Oxygen Concentrator model 8F-3A, Oxygen Concentrator model 8F-5A
Applicant
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Product Code
CAW · Anesthesiology
Decision Date
Apr 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.
Device Story
Oxygen concentrator (models 8F-3A/8F-5A) utilizes molecular sieve and pressure swing adsorption (PSA) to separate nitrogen from room air; produces enriched oxygen output. Used in professional healthcare facilities and home environments; operated by patients or healthcare providers. Device includes compressor, fan, flowmeter, and circuit breaker. Output delivered via nasal cannula. Provides supplemental oxygen to patients; not life-supporting. Healthcare providers monitor output via LCD display and indicator lights; device includes alarms for pressure, flow, current, temperature, and power status. Benefits include portable oxygen therapy for patients with respiratory needs.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (ES 60601-1), EMC (IEC 60601-1-2), alarm systems (IEC 60601-1-8), home healthcare environment safety (IEC 60601-1-11), oxygen concentrator performance (ISO 80601-2-69), biocompatibility (ISO 18562-1/2, ISO 10993-5/10), software verification/validation (IEC 62304), shelf life, transportation (ASTM D4169), and altitude performance.
Technological Characteristics
Pressure swing adsorption (PSA) with molecular sieve. Materials: gas pathway components (compressor, valve, silicone tube, sieve bed, oxygen tank) tested for biocompatibility per ISO 10993 and ISO 18562. Connectivity: standalone. Power: electrical. Form factor: portable unit with LCD display and buttons. Software: moderate level of concern. Standards: ISO 80601-2-69, ES 60601-1, IEC 60601-1-2, ASTM D4169-16.
Indications for Use
Indicated for adult patients requiring supplemental oxygen therapy in professional healthcare facilities or home environments. Not for life-supporting or life-sustaining use.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
April 28, 2023
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. % Randy Jiang Sr. consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K213825
Trade/Device Name: Oxygen Concentrator model 8F-3A, Oxygen Concentrator model 8F-5A Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: March 27, 2023 Received: March 28, 2023
Dear Randy Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ting Song -S
For
James J. Lee, Ph.D. Director, Division of Sleep Disordered Breathing, Respiratory and Anesthesia Office of Health Technologies 1, Ophthalmic, Anesthesia, Respiratory, ENT & Dental Devices Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213825
Device Name
Oxygen Concentrator models 8F-3A&8F-5A
Indications for Use (Describe)
This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K213825
#### 1. Submission Sponsor
Jiangsu Yuyue Medical Equipment& Supply Co., Ltd Yunyang Industrial Park Danyang Jiangsu China Wei Yan, RD/RA Email: weiyan@yuyue.com.cn Phone number: 025-85582701-6225
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX78746 Office Phone:(512) 327-9997 Contact: Randy Jiang Title: Senior Technical Consultant
#### 3. Date Prepared
April 24, 2023
#### 4. Device Identification
| Trade/Proprietary Name: | Oxygen Concentrator models 8F-3A/8F-5A |
|-------------------------|----------------------------------------|
| Common/Usual Name: | Portable oxygen generator |
| Classification Name: | Portable oxygen generator |
| Regulation Number: | 21 CFR 868.5440 |
| Product Code: | CAW |
| Class: | II |
| Classification Panel: | Anesthesiology |
#### 5. Legally Marketed Predicate Device
| Predicate Device: | |
|-------------------------|----------------------------------|
| Trade/Proprietary Name: | L4 Oxygen Concentrator (EverFlo) |
| Common/Usual Name: | Portable oxygen generator |
{4}------------------------------------------------
| Classification Name: | Portable oxygen generator |
|-----------------------|---------------------------|
| Regulation Number: | 21 CFR 868.5440 |
| Product Code: | CAW |
| Class: | II |
| Classification Panel: | Anesthesiology |
| 510(k) Number: | K061261 |
| Manufacturer: | RESPIRONICS, INC. |
## 6. Indication for Use Statement
This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered lifesupporting or life-sustaining.
## 7. Device Description
The Yuyue Oxygen Concentrator models 8F-3A/8F-5A is a portable concentrator which could utilize a molecular sieve and pressure swing adsorption to produce the oxygen. The device could separate nitrogen by absorbing through the molecular sieve when the room air enters the device, and allows the enriched oxygen to be collected. The subject device is intended use for adult population.
The subject device contains primary electrical components including main unit, compressor, fan, flowmeter and circuit breaker, the software is moderate level according to FDA software guideline.
The patient contacting components are listed in following table, the biocompatibility test and gas pathway biocompatibility test was conducted to subject device against ISO 10993-10, ISO 18562-1, ISO 18562-2, ISO 18562-3.
