DIGITAL RADIOGRAPHY CXDI-Elite, DIGITAL RADIOGRAPHY E1
K213780 · Canon, Inc. · MQB · Jan 24, 2022 · Radiology
Device Facts
Record ID
K213780
Device Name
DIGITAL RADIOGRAPHY CXDI-Elite, DIGITAL RADIOGRAPHY E1
Applicant
Canon, Inc.
Product Code
MQB · Radiology
Decision Date
Jan 24, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The DIGITAL RADIOGRAPHY CXDI-Elite / E1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
Solid-state X-ray imager (flat panel detector) using TFT arrays; intercepts X-ray photons via Cesium-Iodide scintillator; converts visible light to electrical signals; generates digital images for display on monitors. Used in clinical settings by healthcare professionals for general radiography. Device communicates via wired or wireless connection to operator console. Software (CXDI Control Software V3.10) manages image acquisition, calibration, and synchronization with X-ray generators. Includes 'Intelligent NR' function and 'Standard Synchronization Mode with Built in AEC Assistance'. Output allows clinicians to view radiographic images for diagnostic decision-making, replacing traditional film/screen systems.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing included functional specifications, software testing, and non-clinical image comparisons. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-54, and IEC 60529 (IP57) confirmed. DQE performance measured at 0.67 (at 3.5uGy, 0.5 lp/mm).
Indicated for digital image capture for conventional film/screen radiographic examinations of human anatomy in general purpose diagnostic procedures. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
January 24, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Canon Inc. % Mr. Gregory Woodard Biomedical Engineer Ken Block Consulting 800 East Campbell Road, Suite 202 RICHARDSON TX 75081
Re: K213780
Trade/Device Name: DIGITAL RADIOGRAPHY CXDI-Elite, DIGITAL RADIOGRAPHY E1 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: December 2, 2021 Received: December 3, 2021
Dear Mr. Woodard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213780
Device Name DIGITAL RADIOGRAPHY CXDI-Elite DIGITAL RADIOGRAPHY E1
#### Indications for Use (Describe)
The DIGITAL RADIOGRAPHY CXDI-Elite / E1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Canon" in a large, bold, red font. Below the word "Canon" is the text "5. 510(k) SUMMARY" in a smaller, black font. The text is left-aligned and the background is white.
| K213780 | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Applicant/<br>Sponsor: | Canon Inc.<br>9-1 Imaikami-cho<br>Nakahara-ku, Kanagawa 211-8501 JAPAN | |
| Contact Person: | Mr. Akira Hirai<br>General Manager<br>TEL: 81-3-3758-2111<br>FAX: 044-739-6695<br>hirai.akira@mail.canon | |
| Date Prepared: | December 2, 2021 | |
| Proposed Device | Manufacturer: Canon Inc.<br>Trade Name: DIGITAL RADIOGRAPHY CXDI-Elite<br>DIGITAL RADIOGRAPHY E1<br>Common Name: Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray System<br>Product Code: MQB<br>Regulation: 892.1680, Stationary X-ray System | |
| Predicate Device: | Clearance: K170332<br>Manufacturer: Canon Inc.<br>Trade Name: DIGITAL RADIOGRAPHY CXDI-710C Wireless<br>Common Name: Solid State X-ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray System<br>Product Code: MQB<br>Regulation: 892.1680, Stationary X-ray System | |
| Reference Device: | Clearance: K192632<br>Manufacturer: Canon Inc.<br>Trade Name: DIGITAL RADIOGRAPHY CXDI-702C Wireless<br>Common Name: Solid State X-ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray System<br>Product Code: MQB<br>Regulation: 892.1680, Stationary X-ray System | |
| Device Description: | The DIGITAL RADIOGRAPHY CXDI-Elite, also called the DIGITAL<br>RADIOGRAPHY E1, (hereinafter referred to as CXDI-Elite) is a solid-state x-ray<br>imager using TFT (thin-film transistor) arrays. The CXDI-Elite is a series of detectors,<br>currently consisting of the CXDI-720C Wireless detector unit, also called the AR-<br>E3543W detector. The detector intercepts x-ray photons, and the Cesium-Iodide<br>scintillator emits visible spectrum photons that illuminate an array of photodetectors<br>that create electrical signals. After the electrical signals are generated, the signals are<br>converted to a digital image, and the images will be displayed on monitors. The digital<br>image can be communicated to the operator console via a wired or wireless connection.<br><br>The monitors used with the CXDI-Elite are not a part of this submission.<br><br>The CXDI Control Software is updated from V2.16 to V3.10 in this submission. The<br>update to the software includes bug fixes, modification to check-in function, | |
