K213768 · Circinus Medical Technology, LLC · OWB · Dec 1, 2022 · Radiology
Device Facts
Record ID
K213768
Device Name
Bolt Navigation System
Applicant
Circinus Medical Technology, LLC
Product Code
OWB · Radiology
Decision Date
Dec 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Bolt Navigation System assists in the accurate placement of pedicle screws when used in conjunction with an intraoperative fluoroscope. It utilizes intraoperative fluoroscopic and pre- operative MRI or CT axial images to provide surgical planning and navigational telemetry relative to gravity, based on a fixed entry point ascertained by the user and validated by intraoperative fluoroscopic imaging. It is not intended to track patient position. The System is indicated for open and minimally invasive pedicle screw placement using a posterior approach in the thoracolumbar and sacral spine (T-9 to S1) where the patients' relevant rigid anatomical structures can be clearly identified on the imaging.
Device Story
BNS provides navigational guidance for pedicle screw placement; utilizes preoperative MRI/CT axial images and intraoperative fluoroscopy. System hardware includes Bolt Navigation Unit (BNU) (iPod touch with proprietary software), single-use case, and sterile drape. Surgeon selects entry point; system calculates angular orientation of surgical instruments (e.g., probe, awl) relative to gravity and planned trajectory. BNU screen attaches to instrument, allowing simultaneous viewing of surgical site and navigation data to reduce attention shift. Used in OR by surgeons. Benefits include improved accuracy of screw placement. Does not track patient position.
Clinical Evidence
Bench testing included sterilization, biocompatibility, electrical safety (IEC 60601-1/1-2), and usability (IEC 62366-1). Accuracy validated via phantom testing (mean 0.35°) and a multi-surgeon cadaveric study (T9-S1). Cadaveric results showed mean accuracy error of 1.59° (95% CI: 1.31°-1.86°), statistically significantly lower than the 3° threshold. Supplementary clinical study data provided.
Technological Characteristics
System comprises iPod touch-based BNU, single-use case, and sterile drape. Sensing principle: gravity-based orientation tracking linked to preoperative/intraoperative imaging. Connectivity: standalone mobile device. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62366-1, ASTM F2554-10. Software: proprietary navigation application.
Indications for Use
Indicated for patients requiring open or minimally invasive pedicle screw placement via posterior approach in the thoracolumbar and sacral spine (T-9 to S1), provided relevant rigid anatomical structures are identifiable on imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 1, 2022
Circinus Medical Technology, LLC % Alex Cadotte Associate Director, Software & Digital Health MCRA, LLC 803 7th St., NW, 3rd Floor WASHINGTON DC 20001
Re: K213768
Trade/Device Name: Bolt Navigation System ("BNS") Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB Dated: October 18, 2022 Received: October 27, 2022
Dear Alex Cadotte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
2022.12.01
11:54:38
-05'00'
*Lu Jiang*
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213768
Device Name Bolt Navigation System ("BNS")
#### Indications for Use (Describe)
The Bolt Navigation System assists in the accurate placement of pedicle screws when used in conjunction with an intraoperative fluoroscope. It utilizes intraoperative fluoroscopic and pre- operative MRI or CT axial images to provide surgical planning and navigational telemetry relative to gravity, based on a fixed entry point ascertained by the user and validated by intraoperative fluoroscopic imaging. It is not intended to track patient position. The System is indicated for open and minimally invasive pedicle screw placement using a posterior approach in the thoracolumbar and sacral spine (T-9 to S1) where the patients' relevant rigid anatomical structures can be clearly identified on the imaging.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | K213768 | | | |
|--------------------|-------------------------------------------------------------------------------------------------------------|--|--|--|
| Device Trade Name: | Bolt Navigation System (“BNS”) | | | |
| Common Name: | Image-intensified fluoroscopic X-ray system | | | |
| Manufacturer: | Circinus Medical Technology, LLC<br>4450 Arapahoe Avenue<br>Suite 1000<br>Boulder, CO 80303 | | | |
| Contact: | Mr. Patrick West<br>Circinus Medical Technology, LLC<br>Phone: (207) 651-1334<br>Email: patrick@boltnav.com | | | |
| Prepared by: | MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: (202) 552-5800 | | | |
| Date Prepared: | Oct 25, 2021 | | | |
| Classification: | 21 CFR §892.1650 | | | |
| Class: | II | | | |
| Product Codes: | OWB | | | |
### Indications for Use:
The Bolt Navigation System assists in the accurate placement of pedicle screws when used in conjunction with an intraoperative fluoroscope. It utilizes intraoperative fluoroscopic and preoperative MRI or CT axial images to provide surgical planning and navigational telemetry relative to gravity, based on a fixed entry point ascertained by the user and validated by intraoperative fluoroscopic imaging. It is not intended to track patient position. The System is indicated for open and minimally invasive pedicle screw placement using a posterior approach in the thoracolumbar and sacral spine (T-9 to S1) where the patients' relevant rigid anatomical structures can be clearly identified on the imaging.
