The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the illiac, femoral, iliofemoral, and popliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Device Story
Over-the-wire (OTW) balloon dilatation catheter for percutaneous transluminal angioplasty (PTA). Features nylon semi-compliant balloon with four embedded stainless steel serrated scoring elements. During inflation, serrated elements create linear, interrupted scoring on endoluminal surface to modify plaque and facilitate arterial expansion. Used by physicians in clinical settings for peripheral vascular interventions. Output is mechanical dilation of stenotic vessels. Benefits include improved vessel patency through controlled plaque modification.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included catheter delivery, deployment, retraction, balloon rated burst, balloon fatigue, catheter torsion, flex/kink, and tensile testing. Results demonstrated performance consistent with predicate and reference devices.
Technological Characteristics
OTW catheter; semi-compliant nylon balloon; 4 stainless steel serrated scoring elements; Pre-Fabricated Cover (PFC) top layer; urethane base layer. Sterilized via Ethylene Oxide gas. Nominal pressure 6 atm; rated burst pressure 10 atm. Dimensions: 4.0-6.0 mm diameter, 40-120 mm length, 150 cm effective length.
Indications for Use
Indicated for patients requiring dilatation of lesions in iliac, femoral, iliofemoral, and popliteal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for coronary or neuro-vasculature use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2021
Cagent Vascular, LLC Carol Burns CEO 150 Strafford Avenue #315 Wayne, Pennsylvania 19087
Re: K213728
Trade/Device Name: Serranator® PTA Serration Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PNO Dated: November 24, 2021 Received: November 26, 2021
Dear Carol Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213728
Device Name Serranator® PTA Serration Balloon Catheter
Indications for Use (Describe)
The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the illac, femoral, and poplited arteries for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Type of Use (Select one or both, as applicable):
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Cagent Vascular. The word "CAGENT" is written in large, dark blue letters. Below it, the word "VASCULAR" is written in smaller, teal letters. A semi-circle made of teal dashes surrounds the logo. The registered trademark symbol is located to the right of the word "CAGENT".
510(k) Summary [as required by 21 CFR 807.92(c)]
# Serranator® PTA Serration Balloon Catheter 510(k) K213728
| DATE PREPARED: | November 24, 2021 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | Carol Burns/ President & CEO<br>150 Strafford Avenue, #315<br>Wayne, PA 19087 |
| CONTACT: | Carol A. Burns, President and CEO<br>Phone: (610) 688-2006<br>Fax: (610) 688-2667<br>Email: cburns@cagentvascular.com |
| TRADE NAME: | Serranator® PTA Serration Balloon Catheter |
| COMMON NAME: | Percutaneous Catheter |
| CLASSIFICATION | 21CFR 870.1250 |
| DEVICE CLASS: | Class II |
| PANEL CODE: | PNO |
| PREDICATE DEVICE: | K163380 |
| REFERENCE DEVICE: | K203437 |
### INTENDED USE/INDICATIONS FOR USE:
The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the illiac, femoral, iliofemoral, and popliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
# DEVICE DESCRIPTION:
The Serranator® PTA Serration Balloon Catheter is an Over-The-Wire (OTW) balloon dilatation catheter designed to perform Percutaneous Transluminal Angioplasty (PTA) for peripheral indications as described in the Indication for Use statement. The Serranator® has a nylon semicompliant balloon with embedded external metal strips or scoring elements. The unique scoring elements are serrated, designed to modify the plaque by creating linear, interrupted scoring along
{4}------------------------------------------------
the endoluminal surface. This occurs during balloon inflation and is designed to aid arterial expansion.
# COMPARISON WITH PREDICATE AND REFERENCE DEVICES:
A detailed comparison of the subject device (Serranator) and the predicate device (K163380) along with the reference device K203437 is shown in the table below demonstrates that the technological characteristics of the subject device such as components, design, sterilization method, shelf life and operating principle are identical or similar to the predicate device (K163380).