| K213780 | modification to communication with X-ray emission devise, calibration support, addition of the Intelligent NR function (previously cleared under K212269), and the addition of Standard Synchronization Mode with Built in AEC Assistance. | |
| Indications for Use: | The DIGITAL RADIOGRAPHY CXDI-Elite / El provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. | |
| Summary of Technological Characteristics: | Comparison with the predicate device shows the technological characteristics of the CXDI-Elite is substantially equivalent to the predicate device. The flat panel detector units are functionally similar, but the proposed device adds Standard Synchronization Mode with Built in AEC Assistance as a photographing mode. | |
| The major differences between the CXDI-Elite and the predicate are differences in standard and optional components, newer version of CXDI control software, difference in case material, and a photographing mode not available in the predicate device. The proposed indications for use statement is identical to the predicate device. | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Canon logo in red, followed by the text "5. 510(k) SUMMARY" in black. The Canon logo is a stylized version of the company's name. The text below the logo appears to be a section heading or title, possibly from a document or report. The text is in a bold font and is left-aligned.
| | Proposed Device | Predicate Device | Reference Devices | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Trade Name | DIGITAL RADIOGRAPHY<br>CXDI-Elite / E1 | DIGITAL RADIOGRAPHY<br>CXDI-710C Wireless | DIGITAL RADIOGRAPHY<br>CXDI-702C Wireless | |
| 510(k)<br>Submitter<br>[Number] | Canon Inc.<br>[K213780] | Canon Inc.<br>[K170332] | Canon Inc.<br>[K192632] | IDENTICAL |
| Indication for<br>Use | The DIGITAL RADIOGRAPHY<br>CXDI-Elite / E1 provides digital<br>image capture for conventional<br>film/screen radiographic<br>examinations. This device is<br>intended to capture, for display,<br>radiographic images of human<br>anatomy, and to replace<br>radiographic film/screen systems<br>in all general purpose diagnostic<br>procedures. This device is not<br>intended for mammography<br>applications. | The DIGITAL RADIOGRAPHY<br>CXDI-710C Wireless provides<br>digital image capture for<br>conventional film/screen<br>radiographic examinations. This<br>device is intended to capture, for<br>display, radiographic images of<br>human anatomy, and to replace<br>radiographic film/screen systems<br>in all general purpose diagnostic<br>procedures. This device is not<br>intended for mammography<br>applications. | The DIGITAL RADIOGRAPHY<br>CXDI-702C Wireless provides<br>digital image capture for<br>conventional film/screen<br>radiographic examinations. This<br>device is intended to capture, for<br>display, radiographic images of human<br>anatomy, and to replace<br>radiographic film/screen systems<br>in all general purpose diagnostic<br>procedures. This device is not<br>intended for mammography<br>applications. | IDENTICAL |
| Application | General Radiography | General Radiography | General Radiography | IDENTICAL |
| Case Material | Magnesium alloy/ Fiberglass | Fiberglass | Magnesium Alloy | MODIFIED |
| Scintillator | CsI(TI)<br>[Cesium Iodide doped with<br>Thallium] | CsI(TI)<br>[Cesium Iodide doped with<br>Thallium] | CsI(TI)<br>[Cesium Iodide doped with<br>Thallium] | IDENTICAL |
| Pixel Pitch | 125μm | 125μm | 125μm | IDENTICAL |
| Spatial<br>Resolution | 45% [MTF@2lp/mm] | 35% [MTF@2lp/mm] | 35% [MTF@2lp/mm] | IDENTICAL |
| IP Level | IP57 | IPX7 | IP54 | MODIFIED |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Canon" in a large, bold, red font. Below the word "Canon" is the text "5. 510(k) SUMMARY" in a smaller, bold, black font. The text is centered below the word "Canon".
