K213800 · Artiglass Srl · QEH · May 19, 2022 · General Hospital
Device Facts
Record ID
K213800
Device Name
Artiglass NRFitTM Tip L.O.R. Glass Syringes
Applicant
Artiglass Srl
Product Code
QEH · General Hospital
Decision Date
May 19, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Artiglass NRFit TM tip Glass Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique as explained in standard textbooks. These syringes are not intended for injection or aspiration.
Device Story
Device is a glass syringe used by clinicians to verify epidural needle placement via Loss of Resistance (LOR) technique. Input is manual tactile feedback from the plunger during needle advancement; device transforms this into a visual confirmation of resistance change. Output is the physical movement of the plunger within the barrel. Used in clinical settings (e.g., anesthesia) by physicians. Benefits include accurate identification of the epidural space, reducing risk of improper needle placement. Device consists of borosilicate glass barrel/plunger and nickel-plated brass NRFit tip (ISO 80369-6 compliant) to prevent misconnections. Not for injection or aspiration.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence is supported by non-clinical bench testing, including leakage, pressure decay, axial load, unscrewing, and overriding resistance tests per ISO 80369-6 and ISO 80369-20.
Technological Characteristics
Materials: Borosilicate neutral glass (barrel/plunger), nickel-plated brass (tip). Dimensions: 5mL and 10mL volumes. Connector: ISO 80369-6 (NRFit) compliant. Sterilization: Steam (provided non-sterile). Biocompatibility: ISO 10993-1, 4, 5, 10, 11. Standards: ISO 595-1, ISO 595-2, ISO 80369-6, ISO 80369-20, ISO 7886-1.
Indications for Use
Indicated for adult patients to verify epidural needle tip placement via the Loss of Resistance technique. Not for injection or aspiration.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 19, 2022
Artiglass Srl Sabrina Baccarin Quality Director Via Piemonte 13 Due Carrare, Padova 35020 Italy
Re: K213800
Trade/Device Name: Artiglass NRFitTM Tip L.O.R. Glass Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QEH Dated: April 15, 2022 Received: April 21, 2022
Dear Sabrina Baccarin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Sub part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213800
Device Name Artiglass NRFit Tip L.O.R. Glass Syringes
Indications for Use (Describe)
The Artiglass NRFit tip Glass Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique as explained in standard textbooks. These syringes are not intended for injection or aspiration.
Type of Use (Select one or both, as applicable):
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <path d="M2 2 L14 2 L14 14 L2 14 Z M3 3 L13 3 L13 13 L3 13 Z" fill="none" stroke="black" stroke-width="1"></path> <path d="M4 7 L6 9 L12 3" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;"> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> Over The Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Applicant:
Artiglass Srl via Piemonte 13, Due Carrare Padova. Italy
| Contact Person: | Sabrina Baccarin |
|-----------------------|-------------------------|
| Contact Title: | Quality Director |
| Contact Phone Number: | +39 049 5290442 |
| Contact Fax Number: | +39 049 5290446 |
| E-mail: | regulatory@artiglass.it |
Official Correspondent: same as Contact Person
### Date Summary Prepared: April 15, 2022
Classification: 21 CFR § 880.5860, Class II Classification name: Piston Syringe Product Code: QEH Trade Names : Artiglass NRFit™ Tip L.O.R. Glass Syringes Generic/Common Name: NRFitTM Tip L.O.R. Glass Syringes
#### Predicate device
Predicate Device: Artiglass L.O.R. Glass Syringes 510(k) Number: K122416 510(k) Clearance Date: May 3, 2013 Regulation Number: 21 CFR § 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF
### Indications for use
The Artiglass NRFit TM tip Glass Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique as explained in standard textbooks. These syringes are not intended for injection or aspiration.
#### Intended population
The syringe is intended for adult patient.
### Product Description
The Artiglass NRFit TM tip Glass Syringes provided not sterile with ISO 80369-6 (NRFit™) compliant fittings. They are available in 5mL and 10mL, lock and slip configurations and 5mL and 10mL lock configurations. The NRFit™ tips allow for connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices. The syringe assembly consists of a borosilicate neutral with a graduated scale in milliliters (mL), a borosilicate neutral glass plunger and a nickel-plated brass tip. The plunger rod is amber colored to designate a device intended to only connect to ISO 80369-6 compatible devices such as spinal or epidural needles.
### Comparison of Technological Characteristics
The comparison table of technological characteristics is documented in the following table Predicate Device Comparison.
