dual-layer sheets, ultrasonically bonded edges. Basis weights: 50, 60, 75 GSM. Sterilization: Pre-vacuum steam (273°F/134°C). Standards: ISO 11607-1, EN 868-2, ISO 10993, ASTM F2101-14, ANSI/AAMI ST79:2017.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization via pre-vacuum steam (273ºF/134ºC, 4 min) and to maintain sterility of the enclosed devices until use. Recommended for trays up to 13.7 lbs (6.2 kg).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 14, 2022
Care Essentials Pty Ltd Abhay Sinha Managing Director 103 Mornington Street North Geelong, Victoria 3215 Australia
Re: K213907
Trade/Device Name: CareGUARD* Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: December 7, 2021 Received: December 14, 2021
## Dear Abhay Sinha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213907
Device Name CareGUARD* Sterilization Wrap
#### Indications for Use (Describe)
CareGUARD* Sterilization Wrap is a single-use, non-woven sterilization wrap intended to be sterilized by a health care provider using pre-vacuum steam at 273ºF / 134ºC for 4 minutes. The CareGUARD* Sterilization Wrap was validated for a dry time of 30 minutes. The CareGUARD* Sterilization wrap is recommended to be used with trays weighing up to 13.7 lbs.
The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used.
Test results validated that CareGUARD* Sterilization wraps allowed sterilization of the enclosed devices by pre-vacuum cycles.
Refer to Table 1 for recommendations for use per Pre-vacuum steam Sterility Maintenance Study.
#### Table 1: Basis weight recommendations for pre-vacuum steam
| Basis Weight | Maximum Wrapped Package Content Weight |
|--------------|----------------------------------------|
| 50 GSM | |
| 60 GSM | 6.2 kg / 13.7 lbs |
| 75 GSM | |
The following load was used in the Pre-vacuum steam Sterility Maintenance Study:
- · 50 GSM: 21 in. x 10 in. x 4 ¾ in. (54 cm x 25 cm x 12 cm) tray containing 13.7 lbs (6.2 kg) of metal mass The metal mass included surgical instruments used for orthopedic joint replacement
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Care Essentials. The logo features a stylized letter "C" in two colors: gold on the inside and blue on the outside. Below the "C" are the words "CARE ESSENTIALS" in a sans-serif font, with "CARE" in a larger font size than "ESSENTIALS".
# K213907 CareGUARD* Sterilization Wrap 510(k) SUMMARY
| 510(k) Applicant: | Abhay Sinha<br>Care Essentials Pty. Ltd.<br>103 Mornington Street<br>North Geelong<br>Victoria, Australia 3215<br>Phone: +61-3-5277-1455<br>Email: queries@careessentials.com.au |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Correspondent: | Abhay Sinha<br>Care Essentials Pty. Ltd.<br>103 Mornington Street<br>North Geelong<br>Victoria, Australia 3215<br>Phone: +61-3-5277-1455<br>Email: queries@careessentials.com.au |
| Date of Summary: | December 7, 2021 |
| Type of 510(k) Submission: | Traditional |
| Device Trade or Proprietary Name:<br>Common name:<br>Classification Name:<br>Product Code:<br>Classification Panel:<br>Regulatory Class:<br>Regulation #: | CareGUARD*TM Sterilization Wrap<br>Sterilization Wrap<br>Wrap, Sterilization<br>FRG<br>General Hospital<br>Class II<br>21 CFR 880.6850 |
| Predicate Device: | Kimberly-Clark* Corporation KIMGUARD<br>Sterilization Wrap - K082554 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Care Essentials. The logo features a stylized letter "e" in two colors: gold and blue. The gold color is used for the inner part of the "e", while the blue color is used for the outer part. Below the stylized "e", the words "CARE ESSENTIALS" are written in blue, with "CARE" on the top line and "ESSENTIALS" on the bottom line.
