← Product Code [FLE](/productcode/FLE) · K213758

# Lexa PLUS RIS-311 (K213758)

_W&H Sterilization S.R.L. · FLE · Aug 19, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213758

## Device Facts

- **Applicant:** W&H Sterilization S.R.L.
- **Product Code:** [FLE](/productcode/FLE.md)
- **Decision Date:** Aug 19, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6880
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

RIS-311 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a healthcare provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of medical and dental instruments that are validated to be sterm. The RIS-311 has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

## Device Story

Lexa PLUS RIS-311 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer. Device uses pressurized steam to sterilize medical and dental instruments. Operated by healthcare providers in clinical settings. Features multiple programs (Wrapped, Low Temperature, Unwrapped) with specific sterilization times, temperatures, and drying cycles. Device ensures sterilization of validated instruments; prevents damage by excluding liquid loads and biomedical waste. Provides effective sterilization for patient-ready instruments.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Dynamic-air-removal (pre-vacuum) table-top steam sterilizer. Operates at 250°F (121°C) or 270°F (132°C). Supports various load types including pouched instruments, dental handpieces, cassettes, and textiles. Maximum load capacity up to 18 lbs.

## Regulatory Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 19, 2022

W&H Sterilization s.r.l. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172

Re: K213758

Trade/Device Name: Lexa PLUS RIS-311 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: August 16, 2022 Received: August 17, 2022

Dear Angela Blackwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K213758

Device Name

Lexa PLUS RIS-311

## Indications for Use (Describe)

RIS-311 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a healthcare provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of medical and dental instruments that are validated to be sterm. The RIS-311 has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.

Key program features, including sterilization time, temperature and recommended load type are listed in the following table:

| Program         | Type of Load and Load<br>Weight                                                                                                                           | Sterilization<br>Temperature | Sterilization Time | Drying Time<br>(Recommended) |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------------------|------------------------------|
| Wrapped         | Pouched Instruments<br>and Dental Handpieces<br>on trays 14 lbs (6.3 kg)<br>(trays excluded).                                                             | 270 °F (132 °C)              | 4 minutes          | 22 minutes                   |
|                 | Instruments and dental<br>handpieces in wrapped<br>cassettes or in trays and<br>wrapped cassettes 16.5<br>lbs (7.5 kg) (trays and<br>cassettes included). |                              |                    |                              |
|                 | Pouched Textile load on<br>trays 4.4 lbs (2.0 kg)<br>(trays excluded).                                                                                    |                              |                    |                              |
| Low Temperature | Pouched Instruments<br>(solid instruments) on<br>trays 14 lbs (6.3 kg)<br>(trays excluded).                                                               | 250 °F (121 °C)              | 30 minutes         | 28 minutes                   |
|                 | Solid instruments in<br>wrapped cassettes or in<br>trays and wrapped<br>cassettes 16.5 lbs (7.5<br>kg) (trays and cassettes<br>included).                 |                              |                    |                              |
|                 | Pouched Textile load on<br>trays 4.4 lbs (2.0 kg)<br>(trays excluded).                                                                                    |                              |                    |                              |
| Unwrapped       | Instruments and dental<br>handpieces, up to 18 lbs<br>(8.1 kg).                                                                                           | 270°F (132°C)                | 3.30 minutes       | 4 minutes                    |

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K213758](https://fda-staging.innolitics.com/device/K213758)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com