K213813 · Neuramedica, Inc. · FZP · Jul 22, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213813
Device Name
DuraFuse Clip and Applier System
Applicant
Neuramedica, Inc.
Product Code
FZP · General, Plastic Surgery
Decision Date
Jul 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DuraFuse Clips are indicated for open and tubular retractor procedures of the spine in the prone position for approximation/attachment and/or closure of the dura mater in neurosurgical and orthopedic spine procedures.
Device Story
DuraFuse Clip and Applier System facilitates rapid dural closure during spine surgery. System consists of small, bioabsorbable, non-penetrating clips and a reusable stainless steel applier. Surgeon uses applier to place clips on everted dural edges; clips provide mechanical pressure to seal spinal column, allowing tissue healing. Device used in OR by neurosurgeons or orthopedic surgeons during open or tubular retractor procedures. Low-profile applier design ensures surgical site visibility. Bioabsorbable material eliminates need for permanent metal implants. Clinical benefit includes effective dural approximation and prevention of CSF leakage.
Clinical Evidence
Evidence includes biocompatibility (ISO 10993-1), bench testing, human factors study, and animal study. Bench testing confirmed closure of 1-10 cm defects without leakage. Human factors study (n=14 surgeons) validated ease of use and visibility in cadaver models. Animal study (swine, n=8) compared DuraFuse to predicate in laminectomy/durotomy model; 14 and 90-day follow-up (CT myelogram, histomorphology) showed favorable tissue response, dural contiguity, and no CSF leakage, comparable to control.
Indicated for patients undergoing open or tubular retractor spine procedures in the prone position requiring dural closure or approximation during neurosurgical or orthopedic surgery.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 22, 2022
NeuraMedica, Inc. Rachel Dreilinger CEO 402 Beavercreek Road, Suite 110 Oregon City, Oregon 97045
Re: K213813
Trade/Device Name: DuraFuse Clip and Applier System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP Dated: June 17, 2022 Received: June 21, 2022
# Dear Rachel Dreilinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213813
Device Name DuraFuse Clip and Applier System
#### Indications for Use (Describe)
DuraFuse Clips are indicated for open and tubular retractor procedures of the spine in the prone position for approximation/attachment and/or closure of the dura mater in neurosurgical and orthopedic spine procedures.
Type of Use (Select one or both, as applicable):
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### Submitter's Information
NeuraMedica Inc. 1515 7th STE A Oregon City, OR 97045 503-914-0723
| Contact Person: | Rachel Dreilinger, CEO |
|-----------------|------------------------|
| | rachel@neuramedica.com |
Date Prepared: June 17, 2022
#### Device Information
| Device: | Trade Name: | DuraFuse Clip and Applier System |
|---------|-----------------------|----------------------------------|
| | Classification Name: | Clip, Implantable |
| | Regulation Number: | 878.4300 |
| | Product Code: | FZP |
| | Classification Panel: | General &<br>Plastic Surgery |
| | Classification: | Class 2 Device |
## Predicate Device
The predicate device with which NeuraMedica DuraFuse Clips claims substantially equivalency is below:
| Company: | United States Surgical<br>Corporation |
|----------|------------------------------------------|
| Device: | Auto Suture Modified VCS Clip<br>Applier |
| 510(k): | K962043, Cleared September 23, 1996 |
#### Device Description
NeuraMedica's DuraFuse Clip is a small, bioabsorbable, non-penetrating clip for rapid closure of the dura mater during spine surgery. DuraFuse Clips are applied individually using the reusable stainless steel NeuraMedica Applier has a low profile that allows visibility for implantation of clips in open or tubular retractor procedures.
NeuraMedica DuraFuse Clips are provided sterile and labeled for single use. Dural Clips are sterilized via a validated e-beam dose.
NeuraMedica's DuraFuse Applier is autoclavable. It is provided non-sterile.
{4}------------------------------------------------
## Indication for Use
DuraFuse Clips are indicated for open and tubular retractor procedures of the spine in the prone position for approximation/attachment and/or closure of the dura mater in neurosurgical and orthopedic spine procedures.