## Device Description
CareGUARD* Sterilization Wrap is made from 100% polypropylene spunbond-meltblownspunbond (SMS) non-woven fabric. The sterilization wrap is provided in three different material basis weights of three product offerings in various dimensions. CareGUARD* Single Layer Sterilization wrap is comprised of a single sheet or one layer of SMS fabric. CareGUARD* Dual Layers Sterilization Wrap and CareGUARD* Contrast Sterilization Wrap, are comprised of two single layer sheets of SMS fabric ultrasonically bonded along two opposing edges. The wrap design allows for use of the sequential or simultaneous double-wrapping technique per recommendations from ANSI/AAMI ST79:2017 and allows for a sterilized pack to be opened aseptically. All basis weights utilize the same material technology. This product is a single-use and a non-sterile medical device.
CareGUARD* Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using pre-vacuum steam at 273°F/134°C for 4 minutes.
| No. | Model | Product Description |
|-----|----------------|---------------------------------------------------------------------------------------------------------------|
| 1 | SW606050-B | CareGUARD* Single Layer, 60cm X 60cm (24" X 24"), 50GSM, Blue |
| 2 | SW606060-B | CareGUARD* Single Layer, 60cm X 60cm (24" X 24"), 60GSM, Blue |
| 3 | SW606075-B | CareGUARD* Single Layer, 60cm X 60cm (24" X 24”), 75GSM, Blue |
| 4 | SW767650-B | CareGUARD* Single Layer, 76cm X 76cm (30” X 30"), 50GSM, Blue |
| 5 | SW767660-B | CareGUARD* Single Layer, 76cm X 76cm (30" X 30"), 60GSM, Blue |
| 6 | SW767675-B | CareGUARD* Single Layer, 76cm X 76cm (30” X 30”), 75GSM, Blue |
| 7 | SW919150-B | CareGUARD* Single Layer, 91cm X 91cm (36" X 36"), 50GSM, Blue |
| 8 | SW919160-B | CareGUARD* Single Layer, 91cm X 91cm (36" X 36"), 60GSM, Blue |
| 9 | SW919175-B | CareGUARD* Single Layer, 91cm X 91cm (36" X 36"), 75GSM, Blue |
| 10 | SW10010050-B | CareGUARD* Single Layer, 100cm X 100cm (40" X 40"), 50GSM, Blue |
| 11 | SW10010060-B | CareGUARD* Single Layer, 100cm X 100cm (40" X 40"), 60GSM, Blue |
| 12 | SW10010075-B | CareGUARD* Single Layer, 100cm X 100cm (40" X 40"), 75GSM, Blue |
| 13 | SW11611650-B | CareGUARD* Single Layer, 116cm X 116cm (45" X 45"), 50GSM, Blue |
| 14 | SW11611660-B | CareGUARD* Single Layer, 116cm X 116cm (45" X 45"), 60GSM, Blue |
| 15 | SW11611675-B | CareGUARD* Single Layer, 116cm X 116cm (45" X 45"), 75GSM, Blue |
| 16 | SW12012050-B | CareGUARD* Single Layer, 120cm X 120cm (48" X 48"), 50GSM, Blue |
| 17 | SW12012060-B | CareGUARD* Single Layer, 120cm X 120cm (48" X 48"), 60GSM, Blue |
| 18 | SW12012075-B | CareGUARD* Single Layer, 120cm X 120cm (48" X 48"), 75GSM, Blue |
| 19 | SW13713750-B | CareGUARD* Single Layer, 137cm X 137cm (54" X 54"), 50GSM, Blue |
| 20 | SW13713760-B | CareGUARD* Single Layer, 137cm X 137cm (54" X 54"), 60GSM, Blue |
| 21 | SW13713775-B | CareGUARD* Single Layer, 137cm X 137cm (54" X 54"), 75GSM, Blue |
| 22 | SW13718250-B | CareGUARD* Single Layer, 137cm X 182cm (54" X 72"), 50GSM, Blue |
| 23 | SW13718260-B | CareGUARD* Single Layer, 137cm X 182cm (54" X 72"), 60GSM, Blue |
| 24 | SW13718275-B | CareGUARD* Single Layer, 137cm X 182cm (54" X 72"), 75GSM, Blue |
| 25 | SW606050-B/B | CareGUARD* Dual Layers, 60cm X 60cm (24" X 24"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 26 | SW606060-B/B | CareGUARD* Dual Layers, 60cm X 60cm (24" X 24"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 27 | SW606075-B/B | CareGUARD* Dual Layers, 60cm X 60cm (24" X 24"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 28 | SW767650-B/B | CareGUARD* Dual Layers, 76cm X 76cm (30” X 30”), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 29 | SW767660-B/B | CareGUARD* Dual Layers, 76cm X 76cm (30” X 30”), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 