K213726 · AVAVA, Inc. · GEX · Jun 28, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213726
Device Name
PL-1 Skin Treatment System
Applicant
AVAVA, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PL-1 Skin Treatment System is a 1064 nm laser system intended to ablate, vaporize, and coagulate soft body tissues for aesthetic and cosmetic applications in the medical specialties of dermatology and plastic surgery. It is also indicated for removal of benign pigmented lesions including; nevus of Ota, café au lait spot, ephelides, solar lentigines), Becker's nevus, nevus spilus and treatment of common nevi. The PL-1 Skin Treatment System is intended to be used by medical professionals and staff who are of lasers and who are familiar with the technology, operation of the system, and safety precautions.
Device Story
1064nm Q-switched laser system; console, tablet, and handpiece components. Operator inputs treatment parameters via tablet touchscreen; places handpiece on skin; initiates laser energy via foot switch. Device uses selective photothermolysis to ablate, vaporize, and coagulate soft tissue. Used in clinical settings by trained medical professionals. Output is microbeam laser energy delivered to skin. Benefits include aesthetic/cosmetic tissue treatment and removal of benign pigmented lesions. Software controls system electronics, power, and laser delivery.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Verification and validation included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), software validation (IEC 62304), and biocompatibility (ISO 10993-1).
Technological Characteristics
1064nm Q-switched flash lamp pumped laser; Class 4; 850mJ max pulse energy; 6-12ns pulse width; 10Hz max repetition rate. Microbeam delivery (60 beams, 8mm spot). Touchscreen interface; foot switch trigger. Software-controlled. No sterile components. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for ablation, vaporization, and coagulation of soft body tissues for aesthetic/cosmetic applications in dermatology and plastic surgery; removal of benign pigmented lesions (nevus of Ota, café au lait spot, ephelides, solar lentigines, Becker's nevus, nevus spilus) and treatment of common nevi.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2022
Avava, Inc Jay Bhawalkar Chief Technology Officer 275 Second Avenue, Floor 3 Waltham, Massachusetts 02451
Re: K213726
Trade/Device Name: PL-1 Skin Treatment System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: May 25, 2022 Received: May 27, 2022
Dear Jay Bhawalkar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213726
Device Name PL-1 Skin Treatment System
#### Indications for Use (Describe)
The PL-1 Skin Treatment System is a 1064 nm laser system intended to ablate, vaporize, and coagulate soft body tissues for aesthetic and cosmetic applications in the medical specialties of dermatology and plastic surgery. It is also indicated for removal of benign pigmented lesions including; nevus of Ota, café au lait spot, ephelides, solar lentigines), Becker's nevus, nevus spilus and treatment of common nevi..
The PL-1 Skin Treatment System is intended to be used by medical professionals and staff who are of lasers and who are familiar with the technology, operation of the system, and safety precautions.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### General Provisions
| 510(k) Owner's Name: | AVAVA, Inc. |
|------------------------|-------------------------------------------------------------------------------------|
| Address: | 245 Second Avenue, Floor 3 Waltham, MA 02451 |
| Contact Person: | Jay Bhawalkar, Ph.D., Chief Technology Officer |
| Phone Number: | Office: 617-377-7961 |
| Fax Number: | Not applicable |
| Classification Name: | Laser Surgical Instrument for Use in General and Plastic<br>Surgery and Dermatology |
| Regulation: | 21 CFR 878.4810 |
| Regulatory Class: | II |
| Product Code: | GEX, ONG |
| Proprietary Name: | PL-1 Skin Treatment System |
| Common Name: | Powered Laser Surgical Instrument with Microbeam Output |
| Date Summary Prepared: | May 26, 2022 |
#### Name of Predicate Device(s)
- Primary Predicate: Quanta Chrome, K202503
#### Intended Use
The PL-1 Skin Treatment System is a 1064 nm laser system intended to ablate, vaporize, and coagulate soft body tissues for aesthetic and cosmetic applications in the medical specialties of dermatology and plastic surgery. It is also indicated for removal of benign pigmented lesions including; nevus of Ota, café au lait spot, ephelides, solar lentigines), Becker's nevus, nevus spilus and treatment of common nevi. The PL-1 Skin Treatment System is intended to be used by medical professionals and staff who are trained in the use of lasers and who are familiar with the technology, operation of the system, and safety precautions.