{4}------------------------------------------------
| Model | Special 510(k) Artiglass<br>NRFitTM Tip L.O.R. Glass<br>Syringes<br>Subject Device | 510(k) Artiglass<br>L.O.R. Glass<br>Syringes |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| | Subject Device | K122416 |
| Product Code | QEH | FMF (1) |
| Common Name | Syringe, Piston | Same |
| Regulation Number | 880.5860 | Same |
| Regulation Description | Piston Syringe | Same |
| Submission Type | 510(k) Special | 510(k) Traditional |
| 510(k) Number | NA | K122416 |
| Intended Use | The Artiglass NRFit ™ tip Glass<br>Syringe is intended for use, in<br>conjunction with an epidural<br>needle, to verify the needle tip<br>placement in the epidural space<br>by the Loss of Resistance<br>technique as explained in<br>standard textbooks. These<br>syringes are not intended for<br>injection or aspiration. The<br>syringe is intended for adult<br>patient. | Same |
| Reusable/Sigle use | Single use (7) | Reusable |
| Sterility | Provided not sterile<br>The non-sterile NRFit ™ syringe<br>is intended to be sterilized prior<br>to use to repackagers/medical<br>device manufacturers. | Same |
| Tip | NRFit ™ Connector, Luer Metal<br>and Luer Lock | Luer Glass, Luer<br>Metal and Luer Lock<br>(2) |
| Tip material | Nickel plated brass | Same |
| Syringe Barrel | Borosilicate neutral glass | Same |
| Plunger | Borosilicate neutral glass | Same |
| Available Volumes | 5 mL, 10 mL | Same |
| Model | Special 510(k) Artiglass<br>NRFitTM Tip L.O.R. Glass<br>Syringes<br>Subject Device | 510(k) Artiglass<br>L.O.R. Glass<br>Syringes<br>K122416 |
| Calibrated Barrel<br>Volume | yes | Same |
| Recommended<br>sterilization method | Steam | Same |
| NRFit Connector | Yes; compliant with ISO<br>80369-6 | No; (2) |
| Package | Individually, in bulk | Same |
| Biocompatibility<br>Compliance | ISO 10993-1 Fifth edition 2018-<br>08 | ISO 10993-1 Fourth<br>edition 2009-10-15<br>(3) |
| | 10993-4 Third edition 2017-04 | ISO 10993-<br>4:2002/amd 1:2006<br>Hemocompatibility<br>(Interaction with<br>blood) (4) |
| | same | ISO 10993-5:2009 |
| | same | ISO 10993-10:2010<br>Irritation and skin<br>sensitization |
| | 10993-11 Third edition 2017-09 | ISO 10993-11:2006<br>Systemic toxicity<br>(5) |
| Standard Compliance | ISO 595-1:1988, ISO 595-2:1987,<br>ISO 80369-6:2016, ISO 80369-<br>20:2015 | ISO 594-1:1986, ISO<br>594-2:1998,<br>ISO 595-1:1988, ISO<br>595-2:1987<br>(6) |
## TABLE: Predicate Device Comparison
{5}------------------------------------------------
K213800
# SECTION 5 510 (K) Summary
{6}------------------------------------------------
### Discussions of differences in technological characteristics
- (1) The Product code is different. The differences were addressed through ISO 80369-20 and performance testing.
- (2) The tip of the subject device is specific. The differences were addressed through ISO 80369-20 performance testing.
- (3) The differences were addressed through the "Artiglass NRFitTM Tip L.O.R. Glass Syringes Biological evaluation plan"
- (4) The tests have been repeated according to 10993-4 Third edition 2017-04 and ASTM F756-17 (Hemolysis)
- (5) The tests have been repeated according to 10993-11 Third edition 2017-09
- (6) The differences were addressed through ISO 80369-6, ISO 80369-20 and ISO 7886-1 performance testing.
- (7) The syringe is intended for a single use. This not affect the effectiveness and safety of the device itself.
### Performance Testing in Support of Substantial Equivalence Determination
The performance of the subject Artiglass NRFitTM Tip L.O.R. Glass Syringes is demonstrated as tested per the applicable requirements of ISO 80369-6, Small bore connectors for liquids and gases in healthcare applications -Part 6: Connectors for neuraxial applications.
Non-Clinical Tests:
- 1. Leakage by pressure decay per ISO 80369-6
- 2. Sub-atmospheric-pressure air leakage per ISO 80369-6
- 3. Resistance to separation from axial load per ISO 80369-6
- 4. Resistance to separation from unscrewing per ISO 80369-6
- 5. Resistance to overriding per ISO 80369-6
- 6. Leakage between piston and barrel per ISO 595-2
- 7. Ink adhesion (permanence of marking) per ISO 595-2
- 8. Identification of burrs, hooks, cracks, foreign contamination, missing components per visual inspection
- 9. Fluid leakage requirement per ISO 80369-6
- 10. Stress Cracking per ISO 80369-6
Connector testing performed on the proposed device included the items listed below, in accordance with ISO 80369-6:2016, using the test methods provided in ISO 80369-20. The testing demonstrates the proposed devices conform to the applicable requirements of ISO 80369-6:2016
| Individual test Defined in ISO 80369-6 | Requirement Defined in<br>ISO 80369-6 | Test Method Defined<br>in ISO 80369-20 |
|---------------------------------------------|---------------------------------------|----------------------------------------|
| Fluid leakage requirement | Clause 6.1.1 | Annex B |
| Leakage by pressure decay | Clause 6.1.2 | Annex B |
| Sub-atmospheric-pressure air leakage | Clause 6.2 | Annex D |
| Stress Cracking | Clause 6.3 | Annex E |
| Resistance to separation from axial load | Clause 6.4 | Annex F |
| Resistance to separation from<br>unscrewing | Clause 6.5 | Annex G |
| Resistance to overriding | Clause 6.6 | Annex H |
{7}------------------------------------------------
Even though the geometry of the tip is slightly different in the subject device manufactured by Artiglass, the results of the biocompatibility testing performed by Artiglass on the previously cleared syringe can be considered adequate also to assess the biocompatibility of the subject device.
The materials of the components of the test article are identical to the correspondent components of the subject medical device in its final finished form in formulation, processing, and no other chemicals have been added. No changes in material have occurred since the biocompatibility testing performed by the manufacturer on the predicate device Artiglass L.O.R. Glass Syringe. The tip shape does not pose new issues of biocompatibility as the test article represents a worst case in respect to the subject device.
#### Clinical and Animal Tests:
Clinical tests were not required to demonstrate performance of between the subject and predicate devices. Product functionality has been adequately assessed by nonclinical tests.
#### Conclusion:
The submitted and the predicate device have the same indications for use and technological characteristics. The test results and comparison results show that the proposed device is substantially equivalent to the predicate devices in performance, materials, intended use, technological and operational characteristics.
Based on the intended use, technological characteristics, and performance testing, the proposed Artiglass NRFitTM Tip L.O.R. Glass Syringes has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.