30 | SW767675-B/B | CareGUARD* Dual Layers, 76cm X 76cm (30” X 30”), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 31 | SW919150-B/B | CareGUARD* Dual Layers, 91cm X 91cm (36" X 36"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 32 | SW919160-B/B | CareGUARD* Dual Layers, 91cm X 91cm (36" X 36"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 33 | SW919175-B/B | CareGUARD* Dual Layers, 91cm X 91cm (36" X 36"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 34 | SW10010050-B/B | CareGUARD* Dual Layers, 100cm X 100cm (40" X 40"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 35 | SW10010060-B/B | CareGUARD* Dual Layers, 100cm X 100cm (40" X 40"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 36 | SW10010075-B/B | CareGUARD* Dual Layers, 100cm X 100cm (40" X 40"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 37 | SW11611650-B/B | CareGUARD* Dual Layers, 116cm X 116cm (45" X 45"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 38 | SW11611660-B/B | CareGUARD* Dual Layers, 116cm X 116cm (45" X 45"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 39 | SW11611675-B/B | CareGUARD* Dual Layers, 116cm X 116cm (45" X 45"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 40 | SW12012050-B/B | CareGUARD* Dual Layers, 120cm X 120cm (48" X 48"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 41 | SW12012060-B/B | CareGUARD* Dual Layers, 120cm X 120cm (48" X 48"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 42 | SW12012075-B/B | CareGUARD* Dual Layers, 120cm X 120cm (48" X 48"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 43 | SW13713750-B/B | CareGUARD* Dual Layers, 137cm X 137cm (54" X 54"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 44 | SW13713760-B/B | CareGUARD* Dual Layers, 137cm X 137cm (54" X 54"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 45 | SW13713775-B/B | CareGUARD* Dual Layers, 137cm X 137cm (54" X 54"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 46 | SW13718250-B/B | CareGUARD* Dual Layers, 137cm X 182cm (54" X 72"), 50GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 47 | SW13718260-B/B | CareGUARD* Dual Layers, 137cm X 182cm (54" X 72"), 60GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 48 | SW13718275-B/B | CareGUARD* Dual Layers, 137cm X 182cm (54" X 72"), 75GSM, Blue/Blue<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 49 | SW606050-B/W | CareGUARD* Contrast, 60cm X 60cm (24" X 24"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 50 | SW606060-B/W | CareGUARD* Contrast, 60cm X 60cm (24" X 24"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 51 | SW606075-B/W | CareGUARD* Contrast, 60cm X 60cm (24" X 24"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 52 | SW767650-B/W | CareGUARD* Contrast, 76cm X 76cm (30" X 30"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 53 | SW767660-B/W | CareGUARD* Contrast, 76cm X 76cm (30" X 30"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 54 | SW767675-B/W | CareGUARD* Contrast, 76cm X 76cm (30" X 30"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 55 | SW919150-B/W | CareGUARD* Contrast, 91cm X 91cm (36" X 36"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 56 | SW919160-B/W | CareGUARD* Contrast, 91cm X 91cm (36" X 36"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 57 | SW919175-B/W | CareGUARD* Contrast, 91cm X 91cm (36" X 36"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 58 | SW10010050-B/W | CareGUARD* Contrast, 100cm X 100cm (40" X 40"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 59 | SW10010060-B/W | CareGUARD* Contrast, 100cm X 100cm (40" X 40"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 60 | SW10010075-B/W | CareGUARD* Contrast, 100cm X 100cm (40" X 40”), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 61 | SW11611650-B/W | CareGUARD* Contrast, 116cm X 116cm (45" X 45"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 62 | SW11611660-B/W | CareGUARD* Contrast, 116cm X 116cm (45" X 45"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 63 | SW11611675-B/W | CareGUARD* Contrast, 116cm × 116cm (45" X 45"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 64 | SW12012050-B/W | CareGUARD* Contrast, 120cm X 120cm (48" X 48”), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 65 | SW12012060-B/W | CareGUARD* Contrast, 120cm X 120cm (48" X 48”), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 66 | SW12012075-B/W | CareGUARD* Contrast, 120cm X 120cm (48" X 48"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 67 | SW13713750-B/W | CareGUARD* Contrast, 137cm X 137cm (54" X 54"), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 68 | SW13713760-B/W | CareGUARD* Contrast, 137cm X 137cm (54" X 54"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 69 | SW13713775-B/W | CareGUARD* Contrast, 137cm X 137cm (54" X 54"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
| 70 | SW13718250-B/W | CareGUARD* Contrast, 137cm X 182cm (54" X 72”), 50GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 100 GSM |
| 71 | SW13718260-B/W | CareGUARD* Contrast, 137cm X 182cm (54" X 72"), 60GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 120 GSM |
| 72 | SW13718275-B/W | CareGUARD* Contrast, 137cm X 182cm (54" X 72"), 75GSM, Blue/White<br>Bonded, Two Layers, Total GSM: 150 GSM |
## TABLE 1: DIMENSIONS, TYPES AND BASIS WEIGHTS
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for Care Essentials. The logo features a stylized letter 'C' in a golden color, with a smaller 'e' nested inside it. Below the graphic, the words 'CARE ESSENTIALS' are printed in a simple, sans-serif font.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Care Essentials. The logo features a stylized letter "C" in blue and yellow, with the words "CARE ESSENTIALS" written in blue below it. The letter "C" is made up of two curved lines that overlap each other. The logo is simple and modern, and it conveys a sense of care and professionalism.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Care Essentials. The logo features a stylized letter 'e' in gold and blue, with the word 'CARE' in blue text above the word 'ESSENTIALS' in a smaller font. The logo is simple and modern, and the colors are bright and inviting.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Care Essentials. The logo features a stylized letter 'e' in blue and yellow, with the word 'CARE' in blue capital letters above the word 'ESSENTIALS' in smaller blue capital letters. The logo is simple and modern, with a focus on the company's name.
#### Indications for Use
CareGUARD* Sterilization Wrap is a single-use, non-woven sterilization wrap intended to be sterilized by a health care provider using pre-vacuum steam at 273°F / 134°C for 4 minutes. The CareGUARD* Sterilization Wrap was validated for a dry time of 30 minutes. The CareGUARD* Sterilization wrap is recommended to be used with trays weighing up to 13.7 lbs.
The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used.
Test results validated that CareGUARD* Sterilization wraps allowed sterilization of the enclosed devices by pre-vacuum cycles.
#### TABLE 2: WRAP BASIS WEIGHT RECOMMENDATIONS FOR PRE-VACUUM STEAM
| Basis Weight | Maximum Wrapped Package Content Weight |
|--------------|----------------------------------------|
| 50 GSM | |
| 60 GSM | 6.2 kg / 13.7 lbs |
| 75 GSM | |
The following load was used in the Pre-vacuum steam Sterility Maintenance Study:
- 50 GSM: 21 in. x 10 in. x 4 ¾ in. (54 cm x 25 cm x 12 cm) tray containing 13.7 Ibs . (6.2 kq) of metal mass
The metal mass included surgical instruments used for orthopedic joint replacement.
