K213770 · Waldemar Link GmbH & Co. KG · LZO · Sep 13, 2022 · Orthopedic
Device Facts
Record ID
K213770
Device Name
SP-CL Hip Stem and LCU Hip System
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LZO · Orthopedic
Decision Date
Sep 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SP-CL Hip Stem and LCU Hip System are indicated for patients with mobility-limiting hip diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The hip systems are indicated for the following conditions: - Primary and secondary coxarthrosis - Osteoarthritis - Necrosis of the femoral head - Femoral neck fractures The stems are indicated for cementless use only.
Device Story
SP-CL Hip Stem and LCU Hip System are orthopedic implants for total hip replacement. SP-CL features anatomically curved stem; LCU features straight stem. Devices are implanted by orthopedic surgeons in clinical settings to replace diseased or fractured hip joints. Stems are manufactured from titanium alloy (Ti6Al4V) with HX (calcium phosphate) coating. Implants are used in conjunction with prosthesis heads and acetabular components. The device modification involves transitioning the HX (CaP) coating application from an external vendor to an in-house process. No changes to fundamental scientific technology, design, or surgical technique. Benefits include restoration of hip function and mobility for patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including shear testing (ASTM F1044-05), tension testing (ASTM F1147-05), stereological evaluation (ASTM F1854-15), infrared spectroscopy, dissolution/solubility rate (ASTM F1926/F1926M-14), elemental/chemical analysis (ASTM F1609-08), shelf life, and endotoxin testing.
Technological Characteristics
Titanium alloy (Ti6Al4V per ISO 5832-3, ASTM F136) stems with HX (CaP) coating (ASTM F1609). Cementless, semi-constrained hip prosthesis. Available in various sizes, lengths, and offsets. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients with mobility-limiting hip diseases, fractures, or defects (e.g., primary/secondary coxarthrosis, osteoarthritis, femoral head necrosis, femoral neck fractures) untreatable by conservative/osteosynthetic procedures. For cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
LINK SP-CL Hip System PoroLink (microporous) and HX (CaP) coated & LINK LCU Hip System PoroLink (microporous) and HX (CaP) coated (K161840)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 13, 2022
Waldemar Link GmbH & Co. KG % Pia Müller Regulatory Affairs Manager Waldemar Link GmbH & Co. KG Oststraße 4-10 Norderstedt, Germany 22844
Re: K213770
Trade/Device Name: SP-CL Hip Stem and LCU Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: August 18, 2022 Received: August 18, 2022
Dear Pia Müller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
### 510(k) Number (if known) K213770
#### Device Name
SP-CL Hip Stem and LCU Hip System
#### Indications for Use (Describe)
The SP-CL Hip Stem and LCU Hip System are indicated for patients with mobility-limiting hip diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures.
The hip systems are indicated for the following conditions:
- Primary and secondary coxarthrosis
- Osteoarthritis
- Necrosis of the femoral head
- Femoral neck fractures
The stems are indicated for cementless use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Waldemar Link GmbH & CO. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration #:3003386935 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Waldemar Link GmbH & Co. KG<br>Pia Müller (Regulatory Affairs Manager)<br>Oststraße 4-10<br>Norderstedt, Germany 22844<br>Phone: +49-40 53995-841<br>Fax: +49-40 53995-174<br>E-Mail: p.mueller@linkhh.de |
| Date Prepared: | August 17th, 2022 |
| Trade Name: | SP-CL Hip Stem and LCU Hip System |
| Common Name: | Hip Prosthesis |
| Classification Name: | Prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous; 21 CFR 888.3353, product code<br>LZO<br><br>Prosthesis, hip, semi-constrained, uncemented, metal/polymer,<br>non-porous, calcium-phosphate; 21 CFR 888.3353, product code MEH |
| Classification and Panel: | Class II, Orthopedic / 87 |
| Predicate Devices: | Waldemar Link; LINK SP-CL Hip System<br>PoroLink (microporous) and HX (CaP) coated & LINK<br>LCU Hip System PoroLink (microporous) and HX (CaP) coated, K161840 |
{4}------------------------------------------------
Device Description: The submission includes a straight stem, LCU Hip Sytem, and an anatomically curved stem, SP-CL Hip Stem. Both stems are comprised of titanium alloy (Ti6A14V acc. to ISO 5832-3, ASTM F136) available with HX (CaP) (ASTM F1609) coating. The stems are available in a range of sizes, lengths, and offsets. In total hip replacement, the SP-CL Hip Stem and LCU Hip System are combined with a corresponding prosthesis head and the acetabular implant components manufactured by Waldemar Link. The change that is subject of this 510(k) is to add calcium phosphate coating (CaP) done by Waldemar Link inhouse to the above listed systems. The devices are already commercially available with CaP coating applied by an external vendor. There is no change to the fundamental scientific technology of the referenced hip systems with the modifications in this 510(k) submission. This includes no changes to materials, design, sterilization, packaging, or method of manufactured. All components are sterile and for single use only. Indications for Use: The SP-CL Hip Stem and LCU Hip System are indicated for patients with mobility-limiting hip diseases, fractures which cannot be treated by conservative or defects or osteosynthetic procedures. The hip systems are indicated for the following conditions: -Primary and secondary coxarthrosis Osteoarthritis -Necrosis of the femoral head --Femoral neck fractures The stems are indicated for cementless use only. Comparison to the predicate: The SP-CL Hip Stem and LCU Hip System with HX (CaP) inhouse coating are substantially equivalent to LINK SP-CL and LINK LCU Hip System PoroLink (microporous) and HX (CaP) coated (K161840).
> The only change of the hip systems is the HX (CaP) coating applied inhouse, which is substantially equivalent to the HX (CaP) coating applied by the external vendor. Features comparable to the predicate devices include the same
{5}------------------------------------------------
indications, dimensions, materials, packaging, sterilization, surgical implantation technique and intended use.
| Performance Testing: | Non-clinical performance testing and analysis were provided, including: |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <ul><li>- Shear Testing (ASTM F1044-05)</li><li>- Tension Testing (ASTM F1147-05)</li><li>- Stereological Evaluation (ASTM F1854-15)</li><li>- Infrared Spectroscopy</li><li>- Dissolution/ Solubility Rate (ASTM F1926/F1926M-14)</li><li>- Elemental Analysis (ASTM F1609-08)</li><li>- Chemical Analysis (ASTM F1609-08)</li><li>- Shelf life</li><li>- Endotoxin testing</li></ul> |
| | The results of non-clinical performance testing demonstrate that the device is suitable for its intended purpose and substantially equivalent to the predicates. |
| Clinical Testing: | Clinical performance testing was not required to demonstrate the substantial equivalence of this device. |
| Conclusion: | The subject devices SP-CL Hip Stem and LCU Hip System with HX (CaP) inhouse coating are substantially equivalent to the predicate device identified in this premarket notification. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.