K213725 · GE Medical Systems SCS · JAK · Mar 11, 2022 · Radiology
Device Facts
Record ID
K213725
Device Name
CardIQ Suite
Applicant
GE Medical Systems SCS
Product Code
JAK · Radiology
Decision Date
Mar 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213725 · Mar 11, 2022
CardIQ Suite
GE Medical Systems SCS
Retrospective CT clinical exam database
A database of retrospective clinical CT exams was used to validate the performance of the automated calcium scoring algorithm, ensuring it functions correctly across representative clinical acquisition protocols and indicators.
Calcium Scoring Algorithm Validation; Retrospective validation study
Clinical CT exams representative of intended use scenarios
Not applicable for this study
Algorithm capability to automatically segment, label, and score calcific regions
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Calcium scoring
Deep learning algorithm
—
Very high correlations between manual and automated methods
Database of retrospective CT exams
—
Representative set of clinical sample images
3 (board certified radiologists)
Indications for Use
CardIQ Suite is a collection of non-invasive software features intended to analyze CT cardiovascular anatomy and pathology and aid in determining treatment paths.
Device Story
CardIQ Suite is a non-invasive software application for analyzing cardiovascular CT DICOM data. It processes 2D/3D cardiac non-contrast and angiography images to provide automated/manual measurements, vessel visualization, and chamber mobility assessment. The device features an automated deep learning algorithm for segmenting and labeling coronary artery calcifications, calculating Agatston/Janowitz 130 (AJ 130) and Volume scores. It also includes a Coronary 2D Review toolset for interactive multi-phase data review, distance measurements, and ROI analysis. Used in clinical settings by radiologists/clinicians, the software integrates into existing workflows to assist in diagnosing cardiovascular disease and determining treatment paths. Output includes quantitative scores and visual data, which are exported as DICOM SR or images for reporting and archiving. The device aids clinical decision-making by providing standardized, reproducible measurements of cardiac structures and calcifications, potentially improving diagnostic efficiency and patient prognosis monitoring.
Clinical Evidence
Clinical evidence consists of a reader study comparing automated calcium scoring output from CardIQ Suite against manual scoring performed by three board-certified radiologists on a representative set of clinical images. Results demonstrated very high correlations between manual and automated methods for total calcium score, confirming equivalent performance to the predicate device.
Technological Characteristics
Software-based analysis tool for DICOM CT cardiac data. Employs deep learning algorithm for automated segmentation/labeling of calcifications and rule-based methods for Agatston/Volume scoring. Complies with NEMA PS 3.1-3.20 (DICOM) standards. Operates on the AW Server platform. Software level of concern: Moderate.
Indications for Use
Indicated for analysis of cardiovascular anatomy and pathology using 2D or 3D CT cardiac non-contrast and angiography DICOM data. Used for visualization/measurement of vessels and chamber mobility; aids diagnosis and treatment planning for coronary artery disease, heart structures, and follow-up for stents, bypasses, and plaque imaging. Includes calcium scoring for non-contrasted cardiac images to evaluate calcified plaques in coronary arteries, heart valves, and great vessels (aorta) and monitor calcium progression for cardiac disease prognosis.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
March 11, 2022
Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GE Medical Systems SCS % Yonghui Han Regulatory Affairs Leader 283, rue de la Miniere Buc, 78530 FRANCE
Re: K213725
Trade/Device Name: CardIQ Suite Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK, LLZ Dated: February 17, 2022 Received: February 18, 2022
Dear Yonghui Han:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K213725
Device Name
CardIQ Suite
#### Indications for Use (Describe)
CardIQ Suite is a non-invasive software application designed to provide an optimized application to analyze cardiovascular anatomy and pathology based on 2D or 3D CT cardiac non contrast and angiography DICOM data from acquisitions of the heart. It provides capabilities for the visualization and measurement of vessels and visualization of chamber mobility. CardIQ Suite also aids in diagnosis and determination of treatment paths for cardiovascular diseases to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging. CardIO Suite provides calcium scoring, a non-invasive software application, that can be used with non-contrasted cardiac images to evaluate calcified plaques in the coronary arteries, heart valves and great vessels such as the aorta. Calcium Scoring may be used to monitor the progression of calcium in coronary arteries overtime, which may aid in the prognosis of cardiac disease.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized, cursive font, enclosed within a circular shape. The circular shape is made up of swirling lines, giving it a dynamic and fluid appearance. The logo is presented in a blue color.