#### Device Description
The PL-1 Skin Treatment System is a 1064nm, Q-switched laser system. The system includes three (3) main components: Console, Tablet, and Handpiece. The Console houses the system control electronics, power distribution, contact cooling, and laser. The primary user interface for controlling the system through a touch screen graphical user interface. The Handpiece contains the optics, focusing components, and reference window assembly.
The PL-1 Skin Treatment System is a software-controlled device. The operator enters treatment parameters on the Tablet and places the Handpiece on the treatment site. A treatment is initiated by the operator to cause laser energy to be projected into the skin of a patient. The device is intended to be used by a suitably trained professional in a clinical setting. There are no sterile components.
{4}------------------------------------------------
#### Technological Characteristics
The PL-1 Skin Treatment System has the same intended use and similar indications for use, technological characteristics and operating principles as the Quanta Chrome system (Quanta systems) operating in the 1064 Q-switched, nanosecond mode with a microbeam handpiece. The design and components are very similar to the Quanta systemas shown in the following table. The differences are minor and do not raise any new issues of safety or efficacy of the PL-1 Skin Treatment System device as shown in the performance testing results.
| Characteristic | Subject Device | Primary Predicate | Substantial<br>Equivalence? |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | PL-1 Skin Treatment System | Chrome | NA |
| Manufacturer | AVAVA | Quanta | NA |
| 510k | K213726 | K202503 | NA |
| Classification | 878.4810 Laser surgical<br>instrument for use in general<br>and plastic surgery and in<br>dermatology | 878.4810 Laser surgical<br>instrument for use in general<br>and plastic surgery and in<br>dermatology | Yes, same classification |
| Product code | GEX, ONG | GEX | Yes, same code |
| Device class | II | II | Yes, same class |
| Intended use | Aesthetic and cosmetic<br>applications requiring ablation,<br>vaporization and coagulation of<br>body soft tissues in the medical<br>specialties of dermatology and<br>plastic surgery | Aesthetic, cosmetic and surgical<br>applications requiring incision,<br>excision, ablation, vaporization and<br>coagulation of body soft tissues in<br>the medical specialties of<br>dermatology, general, plastic and<br>oral surgery | Yes, same intended use<br>for aesthetic and cosmetic<br>applications |
| Indications | The PL-1 Skin Treatment system<br>is indicated for the removal of<br>benign pigmented lesions<br>including; nevus of Ota, café au<br>lait spot, ephelides, solar<br>lentigo (lentigines), Becker's<br>nevus, nevus spilus, and<br>treatment of common nevi. | The Quanta Chrome system<br>operating in 1064nm<br>nanosecond mode with<br>microbeam handpiece is<br>indicated for: Removal of dark<br>ink (black, blue and brown)<br>tattoos. Removal of benign<br>pigmented lesions including;<br>nevus of Ota, Café au lait spot,<br>Ephalides, solar lentigo<br>(lentigines), Becker Nevus,<br>Nevus spilus. Treatment of<br>common nevi. Removal or<br>lightening of unwanted hair Skin<br>resurfacing procedures for the<br>treatment of acne scars and<br>wrinkles. | Yes, same indication<br>for use for BPL based<br>on equivalence in<br>operating principle,<br>laser type, wavelength,<br>treatment parameters<br>and demonstrated<br>performance to<br>specifications |
| Characteristic | Subject Device | Primary Predicate | Substantial<br>Equivalence? |
| Operating Principles | | | |
| Focused<br>microbeams | Yes | Yes | Yes |
| Mechanism | Selective photothermolysis | Selective photothermolysis | Yes |
| Laser | | | |
| CDRH Laser Classification | Class 4 | Class 4 | Yes |
| Q-switched flash lamp<br>pumped laser | Yes | Yes | Yes |
| Wavelength (nm) | 1064 | 1064 | Yes |
| Maximum pulse energy<br>(mJ) | 850 | 800 | Yes, difference of 6% is<br>well under acceptable<br>tolerance of 20% for<br>single system per IEC<br>60601-2-22 |
| Nominal pulse width (ns) | 6 to 12 | 6 to 12 | Yes |
| Pulse repetition rate,<br>maximum (Hz) | 10 | 20 | Yes , repetition rate for<br>the subject device is a<br>subset of the cleared<br>specification for the<br>predicate device |
| Aiming beam | None | 635 nm | Yes, position at<br>intended treatment<br>site achieved through<br>specified means. |
| Treatment | | | |
| Optics | Focusing optics in<br>handpiece | Focusing optics in<br>handpiece | Yes, similar delivery of<br>focused laser energy |
| Tissue contact | Reference window | Distance gauge | Yes, treatment distance is<br>maintained through<br>specified means |
| Cooling mechanism | None | Integrated skin cooler for<br>optional use prior to laser<br>treatment | Yes, optional for<br>predicate system. |