## Technological Characteristics Comparison
## TABLE 3: COMPARISON OF PROPOSE AND PREDICATE DEVICES
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Manufacturer | Care Essentials Pty Ltd | Halyard Health (Formerly<br>Kimberly Clark Health Care) | Different |
| 510(k) Reference | K213907 | K082554 | Different |
| Device Models | 50 gsm<br>60 gsm<br>75 gsm | H300<br>H400<br>H500 | Similar |
| Common name | Sterilization Wrap | Sterilization Wrap | Same |
| Regulation Number | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| Class | II | II | Same |
| Product Code | FRG | FRG | Same |
| Intended Use | To enclose another<br>medical device that is to | To enclose another medical<br>device that is to be sterilized<br>by a healthcare provider. | Same |
| | be sterilized by a<br>healthcare provider. | | |
| Indications for Use | CareGUARD* Sterilization<br>Wrap is a single-use, non-<br>woven sterilization wrap<br>intended to be sterilized by<br>a health care provider<br>using pre-vacuum steam<br>at 273°F / 134°C for 4<br>minutes. The<br>CareGUARD* Sterilization<br>Wrap was validated for a<br>dry time of 30 minutes.<br>The CareGUARD*<br>Sterilization wrap is<br>recommended to be used<br>with trays weighing up to<br>13.7 lbs.<br>The wrap is intended to<br>allow sterilization of the<br>enclosed medical<br>device(s) and maintain<br>sterility of the enclosed<br>device(s) until used.<br>Test results validated that<br>CareGUARD* Sterilization<br>wraps allowed sterilization<br>of the enclosed devices by<br>pre-vacuum cycles. | KimGUARD* Sterilization<br>Wrap is intended to be used<br>to enclose another medical<br>device that is to be sterilized<br>by a healthcare provider<br>using:<br>• Pre-vacuum Steam<br>270°F/132°C for 4 minutes<br>• Ethylene oxide (100%)<br>Concentration: 725 - 735<br>mg/L at 131°F/55°C and<br>40% - 80% relative<br>humidity for 60 minutes<br>The wrap is intended to<br>allow sterilization of the<br>enclosed medical device(s)<br>and maintain sterility of the<br>enclosed device(s) until<br>opened.<br>The wrap was validated for<br>aeration times for<br>• EtO sterilization<br>8 hours at 55°C or 12<br>hours at 43.3°C.<br>• Pre-vacuum steam<br>sterilization<br>20 minutes - Models 100,<br>200 and 300<br>30 minutes – Models 400,<br>500, 600 | Similar |
| Device Design | Two sheets of nonwoven<br>polypropylene fabric. Each<br>sheet is composed of<br>three thermally-bonded<br>layers consisting of a<br>meltblown polypropylene<br>layer surrounded by<br>spunbond polypropylene<br>layers (SMS). | Two sheets of nonwoven<br>polypropylene fabric. Each<br>sheet is composed of three<br>thermally-bonded layers<br>consisting of a meltblown<br>polypropylene layer<br>surrounded by spunbond<br>polypropylene layers (SMS). | Same |
| Wrapping Technique | Sequential/Simultaneous<br>double wrapping | Sequential/Simultaneous<br>double wrapping | Same |
| Method for bonding SMS<br>layers | Thermal bonding with oval<br>shape pin in angular<br>pattern. | Thermal bonding with round<br>pin, hexagonal, triangle<br>bond pattern ("daisy"<br>pattern). | Similar |
| Method for bonding<br>between sterilization<br>wraps | If applicable, ultrasonically<br>seamed along two edges. | If applicable, ultrasonically<br>seamed along two edges. | |
| Material Composition | Polypropylene with blue<br>and white pigments. | Polypropylene with blue and<br>white pigments. | Same |
| Single Use Device | Yes | Yes | Same |
| Shelf Life | 2 years from date of<br>manufacture | NIL | Similar |
| Biocompatibility | Applicable parts of ISO<br>10993 — Biological<br>evaluation of medical<br>devices (see Table 4) | Applicable parts of ISO<br>10993 — Biological<br>evaluation of medical<br>devices | Same |
| Maintenance of Package<br>Sterility | Real-time testing following<br>sterilization using pre-<br>vacuum steam supports<br>maintenance of package<br>sterility for 30 days. | Real-time testing following<br>sterilization using pre-<br>vacuum steam or Ethylene<br>Oxide supports<br>maintenance of package<br>sterility for 30 days. | Similar |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Care Essentials. The logo features a stylized letter "e" in blue and gold, with the words "CARE ESSENTIALS" written in blue below it. The letter "e" is designed with a flowing, curved shape, and the text is in a simple, sans-serif font.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo for a company called "Care Essentials". The logo features two overlapping letter "C" shapes, with the inner "C" being gold and the outer "C" being blue. The words "CARE ESSENTIALS" are written in blue below the "C" shapes.
