K213846 · Quest Medical, Inc. · FPA · Jun 5, 2022 · General Hospital
Device Facts
Record ID
K213846
Device Name
Q2 Blood Administration Sets
Applicant
Quest Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Jun 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Blood Administration Sets are used to deliver blood, blood components, and IV fluids from a container to a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and the duration of therapy.
Device Story
Q2 Blood Administration Sets are single-use, disposable intravenous sets for delivering blood, blood components, and IV fluids from a container to a patient's vascular system. Operation occurs via gravity flow or manual hand pump activation. Components include tubing, bag spike, 200-micron blood filter drip chamber, luer connectors, clamps, stopcocks, and needleless connectors. Used in clinical settings by healthcare providers to manage fluid/blood resuscitation. The device facilitates controlled infusion; the hand pump allows for active fluid delivery. Benefits include reliable, filtered delivery of blood products and fluids, ensuring patient safety through standardized flow control and particulate filtration.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 1135-4 (transfusion equipment), ISO 80369-7 (small-bore connectors), USP 788 (particulate matter), mechanical hemolysis testing, and comprehensive biocompatibility testing (cytotoxicity, sensitization, irritation, hemocompatibility, systemic toxicity, genotoxicity). Sterility, shelf-life, and shipping validation performed per ISO 11135, ASTM F1980, and ASTM D4169.
Technological Characteristics
Single-use, disposable IV administration set. Materials: LDPE, ABS, DEHP-free PVC, polypropylene, polycarbonate, silicone. Features: 200-micron filter, hand pump, luer connectors, clamps. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: None. Standards: ISO 1135-4, ISO 80369-7, USP 788, ISO 10993-1, ASTM F1140, ASTM F1929, ASTM F1980, ASTM D4169.
Indications for Use
Indicated for delivery of blood, blood components, and IV fluids to patients of any age/gender, provided vascular anatomy and therapy duration are appropriate.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
IV Administration Sets with 200um Blood Filter (K143082)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
Quest Medical, Inc. Tosan Eweka Regulatory Affairs Manager One Allentown Parkway Allen, Texas 75002
Re: K213846
Trade/Device Name: 02 Blood Administration Sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: May 5, 2022 Received: May 6, 2022
# Dear Tosan Eweka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213846
Device Name Q2 Blood Administration Sets
#### Indications for Use (Describe)
The Blood Administration Sets are used to deliver blood, blood components, and IV fluids from a container to a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and the duration of therapy.
Type of Use (Select one or both, as applicable)
| <span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:120%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K213846-510K Summary
#### GENERAL INFORMATION
#### Date Prepared: June 6, 2022
#### Applicant:
Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 972-390-2881 Fax:
#### Contact Person:
Tosan Eweka Regulatory Affairs Manager Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 Fax: 972-390-2881 Email: teweka@questmedical.com
### DEVICE INFORMATION
Trade Name: Q2 Blood Administration Set
Generic/Common Name: Intravascular Administration Set
Classification: Class II Regulation Number: 21CFR§880.5440
FPA Product Code:
{4}------------------------------------------------
# PREDICATE DEVICE(S)
K143082, B. Braun Medical IV Administration Sets with 200um Blood Filter.
# DEVICE DESCRIPTION
Quest Medical's Q2 Blood Administration Sets are single use, disposable intravenous administration sets used to deliver blood, blood components and IV fluids from a container to a patient's vascular system through the use of a hand pump or through gravity flow. The sets include tubing, bag spike, 200 micron blood filter drip chamber, luer connectors, clamps, hand pump, stopcock and needleless connector.
