K213840 · MolecuLight, Inc. · QJF · May 18, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213840
Device Name
MolecuLight I:X
Applicant
MolecuLight, Inc.
Product Code
QJF · General, Plastic Surgery
Decision Date
May 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4550
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213840 · May 18, 2022
MolecuLight I:X
MolecuLight, Inc.
Retrospective clinical patient data (post hoc analysis of 78 patients)
Retrospective analysis of clinical data was used to demonstrate the effectiveness of the MolecuLight i:X in guiding wound sampling to detect bacterial burden and pathogens of interest compared to standard-of-care sampling.
Post hoc retrospective analysis; Retrospective comparative analysis
78 patients with wounds; Sample Size: 78
Standard of Care (SoC) method of sampling (center of the wound)
Sensitivity to detect bacterial load ≥ 10^4 CFU/g; count of bacterial species detected; count of pathogens of interest detected
Indications for Use
The MolecuLight i:X is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to (i) View and digitally record images of a wound, (ii) Measure and digitally record the size of a wound, and (iii) View and digitally record images of fluorescence emitted from a wound when exposed to an excitation light. The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >10^4 CFU per gram) as compared to examination of clinical signs and symptoms alone. The MolecuLight i:X device should not be used to rule-out the presence of bacteria in a wound. The MolecuLight i:X does not diagnose or treat skin wounds.
Device Story
Handheld imaging device; uses 405 nm LED excitation light to induce autofluorescence in wounds; captures standard color and fluorescence images via integrated optical/microelectronic components; displays images on high-resolution LCD touch screen. Used by clinicians at point-of-care for wound assessment. Fluorescence imaging identifies regions with elevated bacterial loads (>10^4 CFU/g) and specific bacterial species (Gram+/-, aerobic/anaerobic). Clinicians use visual fluorescence cues in conjunction with clinical signs/symptoms to guide wound sampling (biopsy) to more representative areas, improving detection of pathogens compared to standard-of-care center-of-wound sampling. Benefits include enhanced identification of bacterial burden and pathogens of interest, potentially informing clinical management.
Clinical Evidence
Retrospective analysis of 78 patients comparing fluorescence-guided (FL-guided) biopsy to standard-of-care (SoC) biopsy. FL-guided biopsy demonstrated superior sensitivity (98.7%, 95% CI: 93.06%-99.97%) for detecting bacterial loads ≥10^4 CFU/g compared to SoC (87.2%, 95% CI: 77.7%-93.7%, p=0.012). FL-guided sampling also detected a higher mean number of bacterial species (3.03 vs 2.23, p<0.001) and pathogens of interest (1.73 vs 1.42, p=0.002).
Technological Characteristics
Handheld imaging device; 405 nm LED excitation; emission detection 500-545 nm and 600-665 nm. 5-megapixel resolution; 30 fps frame rate; 1136 x 640 pixel image size. Non-patient contacting (8-12 cm working distance). Battery/wall powered. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, and IEC 62471. Software-based image processing and measurement.
Indications for Use
Indicated for adult patients with skin wounds. Used by clinicians at the point of care to visualize, record, and measure wounds and to visualize wound fluorescence to assist in identifying bacterial loads >10^4 CFU/g. Not for ruling out bacterial presence or diagnosing/treating wounds.
Regulatory Classification
Identification
An autofluorescence detection device for general surgery and dermatological use is an adjunct tool that uses autofluorescence to detect tissues or structures. This device is not intended to provide a diagnosis.
Special Controls
In combination with the general controls of the FD&C Act, the autofluorescence detection device for general surgery and dermatological use is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo testing under anticipated conditions of use must characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical, and thermal safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance testing must demonstrate the sterility of patient-contacting components of the device.
(6) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(7) Performance testing must demonstrate laser and light safety for eye, tissue, and skin.
(8) Labeling must include the following:
(i) Instructions for use;
(ii) The detection performance characteristics of the device when used as intended; and
(iii) A shelf life for any sterile components.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 20, 2022
MolecuLight, Inc. Jordan John Director. Quality Assurance & Regulatory Affairs Suite 700, 425 University Avenue Toronto, Ontario M5G 1T6 Canada
Re: K213840
Trade/Device Name: MolecuLight I:X Regulation Number: 21 CFR 878.4550 Regulation Name: Autofluorescence detection device for general surgery and dermatological use Regulatory Class: Class II Product Code: QJF, FXN
Dear Jordan John:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 18, 2022. Specifically, FDA is updating this SE Letter as an administrative correction for an inappropriate product code that does not categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Office of Surgical and Infection Control Devices at 301-796-7674 Jianting Wang, or Jianting.wang(@fda.hhs.gov.
