The G7 dental chair with operative unit is intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental chair and attached dental devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
Device Story
G7 dental chair and operative unit; provides user interface for dental practitioner to control chair and attached instruments. Inputs: electricity, air, water, vacuum. Outputs: mechanical chair positioning (headrest, backrest, height), instrument delivery (3-way syringe, saliva ejector), and illumination. Used in dental operatory; operated by dentist/staff. Features hydraulic chair system, programmable positions, and safety switches. Benefits: provides intuitive control center for patient treatment procedures.
Clinical Evidence
Bench testing only. Compliance verified with ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 80601-2-60 (dental equipment safety), and ISO 7494-1/2 (dental unit performance). Software verification and validation testing performed per FDA guidance.
Technological Characteristics
Dental operative unit with hydraulic chair system. Materials include aluminum (DCI components). Power: AC110V/115V, 50/60Hz. Connectivity: wireless foot controller (PCB MPU boards). Standards: ISO 7494-1/2, IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60. Sterilization: N/A (surface disinfection).
Indications for Use
Indicated for use in general dental applications for patients requiring dental diagnosis, treatment, or operations. Used by dental practitioners to control dental chair and attached devices.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 3, 2023
Shinhung Company Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Suite 610 Houston, Texas 77054
Re: K213749
Trade/Device Name: G7, Dental Operative Unit and Accessories Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit And Accessories Regulatory Class: Class I, reserved Product Code: EIA Dated: March 6, 2023 Received: March 6, 2023
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213749
Device Name G7 Dental System
### Indications for Use (Describe)
The G7 Dental System and accessories are intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental chair and attached dental devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K213749
510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
| Date 510k summary prepared: | 3/17/2022 |
|-----------------------------|----------------------------------------------------------------------------------|
| I.<br>Submitter | |
| Submitter's Name : | Shinhung Co., Ltd. |
| Submitter's HQ Address: | Shinhung Bldg, 450, Cheongpa-ro, Jung-gu,<br>Seoul, 100-858, Korea |
| Factory Address: | 42-25, 27beon-gil, Dongsan-ro, Danwon-gu, Ansan-si<br>Gyeonggi-Do, 425-852 Korea |
| Submitter's Telephone: | +82 (2) 6366-2175 |
| Contact person: | Mr. Sam Su Kyong <sskyong@shinhung.co.kr> /<br>RA Manager |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| (U.S. Designated agent) | Mtech Group |
| Address: | 7505 Fannin St. Ste 610, Houston, TX 77054 |
| Telephone: | +713-467-2607 |
#### II. Subject Device
| Trade/proprietary name | : G7 |
|------------------------|-----------------------------------------|
| Regulation Name | : Dental Operative Unit and Accessories |
| Regulation Number | : 21 CFR 872.6640 |
| Product Code | : EIA |
| Regulatory Class | : Class I |
{4}------------------------------------------------
Prescription Use.
#### III. Predicate Device
| 510(k) Number | K151996 (Jan. 22, 2016) |
|----------------------|---------------------------------------|
| Primary Manufacturer | SHINHUNG. CO,.LTD |
| Device Name | TAURUS C1 |
| Regulation Name | Dental Operative Unit and Accessories |
| Regulation Number | 21 CFR 872.6640 |
| Product Code | EIA |
| Regulatory Class | Class I |
| Prescription Use | |
This predicate has not been subject to a design-related recall.
#### Reference Device IV.
| 510(k) Number | K032756 (Dec. 04, 2003) |
|----------------------|---------------------------------------|
| Primary Manufacturer | A-DEC, Inc |
| Device Name | A-DEC 532/533 |
| Regulation Name | Dental Operative Unit and Accessories |
| Regulation Number | 21 CFR 872.6640 |
| Product Code | EIA |
| Regulatory Class | Class I |
Prescription Use
#### V. Device Description
The G7 dental chair with operative unit is a dental chair with an operative unit attached for patient
{5}------------------------------------------------
to sit during the dental diagnosis, treatment and /or operation. The dental chair has adjustable headrest, backrest, seat height and armrests. The proposed device also consists of dentist table with instrument holders, assistant table, water unit, arm system, cuspidor (spittoon), dental light and foot control. The instrument holder includes a 3-way syringe and salvia ejector supplied by DCI which are cleared by FDA (K151987).