#### K213780
| Components | Detector (Sensor) | Detector (Sensor) | Detector (Sensor) | MODIFIED |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------|
| | Battery Pack (LB-4A) | Battery Pack (LB-4A) | Battery Pack (LB-4A) | |
| | Battery Charger (BC-01/BC-1A) | Battery Charger (BC-1A) | Battery Charger (BC-1A) | |
| | Multi Box (MB-02) | Multi Box (MB-4A) | Multi Box (MB-4A) | |
| | Status Indicator (SI-01/SI-4A) | Status Indicator (SI-4A) | Status Indicator (SI-4A) | |
| | Wiring Cable (WC-01) | Wiring Cable (WC-4A) | Wiring Cable (WC-4A) | |
| | PC Connection Cable (CP-01) | PC Connection Cable (CP-4A) | PC Connection Cable (CP-4A) | |
| | Ready Indicator (RI-3A) | Ready Indicator (RI-3A) | Ready Indicator (RI-3A) | |
| | X-ray Interface Box (XB-1A) | Docking Station (DS-4A) | X-ray Interface Box (XB-1A)<br>Docking Station (DS-4A) | |
| Software | CXDI Control Software V3.10 | CXDI Control Software V2.16 | CXDI Control Software V2.19 | MODIFIED |
| External<br>Dimensions | 384 x 460 x 15.5 mm | 384 x 460 x 15.7 mm | 384 x 460 x 15.7 mm | MODIFIED |
| DQE | A typical DQE value at 3.5uGy in<br>0.5 lp/mm is 0.67, with<br>measurement error of less than<br>±10%. | A typical DQE value at 4 $\mu$ Gy in 0<br>lp/mm is 0.6. The level of<br>uncertainty is estimated as less<br>than ±10%. | A typical DQE value at 1 mR in<br>0.5 lp/mm is 0.58. The level of<br>uncertainty is estimated as less<br>than ±10%. | MODIFIED |
| Detector<br>Technology | TFT | TFT | TFT | IDENTICAL |
| Photographing<br>Mode | Standard Synchronization Mode,<br>Standard Synchronization Mode<br>with Built in AEC Assistance,<br>Non Generator Connection Mode,<br>Standalone Mode | Standard Synchronization Mode,<br>Non Generator Connection Mode,<br>Standalone Mode | Standard Synchronization Mode,<br>Non Generator Connection Mode | MODIFIED |
| Wireless<br>Communication | IEEE 802.11a/b/g/n/ac<br>2.4GHz/5GHz<br>Bluetooth Low Energy | IEEE 802.11n<br>2.4GHz/5GHz | IEEE 802.11n/a/g/b<br>2.4GHz/5GHz | MODIFIED |
## Summary of Non-Clinical / Test Data:
Tests were performed on the device which demonstrated that the device is safe and effective, performs comparably to and is substantially equivalent to the predicate device. Tests included verification/validation testing to internal functional specifications (including software) and non-clinical image comparisons. Documentation was provided demonstrating compliance of the CXDI-Elite to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation plus traceability of verification/validation tests to software requirements and software risk hazards. Other FDA guidance documents used in development include Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
Documentation was provided demonstrating that the device complies with the FDA requirements stated in Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices. The evaluations of the device, compared to the predicate, show the device to be equivalent to the predicate.
Testing confirmed that the device complies with the U.S. Performance Standard for radiographic equipment and with relevant voluntary safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, and 60601-2-54. An IP Code of IP57 has been established for the CXDI-Elite per IEC 60529.
Clinical testing is not necessary for the current submission, based on the minor differences from the predicate device. Adequate detector bench testing should be sufficient to demonstrate that the subject detector CXDI-Elite / E1 works as intended.
Together, these verification/validation activities successfully demonstrated that the device correctly performs as designed, has been validated for its intended use, and raises no new questions regarding either safety or effectiveness when compared to the
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Canon" in a large, red, bold font at the top. Below the word "Canon" is the text "5. 510(k) SUMMARY" in a smaller, black, bold font. The text is aligned to the left and appears to be part of a document or report. The image is simple and contains only text.
## K213780
predicate device. Therefore, the verification/validation testing conducted supports a determination of substantial equivalence for the device.
- Company considers the DIGITAL RADIOGRAPHY CXDI-Elite / E1 device to be Conclusion: substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.