Sincerely,
Digitally signed by Jianting Wang
Jianting Wang -S
Date: 2022.07.20 11:11:28 -04'00'
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 18, 2022
MolecuLight, Inc. Jordan John Director. Quality Assurance & Regulatory Affairs Suite 700, 425 University Avenue Toronto, Ontario M5G 1T6 Canada
Re: K213840
Trade/Device Name: MolecuLight i:X Regulation Number: 21 CFR 878.4550 Regulation Name: Autofluorescence Detection Device For General Surgery And Dermatological Use Regulatory Class: Class II Product Code: QDG, QJF Dated: April 13, 2022 Received: April 20, 2022
Dear Jordan John:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
Page 2
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Purva U. Pandya -S
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213840
Device Name MolecuLight i:X
Indications for Use (Describe)
The MolecuLight i:X is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to
(i) View and digitally record images of a wound,
(ii) Measure and digitally record the size of a wound, and
(iii) View and digitally record images of fluorescence emitted from a wound when exposed to an excitation light.
The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >10^4 CFU per gram) as compared to examination of clinical signs and symptoms alone. The MolecuLight i:X device should not be used to rule-out the presence of bacteria in a wound.
The MolecuLight i:X does not diagnose or treat skin wounds.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
# MolecuLight i:X
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
MolecuLight Inc. Suite 700, 425 University Avenue Toronto, ON, Canada M5G 1T6 Phone: 647-362-4684 Contact Person: Jordan John
Date Prepared: May 13, 2022
#### Name of Device
MolecuLight i:X
#### Device Classification and Product Code
Autofluorescence detection device, 21 CFR 878.4550, Class II, QJF Tape, Camera, Surgical, 21 CFR 878.4160, Class I, FXN
#### Predicate Devices
MolecuLight i:X (K210882)
#### Indications for Use
The MolecuLight i:X is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to
View and digitally record images of a wound, (i)
(ii) Measure and digitally record the size of a wound, and
View and digitally record images of fluorescence emitted from a wound when exposed to an (iii) excitation light.
The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >104 CFU per gram as compared to examination of clinical signs and symptoms alone. The MolecuLight i:X device should not be used to rule-out the presence of bacteria in a wound.
The MolecuLight i:X does not diagnose or treat skin wounds.
{5}------------------------------------------------
# Device Description
The MolecuLight i:X Imaging Device is a handheld medical imaging device comprised of a high-resolution color LCD display and touch-sensitive screen with integrated optical and microelectronic components. MolecuLight i:X uses its patented technology to enable real-time standard digital imaging and fluorescence (FL) imaging in wounds and surrounding healthy skin of patients as well as wound area measurements.
# Comparison of Intended Use, Indications for Use and Technological Characteristics with the Predicate Device
- The intended use and technological characteristics of the subject MolecuLight i:X are I. identical to the previously cleared MolecuLight i.X. The only difference between the subject and predicate device are additional statements in the device's labeling clarifying: i) that the fluorescence in a region of a wound corresponds to the presence of elevated bacterial loads (>104 CFU/g); ii) that fluorescence in a region of a wound corresponds to the presence of more bacterial species and bacterial species of interest; iii) that species that produce red fluorescence is detectable by the MolecuLight i.X in vitro; iv) that Gram(+), Gram(-), aerobic and anaerobic species produce fluorescence detected by the MolecuLight i:X.
These statements do not change the indications for use of the device, and does not raise any new questions of safety or efficacy. The statement is supported by additional analysis of the clinical study reported in support of K191371.
