K213876 · Spineology, Inc. · OLO · Aug 5, 2022 · Neurology
Device Facts
Record ID
K213876
Device Name
Spineology Navigation Instruments
Applicant
Spineology, Inc.
Product Code
OLO · Neurology
Decision Date
Aug 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Spineology Navigation Instruments are indicated for use during the preparation and placement of. Spineology's Fortress™, Threshold™, Threshold™ V2, and Palisade™ pedicle screws . Spineology's OptiMesh® Portal Assembly . during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous orthopedic procedures. These instruments are designed for use with the Medtronic System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy.
Device Story
Spineology Navigation Instruments are non-sterile, reusable, manually operated stainless steel surgical instruments. They include pedicular fixation tools (awls, taps, drills, screwdrivers) and portal placement tools (dilator, portal, tightener, wrench, surrogate). Instruments feature geometric interfaces for connection to Medtronic NavLock or SureTrak II Universal Trackers. Used in OR by surgeons during spinal procedures; instruments integrate with Medtronic StealthStation (S7/S8) optical navigation systems. Navigation allows real-time tracking of instrument position relative to patient anatomy on system displays, assisting precise anatomical localization for implant placement. Benefits include improved surgical accuracy and guidance during spinal fixation and portal access.
Clinical Evidence
Bench testing only. Verification testing included instrument mating, registration, and accuracy testing to confirm compatibility with the Medtronic StealthStation S8 Spine System. Testing confirmed that the instruments can be registered and optically navigated accurately under expected-use conditions in a simulated environment.
Technological Characteristics
Non-sterile, reusable, manually operated stainless steel instruments. Features dimensional interfaces for connection to Medtronic NavLock or SureTrak II Universal Trackers. Compatible with Medtronic StealthStation S7 and S8 optical navigation systems. No electronic components; passive tracking via external optical navigation system.
Indications for Use
Indicated for patients undergoing spinal surgery requiring preparation and placement of pedicle screws or OptiMesh Portal Assembly. Used in open, minimally invasive, or percutaneous orthopedic procedures where stereotactic surgery is appropriate and rigid anatomical structures (spine/pelvis) can be referenced to images.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
OptiMesh® Expandable Interbody Fusion System (DEN200010)
Submission Summary (Full Text)
{0}------------------------------------------------
August 5, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Spineology Inc. Andrew Adams Group Director of Regulatory & Ouality Affairs 7800 3rd Street North, Suite 600 Saint Paul, Minnesota 55128
Re: K213876
Trade/Device Name: Spineology Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: June 24, 2022 Received: June 27, 2022
Dear Andrew Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known)
K213876
## Device Name
Spineology Navigation Instruments
### Indications for Use (Describe)
Spineology Navigation Instruments are indicated for use during the preparation and placement of.
- Spineology's Fortress™, Threshold™, Threshold™ V2, and Palisade™ pedicle screws .
- Spineology's OptiMesh® Portal Assembly .
during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous orthopedic procedures.
These instruments are designed for use with the Medtronic System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Spineology.
| Date Prepared: | August 5, 2022 |
|-----------------|-------------------------------------------------------------------------|
| Submitter: | Spineology Inc. |
| | 7800 3rd Street North |
| | Suite 600 |
| | Saint Paul, MN 55128 |
| | Establishment Registration Number: 2135156 |
| Contact Person: | Andrew Adams |
| | Group Director of Regulatory & Quality Affairs |
| | Phone: 651.256.8500 |
| | Fax: 651.256.8505 |
| | Email: <a href="mailto:aadams@spineology.com">aadams@spineology.com</a> |
# Device Name and Classification
| Predicate Device | | |
|------------------|-----------|----------------------------------------------------|
| Primary: | K182345 | Spineology Navigation Instruments |
| Reference: | K170011 | Medtronic StealthStation™ S8 Spine Software v1.0.0 |
| Reference: | DEN200010 | OptiMesh® Expandable Interbody Fusion System |
#### 1. Purpose
The purpose of this premarket notification is to obtain FDA clearance for the introduction of new instruments to the Spineology Navigation Instruments family. These subject instruments are for Portal Placement navigation, specifically Spineology's OptiMesh Portal Assembly.
#### 2. Device Description
Spineology Navigation Instruments are non-sterile, reusable surgical instruments that are manufactured from stainless steel and operated manually.
Spineology Navigation Instruments for Pedicular Fixation include Awls, Bone Taps, Drills, and Screwdrivers. The navigable instruments are equipped with a dimensional feature that allows
{4}------------------------------------------------
connection to Medtronic's NavLock Trackers. A Navigation Adapter is also available with dimensional features that allows the same instruments to Medtronic's SureTrak" II Universal Trackers. These instruments are compatible with Spineology's Fortress", Threshold" Threshold™ V2, and Palisade™ Pedicular Fixation Systems and Medtronic's StealthStation® S7 (v2.1.0) and S8 (v1.0.0) Spine System.
Spineology Navigation Instruments for Portal Placement include a Dilator, Portal, Tightener, Wrench, and a Portal Surrogate. The navigable instruments are equipped with dimensional features that allow connection to Medtronic's SureTrak II Universal Trackers. These instruments are compatible with Spineology's OptiMesh Expandable Interbody Fusion System for Access Portal Placement and Medtronic's StealthStation® S8 (v1.2.0) Spine System.
Spineology Navigation Instruments are intended to be used during the preparation and placement of spinal implants in optically navigated procedures.
#### 3. Indications for Use
Spineology Navigation Instruments are indicated for use during the preparation and placement of:
- Spineology's Fortress™, Threshold™, Threshold™ V2, and Palisade™ pedicle screws ●
- . Spineology's OptiMesh® Portal Assembly
during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous orthopedic procedures.
These instruments are designed for use with the Medtronic StealthStation® Spine System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy.
#### Technological Characteristics 4.
When compared to the predicate device, Spineology Navigation Instruments for Portal Placement have the same intended use and the same technological characteristics, including:
●
- Primary Design Features
- Materials of Construction ●
- Function / Performance
- Risk Profile ●
Principle of Operation
- Use with Spineology's Instrument Systems ●
- Use with Medtronic's StealthStation System .
- Fundamental Scientific Technology ●
{5}------------------------------------------------
#### 5. Non-Clinical Testing
Design verification testing, including mating, registration, and accuracy, was conducted to ensure that Spineology Navigation Instruments for Portal Placement are safe and effective for their intended use, to ensure functionality and compatibility with the Medtronic StealthStation S8 Spine System, and to demonstrate substantial equivalence to the predicate device.
- Instrument mating testing was performed to ensure that Spineology Navigation Instruments . for Portal Placement can be mated with the Medtronic SureTrak II Universal Tracker to allow for instrument registration and optical navigation. The ability to withstand expecteduse conditions was also confirmed.
- Registration testing was performed to ensure that Spineology Navigation Instruments for ● Portal Placement can be registered with the Medtronic StealthStation S8 Spine System to allow for optical navigation.
- . Accuracy testing was performed to ensure that Spineology Navigation Instruments for Portal Placement can be optically navigated accurately, as compared to predicate instruments, under expected-use conditions in a simulated environment with the Medtronic StealthStation S8 Spine System.
#### 6. Conclusion
Based on the intended use, technological characteristics, and comparison to the predicate device, Spineology Navigation Instruments for Portal Placement have been shown to be substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.