#### Indications for Use VI.
The G7 dental chair with operative unit is intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental chair and attached dental devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
| Model | TAURUS C1 (K151996) | G7 (K213749) | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The TAURUS C1 Dental<br>System and accessories are<br>intended for use in general<br>dental applications by<br>providing the dental<br>practitioner a user interface to<br>control operation of the dental<br>chair and attached dental<br>devices. The system delivers<br>air, water, vacuum and<br>electricity to allow the dental<br>practitioner an intuitive<br>control center for all common<br>and normal patient treatment<br>procedures performed in the<br>dental operatory. | The G7 dental chair with<br>operative unit is intended for<br>use in general dental<br>applications by providing the<br>dental practitioner a user<br>interface to control operation of<br>the dental chair and attached<br>dental devices. The system<br>delivers air, water, vacuum and<br>electricity to allow the dental<br>practitioner an intuitive control<br>center for all common and<br>normal patient treatment<br>procedures performed in the<br>dental operatory. | |
| Construction | CHAIR / UNIT / DOCTOR<br>TABLE /ASSISTANT<br>TABLE / OPERATING<br>LIGHT / FOOT CONTROL /<br>STOOL | CHAIR /UNIT/ DOCTOR<br>TABLE /ASSISTANT<br>TABLE / OPERATING<br>LIGHT / FOOT CONTROL /<br>STOOL | |
| Power Supply | AC110V/115V, 50/60Hz | AC110V/115V,50/60Hz | |
| | | | |
| Description | Power Shock<br>Protection | Class 1 | Class 1 |
| | Electric<br>Protection | B Type | B Type |
| | EMC Standard | EN 60601-1-2<br>IEC 61000-3-2<br>IEC 61000-3-3 | EN 60601-1-2<br>IEC 61000-3-2<br>IEC 61000-3-3 |
| | Dental units<br>Standard | ISO 7494-1<br>ISO 7494-2 | ISO 7494-1<br>ISO 7494-2 |
| | Dental unit&<br>chair particular<br>standard | - | IEC 80601-2-60 |
| | Safety Standard | IEC 60601-1 | IEC 60601-1 |
| | Chair Operating<br>system | Hydraulic system | Hydraulic system |
| | UNIT/CHAIR<br>form type | Over-arm Contour type | Over-arm Contour type |
| | Inlet Air<br>Pressure (Mpa) | 0.6 ~ 0.78MPa | 0.6 ~ 0.78MPa |
| | Inlet Water<br>Pressure (Mpa) | 0.2 ~ 0.59MPa | 0.2 ~ 0.59MPa |
| | Allowable<br>Patient Load | 150 Kgf | 150Kgf |
| | Vacuum system | Central Vacuum system or<br>Air Vacuum system | DCI vacuum system with HVE<br>and saliva ejectors (K151987) |
| | Flushing system | N/A | Included |
| | Water treatment<br>with ICX tablet | N/A | Included |
| | 3-Way Syringe | Adjusting the Water / Air /<br>Spray function in the syringe | Adjusting the Water / Air /<br>Spray function in the syringe<br>with DCI tip (K151987) |
| | Vacuum syringe<br>tip | Disposable tip | Not provided |
| | Vacuum syringe | POLY AMID, Durr | Aluminum, DCI |
| | Saliva-ejector | POLY AMID, Durr | Aluminum, DCI |
| | | | |
| | Low-speed /<br>High-speed<br>handpiece | Not provided | Not provided |
| | Scaler | Piezon kit, EMS | Piezon kit, EMS |
| | Cuspidor | Spittoon bowl / Rinsing pipe /<br>Tumbler filler | Spittoon bowl / Rinsing pipe /<br>Tumbler filler |
| | Water System | City Water / Distilled Water | City Water / Distilled Water |
| Operating<br>Light | Type | Halogen (Shin Hung) | EDI LED |
| | Light head<br>structure | 3-axis head adjustment<br>method | 3-axis head adjustment<br>method |
| | ON/OFF Control | Auto and manual ON/OFF | Auto and manual ON/OFF |
| | Brightness<br>control | 2 stage control | Continuous control |
| | Light pattern | 130mmX70mm @ 700mm | 130mmX70mm @ 700mm |
| | Loading<br>Capacity | 150kgf | 150kgf |
