ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays or veneers. ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.
Device Story
Dental glass ceramics (powders/pastes) used to coat, stain, and glaze final-sintered zirconia dental restorations; includes ICE Ceramics, ICE Stains, and Fresco product lines. Applied in layers by trained dental personnel in a laboratory setting; requires firing processes followed by manual grinding and polishing. Accessories (liquids/gels) adjust viscosity or refresh materials. Output is a customized, aesthetic dental restoration. Benefits include improved appearance and surface finish of zirconia-based dental prosthetics.
Clinical Evidence
Bench testing only. Physical properties evaluated per ISO 6872 (flexural strength ≥ 50 MPa; chemical solubility < 100 µg/cm²). Biocompatibility established per ISO 10993-1 and ISO 10993-5.
Technological Characteristics
Glass ceramic composition with metallic oxides and pigments. Supplied as powders, pastes, and liquids. Classified as Type 1, class 1b per ISO 6872. Biocompatibility per ISO 10993. No electronic components or software.
Indications for Use
Indicated for patients requiring dental restorations in zirconia, including single crowns, bridges, inlays, onlays, or veneers, for the purpose of coating, staining, and glazing.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2022
Zirkonzahn srl Sandra Leitner Regulatory Affairs Responsible Via An der Ahr 7 Gais, BZ 39030 Italy
Re: K213722
Trade/Device Name: ICE Ceramics, ICE Stains, Fresco Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: November 10, 2022 Received: November 14, 2022
Dear Sandra Leitner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213722
Device Name ICE Ceramics, ICE Stains and Fresco
#### Indications for Use (Describe)
ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays or veneers.
ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 05 510(k) Summary - K213722
#### 510(k) Summary
## APPLICANT
Zirkonzahn srl Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: info@zirkonzahn.com
#### CONTACT PERSON
Julian Steger Chief Executive Officer Zirkonzahn srl Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: julian.steger@zirkonzahn.com
#### DATE SUMMARY PREPARED: December 13, 2022
#### DEVICE IDENTIFICATION
| Trade/Proprietary Name: | ICE Ceramics, ICE Stains and Fresco |
|-------------------------|-------------------------------------|
| Regulation Number: | 21 CFR 872.6660 |
| Common Name: | Powder, Porcelain |
| Classification Name: | Porcelain Powder for Clinical use |
| Class: | II |
| Product Code: | EIH |
| Panel: | Dental |
{4}------------------------------------------------
#### LEGALLY MARKETED PREDICATE DEVICE
| Company: | Zirkonzahn srl |
|----------------|----------------|
| Device Name: | Zirkonzahn Ice |
| Product Code: | EIH |
| 510(k) Number: | K061851 |
#### INDICATIONS FOR USE
ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.
ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.
#### DEVICE DESCRIPTION
All devices included in this submission are dental glass ceramics. There are different
device types from three device groups:
#### ICE Ceramics Group:
- ICE Ceramics
- ICE Ceramics Tissue
- ICE Ceramics Dentine +
- . ICE Ceramics Dynamic Dentine
#### ICE Stains Group:
- . ICE Stains
- ICE Stains Prettau® ●
- ICE Stains 3D ●
- ICE Stains Glaze Fluo .
#### Fresco Group:
- Fresco ●
- Fresco Gingiva
- Fresco Enamel ●
- Fresco Dentine +
- Fresco Dynamic Dentine ●
The above mentioned devices are supplied in form of powders or pastes and are available in different colors and/ or in different quantity sizes. The users can choose among the different options according to their requirements and preferences. Furthermore specific accessories are supplied to optimize the working procedure:
{5}------------------------------------------------
ICE BUILD UP LIQUID is an accessory used for mixing of ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE + and ICE CERAMICS DYNAMIC DENTINE. FRESCO LIQUID and FRESCO GEL are accessories used to change the viscosity of FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE products.
ICE STAIN LIQUID is an accessory used to freshen up ICE STAINS und ICE STAINS PRETTAU®; REFRESHING LIQUID 3D is an accessory used to freshen up ICE STAINS 3D; REFRESHING LIQUID GLAZE FLUO is an accessory used to freshen up ICE STAINS GLAZE FLUO.