{5}------------------------------------------------
| Name of Component | Type and Duration of Contact | Direct or Indirect<br>Contact |
|-------------------|--------------------------------------------------------|-------------------------------|
| Compressor | External communicating Device,<br>Long term, > 30 days | Indirect (gas pathway) |
| Combination valve | External communicating Device,<br>Long term, > 30 days | Indirect (gas pathway) |
| Silicone tube | External communicating Device,<br>Long term, > 30 days | Indirect (gas pathway) |
| Sieve bed | External communicating Device,<br>Long term, > 30 days | Indirect (gas pathway) |
| Oxygen tank | External communicating Device,<br>Long term, > 30 days | Indirect (gas pathway) |
# 8. Substantial Equivalence Discussion
The following table compares the Oxygen Concentrator to the predicate device with respect to indications for use, and technological characteristics, forming the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
{6}------------------------------------------------
| Characteristics | Subject Device | Predicate Device K061261 | Comparison |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Device name | oxygen concentrator models<br>8F-3A/8F-5A | L4 Oxygen<br>Concentrator(EverFlo) | N/A |
| Classification | 2 | 2 | Same |
| Product Code | CAW | CAW | Same |
| Regulation<br>Number | 21 CFR 868.5440 | 21 CFR 868.5440 | Same |
| Intended Use | Provide supplemental oxygen | Provide supplemental oxygen | Same |
| Indication for<br>Use | This oxygen concentrator is<br>intended for using as an<br>oxygen supplement device in<br>the professional healthcare<br>facility and home healthcare<br>environment. It provides high<br>concentration of oxygen to<br>persons requiring oxygen<br>therapy. This device is to be<br>used as an oxygen supplement<br>and is NOT considered life-<br>supporting or life-sustaining. | The Respironics L4 Oxygen<br>Concentrator is intended to<br>provide supplemental oxygen<br>to persons requiring oxygen<br>therapy. The device is not<br>intended to be life supporting<br>or life sustaining.<br>The Respironics L4 Oxygen<br>Concentrator is intended for<br>use in the home or<br>hospital/institutional<br>environment. | Equivalent |
| Environment of<br>Use | professional healthcare facility<br>and home healthcare<br>environment. | home or hospital /<br>institutional environment. | Same |
| IP classification | IP21 | IP21 | Same |
| Patient<br>Interface | nasal oxygen cannula | nasal oxygen cannula | Same |
| Technology | Pressure Swing Adsorption<br>with molecular sieve | Pressure Swing Adsorption<br>with molecular sieve | Same |
| Oxygen<br>Concentration | (8F-3A): ≥90%;<br>(8F-5A): 95.5%-87% | 90-96% from 1 to 5 LPM | Equivalent |
| Characteristics | Subject Device | Predicate Device K061261 | Comparison |
| Flow Rate | (8F-3A): 1-3 LPM;<br>(8F-5A): 1-5 LPM | from 1 to 5 LPM | Equivalent |
| Output<br>Pressure | ≤70 kPa | 44.8 kPa (6.5PSIG) | Equivalent |
| User Interface | Buttons | Buttons | Same |
| | LCD Display and Indicator light | LCD Display and Indicator light | |
| Alarms | E 1 low pressure | System malfunction | Similar |
| | E 2 high pressure | | |
| | E 3 low current | | |
| | E 4 high current | | |
| | Low oxygen concentration<br>alarm | Low oxygen condition (opi unit only) | |
| | E 5 high temperature inside<br>the device | N/A | |
| | N/A | high oxygen flow condition | |
| | LL low flowrate | Low oxygen flow condition | |
| | Power interrupt alarm | Power interrupt alarm | |
| Complies with<br>ISO 10993-1 | Yes | Yes | Same |
| Complies with<br>ISO 80601-2-69 | Yes | Yes | Same |
| Complies with<br>ASTM D4169-16 | Yes | Yes | Same |
| Complies with<br>ES 60601-1<br>Electrical Safety | Yes | Yes | Same |
| Characteristics | Subject Device | Predicate Device K061261 | Comparison |
| Complies with<br>IEC 60601-1-2<br>EMC Safety | Yes | Yes | Same |
# Table 5A – Comparison of Characteristics
{7}------------------------------------------------
{8}------------------------------------------------
#### Comparison Discussion Summary:
The Oxygen Concentrator models 8F-3A&8F-5A has the same intended use and the same or similar technological characteristics and functionality as the predicate, the slight differences do not raise new questions of safety and effectiveness as compared to the predicate device.
#### 9. Non-Clinical Performance Data
To demonstrate safety and effectiveness of Oxygen Concentrator models 8F-3A/8F-5A and to show substantial equivalence to the original device, Jiangsu Yuyue Medical Equipment& Supply Co., Ltd completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The device passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Electrical safety testing per ES 60601-1
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2
- . General requirements, tests and guidance for alarm systems per IEC 60601-1-8
- General requirements for basic safety and essential performance in the home healthcare environment per IEC 60601-1-11
- Particular requirements for the basic safety and essential performance of oxygen concentrator equipment per ISO 80601-2-69
- Biocompatibility in gas pathway per ISO 18562-1, ISO 18562-2, ISO 10993-5, ISO 10993-10.
- Software verification and validation per IEC 62304/FDA Guidance – results /conclusion
- . Shelf Life Testing - Supports the claimed shelf life
- . Transportation Testing per ASTM D4169 – Demonstrates package integrity maintained
- . Altitude limit 2000m device performance test
- Accessory compatibility test including connect security, accessory performance
#### 10. Statement of Substantial Equivalence
The Oxygen Concentrator models 8F-3A/8F-5A has the same intended use as the predicate device, and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated the Oxygen Concentrator models 8F-3A/8F-5A is as safe and effective as the predicate device. Therefore, the Oxygen Concentrator models 8F-3A/8F-5A is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.