K213745 · Shenzhen Future Electronic Co., Ltd. · IRP · Mar 31, 2022 · Physical Medicine
Device Facts
Record ID
K213745
Device Name
Air Compression Therapy Device, model: ST-502
Applicant
Shenzhen Future Electronic Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Mar 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Air Compression Therapy Device(model: ST-502) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in good health. The Air Compression Therapy Device (model: ST-502) simulates kneading and stroking of tissues by using an inflatable garment.
Device Story
Air Compression Therapy Device (ST-502) is a powered inflatable tube massager for home, clinic, or athlete training use. Device consists of a control unit (air pump, pressure sensor, microprocessor) connected via hoses to leg sleeves with three compression chambers. User operates via interface buttons (Power, Intensity, Mode) to select therapy parameters. Device inflates/deflates chambers sequentially or peristaltically to simulate kneading/stroking of tissues. Microprocessor controls timing and pressure (0-240 mmHg) based on sensor feedback. Once target pressure is reached, device deflates via internal valve. Intended to provide temporary relief of minor muscle aches and pains and increase local circulation. Benefits include non-invasive muscle recovery and circulation support.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (ANSI/AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Powered inflatable tube massager; 3-chamber leg sleeves; ABS plastic enclosure; microprocessor-controlled air pump; 100-240V AC power; 0-240 mmHg pressure range; 3 modes (sequential/peristaltic); software-controlled timing/pressure; biocompatible sleeve materials (polyester/nylon); compliant with ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 10993, and IEC 62304.
Indications for Use
Indicated for adults in good health for temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas (low limbs).
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Air Compression Therapy Device, model: S9019 (K193354)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2022
Shenzhen Future Electronic Co., Ltd % You Yijie Manager Qimmiq Medical Consulting Service Co., Ltd. RM.1711, Building K NO.101 Science Ave International Creative Valley Guangzhou, Guangdong 510663 China
Re: K213745
Trade/Device Name: Air Compression Therapy Device, model: ST-502 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: February 2, 2022 Received: February 2, 2022
Dear You Yijie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K213745
Device Name
Air Compression Therapy Device, model: ST-502
Indications for Use (Describe)
The Air Compression Therapy Device(model: ST-502) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in good health. The Air Compression Therapy Device (model: ST-502) simulates kneading and stroking of tissues by using an inflatable garment.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 510(k) Summary — K213745
### 1. Submitter's Information
#### Establishment Registration Information
Name: SHENZHEN FUTURE ELECTRONIC CO., LTD. Address: Bldg B. Qiangchengda Industrial Park, No. 30, Youtian Road, Longgang SHENZHEN Guangdong, CN 518116
#### Contact Person of applicant
Name: Zhang Benrong Address: Room 301-302, No.15 HanchengRoad, Qingdao Free Trade Zone, Shandong, China, 266555 TEL: +86-13410947296 Email: zhaoqihong@futuresz.com.cn
#### Contact Person of the Submission:
Name: Yijie You Address: RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone, Guangzhou China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@qimmiq-med.com
Date prepared: Nov. 15, 2021
### 2. Device Information
| Trade Name: | Air Compression Therapy Device |
|----------------------|-----------------------------------|
| Model: | ST-502 |
| Classification name: | Massager, Powered Inflatable Tube |
| Review panel: | Physical Medicine |
| Product code: | IRP |
| Regulation Class: | II |
| Regulation Number: | 21CFR890.5650 |
### 3. Predicate Device Information
#### Predicate device:
510(k) submitter/holder: Shenzhen Dongjilian Electronics Co., Ltd.
{4}------------------------------------------------
| 510(K) Number: | K193354 |
|----------------------|-----------------------------------|
| Trade Name: | Air Compression Therapy Device |
| Model: | S9019 |
| Classification name: | Massager, Powered Inflatable Tube |
| Review panel: | Physical Medicine |
| Product code: | IRP |
| Regulation Class: | II |
| Regulation Number: | 21CFR890.5650 |
### 4. Device description
Air Compression Therapy Device, model: ST-502 is a powered inflatable tube massager. The Air Compression Therapy Device, model: ST-502 is consists of an air pump, air pressure sensor, leg sleeves and adapter. The air pump, air pressure sensor and control components are protectively housed in a plastic case of the control unit. The control unit is connected to the dedicated leg sleeves via a series of hoses; each leg sleeve has 3 compression chambers. The Air Compression Therapy Device, model: ST-502 utilizes the pneumatically controlled leg sleeves actuated by an electronically controlled air pump unit.