| Dental<br>Chair | Movement<br>Range (Chair) | 530 ~825mm(± 20 mm) | 400~760mm(± 20 mm) |
| | Movement<br>Range (Backrest) | $0° ~ 63° (± 3°)$ | $0° ~ 70° (± 3°)$ |
| | Headrest | Double-articulating headrest | Double-articulating headrest |
| | Table arm | Manual system | Air brake system |
| | Safety switch<br>function | BACKREST and auxiliary<br>arm, 2 safety switches | Rink cover and swing cover<br>safety switch, emergency stop<br>button located on the back of<br>the chair seat, 3 safety<br>switches. |
| | | The safety switch stops the<br>chair operation. | The safety switch stops the<br>chair operation. |
| | Programmable<br>chair position | Programmable chair position:<br>Total 5 positions | Programmable chair position:<br>Total 5 positions |
| Delivery<br>system | Handpiece<br>holder | 6 integrated rotation type<br>holder | 6 integrated rotation type<br>holder |
| | Handle | Fixed handle | Fixed handle |
| | Touchpad | Membrane panel | Membrane Panel |
| | Handpiece water | Manually adjustable | Manually adjustable handpiece<br>water / air cooling. |
| | | | |
| | Flexarm tension | Control the tension to adjust<br>the shift of TABLE ARM<br>rotation. | Control the tension to adjust<br>the shift of TABLE ARM<br>rotation. |
| Accessories | Instrument<br>holders | HVE, SE, 3-WAYsyringe<br>attached | HVE, SE, 3-WAYsyringe<br>attached |
| | Foot control | Standard, all-in-one | Standard, all-in-one, |
| | Cuspidor | Attach/detachable GLASS<br>disassemble for cleaning. | Magnetic cuspidor disassemble<br>for cleaning |
| | Solids collector | Equipped | Equipped |
| Support<br>center | Distilled water<br>reservoir | Equipped | Equipped |
| | Cup water<br>system | Automatic and manual water<br>supply system. | Automatic and manual water<br>supply system. |
| | | Watering time can be set by<br>the user. | Watering time can be set by<br>the user. |
| | | | |
| | | | |
| Operation<br>Method | Control panel / Assistant control panel / Foot controller | | |
VII. Comparison of Technological Characteristics with the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# VIII. Comparison of Technological Characteristics with the Reference Device
| Model | A-DEC 532/533 (K032756) | G7 (K213749) | |
|-------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Indications for<br>Use | The A-dec 532/533 Delivery<br>System Dental Unit and<br>accessories are intended for<br>use in general dental<br>applications by providing the<br>dental practitioner a user<br>interface to control operation<br>of the dental chair and<br>attached dental devices. The<br>system delivers air, water,<br>vacuum and electricity to<br>allow the dental practitioner<br>an intuitive control center for<br>all common and normal<br>patient treatment procedures<br>performed in the dental<br>operatory. | The G7 dental chair with<br>operative unit is intended for<br>use in general dental<br>applications by providing the<br>dental practitioner a user<br>interface to control operation of<br>the dental chair and attached<br>dental devices. The system<br>delivers air, water, vacuum and<br>electricity to allow the dental<br>practitioner an intuitive control<br>center for all common and<br>normal patient treatment<br>procedures performed in the<br>dental operatory. |
{9}------------------------------------------------
| | Construction | CHAIR / UNIT / DOCTOR<br>TABLE / ASSISTANT<br>TABLE / OPERATING<br>LIGHT / FOOT CONTROL /<br>STOOL | CHAIR /UNIT/ DOCTOR<br>TABLE / ASSISTANT<br>TABLE / OPERATING<br>LIGHT / FOOT CONTROL /<br>STOOL |
|--|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.