| | SUBJECT DEVICE | Predicate Device |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MolecuLight i:X | MolecuLight i:X (K210882) |
| Device Name | MolecuLight i:X | MolecuLight i:X |
| Manufacturer | MolecuLight Inc. | MolecuLight Inc. |
| 510(k) Number | - | K191371 |
| Regulatory Class | Class II | Class II |
| Regulation Number | QJF | QJF |
| Product Classification | 21 CFR 878.4550 | 21 CFR 878.4550 |
| Classification Name | Autofluorescence detection device for<br>general surgery and dermatological use | Autofluorescence detection device for<br>general surgery and dermatological use |
| Intended Use | Intended for general surgery and<br>dermatological use as an adjunct tool<br>that uses autofluorescence to detect<br>tissues or structures. This device is not<br>intended to provide a diagnosis. | Intended for general surgery and<br>dermatological use as an adjunct tool<br>that uses autofluorescence to detect<br>tissues or structures. This device is not<br>intended to provide a diagnosis. |
| Indications for Use | The MolecuLight i:X is a handheld<br>imaging tool that allows clinicians<br>diagnosing and treating skin wounds,<br>at the point of care, to<br><br>(i) View and digitally record<br>images of a wound. | The MolecuLight i:X is a handheld<br>imaging tool that allows clinicians<br>diagnosing and treating skin wounds,<br>at the point of care, to<br><br>(i) View and digitally record<br>images of a wound. |
| | | | | Table 1: Comparison of Technological Characteristics for Fluorescence Imaging | | |
|--|--|--|--|-------------------------------------------------------------------------------|--|--|
| | | | | | | |
{6}------------------------------------------------
| | SUBJECT DEVICE<br>MolecuLight i:X | Predicate Device<br>MolecuLight i:X (K210882) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (ii) Measure and digitally record<br>the size of a wound, and<br>(iii) View and digitally record<br>images of fluorescence emitted from a<br>wound when exposed to an excitation<br>light. | (ii) Measure and digitally record<br>the size of a wound, and<br>(iii) View and digitally record<br>images of fluorescence emitted from a<br>wound when exposed to an excitation<br>light. |
| | The fluorescence image, when used in<br>combination with clinical signs and<br>symptoms, has been shown to increase<br>the likelihood that clinicians can<br>identify wounds containing bacterial<br>loads >104 CFU per gram as compared<br>to examination of clinical signs and<br>symptoms alone. The MolecuLight<br>i:X device should not be used to rule-<br>out the presence of bacteria in a<br>wound.<br>The MolecuLight i:X does not<br>diagnose or treat skin wounds. | The fluorescence image, when used in<br>combination with clinical signs and<br>symptoms, has been shown to increase<br>the likelihood that clinicians can<br>identify wounds containing bacterial<br>loads >104 CFU per gram as compared<br>to examination of clinical signs and<br>symptoms alone. The MolecuLight<br>i:X device should not be used to rule-<br>out the presence of bacteria in a<br>wound.<br>The MolecuLight i:X does not<br>diagnose or treat skin wounds. |
| Labelled relationship<br>between Cyan<br>fluorescence and<br><i>Pseudomonas<br/>aeruginosa</i> | Yes | Yes |
| Labelled relationship<br>between region of a<br>wound and presence of<br>elevated bacterial<br>loads (>104 CFU/g) | Yes | No |
| Labelled relationship<br>between fluorescence<br>in a region of a wound<br>and the presence of<br>more bacterial species<br>and bacterial species of<br>interest | Yes | No |
| Labelled relationship<br>between species that<br>produce red<br>fluorescence and red<br>fluorescence signature | Yes | No |
| Labelled relationship<br>between fluorescence<br>imaging and<br>identifying wounds<br>with elevated bacterial<br>load including Gram | Yes | No |
| | SUBJECT DEVICE<br>MolecuLight i:X | Predicate Device<br>MolecuLight i:X (K210882) |
| (+) Gram (-), aerobic<br>and anaerobic species. | | |
| Target Organ | Wounds | Wounds |
| Patient Population | Adult patients | Adult patients |
| Operating Modes | Standard and fluorescence imaging, video and image capture | Standard and fluorescence imaging, video and image capture |
| Excitation Light | 405 nm light emitted from light emitting diodes (LED)s | 405 nm light emitted from light emitting diodes (LED)s |
| Laser Power Density | N/A | N/A |
| Infrared LED | N/A | N/A |
| White LED | N/A | N/A |
| Emission Wavelength | 500-545 nm and 600-665 nm | 500-545 nm and 600-665 nm |
| Contrast agent | Not required - autofluorescent target | Not required - autofluorescent target |
| Working Distance | 8-12 cm | 8-12 cm |
| Resolution (focal plane) | 5 megapixels | 5 megapixels |
| Magnification | N/A | N/A |
| Maximum Frame Rate | 30 images/sec | 30 images/sec |
| Camera Bit Depth | 8 bits | 8 bits |
| Image Size (Pixels) | 1136 x 640 pixels | 1136 x 640 pixels |
| Image Format | JPEG | JPEG |
| Video Format | MOV | MOV |
| Software Operating System (OS)<br>Compatibility | Apple iOS 9.3.5 | Apple iOS 9.3.5 |
| Measurement Functionality | Wound length, width, and area measurements | Wound length, width, and area measurements |
| Power Supply | Battery and Wall | Battery and Wall |
| Display | Handheld device; no remote display | Handheld device; no remote display |
| Shelf-Life | 2 Years | 2 Years |
| Lifetime | 5 Years | 5 Years |
| Patient Contacting Materials | Non-patient contacting device (held 8-12 cm from skin) | Non-patient contacting device (held 8-12 cm from skin) |
| Sterility | Used non-sterile | Used non-sterile |
| Electrical Safety | Compliance to IEC 60601-1 | Compliance to IEC 60601-1 |
| Mechanical Safety | Compliance to IEC 60601-1 | Compliance to IEC 60601-1 |
| Chemical Safety | No chemical delivered or used as part of the system | No chemical delivered or used as part of the system |
| Standards with which the Device Complies | IEC 60601-1-2<br>IEC 60601-1<br>IEC 60601-2-57<br>IEC 62471 | IEC 60601-1-2<br>IEC 60601-1<br>IEC 60601-2-57<br>IEC 62471 |
{7}------------------------------------------------
{8}------------------------------------------------
In summary, the modified MolecuLight i:X with the additional labeling statements is substantially equivalent to the legally marketed MolecuLight i:X. The intended use of the i:X device is the same as the predicate, and there are no differences in technological characteristics. The additional labeling statements do not raise different questions of safety or efficacy. Retrospective analysis has demonstrated the safety and effectiveness of MolecuLight i:X with regards to the additional labeling statements. Thus, the MolecuLight i:X is substantially equivalent to the previously cleared MolecuLight i:X.