# 510(k) Summary
K213725
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | November 23, 2021 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems SCS<br>Establishment Registration Number - 9611343<br>283 rue de la Miniere<br>78530 Buc, France |
| Primary Contact Person: | Yonghui Han<br>Regulatory Affairs Leader<br>GE Healthcare<br>(86)10 57083350/(+86) 13311387032<br>Yonghui.Han@ge.com |
| Secondary Contact Person: | Elizabeth Mathew<br>Senior Regulatory Affairs Manager<br>GE Healthcare<br>Tel:(262)424-7774<br>Email: Elizabeth.Mathew@ge.com |
| Device Trade Name: | CardIQ Suite |
| Common/Usual Name: | CardIQ Suite |
| Primary Regulation Number: | Computed Tomography X-Ray System (21 CFR 892.1750) |
| Primary Product Code: | JAK |
| Secondary Product Code: | LLZ |
| Classification: | Class II |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by a series of swirling lines, resembling water droplets or waves, which are also blue. The overall design is simple and iconic, representing the company's brand identity.
| Predicate Device | |
|--------------------|--------------------------------------------------|
| Device name: | SmartScore 4.0 |
| Manufacturer: | GE Medical Systems SCS |
| 510(k) number: | K020929 |
| Regulation Number: | 21 CFR 892.1750 Computed tomography X-Ray System |
| Product Code: | JAK |
| Classification: | Class II |
| Reference Devices | |
| Device name: | CardIQ Xpress 2.0 |
| Manufacturer: | GE Medical Systems SCS |
| 510(k) number: | K073138 |
| Regulation Number: | 21 CFR 892.1750 Computed tomography X-Ray System |
| Product Code: | JAK |
| Classification: | Class II |
| Device name: | Syngo.CT CaScoring |
| Manufacturer: | Siemens Medical Solutions USA, Inc. |
| 510(k) number: | K201034 |
| Regulation Number: | 21 CFR 892.1750 Computed tomography X-Ray System |
| Product Code: | JAK |
| Classification: | Class II |
## Device Description:
CardlQ Suite is a non-invasive software application designed to work with DICOM CT data acquisitions of the heart. It is a collection of tools that provide capabilities for generating measurement's both automatically and manually, displaying images and associated measurements in an easy-to-read format and tools for exporting images and measurements in a variety of formats.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are white swirls around the outside of the circle. The logo is simple and recognizable, and it is often used to represent the company.
CardIQ Suite provides an integrated workflow to seamlessly review calcium scoring and coronary CT angiography (CCTA) data. Calcium Scoring has the capability to automatically segment and label the calcifications within the coronary arteries, and then automatically compute a total and per territory calcium score. The calcium segmentation/labeling is using algorithm. The calcium scoring is based on the standard Agatston/Janowitz 130 (AJ 130) and Volume scoring methods for the segmented calcific regions. The software also provides the users a manual calcium scoring capability that allows them to edit (add/delete or update) auto scored lesions. It also allows the user to manually score calcific lesions within coronary arteries, aorta, aortic valve as well as other general cardiac structures. Calcium scoring offers quantitative results in the AJ 130 score, Volume and Adaptive Volume scoring methods.
Calcium Scoring results can be exported as DICOM SR to assist with integration into structured reporting templates. Images can be saved and exported for sharing with referring physicians, incorporating into reports and archiving as part of the CT examination.
CardlQ Suite provides the Coronary 2D Review toolset which allows interactive review of cardiac exams. Coronary CTA datasets can be reviewed utilizing the double oblique angles to visually track the path of the coronary arteries as well as to view the common cardiac chamber orientations. Cine capability for multi-phase data may be useful for visualization of cardiac structures such as chambers, valves and arteries in motion. Distance measurement and ROI tools are available for quantitative evaluation of the anatomy.
#### Intended Use:
CardIQ Suite is a collection of non-invasive software features intended to analyze CT cardiovascular anatomy and pathology and aid in determining treatment paths.