All these devices are used by specifically trained personnel to manufacture customized dental restorations.
### DISCUSSION OF NON CLINCAL TESTS
Non clinical testing was performed to determine the physical properties of the subject device. Performance testing was conducted in accordance with the International Standard ISO 6872 and the results show that the devices are well suited for their intended use. Furthermore, biocompatibility was established in according to ISO 10993-1 and ISO 10993-5.
{6}------------------------------------------------
# COMPARISON TABLE
| Devices | New Devices:<br>ICE Ceramics, ICE Stains and Fresco<br>K213722 | Predicate devices:<br>Zirkonzahn Ice<br>K061851 | Comparison |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Zirkonzahn srl | Zirkonzahn srl | |
| Product Code | EIH | EIH | Same |
| Regulation<br>Number | 872.6660 | 872.6660 | Same |
| Regulation Name | Porcelain powder for clinical use | Porcelain powder for clinical use | Same |
| Indications for<br>use | ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS<br>DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO<br>ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and<br>FRESCO DYNAMIC DENTINE are type I ceramics for coating,<br>staining and glazing dental restorations made of final-sintered<br>zirconia. They are conceived for patients who require dental<br>restorations in zirconia, such as single crowns, bridges, inlays, onlays<br>or veneers.<br>ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO<br>and ICE STAINS 3D are type I ceramics for<br>coating, staining and glazing dental restorations made of final-sintered<br>zirconia. They are conceived for patients who<br>require dental restorations in zirconia, such as single crowns, bridges,<br>inlays, onlays or veneers. | Metal free single posterior crowns,<br>multiple unit anterior<br>crowns/bridges, inlays, onlays<br>bonded dental restorations. | Similar<br>Zirkonzahn Ice contains a<br>system of devices<br>(including zirconia blanks,<br>coloring liquids), not only<br>dental ceramics, but the<br>new as well as the<br>predicate devices are<br>intended to veneer /<br>individualize dental<br>frameworks. |
| Sizes of dental<br>frameworks | All, dependent on zirconia device where applied | All, dependent on zirconia device<br>where applied | Same |
| Chemical<br>composition | Glass ceramic, other metallic oxides, with pigments | Glass ceramic, other metallic oxides, with pigments | Similar |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5 | Not specified in K061851 | Similar |
| Tested according<br>to ISO 6872 | Flexural strength ≥ 50 MPa<br>Chemical solubility < 100 µg/cm² | Flexural strength ≥ 50 MPa<br>Chemical solubility < 100 µg/cm² | Similar |
| Classification<br>acc. to ISO 6872 | Type 1, class 1b | Type 1, class 1b | Same |
| Principles of<br>Operation | The devices can be applied in several layers on final-sintered<br>zirconia structures followed by firing processes. Subsequently,<br>the dental structures are finalized by manual grinding and<br>polishing. | The devices can be applied in several layers on final-sintered<br>zirconia structures followed by firing processes. Subsequently, the<br>dental structures are finalized by manual grinding and polishing. | Same |
| Shades | Dentine shades (Vita dentine shades A1 –D4 etc.)<br>Intensive colors<br>Glazes | Dentine shades (Vita dentine shades A1 -D4 etc.)<br>Intensive colors<br>Glazes | Similar |
| Form | Powders, pastes and liquids | Powders, pastes and liquids | Same |
| Storage<br>Conditions | In tightly sealed original packaging with a maximal temperature<br>of 35 °C for ICE Stains, Fresco and accessory liquids.<br><br>If stored too cold, the liquid may freeze and must be defrosted<br>before use.<br><br>If stored too warm, the pastes may dry out. | Storage temperature between<br>10 °C and 30 °C | Differences do not have<br>an effect on final medical<br>device, therefore<br>difference not significant. |
{7}------------------------------------------------
{8}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The new devices and the predicate devices are similar in function and intended use. The differences noted do not raise any new questions about safety and effectiveness. Therefore, Zirkonzahn concludes that the proposed devices are substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.