{5}------------------------------------------------
| Table 1. Comparison with Predicate Device: |
|--------------------------------------------|
|--------------------------------------------|
| | Subject Device | Predicate Device<br>Serranator Alto1 PTA Serration Balloon Catheter<br>(Not Commercialized) | Reference Device<br>Serranator PTA Serration<br>Balloon Catheter<br>(Commercialized) |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Serranator® PTA Serration<br>Balloon Catheter | Same | Same |
| Manufacturer | Cagent Vascular, Inc. | Cagent Vascular, Inc. | Cagent Vascular, Inc. |
| 510(k) | - | K163380 | K203437 |
| Intended<br>Use/Indication for<br>Use | The Serranator® PTA<br>Serration Balloon Catheter<br>is intended for dilatation of<br>lesions in the iliac, femoral,<br>iliofemoral, and popliteal,<br>arteries for the treatment of<br>obstructive lesions of native<br>or synthetic arteriovenous<br>dialysis fistulae. Not for use<br>in the coronary or neuro-<br>vasculature. | The Serranator® Alto PTA Serration Balloon Catheter is<br>intended for dilatation of lesions in the iliac, femoral,<br>iliofemoral, and popliteal arteries and for the treatment of<br>obstructive lesions of native or synthetic arteriovenous<br>dialysis fistulae. Not for use in the coronary or neuro-<br>vasculature. | The Serranator® PTA<br>Serration Balloon<br>Catheter is intended for<br>dilatation of lesions in<br>infrapopliteal arteries. Not<br>for use in the coronary or<br>neuro-vasculature. |
| Regulation | 21CFR870.1250 | 21CFR870.1250 | 21CFR870.1250 |
| Regulation Name | Percutaneous catheter | Percutaneous catheter | Percutaneous catheter |
| Regulation Class | Class II | Class II | Class II |
| Product Code | PNO | PNO | PNO |
| Prescription/OTC | Prescription | Prescription | Prescription |
| Catheter Design | OTW | OTW | OTW |
| Balloon diameter<br>(mm) | 4.0, 5.0, 6.0 | 4.0, 5.0, 6.0 | 2.5, 3.0, 3.5 |
| Balloon length (mm) | 40, 120 | 40, 80, 120 | 40, 80, 120 |
| Number of Strips | 4 | 4 | 3 |
| Strip material | Stainless Steel | Stainless Steel | Stainless Steel |
| Strip coating | Plasma with Silane | n.a. | Plasma with Silane |
| Effective Length (cm) | 150 | 150 | 150 |
| Balloon Compliance | Semi Compliant | Semi Compliant | Semi Compliant |
| | Subject Device | Predicate Device<br>Serranator Alto1 PTA Serration Balloon Catheter<br>(Not Commercialized) | Reference Device<br>Serranator PTA Serration<br>Balloon Catheter<br>(Commercialized) |
| Balloon top layer | Pre-Fabricated Cover<br>(PFC) | Urethane (dip coat) | Pre-Fabricated Cover<br>(PFC) |
| Balloon base layer | Urethane | Urethane | Urethane |
| Balloon base | Nylon | Nylon | Nylon |
| Sterilization Method | Ethylene Oxide Gas | Ethylene Oxide Gas | Ethylene Oxide Gas |
| Nominal Pressure<br>(atm) | 6 | 6 | 6 |
| Hydrophilic Coating | No | Yes | No |
| Rated Burst Pressure<br>(atm) | 10 | 12 | 12 |
| Packaging | Tyvek® Pouch and Carton<br>Packaging Hoop/Hoop<br>Attachment | Tyvek® Pouch and Carton<br>Packaging Hoop/Hoop Attachment | Tyvek® Pouch and Carton<br>Packaging Hoop/Hoop<br>Attachment |
{6}------------------------------------------------
{7}------------------------------------------------
# NON-CLINICAL TESTING/PERFORMANCE DATA:
The following bench tests were performed based on a risk assessment of the subject device and its alignment with the cleared predicate (K163380) and the reference device (K203437) catheter design, materials, and processes,
Non-clinical bench testing was performed on the subject device to determine substantial equivalence. The following tests were performed:
- . Catheter Delivery. Deployment and Retraction
- . Balloon Rated Burst Test
- Balloon Fatigue
- . Catheter Torsion
- . Catheter Flex/Kink
- . Catheter Tensile Test
In vitro bench testing demonstrated that the subject device performed as intended and similar to the predicate device.
### BIOCOMPATIBILITY:
Biocompatibility testing was not repeated for the subject device as there were no new materials or manufacturing process changes as compared to the identical fabrication methods of the previously cleared reference device K203437.
#### SHELF LIFE:
Shelf life testing was not repeated for the subject device as there were no new materials or manufacturing process changes as compared to the identical fabrication methods of the previously cleared reference device K203437. The testing for the subject device of K203437 indicated that there were no differences in performance of unaged and aged devices.
#### CONCLUSION:
The Serranator® PTA Serration Balloon Catheter included in this notification is substantially equivalent to the previously cleared predicate device in terms of intended use and technological characteristics as demonstrated by bench testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.