# Summary of Non-clinical Testing
# TABLE 4: SUMMARY OF NON-CLINICAL TESTING
| Test | Relevant Standard(s) | Acceptance Criteria | Results |
|------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Physical Properties | ISO 11607-1<br>EN 868-2 | Comply with the physical<br>property requirements of ISO<br>11607-1 and EN 868-2. | The processed and<br>unprocessed<br>CareGUARD*<br>Sterilization Wrap<br>met the<br>acceptance criteria<br>and demonstrated<br>passing results. |
| Chemical Properties | ISO 11607-1<br>EN 868-2 | Comply with the chemical<br>property requirements of ISO<br>11607-1 and EN 868-2. | The chemical<br>properties testing<br>met the<br>acceptance criteria<br>and demonstrated<br>passing results. |
| Biocompatibility | ISO 10093-5<br>ISO 10993-10 | Determine whether the use of<br>the medical device is cytotoxic,<br>skin irritating or skin-<br>sensitizing. | Based on the<br>study, the medical<br>device was found<br>to be non-cytotoxic,<br>non-skin irritating<br>and non-skin<br>sensitizing. |
| Bacterial Filtration<br>Efficiency (BFE) | ASTM F2101-14 | Determining the bacterial<br>filtration efficiency of the<br>sterilisation wrap | Pass |
| Pre-vacuum Steam | | | |
| Sterilization<br>Validation | AS/NZS 4187:2014<br>ANSI/AAMI/ST8:2013<br>(R2018)<br>ANSI/AAMI ST79:2017 | A pre-vacuum steam<br>sterilization method was<br>validated to a sterility<br>assurance level (SAL) of 10-6 | Pass |
| Sterilant Penetration | AS/NZS 4187:2014<br>ANSI/AAMI/ST8:2013<br>(R2018)<br>ANSI/AAMI ST79:2017 | The testing details a method<br>used in determining the<br>internal temperature profile for<br>wrapped sterilization packs<br>when processed in a steam<br>sterilization pre-vacuum cycle<br>at 273°F/134°C for 4 minutes. | Testing has<br>demonstrated<br>adequate sterilant<br>penetration |
| Validation - Dry Time | AS/NZS 4187:2014<br>ANSI/AAMI/ST8:2013<br>(R2018)<br>ANSI/AAMI ST79:2017<br>EN 285:2015+A1:2021 | Determination of the proper<br>drying time required. | Test samples meet<br>or exceed the<br>minimum criteria<br>for dry time |
| Maintenance of<br>Sterility | ISO 11607-1:2019<br>ANSI/AAMI ST79:2017<br>AS/NZS 4187:2014<br>ANSI/AAMI/ST8:2013<br>EN 285:2015+A1:2021 | Determination of the<br>maintenance of sterility for<br>the wrapped loads. | Real-time testing<br>following<br>sterilization using<br>pre- vacuum<br>steam supports<br>maintenance of<br>package sterility<br>for 30 days. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for "CARE ESSENTIALS". The logo features a stylized letter "e" in a golden color, with a curved design. Below the "e", the words "CARE ESSENTIALS" are written in a blue sans-serif font, with "CARE" on top of "ESSENTIALS".
# Summary of Clinical Testing
Clinical testing is not applicable to the CareGUARD* Sterilization Wrap.
## Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the CareGUARD* Sterilization Wrap (K213907), is as safe, as effective and performs as well as or better than the legally marketed device identified [Kimberly-Clark* Corporation KIMGUARD* Sterilization Wrap (K082554)].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.