K213894 · Dwfritz Automation, Inc. · FXX · Mar 14, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213894
Device Name
DWFritz ASM2000
Applicant
Dwfritz Automation, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Mar 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
When properly worn, DWFritz ASM2000 surgical face mask is intended to protect both the patient and healthcare worker from transfer of microorganisms, body fluids and particulate matter. The device is non-sterile and for single use only.
Device Story
DWFritz ASM2000 is a 3-ply, flat-pleated surgical face mask; designed for single-use, non-sterile application. Construction includes outer and inner layers of spunbond polypropylene and a middle layer of meltblown polypropylene filter. Features include polyester/spandex ear loops and a galvanized wire nose bridge with polyester coating. Device is worn over the face and mouth to provide a physical barrier against microorganisms, body fluids, and particulate matter. Used in clinical or general settings by healthcare workers and patients; provides protection for both parties by reducing transfer of contaminants. No software or electronic components.
Clinical Evidence
Bench testing only. Performance evaluated against ASTM F2100-2019 Level 2 standards. Results: Bacterial Filtration Efficiency (BFE) 99.9%; Particulate Filtration Efficiency (PFE) 98.8-99.7%; Fluid Resistance 94/96 passed at 120 mmHg; Differential Pressure 2.9-4.7 mmH2O/cm2; Flammability Class 1. Biocompatibility testing (ISO 10993-05, 10993-10) confirmed the device is not cytotoxic, not skin sensitive, and not irritating.
Technological Characteristics
3-ply flat-pleated mask. Materials: Spunbond polypropylene (inner/outer layers), meltblown polypropylene (filter layer), polyester/spandex (ear loops), galvanized wire with polyester coating (nose bridge). Dimensions: 9.8 x 17.2 ± 0.5 cm. Performance: ASTM F2100 Level 2. Biocompatibility: ISO 10993-05 (cytotoxicity), ISO 10993-10 (sensitization/irritation). Flammability: 16 CFR 1610 Class 1. Non-sterile, single-use, latex-free.
Indications for Use
Indicated for use by patients and healthcare workers to protect against transfer of microorganisms, body fluids, and particulate matter. Intended for OTC use.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 14, 2022
DWFritz Automation, Inc. % John Gillespy President FDA 510k Consultants, LLC 1100 Del Lago Circle Palm Beach Gardens, Florida 33410
Re: K213894
Trade/Device Name: DWFritz ASM2000 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 10, 2021 Received: December 14, 2021
Dear John Gillespy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bifeng Qian Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213894
Device Name DWFritz ASM2000
Indications for Use (Describe)
When properly worn, DWFritz ASM2000 surgical face mask is intended to protect both the patient and healthcare worker from transfer of microorganisms, body fluids and particulate matter. The device is non-sterile and for single use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary_K213894
Prepared in accordance with the requirements of 21 CFR Part 807.92.
| 1. 510(k) Submitter: | DWFritz Automation, Inc.<br>9600 SW Boeckman Rd., Wilsonville OR 97070<br>Phone: 503-885-7030<br>Email: cpsmith@dwfritz.com | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 2. Company Contact: | Curtis Smith, Senior Quality Manager | |
| 3. Date of Submission: | December 10, 2021 | |
| 4. 510(k) Preparer: | John F. Gillespy, MBA<br>FDA 510k Consulting, LLC<br>Palm Beach Gardens, FL 33410<br>Phone: 386-243-4332<br>Email: john@fda510kconsultants.com | |
| 5. Device Classification: | Trade name:<br>Common name:<br>Device:<br>Class:<br>Regulation #:<br>Product Code: | DWFritz ASM2000<br>Surgical Face Mask<br>Surgical Mask<br>II<br>878.4040<br>FXX |
| 6. Predicate: | Applicant:<br>Device:<br>510(k) Number: | Mexpo International, Inc.<br>Avianz Surgical Face Mask<br>K200847 |
#### 7. Device Description
DWFritz ASM2000 is a 3-ply, flat-pleated style surgical face mask. The device is manufactured with three layers. The inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. Each mask contains ear loops made of polyester filament and spandex fiber to secure the mask over the user's face and mouth with galvanized wire nose bridge to fit firmly over the nose.
This device is not made from any natural rubber latex.
The product is sold non-sterile and intended to be disposable for single-use.
{4}------------------------------------------------
### 8. Indications For Use
When properly worn, DWFritz ASM2000 surgical face mask is intended to protect both the patient and healthcare worker from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.
The device is intended for OTC use.