### INDICATIONS FOR USE
| Characteristics | Predicate Device | Subject Device |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | IV Administration Sets with 200µm<br>Blood Filter<br>K143082 | Q2 Blood Administration Sets<br>K213846 |
| Indication for Use | The IV Administration Sets with 200µm Blood Filter are used to deliver blood, blood components, and IV fluids from a container to a patient's vascular system. When the hand pump component is activated, the device is intended to deliver blood, blood products and crystalloid and colloid resuscitative fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and the duration of therapy. | The Blood Administration Sets are used to deliver, blood components, and IV fluids from container to a patient's vascular system. The devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and the duration of therapy. |
| Prescription Only<br>or over the Counter | Prescription Only | Prescription Only |
# Discussions of differences in Indications for Use statement
The subject device indications does not include the sentence "When the hand pump component is activated, the device is intended to deliver blood, blood products and crystalloid and colloid resuscitative fluids" which is included in the indications for the predicate device. The sentence describes the fluids delivered when using the hand pump component. It specifies crystalloid and colloid resuscitative fluids. Quest does not intend to make any compatibility claims on these fluids, hence it has been deleted from the indications statement of the subject device. Deleting this sentence does not change the overall intended use of both devices, which is administration of blood, blood components, and IV fluids to a patient. The difference in indications statements does not create a new intended use.
{5}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The Q2 Blood Administration Sets has the same intended use, indications for use, and principles of operation as the predicate device, and has similar design features and technological characteristics as the predicate device. The table below outlines and differences between the subject device and the predicate device.
| Element of Comparison | Predicate Device | Subject Device | Analysis of |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| | IV Administration Sets with 200um Blood<br>Filter | Q2 Blood administration Set | Differences |
| | K143082 | K213846 | |
| Product Code | FPA | Same | Same |
| Regulation Number | 21 CFR 880.5440 | Same | Same |
| Regulation Name | Intravascular Administration Set | Same | Same |
| Class | II | Same | Same |
| Mode of Fluid<br>Delivery | Gravity or Hand Pump Activation | Same | Same |
| Hand Pump Design | Cylindrical shape with ball check valves | Same | Same |
| Components | Tubing, drip chamber with 200 micron filter, bag<br>spike, luer connector, roller clamp, slide clamp,<br>hand pump, stopcocks, manifolds and/or luer<br>access devices. | Tubing, drip chamber with 200 micron filter,<br>bag spike, luer connector, roller clamp, slide<br>clamp, pinch clamp, hand pump, stopcocks and<br>luer access device | Different<br>See Comment #1 |
| Element of<br>Comparison | Predicate Device - IV Administration<br>Sets with 200um Blood<br>Filter(K143082) | Subject Device | Analysis of Differences |
| Device Materials | Unknown | Vented Spike - LDPE Riblene FH20,<br>Polyethylene and ABS<br>Roller Clamp - ABS<br>Tubing – Alpha Gary 2235L-78 not made<br>with DEHP PVC<br>200 Micron Blood Filter Drip Chamber -<br>DEHP Free PVC, Purell HP Polypropylene<br>and PA<br>Hand Pump - not made with DEHP PVC<br>and Polypropylene<br>Slide Clamp - Purell HP Polypropylene<br>Pinch Clamp – Polypropylene, Flint Hills<br>P5M6K-080<br>Spin Lock Male Luer - Terlux 2802 HD<br>ABS<br>Stopcock - HDPE, RX1805451118<br>Polycarbonate, LR 3003/40 Wacker<br>Silicone, Blue Colorant K-75238<br>Female Luer - PVC AM22W17 Female<br>Luer Cap - HDPE Chevron HD 9012<br>Y-Site with needleless connector-<br>Polycarbonate and Silicone<br>Solvent Code L – Dichloromethane and<br>Cyclohexanone<br>Solvent Code R – Methyl-Ethyl Ketone<br>and Cyclohexanone | Different<br>See Comment #2 |
| Patient Contact<br>Category/Duration | Externally Communicating, Blood Path<br>Indirect, Prolonged contact | Same | Same |
| Sterilization | Ethylene Oxide | Same | Same |
| Tubing Length | Unknown | 114" — 129" | See comment# 3 |
| Priming Volume | Unknown | 74mL - 76mL | See comment# 3 |
| Needleless<br>Connector Residual<br>Volume | Unknown | 0.09mL | See comment# 3 |
{6}------------------------------------------------
{7}------------------------------------------------
### 510(k) Summary
# Discussions of differences in Technological Characteristics
# Comment # 1: Components
The predicate device and subject device are comprised of the same components with the exception of a manifold in some configurations of the predicate device, which is not included in the subject device and use of pinch clamp in the subject device. Configurations of the predicate device without the manifold are therefore equivalent to the subject device. The additional pinch clamp used in the subject device performs the same functions as the slide clamp. The results from performance testing conducted on the subject device demonstrates that the difference in components between the subject device and predicate device do not raise different questions of safety and effectiveness. The subject device met all performance specifications necessary to fulfil its intended use.