The "Power" button, "Intensity" button, "Mode" button and LED indicators/displays provide for user interface. There is also a port for connecting the AC adapter plug.
The leg sleeves consist of air pipes and three airbags encased inside a soft medical fabric for increased patient comfort and biocompatibility compliance.
In operation, the user simply turns the power on via the "Power" button. And the user presses the "Intensity" button, "Mode" button to adjust the pressure intensity and working mode of the device as the needs of users.
The compression massage direction is from foot room and push up the thighs. By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The leg sleeves work under the action of sensor and microprocessor. Software controls the timing and pressure reflected by the sensor, cycling airflow into and out of the sleeves to compress body.
During in the working condition, once the leg sleeves pressure of the airbags reach the proper level(240mmHg), the device is turned off, and the leg sleeves deflates to the ambient pressure through a valve inside the plastic case.
Environment of use of the device: Clinics, hospital, athlete training, and home environments
#### Principle of operation:
The control unit is connected to the leg sleeves via hose. The control unit contains an air pump, pressing "Power" button to begin normal treatment status, the air pump will pump air into leg sleeves with stetted rhythm though air pipe, the sleeve fits on the low limbs and provide therapy.
### 5. Indications for Use
The Air Compression Therapy Device (model: ST-502) is indicated for the temporary relief of minor
{5}------------------------------------------------
muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Compression Therapy Device (model: ST-502) simulates kneading and stroking of tissues by using an inflatable garment.
## 6. Summary of technological characteristics of device compared to the
| SE<br>Comparisons | Subject device<br>(Air Compression Therapy<br>Device,<br>model: ST-502) | Primary predicate device<br>(Shenzhen Dongjilian<br>Electronics Co.,Ltd.<br>Air Compression Therapy Device,<br>Model: S9019) | Discussion of<br>difference | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| 510K Number | K213745 | K193354 | / | |
| Classification | 21 CFR890.5650 | 21 CFR890.5650 | Same | |
| Product Code | IRP | IRP | Same | |
| FDA Class | II | II | Same | |
| Indications for Use | The Air Compression Therapy<br>Device (model: ST-502) is indicated<br>for the temporary relief of minor<br>muscle aches and pains and for<br>temporary increase in circulation to<br>the treated areas in people who are<br>in good health. The Air<br>Compression Therapy Device<br>(model: ST-502) simulates kneading<br>and stroking of tissues by using an<br>inflatable garment. | The Air Compression Therapy<br>Device is indicated for the<br>temporary relief of minor muscle<br>aches and pains and for<br>temporary increase in circulation to<br>the treated areas in people who are<br>in good health. The Air<br>Compression Therapy<br>Device simulates kneading and<br>stroking of tissues by using an<br>inflatable garment. | Same | |
| Over-The-Counter<br>Use | Yes | Yes | Same | |
| Model | ST-502 | S9019 | / | |
| Treatment<br>area/Structure of<br>Sleeves | Low limbs (Foot, calf and upper leg) | Low limbs (Foot, calf and upper leg) | Same | |
| Environment of Use | Clinics, hospital, athlete training,<br>and home environments | Clinics, hospital, athlete training,<br>and home environments | Same | |
| Patient Population | Adult | Adult | Same | |
| Power source | 100-240V, 50-60Hz | 100-240V 50/60Hz | Same | |
| Power Consumption | 12w | 12W | Same | |
| SW/Firmware/<br>Microprocessor<br>Control | Microprocessor | Microprocessor | Same | |
| Therapy Time | 20 mins | 20 minutes | Same | |
