# Cyclops™ Anterior Cervical Plate System (K213901)

_Degen Medical · KWQ · Jan 7, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213901

## Device Facts

- **Applicant:** Degen Medical
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Jan 7, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DeGen Medical Cyclop™ Anterior Cervical Plate System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation. The Cyclops™ Anterior Cervical Plate system is intended for anterior intervertebral screw fixation of the cervical spine at levels C2 to T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) Spinal Stenosis Trauma (i.e. fracture or dislocation) Deformity or curvatures (including scoliosis, kyphosis, or lordosis) Spinal Tumors Pseudoarthrosis or failed previous fusion Spondylolisthesis Decompression of the spinal following total or partial cervical vertebrectomy

## Device Story

Cyclops Anterior Cervical Plate System provides temporary anterior spinal stabilization during fusion; consists of cervical plates and bone screws. Implants allow fixed or variable angle constructs; screws available in self-tapping or self-drilling options. Used by surgeons in clinical settings for cervical spine fixation (C2-T1). System provides mechanical support to facilitate bone fusion; aids in correcting spinal alignment and stabilizing segments post-decompression or trauma. Submission adds two-level plate lengths to existing system.

## Clinical Evidence

No clinical data. Bench testing and risk review of existing verification/validation data performed.

## Technological Characteristics

Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Fixed or variable angle construct. Self-tapping or self-drilling bone screws. Non-sterile. Mechanical fixation device.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Cyclops™ Anterior Cervical Plate System ([K191786](/device/K191786.md))
- DeGen Medical Hyper-C Anterior Cervical Plate System ([K150759](/device/K150759.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

January 7, 2022

DeGen Medical % Linda Braddon CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock, Georgia 30188

Re: K213901

Trade/Device Name: CyclopsTM Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 13, 2021 Received: December 14, 2021

Dear Linda Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K213901

Device Name Cyclops™ Anterior Cervical Plate System

#### Indications for Use (Describe)

The DeGen Medical Cyclop™ Anterior Cervical Plate System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation. The Cyclops™ Anterior Cervical Plate system is intended for anterior intervertebral screw fixation of the cervical spine at levels C2 to T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with following indications:

- Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Spinal Stenosis
- · Trauma (i.e. fracture or dislocation)
- · Deformity or curvatures (including scoliosis, kyphosis, or lordosis)
- Spinal Tumors
- Pseudoarthrosis or failed previous fusion
- · Spondylolisthesis
- · Decompression of the spinal following total or partial cervical vertebrectomy

| Type of Use (Select one or both, as applicable) |                                               |
|-------------------------------------------------|-----------------------------------------------|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D)  | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DeGen Medical. The logo features three curved lines emanating from a point, each ending in a blue circle. The lines are also blue and appear to be thicker in the middle. Below the graphic is the text "DeGen" in a larger, bolder blue font, with the word "Medical" underneath in a smaller, gray font.

# 510(k) SUMMARY: Cyclops™ Anterior Cervical Plate System

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

| Date Prepared        | December 13, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor              | DeGen Medical, Inc.<br>1321-C North Cashua Drive<br>Florence, SC 29501<br>Phone 877-240-7838<br>Fax 843-407-0545                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 510(k) Contact       | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681<br>Regulatory@SecureBME.com                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name           | Cyclops™ Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name          | Spinal intervertebral body fixation orthosis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code         | KWQ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification       | 21 CFR 888.3060, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Predicate    | K191786 Cyclops™ Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Additional Predicate | K150759 DeGen Medical Hyper-C Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description   | The DeGen Medical Cyclops™ Anterior Cervical Plate Systems (CACP) consists of cervical plates and bone screws. All implants are intended to provide stabilization of the cervical vertebrae. The CACP provides anterior fixation from either fixed or variable angle construct in self-tapping or self-drilling bone screw options. The system is provided non-sterile and is constructed from Titanium alloy (Ti-6Al-4V ELI) per ASTM F136.<br><br>The purpose of this submission is to add additional two-level plate lengths to the Cyclops™ Anterior Cervical Plate System |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for DeGen Medical. The logo features three curved lines emanating from three blue circles at the top, resembling a stylized representation of motion or technology. Below the graphic is the text "DeGen" in a bold, blue font, with the word "Medical" in a smaller, gray font underneath.

| Indications for Use<br>Statement                      | The DeGen Medical Cyclops™ Anterior Cervical Plate system is intended for<br>use as an aid in cervical spinal fusion and is intended for unilateral fixation.<br>The Cyclops™ Anterior Cervical Plate system is intended for anterior<br>intervertebral screw fixation of the cervical spine at levels C2 to T1. The<br>system is indicated for temporary stabilization of the anterior spine during the<br>development of cervical spine fusions in patients with following indications:<br>Degenerative Disc Disease (as defined by neck pain of discogenic<br>origin with degeneration of the disc confirmed by patient history and<br>radiographic studies) Spinal Stenosis Trauma (i.e., fracture or dislocation) Deformity or curvatures (including scoliosis, kyphosis, or lordosis) Spinal Tumors Pseudoarthrosis or failed previous fusion Spondylolisthesis Decompression of the spinal following total or partial cervical<br>vertebrectomy |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics                      | As was established in this submission, the subject Cyclops™ Anterior<br>Cervical Plate System is substantially equivalent to other predicate devices<br>cleared by the FDA for commercial distribution in the United States. The<br>subject device was shown to be substantially equivalent and has the same<br>technological characteristics to its predicate devices through comparison in<br>areas including design, intended use, material composition, function, and<br>range of sizes.                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Non-Clinical<br>Performance<br>Testing Conclusion     | A risk review was conducted on existing verification and validation data for<br>the Cyclops™ Anterior Cervical Plate System regarding the additional two-level<br>plates. The results of this risk review indicated that there would be no impacts<br>to safety or effectiveness of the system by including this device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Substantial<br>Equivalence<br>Summary<br>(Conclusion) | Based on the indications for use, technological characteristics, performance<br>testing, and comparison to predicate devices, the subject DeGen Medical<br>Cyclops™ Anterior Cervical Plate System (CACP) has been shown to be<br>substantially equivalent to legally marketed predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

---

**Source:** [https://fda-staging.innolitics.com/device/K213901](https://fda-staging.innolitics.com/device/K213901)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).