### Comment # 2: Materials
The materials used on the predicate device is unknown. Materials used in the subject device are broadly used in medical devices and have been used on IV Admin and Extension Sets currently marketed by Quest Medical. Results from biocompatibility testing conducted on the subject device demonstrate that differences (if any) in the materials used in the subject device and predicate device do not raise different questions of safety and effectiveness.
# Comment # 2: Tubing Length, Priming Volume and Residual Volume
The tubing length, priming volume and residual volume of the predicate device is unknown. Results from performance qualification testing conducted on the subject device demonstrate that these differences (if any) between the predicate device and the subject device do not raise different questions of safety and effectiveness as all acceptance criteria were met.
### PERFORMANCE DATA
All necessary bench and nonclinical testing was conducted on the Q2 Blood Administration Sets to support a determination of substantial equivalence to the predicate device.
# Nonclinical Testing Summary
# A. Performance Test- Bench
The bench and nonclinical tests conducted on the Q2 Blood Administration Sets in this submission are summarized below. The subject device met the applicable test specifications/acceptance criteria as described in the submission.
- ISO 1135-4:2015 Transfusion equipment for medical use-Part 4: Transfusion . sets for single use, gravity feed
- ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for Intravascular or hypodermic applications
- Particulate matter testing was conducted in accordance with USP <788>Particulate Matter in Injections and met the USP acceptance criteria.
- Mechanical Hemolysis Test
{8}------------------------------------------------
#### 510(k) Summary
### B. Biocompatibility
In accordance with ISO 10993-1, the Q2 Blood Administration Set, the proposed device were classified as: Externally Communicating Device with Direct Blood contact for Prolonged Contact Duration (>24hrs -30days.). The following Testing was conducted:
- Cytotoxicity
- Sensitization ●
- Irritation
- Hemocompatibility
- Acute Systemic Toxicity
- Sub-Chronic Toxicity
- . Genotoxicity
# C. Sterility, Shipping, and Shelf-Life
The subject device is intended to be sterilized with Ethylene Oxide procedures validated per ISO 11135:2014. The subject device met applicable acceptance criteria for EO residuals per ISO 10993-7:2008 and pyrogenicity (LAL method) per USP <161>. The following additional tests were conducted:
- . Sterile Barrier Packaging Testing (ASTM F1140 Burst Test, Peel Test and ASTM F1929 Dye Penetration Test).
- 3 year shelf-life test per ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- Simulated Shipping per ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and System.
Microbial ingress data for the needleless connector component used on the Q2 Blood Administration Sets was referenced to the previously cleared K002689.
### Clinical Testing Summary
Not applicable. Clinical testing was not performed to support these 510(k) submissions.
### CONCLUSIONS
The collective results of the performance testing demonstrate that the Q2 Blood Administration Sets meets the established specifications necessary for consistent performance during its intended use. In addition, the collective performance testing demonstrate that the Q2 Blood Administration Sets do not raise different questions of safety or effectiveness when compared to the predicate device. The results from the performance testing support the conclusion that the Q2 Blood Administration Sets is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.