| Output pressure<br>range | 0~240 mmHg | 0~240 mmHg | Same | |
| Air pressure Level<br>/Compression levels | 3 levels settings:<br>low level:150mmHg;<br>Medium level:185mmHg;<br>High Level: 215mmHg | 3 levels settings:<br>low level:150mmHg;<br>Mid level:185mmHg;<br>High Level: 215mmHg | Same | |
| Pressure error<br>range | $\pm$ 25mmHg | $\pm$ 25mmHg | Same | |
| Inflation time | 3-30s | 3-30s | Same | |
| Keep time | 1-5s | 1-5s | Same | |
| Deflation time | 1-5s | 1-5s | Same | |
| Mode types | Sequential/ Peristaltic | Sequential/ Peristaltic | Same | |
| Cycle time | 25 seconds to 3 minutes and 40<br>seconds | Range of 25 sec to 3 min 40 sec | Same | |
| Number of<br>chambers | 3 Chambers | 3 Chambers | Same | |
| Number of<br>treatment mode | 3 modes | 3 modes | Same | |
| Modes (visual<br>description) | M1:<br>Start in the foot room and push<br>up the thighs, the pressure<br>gradually rises to a<br>predetermined pressure level<br>with each compression, then<br>depressurizes and the pressure<br>drops. Once the thighs are<br>decompressed, the circulation<br>begins again.<br>Pattern 1 follow this stress<br>sequence:<br>Image: Pattern 1 stress sequence<br>M2:<br>Starting in the foot chamber,<br>push up the thighs, and with<br>each compression, the pressure<br>gradually rises to a<br>predetermined air pressure level,<br>holding the air until the entire<br>garment is compressed. All three<br>parts decompress | Mode 1:<br>Starting with the foot chamber and<br>progressing up the thigh chamber,<br>each section compresses and the<br>pressure gradually rises to the pre-<br>determined air pressure level, then<br>decompresses and the air pressure<br>drops.<br>Once the thigh section<br>decompresses, the cycle begins<br>again.<br>Mode 1 follows this pressure<br>sequence:<br>Image: Mode 1 pressure sequence<br>Mode 2:<br>Starting with the foot chamber and<br>progressing up the thigh, each<br>section compresses and the<br>pressure gradually rises to the pre-<br>determined air pressure level, holds<br>the air until the entire garment is<br>compressed. All three sections then | Same | |
| | | | | |
| | | simultaneously, air pressure<br>drops, and the circulation begins<br>again.<br>Pattern 2 follow this pressure<br>sequence:<br>Image: Pressure sequence diagram | decompress simultaneously and the<br>air pressure drops, then cycle begins<br>again.<br>Mode 2 follows this pressure<br>sequence<br>Image: Pressure sequence diagram | |
| | | M3: It consists of two phase.<br>Phase 1: It works as a pattern 1<br>method, and after phase 1 is<br>completed, it continues to phase 2<br>(working as a pattern 2 method)<br>until phase 2 is completed, and<br>then continues to the next cycle<br>without interruption.<br>The pressure sequence of mode 3<br>is combined with mode 1 and mode<br>2. | Mode 3:<br>include two stage, stage 1: it works<br>according to the method of mode 1,<br>after the stage 1 is completed, it go<br>to stage 2(working according to the<br>method of mode 2) without<br>interruption time until finish the stage<br>2, then enter next cycle without<br>interruption .<br>Mode1 $\rightleftharpoons$ Mode2<br>The pressure sequence of mode 3<br>combines mode 1 and mode 2. | |
| Noise level | $\leq$ 65db | | $\leq$ 65dB | Same |
| Sleeve Material | Material of memory cloth: 100%<br>polyester<br>material Velcro(loops):<br>nylon70%, polyester 30%<br>Velcro (hooks): 100% nylon | | Nylon with a Polyurethane laminate | Different<br>(Discussion is<br>indicated in D1) |
| Housing Materials | Molded ABS enclosure | | Molded ABS enclosure | Same |
| Patient contact | Non-conductive attachments | | Non-conductive attachments | Same |
| Size and<br>appearance | L215*W65*H51mm | | 10.2x5.9×25.6<br>(in ) | Different<br>(Discussion is<br>indicated in D2) |
| | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.