### Device Description:
The BNS is comprised of the Bolt Navigation Unit (BNU) (an iPod touch® mobile digital device with the Bolt navigation software loaded on it), the Bolt single use case, and sterile drape. The
{4}------------------------------------------------
BNS is intended to provide navigational guidance during spine surgery. The system uses preand perioperative imaging data, and input from the surgeon via the BNU touchscreen to construct the proper angular position of the instrumentation and implants relative to gravity, and communicates this information to the surgeon via the BNU screen attached to the instrument allowing the surgeon to look at both the surgical site and the navigation data at the same time, thus attenuating the risk of attention shift.
The BNS provides guidance data by displaying the angular orientation of a surgical instrument (such as a pedicle probe or awl) relative to a surgeon selected entry point on the patient and gravity. Angular orientation of the instruments is linked to the imaging data via the BNS.
The system is intended to be used for both image fusion and navigation for spine surgery applications where reference to relevant rigid structures can be identified relative to a perioperative image data of the anatomy and the gravity vector.
### Predicate Device:
The Track X Technology, TrackX Device (K200360) serves as the primary predicate device. The Augmedics xvision Spine System (K190929) and the Medtronic StealthStation (K133444, K170011) serve as reference devices.
#### Technological Characteristics Comparison:
The BNS and its primary predicate technology are designed to aid the physician in positioning surgical instruments and placing implants by providing visual feedback on the movement of instruments relative to imaging data. Both systems use software platforms that allow for the surgeon to track the surgical instruments relative to the perioperative fluoroscopic images.
Both systems enable the surgeon to view the orientation of surgical instruments relative to that planned by the surgeon on preoperative image data while performing the surgical procedure with the aid of a fluoroscope. The systems provide guidance data by tracking and displaying orientation of the Spinal Instruments relative to a surgeon identified entry point and the gravity vector. Tracking data is linked to the preoperative imaging data.
Both systems are designed to work effectively with a range of 3rd party surgical instruments.
The BNS and the reference devices are designed to provide trajectory orientation guidance to the surgeon relative to relevant rigid anatomical structure with a performance goal of < 3°. These systems use software platforms that allow for the surgeon to match intraoperative trajectory to pre- and perioperative images and planning based on a surgeon identified entry-point. Both systems enable the surgeon to view the angular orientation of surgical instruments relative to that planned by the surgeon on preoperative image data while performing the surgical procedure. The BNS provides navigational guidance in three-degrees of freedom relative to a chosen entry point and the gravity vector while the reference devices (K190929, K133444, K170011), provide navigational guidance relative to a frame attached to the patient in six-degrees of freedom.
{5}------------------------------------------------
#### Nonclinical Testing:
Verification/Validation activities and performance testing have been conducted to support that the subject device meets the performance requirements under the indications for use conditions and to support substantial equivalence to the predicate devices, including the following:
- Sterilization, Cleaning, and Shelf-life Validation ●
- Biocompatibility Evaluation ●
- Non-Clinical Software Documentation and Verification
- Non-Clinical Cybersecurity Evaluation
- Electrical safety and EMC testing (IEC 60601-1 and IEC 60601-1-2) ●
- Usability Validation testing per IEC 62366-1: 2015, IEC 62366-2ANSI/AAMA HE75 ● and FDA Guidance for Industry and FDA Staff, including a Surgeon- Performed Cadaveric Study
- Non-Clinical Accuracy Testing per ASTM F2554-10 ●
- Cadaveric Trajectory Accuracy Study, based on a protocol that was discussed and reviewed by FDA via Submission Issue Request on July 12, 2022(Q220500/S001)
Clinical phantom accuracy results
| Mean | Std.<br>Deviation | 95% CI<br>of Mean | 95% CI<br>of Individuals | 99% CI<br>of Mean | 99% CI<br>of Individuals |
|-------|-------------------|-------------------|--------------------------|-------------------|--------------------------|
| 0.35° | 0.20° | 0.39° | 0.69° | 0.41° | 0.82° |
CI (Confidence Interval)
Cadaveric accuracy results
Planned trajectory vs actual placement accuracy (< 3 ° / 95% CI)
| | Estimate | 2-sided<br>95% CI LB | 2-sided<br>95% UB | 1-sided<br>95% UB |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------------------|-------------------|-------------------|
| Parametric* | 1.59 | 1.31 | 1.86 | 1.81 |
| Non-Parametric+ | 1.78 | 1.62 | 2.12 | 2.07 |
| Notes:<br>* Overall mean accuracy error estimate and 2-sided 95% CI and 1-sided 95% CI based on t-student<br>distribution.<br>+ Overall median accuracy error estimate and non-parametric distribution-free (bootstrap) 2-sided 95%<br>CI and 1-sided 95% CI.<br>Testing included levels T9 to S1. | | | | |
CI (Confidence Interval)
The analysis demonstrates that source-specific variability is small and the overall angle accuracy performance with a mean of 1.59° with a 95% CI of being statistically significantly lower than 3 degrees.
{6}------------------------------------------------
## Clinical Testing:
Circinus has conducted a multi-surgeon clinical study. A summary report for this study has been included at the request of the FDA and is intended as supplementary information for the Agency's reference and to provide the Agency further evidence that the subject device performs as intended in its intended clinical application.
#### Conclusion:
The BNS is substantially equivalent to the predicate device(s) with respect to its indications for use, design, and function.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.