#### Indication for Use:
CardIQ Suite is a non-invasive software application designed to provide an optimized application to analyze cardiovascular anatomy and pathology based on 2D or 3D CT cardiac non contrast and angiography DICOM data from acquisitions of the heart. It provides capabilities for the visualization and measurement of vessels and visualization of chamber mobility. CardIQ Suite also aids in diagnosis and determination of treatment paths for cardiovascular diseases to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging.
CardIQ Suite provides calcium scoring, a non-invasive software application, that can be used with noncontrasted cardiac images to evaluate calcified plaques in the coronary arteries, heart valves and great vessels such as the aorta. Calcium Scoring may be used to monitor the progression of calcium in coronary arteries overtime, which may aid in the prognosis of cardiac disease.
#### Technology:
The proposed device CardIQ Suite employs the same fundamental scientific technology as its predicate device and reference devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle has a swirling design around the letters. The logo is simple and recognizable.
## Comparison:
The table below summarizes the key feature/technological differences and similarities between the predicate device and the proposed device:
| Specification | Predicate<br>Device:<br>SmartScore 4.0<br>(K020929) | Proposed<br>Device:<br>CardIQ Suite | Comparison |
|--------------------------------------------------------------------------|-----------------------------------------------------|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Segmentation<br>and labeling<br>calcific regions<br>in the<br>coronaries | Manual | Automated | Substantial Equivalent<br>Deep Learning Algorithm is incorporated in CardIQ Suite to<br>automatically segment and label the calcific regions in the<br>coronary arteries in order to improve workflow efficiency<br>over the manual approach that exists in the predicate<br>device.<br>Automated calcium scoring evaluation already exists in the<br>reference device Syngo.CT CaScoring (K201034). |
| Manual<br>Segmentation<br>and labeling of<br>calcific regions | Yes | Yes | Identical |
| Computation<br>of Agatston<br>score | Yes | Yes | Identical |
| Coronary 2D<br>Review | Not Available | Yes | Substantial Equivalent<br>Coronary 2D Reviews Toolset is available in CardIQ Suite to<br>assist users for a seamless integrated review of coronary<br>artery imaging along with automated calcium scoring<br>feature.<br>This tool set contains identical features to what already<br>exists in the reference device CardIQ Xpress 2.0 (K073138) |
## Determination of Substantial Equivalence:
#### Summary of Non-Clinical, Design Control Testing
CardIQ Suite has successfully completed the design control testing per GE's quality system. It was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. No additional hazards were identified, and no unexpected test results were observed. The proposed
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular shape. The color of the logo is a bright blue. The logo is simple and recognizable, representing the General Electric brand.
device complies with NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard.
The following quality assurance measures were applied to the development of the device:
- Requirements Definition
- Risk Analysis
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- Performance testing (Verification, Validation)
- Safety Testing (Verification)
The proposed CardIQ Suite has been successfully verified on the AW Server platform. All the testing and results did not raise new or different questions of safety and effectiveness other than those already associated with predicate device. Software documentation is for a MODERATE level of concern.
In addition, Engineering has validated CardIQ Suite Calcium Scoring algorithm's capability to automatically segment, label and score the calcific regions in the coronary arteries using a database of retrospective CT exams. This database of exams is representative of the clinical scenarios where CardlQ Suite is intended to be used, with consideration of acquisition protocols and clinical indicators. The result of the algorithm validation showed that the algorithm successfully passed the defined acceptance criteria.
#### Summary of Clinical Testing
Three board certified radiologists manually scored a representative set of clinical sample images using the predicate device and the results were compared with the automated score outputted by CardIQ Suite. Very high correlations were found between manual and automated methods for computing the total calcium score, demonstrating equivalent performance of the CardIQ Suite software to the predicate device SmartScore 4.0.
#### Conclusion:
CardIQ Suite has substantial equivalent technological characteristics as its predicate device.
GE's quality system's design, verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Based on development under GE Healthcare's quality system, successful design verification, software documentation for a "Moderate" level of concern, along with the engineering bench testing and sample reader study GE Healthcare believes that the proposed CardIQ Suite is substantially equivalent to, and hence as safe and as effective for its Intended Use as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.