### Non-Clinical Testing
Each species listed produced red fluorescence that was detectable through fluorescence imaging with the MolecuLight i:X. To demonstrate this, all species were sub-cultured from frozen isolated on commercially available Porphyrin Test Agar (PTA). Porphyrins fluoresce red under violet light illumination. Aerobic bacterial species were cultured and imaged at 24 and 40 hours, while slower growing anaerobic bacterial species were cultured and imaged at 40 and 120 hours. Negative controls were included and imaged at all time points. The fluorescence images taken with the MolecuLight i:X of each bacterial species were analyzed to determine the presence of red fluorescence using a custom algorithm.
# Clinical Performance Testing
Data from post hoc retrospective analysis of 78 patients were analyzed to evaluate the effectiveness of MolecuLight i:X to guide wound sampling to detect bacterial burden, including pathogens of interest, as defined by the CDC, compared to the Standard of Care (SoC) method of sampling, which is collected from the center of the wound.
Data from all wounds that had two samples obtained in the study were analyzed (N = 78). Samples targeted to the brightest location of fluorescence (FL-guided) were more likely to contain elevated bacterial load (≥ 10ª CFU/g) compared to SoC-guided sampling at the center of the wound (See Table 2).
Table 2: Sensitivity of Soc-Guided Biopsy and FL-Guided Biopsy to Detect Any Species at Levels ≥ 104 CFU/G in Patients with Two Biopsies Obtained
| | SoC-guided sample | FL-guided sample | P-value |
|-------------|---------------------------------|-----------------------------------|-----------|
| Sensitivity | 87.2%<br>(95% CI: 77.7%, 93.7%) | 98.7%<br>(95% CI: 93.06%, 99.97%) | P = 0.012 |
Fluorescence-guided biopsies detected a higher number of species compared to SoC-guided biopsies as described in Table 3.
{9}------------------------------------------------
| | Number of Species<br>by FL-guided Biopsy<br>(N=78) | Number of Species<br>by SoC-guided<br>Biopsy<br>(N=78) | Difference<br>in<br>Number<br>of<br>Pathogens<br>(FL<br>Biopsy # -<br>SoC<br>Biopsy #) | P-value<br>(paired t-<br>test) | 95%<br>Confidence<br>Interval |
|-------------------|----------------------------------------------------|--------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------|-------------------------------|
| Mean (SD) | 3.026 (1.667) | 2.231 (1.528) | 0.795<br>(1.804) | P<0.001 | 0.388,<br>1.202 |
| Med (Min,<br>Max) | 3.0 (0.0, 8.0) | 2.0 (0.0, 6.0) | | | |
Table 3: Count of Bacterial Species Detected by SoC and FL-guided Biopsies
FL-guided biopsies detected a higher number of pathogens of interest, defined by the CDC as increasing risk to develop antibiotic resistance, compared to SoC-guided biopsies as described in Table 4.
| | | | Table 4: Count of pathogens of interest detected by SoC and FL-guided biopsies |
|--|--|--|--------------------------------------------------------------------------------|
|--|--|--|--------------------------------------------------------------------------------|
| | Number of<br>Pathogens of<br>Interest<br>by FL-guided<br>Biopsy<br>(N=78) | Number of Pathogens<br>of Interest<br>by SoC-guided<br>Biopsy<br>(N=78) | Difference<br>in<br>Number<br>of<br>Pathogens<br>of Interest<br>(FL<br>Biopsy # -<br>SoC<br>Biopsy #) | P-value<br>(paired<br>t-test) | 95%<br>Confidence<br>Interval |
|-------------------|---------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------|
| Mean (SD) | 1.731<br>(1.124) | 1.423<br>(1.134) | 0.308<br>(0.916) | P=0.002 | 0.101,<br>0.514 |
| Med (Min,<br>Max) | 2.0 (0.0, 5.0) | 1.0 (1.0, 5.0) | | | |
# Conclusion
The modified MolecuLight i:X is substantially equivalent to the cleared MolecuLight i:X.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.