| Energy delivery | Microbeam | Microbeam | Yes |
| Number of microbeams | 60 | 66 | Yes, similar number of<br>microbeams covering<br>same spot dimension |
| Spot diameter (mm) | 8 mm round | 8 mm round | Yes |
| Max fluence per<br>microbeam (J/cm²) | Up to 45 | Up to 44 | Yes |
| Output verification | Bench testing | Bench testing | Yes |
| Device<br>Characteristics | | | |
| User interface | Touchscreen | Touchscreen | Yes |
| Characteristic | Subject Device | Primary Predicate | Substantial<br>Equivalence? |
| Software Control | Yes | Yes | Yes |
| Laser trigger | Foot switch | Foot switch | Yes |
| Emergency shut-off | Yes | Yes | Yes |
| Safety | | | |
| FDA CDRH 21 CFR 1040 | Yes | Yes | Yes |
| Electrical safety<br>(IEC 60601-1) | Yes | Yes | Yes |
| Electromagnetic<br>compatibility<br>(IEC 60601-1-2) | Yes | Yes | Yes |
{5}------------------------------------------------
{6}------------------------------------------------
#### Risk Analysis
Risk analysis was performed according to IEC 14971:2007 Medical Devices- Application of Risk Management to Medical Devices, reviewed by a nationally recognized testing laboratory (NRTL), and found to be in compliance.
#### Summary of Performance Testing
Electrical Safety, Electromagnetic Compatibility, Biocompatibility and Laser Safety
Evaluation of the device was conducted by a nationally recognized testing laboratory (NRTL) and was found in compliance with the following standards:
- IEC 60601-1-2:2014 Ed 4.0 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 Medical electrical . equipment - Part 1: General requirements for basic safety and essential performance
- IEC 62304:2006 + A1:2015, Ed. 1.0 Medical device software Software life cycle ● processes
- IEC 60601-1-6:2010, AMD1:2013 Medical electrical equipment Part 1-6: . General requirements for basic safety and essential performance - Collateral standard: Usability
{7}------------------------------------------------
- IEC 62366-1: 2015 Medical devices Part 1: Application of usability engineering . to medical devices
- IEC 60601-2-22: 2019 Medical electrical equipment Part 2-22: Particular . requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- . IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation . and testing within a risk management process
## Non-Clinical Performance Data
The PL-1 Skin Treatment System performance characteristics have been evaluated through usability testing compliant to FDA Guidance Applying Human Factors and Usability Enqineering to Medical Devices (2016) and IEC 62366-1, and verification and validation of the PL-1 Skin Treatment System's laser energy, performance and control mechanisms. The performance of the PL-1 Skin Treatment System and related parameters of the predicate devices are substantially equivalent.
Software verification and validation testing was conducted, and documentation provided as recommended by FDA Guidance Content of Premarket Submissions for Software Contained in Medical Devices (2005) and IEC 62304. Device software verification and validation results were found acceptable for software release.
Biocompatibility of the patient contacting materials was established per FDA Guidance Use of International Standard ISO 10993-1 Biological evaluation of medical devices –Part 1: Evaluation and Testing Within a Risk Management Process (2018) and the referenced standard.
Animal studies and clinical performance testing were not deemed necessary as the device is using the same key technology, operating principles, and intended use as the predicate device.
{8}------------------------------------------------
### Summary of Substantial Equivalence
The PL-1 Skin Treatment System and the predicate devices have the same intended use for the 1064 Q-switched nanosecond mode with microbeam handpiece. The PL-1 Skin Treatment System presents similar technological characteristics as its predicate devices, including the laser type, wavelength, and device design. Although there are minor differences in the details of the device design, they are not sufficient to raise new questions of safety or efficacy.
#### Conclusion
The PL-1 Skin Treatment System is as safe and effective as the predicate devices operating in the 1064 Q-switched nanosecond mode with a microbeam handpiece. The new device has similar intended use and indications for use, technological characteristics and the same principle of operation as the predicate devices. The minor design differences raise no new issues of safety or efficacy as demonstrated by the performance data. The data show that the PL-1 Skin Treatment System performs in accordance with its specifications and requirements for both safety and effectiveness in similarity to the predicate devices. Thus, the PL-1 Skin Treatment System and the predicates are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.