## Comparison of Indications for Use
The predicate device is indicated for use in vascular anastomosis as well as approximation of dural tissue. DuraFuse Clips have limited indications for use to approximation of dural tissue.
The predicate device does not specify the type of procedure, open or minimallyinvasive, while the indication for DuraFuse Clips specifies the types of procedures that were tested during development, open and tubular-retractor.
The differences in Indications for Use Statements do not alter the therapeutic use of the device in spine surgery to approximate dura tissue nor do they raise new questions of safety or effectiveness.
## Comparison of Technological Characteristics:
The predicate and DuraFuse Clips function the same, they apply pressure to the everted edges of the dura to seal the spinal column and allow the dura to heal. At a high level, the subject and predicate devices are based on the following same technological elements:
- Similar shape and size
- . Clips with rounded tips to avoid tissue penetration and reduce the potential for CSF leakage
- . Hold everted edges of dura together to allow healing
The following technological differences exist between the subject and predicate device:
- Predicate clips are metal while the subject device is a bioabsorbable material
- Predicate clips are provided pre-loaded in an applier, while the subject device is supplied separately from a reusable applier
- . Predicate device has an accessory tool for removal or repositioning of clips, while the subject device has a reusable applier that can be used for removing or repositioning clips
{5}------------------------------------------------
## Performance Data - Biocompatibility
Biocompatibility evaluation was performed on the DuraFuse Clips consistent with ISO 10993-1. Testing demonstrated passing results for cytotoxicity, irritation, and sensitization. Implantation testing demonstrated that long-term tissue responses are not expected to occur from the use of DuraFuse Clips.
## Performance Data – Non-Clinical
Bench top testing using a synthetic dural model was performed and demonstrated that the DuraFuse Clips can successfully approximate synthetic dural material in defects from 1 cm to 10 cm without fluid leakage or damage to dura. The bench top data support the equivalent performance to the predicate device at higher pressures than could be evaluated in the in vivo animal model.
Human Factors testing in a human cadaver model was completed in open and tubular retractor procedures. Fourteen experienced surgeons evaluated the DuraFuse Clip and Applier and found that:
- All components are easy to use
- Applier can be used with one hand
- Applier allows visibility of the surgical site in open and tubular retractor procedures
- Packaging allows the user to easily remove clips for implantation
- Clips can be applied successfully based on the information in the Instructions For Use
## Performance Data – Animal Study
NeuraMedica has demonstrated equivalence of the proposed device to the predicate device in a swine model.
Eight swine underwent creation of an approximately 1 cm durotomy overlying the L2/3 intervertebral discs and was then treated with the test or control article. Animal health was monitored, including incision site and clinical observations, neurological exams, body weight/condition, and clinical pathology, at pre-determined, regular intervals. CT scans of the head/brain and lumbar spine were performed before and before necropsy. Blood samples for clinical pathology were collected prior to surgery and prior to necropsy. On Day 14 or 90, a CSF sample was collected, a CT myelogram was performed, the animals were euthanized, and a comprehensive necropsy followed by histomorphological examination was performed. Assessment of dural contiguity at
{6}------------------------------------------------
treatment sites was performed, specifically including indication of any presence of meningocoele/pseudomeningocoele.
After 14 or 90 days in the porcine spinal laminectomy and durotomy model, the DuraFuse Clip was associated with favorable tissue responses and dural closure with no evidence of CSF leakage. All parameters evaluated were comparable between the test and control groups.
#### Conclusion
In summary, NeuraMedica's DuraFuse Clip and Applier System has the same intended use as the predicate device, are similar in size, and are used in a cleared anatomical location, spine, as the currently cleared device The Dural Clips are sterile and nonpyrogenic. Biocompatibility testing, bench testing and comparative animal testing demonstrate the equivalence of the NeuraMedica DuraFuse Clip and Applier System to the predicate device and that the technological differences do not raise any new concerns of safety and effectiveness.
Based on the 510(k) summaries and the information provided, we conclude that NeuraMedica DuraFuse Dural Clip and Applier System is substantially equivalent to the predicate device, Auto Suture Modified VCS Clip Applier (K962043).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.