#### 9. Comparison To Predicate Device
DWFritz ASM2000 is compared to a predicate device, Avianz Surgical Face Mask (K200847).
| Device | Proposed Device | Predicate Device | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | DWFritz Automation, Inc. | Mexpo International, Inc. | - |
| 510K Number | K213894 | K200847 | - |
| Product Name | Surgical Face Mask | Surgical Face Mask | Same |
| Product Code | FXX | FXX | Same |
| Classification | Class II (21 CFR 878.4040) | Class II (21 CFR 878.4040) | Same |
| Intended Use | When properly worn,<br>DWFritz ASM2000 surgical<br>face mask is intended to<br>protect both the patient<br>and healthcare worker from<br>transfer of micro-<br>organisms, body fluids and<br>particulate matter. This<br>device is non-sterile and for<br>single use only. | When properly worn, the<br>surgical face masks are<br>intended to protect both<br>patient and healthcare<br>workers from transfer of<br>microorganisms, body fluids<br>and airborne particles. This<br>device is non-sterile and for<br>single use only. | Same |
| Model | 3-ply, flat-pleated style with<br>ear loops and nose bridge | 3-Ply, Flat-Pleated Style with<br>ear loops and nose bridge | Same |
| Materials, Dimensions, & Other Features | | | |
| Outer Facing<br>Layer | Spunbond Polypropylene | Spunbond Polypropylene | Same |
| Middle Layer | Meltblown Polypropylene | Meltblown Polypropylene | Same |
| Inner Facing Layer | Spunbond Polypropylene | Spunbond Polypropylene | Same |
| Nose Bridge | Galvanized wire with<br>polyester coating | Galvanized wire with<br>polyester coating | Same |
## Comparison Table
{5}------------------------------------------------
| Ear Loops | Not made with natural rubber<br>latex | Not made with natural rubber<br>latex | Same |
|-----------------------|---------------------------------------|---------------------------------------|-----------|
| Color | Blue (inner facing) & white | White | Similar |
| Dimension (W X L) | 9.8 X 17.2 ± 0.5cm | 9.5 X 17.5 ± 0.5cm | Similar |
| Use | OTC, single use, disposable | OTC, single use, disposable | Same |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Testing | | | |
| Performance Level: | | | |
| ASTM F2100:2019 | Level 2 | Level 2 | Same |
| Bacterial Filtration | 99.9% | 98% | Both Pass |
| Particle Filtration | 98.4% | 98% | Both Pass |
| Fluid Resistance | 94 of 96 pass @ 120 mmHg | 29 of 32 pass @ 120 mmHg | Both Pass |
| Differential Pressure | < 4.7 mmH2O/cm2 | < 5.0 mmH2O/cm2 | Both Pass |
| Flammability | Class 1 | Class 1 | Same |
| Biocompatibility: | | | |
| Cytotoxicity | Not Cytotoxic | Not Cytotoxic | Same |
| Sensitization | Not Skin Sensitive | Not Skin Sensitive | Same |
| Irritation | Not Skin Irritating | Not Skin Irritating | Same |
DWFritz ASM2000 is comparable to the predicate device (K200847) in terms of materials, performance, and biocompatibility.
## 10. Summary Of Non-Clinical Testing
The following performance data has been provided to demonstrate that DWFritz ASM2000 meets the acceptance criteria in the standard.
### Non-Clinical Testing Table
| Test Standard | Purpose | Acceptance Criteria | Results |
|---------------|--------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------|
| ISO 10993-05 | Cytotoxicity | < Grade 2<br>cytotoxic effect | Grade 1<br>(slight reactivity) |
| ISO 10993-10 | Skin Sensitization | No sensitization<br>reaction | No sensitization<br>reaction |
| ISO 10993-10 | Skin Irritation | No irritation reaction | No irritation reaction |
| ASTM F2101 | Bacterial Filtration<br>Efficiency (BFE) | 98% | 99.9% |
| ASTM F2299 | Particulate Filtration<br>Efficiency (PFE) | 98% | 98.8-99.7% in 32 test<br>articles in each of<br>three lots (96 total<br>test articles) |
{6}------------------------------------------------
| ASTM F1862 | Synthetic Blood Fluid<br>Resistance | 29 out of 32 pass at<br>120 mmHg | 30, 31, and 32 pass at<br>120 mmHg, out of 32<br>test articles in each of<br>three lots (96 total<br>test articles) |
|--------------|-------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| MIL-M-36954C | Differential Pressure<br>(Delta P) | < 5.0 mmH20/cm2 | 2.9 - 4.7 mmH20/cm2<br>in 32 test articles in<br>each of three lots (96<br>total test articles) |
| 16 CFR 1610 | Flammability | IBE = Test article<br>ignited but exting-<br>uished, burn time ><br>3.5 sec (Class 1) | IBE for all 14 test<br>articles in each of 3<br>lots (42 total test<br>articles) |
### 11. Patient-Contacting Materials
Materials composition follows:
- Outer facing layer – spunbond polypropylene (blue)
- Middle layer meltblown polypropylene filter (white) ●
- Inner facing layer spunbond polypropylene (white) ●
- Nose bridge – Galvanized wire with polyester coating
- Ear loops – polyester filament & spandex fiber, not made from natural rubber latex (white)
The outer & inner layers and ear loops touch the patient's face and hands. Type and duration of contact with patient follows:
- Surface contact — intact skin
- Limited duration - less than 24 hours
## 12. Software Verification and Validation... The mask contains no software or firmware.
The conclusion drawn from the nonclinical tests demonstrates that the subject device, DWFritz ASM2000, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K200847.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.